AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K211788 · Jul 8, 2021
HALO
Nicolab B.V.
Retrospective consecutive patient cohort from US comprehensive stroke centers
A retrospective multi-center study was conducted to evaluate the sensitivity, specificity, and notification time of the HALO algorithm for detecting Large Vessel Occlusion (LVO) in patients suspected of stroke.
Retrospective cohort; LVO detection; Stroke; Multi-center study
Patients suspected of stroke admitted to US comprehensive stroke centers; Sample Size: 427 patients (after exclusion from 434 scans); Number of Sites: Multi-center
Not applicable for this study
Sensitivity, specificity, and notification time for LVO detection
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Intracranial Large Vessel Occlusion (LVO) detection
HALO is a notification only cloud-based image processing software application using artificial intelligence algorithms to analyze patient imaging data in parallel to the standard of care imaging interpretation. Its intended use is to identify suggestive imaging patterns of a pre-specified clinical condition and to directly notify an appropriate medical specialist. HALO's indication is to facilitate the evaluation of the brain vasculature on patients suspected of stroke by processing and analyzing CT angiograms of the brain acquired in an acute setting. After completion of the data analysis, HALO sends a notification if a pattern suggestive for a suspected intracranial Large Vessel Occlusion (LVO) of the anterior circulation (ICA, M1 or M2) has been identified in an image. The intended users of HALO are defined as medical specialists or a team of specialists that are involved in the diagnosis and care of stroke patients at emergency departments or other departments where stroke patients are administered. They include physicians such as neurologists, radiologists, and/or other emergency department physicians. HALO's output should not be used for primary diagnosis or clinical decisions; the final diagnosis is always decided upon by the medical specialist. HALO is indicated for CT scanners from GE Healthcare and Philips.
Device Story
Cloud-based clinical support tool; processes CT angiograms of brain; uses AI algorithms to detect patterns suggestive of intracranial Large Vessel Occlusion (LVO) of anterior circulation. Input: CTA images from GE Healthcare or Philips scanners; Output: notification to medical specialist (neurologist, radiologist, ED physician) via email; includes link to DICOM web viewer for remote review. Operates in parallel to standard-of-care workflow; does not remove cases from worklist; does not perform image segmentation or highlighting. Specialist makes final clinical diagnosis; device output intended for triage/notification only. Benefits: facilitates faster identification of LVO cases, potentially reducing time to treatment for stroke patients.
Clinical Evidence
Retrospective multi-center study; 427 patients (434 scans). Ground truth established by 3 neuroradiologists. Primary endpoint: LVO detection performance; sensitivity 91.3% (95% CI: 86.6%-94.8%), specificity 85.9% (95% CI: 80.6%-90.2%), AUC 0.97. Secondary endpoint: median notification time 4 minutes 29 seconds (range 3:47-7:12).
Technological Characteristics
Cloud-based software; AI-based image processing; DICOM-compatible; supports GE Healthcare and Philips CT scanners. No image segmentation or highlighting. Triage notification via email with web-viewer link.
Indications for Use
Indicated for patients suspected of stroke requiring evaluation of brain vasculature via CT angiograms in acute settings. Identifies patterns suggestive of intracranial Large Vessel Occlusion (LVO) of the anterior circulation (ICA, M1, or M2).
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
July 8, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an eagle emblem, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" in a larger font size and a blue square behind it.
Nico.Lab B.V. % Ruojuan Zhang RA Consultant MD Squared B.V. High Tech Campus 29 Eindhoven. Noord Brabant 5656 AE THE NETHERLANDS
### Re: K211788
Trade/Device Name: HALO Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS Dated: June 5, 2021 Received: June 9, 2021
### Dear Ruojuan Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K211788
Device Name HALO
#### Indications for Use (Describe)
HALO is a notification only cloud-based image processing software artificial intelligence algorithms to analyze patient imaging data in parallel to the standard of care imaging interpretation. Its intended use is to identify suggestive imaging patterns of a pre-specified condition and to directly notify an appropriate medical specialist.
HALO's indication is to facilitate the evaluation of the brain vasculature on patients suspected of stroke by processing and analyzing CT angiograms of the brain acquired in an acute setting. After completion of the data analysis, HALO sends a notification if a pattern suggestive for a suspected intracranial Large Vessel Occlusion (LVO) of the anterior circulation (ICA, M1 or M2) has been identified in an image.
The intended users of HALO are defined as medical specialists or a team of specialists that are involved in the diagnosis and care of stroke patients at emergency department where stroke patients are administered. The include physicians such as neurologists, radiologists, and/or other emergency department physicians.
HALO's output should not be used for primary diagnosis or clinical decisions; the final diagnosis is always decided upon by the medical specialist. HALO is indicated for CT scanners from GE Healthcare and Philips.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <div> <span> <span style="font-size:16px">✓</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | Over-The-Counter Use (21 CFR 801 Subpart C) |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | |
| Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br><a href="mailto:PRAStaff@fda.hhs.gov">PRAStaff@fda.hhs.gov</a> | |
| <i>"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."</i> | |
{3}------------------------------------------------
# 510(k) Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Submission Number: | K211788 | |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| Date Prepared: | June 4, 2021 | |
| Manufacturer: | NICO.LAB B.V.<br>Paasheuvelweg 25, 1105 BP Amsterdam<br>The Netherlands | |
| Primary Contact<br>Person: | Merel Boers<br>CEO<br>Phone: +31 6 22177913<br>E-mail: mboers@nico-lab.com | |
| Device: | | |
| | Trade Name: | HALO |
| | Classification Name: | Radiological Computer-Assissted Triage and<br>Notification Software |
| | Classification Regulation: | 21CFR §892.2080 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Primary Product Code: | QAS |
| Primary Predicate<br>Device: | | |
| | Trade Name: | HALO |
| | Manufacturer: | NICO.LAB B.V. |
| | 510(k) Clearance: | K200873 (November 20, 2020) |
| | Classification Name: | Radiological Computer-Assissted Triage and<br>Notification Software |
| | Classification Regulation: | 21CFR §892.2080 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Product Code: | QAS |
| Device description: | HALO is a notification only, cloud-based clinical support tool which<br>identifies image features and communicates the analysis results to a<br>specialist in parallel to the standard of care workflow.<br><br>HALO is designed to process CT angiograms of the brain and facilitate<br>evaluation of these images using artificial intelligence to detect patterns<br>suggestive of an intracranial large vessel occlusion (LVO) of the anterior<br>circulation.<br><br>A copy of the original CTA images is sent to HALO cloud servers for<br>automatic image processing. After analyzing the images, HALO sends a<br>notification regarding a suspected finding to a specialist, recommending<br>review of these images. The specialist can review the results remotely in<br>a compatible DICOM web viewer. | |
| Indications for Use: | HALO is a notification only cloud-based image processing software<br>application using artificial intelligence algorithms to analyze patient<br>imaging data in parallel to the standard of care imaging interpretation. Its<br>intended use is to identify suggestive imaging patterns of a pre-specified<br>clinical condition and to directly notify an appropriate medical specialist.<br><br>HALO's indication is to facilitate the evaluation of the brain vasculature<br>on patients suspected of stroke by processing and analyzing CT<br>angiograms of the brain acquired in an acute setting. After completion of<br>the data analysis, HALO sends a notification if a pattern suggestive for a<br>suspected intracranial Large Vessel Occlusion (LVO) of the anterior<br>circulation (ICA, M1 or M2) has been identified in an image.<br><br>The intended users of HALO are defined as medical specialists or a team<br>of specialists that are involved in the diagnosis and care of stroke patients<br>at emergency departments or other departments where stroke patients are<br>administered. They include physicians such as neurologists, radiologists,<br>and/or other emergency department physicians.<br><br>HALO's output should not be used for primary diagnosis or clinical<br>decisions; the final diagnosis is always decided upon by the medical<br>specialist. HALO is indicated for CT scanners from GE Healthcare and<br>Philips. | |
| Technological<br>characteristics: | The subject device, HALO, is substantially equivalent to the predicate<br>device, the previously cleared version of the HALO (K200873). In<br>comparing the technological characteristics, both the subject and<br>predicate devices process CT angiograms of the brain and facilitate<br>evaluation of these images using artificial intelligence to detect patterns<br>suggestive of an intracranial large vessel occlusion (LVO) of the anterior | |
{4}------------------------------------------------
{5}------------------------------------------------
circulation. After analyzing the images, the device sends a notification regarding a suspected finding to a specialist, recommending review of these images. Where the subject and predicate only differ is that the subject device not only process images from GE Healthcare, but also Philips scanners.
The technological characteristic of HALO and how it is comparable to the predicate device are summarized below.
| | Predicate<br>HALO | Subject<br>HALO |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical condition | Large vessel<br>occlusion | Large vessel<br>occlusion |
| Anatomical region<br>of interest | Head | Head |
| Data acquisition<br>protocol | CT angiogram images<br>of the brain | CT angiogram images<br>of the brain |
| Segmentation of<br>region of interest | No; device does not<br>mark, highlight, or<br>direct users' attention<br>to a specific location<br>in the original image. | No; device does not<br>mark, highlight, or<br>direct users' attention<br>to a specific location<br>in the original image. |
| Core Algorithm | Artificial intelligence<br>algorithm with<br>database of images | Artificial intelligence<br>algorithm with<br>database of images |
| Device Output/<br>Notification | The software sends a<br>notification email to<br>the specialist<br>identifying the study<br>of interest.<br>Additionally, the<br>device provides user<br>with a link to DICOM<br>Web viewer allowing<br>users review the<br>images. | The software sends a<br>notification email to<br>the specialist<br>identifying the study<br>of interest.<br>Additionally, the<br>device provides user<br>with a link to DICOM<br>Web viewer allowing<br>users review the<br>images. |
| Triage effectiveness | Notification time is<br>defined as the time<br>between upload of<br>CTA (CTA images<br>are received in the<br>cloud) to the<br>availability of AI<br>results (user receives<br>a notification to view | Notification time is<br>defined as the time<br>between upload of<br>CTA (CTA images<br>are received in the<br>cloud) to the<br>availability of AI<br>results (user receives<br>a notification to view |
| | the suspected<br>finding). | the suspected<br>finding). |
| Independent of<br>standard of care<br>workflow | No cases are removed<br>from worklist | No cases are removed<br>from worklist |
{6}------------------------------------------------
# Performance Data:
Summary of Clinical In a multi-center clinical study, the performance of the HALO clinical decision support algorithm for LVO detection was retrospectively evaluated in a consecutive patient cohort admitted to US comprehensive stroke centers.
> 434 CTA scans of the brain were collected. After exclusion, 427 patients were included for further analyses. Ground truth was established by an expert panel consisting of 3 neuro radiologists. For the primary endpoint: calculation of the performance of the HALO algorithm showed a sensitivity and specificity for LVO detection of respectively 91.3% (95% CI, 86.6%-94.8%) and 85.9% (95% CI, 80.6%-90.2%). The area under the curve (AUC) is 0.97.
> For the secondary endpoints the median notification time for the detected LVO cases was 4 minutes 29 seconds, with a minimum of 3:47 and maximal 7:12.
The HALO performance with regard to sensitivity and specificity, and the notification time are both equivalent to that of the selected predicate device. Therefore, the HALO algorithm fulfils the requirement of a suitable screening tool to support diagnosis of LVOs by flagging these scans as requiring urgent radiologist review.
Substantial The clinical performance tests provided in this 510(k) premarket Equivalence notification demonstrate HALO is substantial equivalence to its predicate Conclusion: device, the previously cleared HALO (K200873). HALO has the same intended use and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the triage use of the device and do not affect its safety and effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.