K211746 · Guangdong Youkey Medical Co., Ltd. · IYN · Feb 18, 2022 · Radiology
Device Facts
Record ID
K211746
Device Name
Pocket Ultrasound System
Applicant
Guangdong Youkey Medical Co., Ltd.
Product Code
IYN · Radiology
Decision Date
Feb 18, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Indications for Use
Pocket Ultrasound System(Model:P50,P51,P52,P53,P54,P55,P56) is intended for use by an appropriately trained, qualified physician for ultrasound evaluation. Specific clinical applications and exam types include: Abdominal, Obstetrics, Gynaecology, Small Parts (breast, thyroid), Peripheral Vascular, Urology. The system is suitable for use in hospital and clinical settings.
Device Story
Pocket Ultrasound System consists of wireless ultrasound probe and companion APP software; probe emits/receives ultrasonic waves; host processes echo signals to generate anatomical/blood flow images; APP deployed on Windows/Android devices via Wi-Fi; system provides B-mode, M-mode, Color/Power Doppler, and PW-mode; used by physicians in hospital/clinic for diagnostic evaluation; provides real-time visualization and measurement tools; aids clinical decision-making by offering detailed anatomical/vascular information; benefits patients through non-invasive, portable diagnostic imaging.
Clinical Evidence
No clinical studies were required. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), biocompatibility (ISO 10993-5/10), acoustic output (FDA Track 3), and software verification.
Technological Characteristics
Portable ultrasound system; wireless transducers (convex, linear, phased array, intracavity); frequency 3.5-7.5MHz; Wi-Fi connectivity to Windows/Android host; removable Li-ion battery; B/M/Color/Power/PW Doppler modes; biocompatible materials per ISO 10993; safety compliance per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-37.
Indications for Use
Indicated for ultrasound evaluation in patients requiring abdominal, obstetric, gynecological, small parts (breast, thyroid), peripheral vascular, or urological imaging. Suitable for use by trained physicians in hospital and clinical settings.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
GuangDong Youkey Medical Co., Ltd. % Hu Zeyi General Manager Unit 601, 6/F, Block B, Building 1, B1 District, Hantian Technology City, Dongping Road, Pingxi Shanghain Village, Guichen Street, Nanhai District Foshan City, Guangdong 528203 CHINA
#### Re: K211746
February 18, 2022
Trade/Device Name: Pocket Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: January 21, 2022 Received: January 21, 2022
Dear Hu Zeyi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K211746
Device Name Pocket Ultrasound System
Indications for Use (Describe)
Pocket Ultrasound System(Model:P50,P51,P52,P53,P54,P55,P56) is intended for use by an appropriately trained, qualified physician for ultrasound evaluation. Specific clinical applications and exam types include: Abdominal, Obstetrics, Gynaecology, Small Parts (breast, thyroid), Peripheral Vascular, Urology. The system is suitable for use in hospital and clinical settings.
Type of Use (Select one or both, as applicable):
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M5.44433 11.6667L14.6667 2.44433M14.6667 14.6667L2.44433 2.44433" stroke="black" stroke-linecap="round" stroke-linejoin="round" stroke-width="2"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="12" stroke="black" stroke-width="2" width="12" x="2" y="2"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K211746
# 510(k) Summary
| Submitter and Address of<br>Manufacturing Facility | GuangDong Youkey Medical Co., Ltd.<br>Unit 601, 6/F, Block B, Building 1, B1 District, Hantian<br>Technology City, Dongping Road, Pingxi Shanghai<br>Village, Guichen Street, Nanhai District, Foshan City,<br>Guangdong, 528203, CHN |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Phone: +86-0757-86258600 |
| | FAX: 0757-86258600 |
| Date: | 2021-5-20 |
| Contact: | Hu Zeyi |
| Trade name: | Pocket Ultrasound System |
| Common Name: | Diagnostic Ultrasound System and Transducer |
| Classification Regulation/<br>Product Code: | 21 CFR 892.1550, Class II/ IYN, IYO, ITX |
| Establishment Registration<br>Number: | NA |
| Reason for Premarket<br>Notification: | New Device |
| Predicate Devices: | K192107 Clarius Ultrasound Scanner |
## Device Description
The Pocket Ultrasound System is designed to be high-end Pocket Ultrasound Systems and mainly include the ultrasound equipment and APP software components.6 models the the the main units are included in this submission, for including P50,P51,P52,P53,P53,P55,P56.
The equipment to complete the ultrasonic imaging function; APP complete the image display, measurement, patient information and image storage and other applications. APP can be deployed on Windows or Android devices which supported WIFI. There are two free APPs to support different systems. The communication protocol between APP and the system is specially defined by ourself, and other manufacturers can not use this protocol. The ultrasound device and the APP software transfer images and control instructions via WIFI.
{4}------------------------------------------------
Different type transducers are available for different model of Pocket Ultrasound System, the frequency scope of different transducer are from 3.5MHz to 6.5MHz, refer to the following table .
| Model | Is the<br>transducer<br>replaceable | Support transducer | Intended use |
|-------|-------------------------------------|---------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| P50 | Yes | C5-2Fs, C5-2Ks, L11-4Ks, L11-4Gs, C8-5Ks, E10-4Ks | Abdominal, Obstetrics, Gynaecology,<br>Small Parts (breast, thyroid),<br>Peripheral Vascular, Urology |
| P51 | No | C5-2Fs | Gynecology and obstetrics |
| P52 | No | C5-2Ks | abdominal kidney |
| P53 | No | L11-4Ks | |
| P54 | No | L11-4Gs | |
| P55 | No | C8-5Ks | Small organ carotid artery |
| P56 | No | E10-4Ks | Department of Obstetrics and<br>Gynecology, urinary system |
The system provides a variety of modes for flexibly selecting, which include B-mode, M-mode, Color-mode, Power-mode, PW-mode, and combined modes (i.e. B/M- mode). Real time image and dynamic changes of tissues, organs or vessels will be showed in the image under B-mode, while coloring blood flow can be seen under Color-mode, PW-mode or Power-mode. M-mode is defined as time motion display of the ultrasound wave along a chosen ultrasound line, so it 's usually applied in diagnosis of cardiovascular diseases.
The system not only provides real-time tissues, organs or blood flow images, but also contains measurement of anatomical structures and analysis packages, which can offer diagnosticians detailed and useful information. The embedded software system along with the touch screen offer users a friendly and convenient graphical interface, which make users be able to easily control the whole system.
### Indications for Use
Pocket Ultrasound System(Model:P50.P51.P52.P53.P54.P55.P56) is intended for use by an appropriately trained, qualified physician for ultrasound evaluation. Specific clinical applications and exam types include: Abdominal, Obstetrics, Gynaecoloqy, Small Parts (breast, thyroid, etc), Peripheral Vascular, Urology. This system is suitable for use in hospital and clinical settings.
## Technological Principle
The transducers inside the probe are responsible to emitting and receiving ultrasonic waves. When the host pushes pulse with a certain frequency and magnitude onto the transducers, a short burst of ultrasound is emitted from transducers and directed into tissue. Echoes are produced as a result of the interaction of sound with tissue, and some of back to the transducers. Echoes come back is then converted into electric signals by transducers and handed to host for further process.
{5}------------------------------------------------
The host is mainly responsible for the following: data processing, in response to user instructions. By receiving the echo pulse transmitting time and timing between the transducer and the echo signal can be calculated by the distance between the structure, form the brightness degree of adjacent image materials. The inherent acoustic impedance difference in the image are different. And through the WIFI transmission and instruction imaging data.
Frequencies vary from about 3MHz for some cardiac, transcranial and deep abdominal applications, to 7MHz for intravascular imaging. Penetration depth and resolution are both associated with the frequency used for imaging, and they are a trade-off as well. The higher frequency is used, better resolution will get, but absorption of the sound energy by tissue also increases which means penetration depth reduces. Proper probes and imaging frequency should be chosen to obtain optimum imaging according to specific application.
Color Doppler imaging takes advantage of the Doppler effect to create a moving image, typically blood flow, of the inside of the body. When the sound bounces off a moving target like a blood vessel, the frequency changes as a result of the Doppler effect. The transducer can detect very subtle frequency changes and record them visually, creating an image which shows where blood is flowing, and in which direction.
Moreover, although the probes need to contact with human body surface in use, the system will not cause any invasive result to or thermal radiation into human body.
### Substantial Equivalence Comparison
Equivalent devices are referred to as predicate devices in alignment with the FDA ' s standard terminology for comparable devices. The predicate devices elected to demonstrate equivalence is the Clarius Ultrasound Scanner(K192107). A comparison of the subject device and the predicate are provided in the table below:
| Criteria for<br>comparison | Subject device | Predicate device (K192107) |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | Pocket Ultrasound System | Clarius Ultrasound Scanner |
| Model | P50,P51,P52,P53,P54,P55,<br>P56 | C3, C7, L7, EC7, C3 HD, L7<br>HD, C7 HD, EC7 HD, L15 HD,<br>PA HD |
| Intended Use and Classification | | |
| Classification | Class II | Class II |
| | The Pocket Ultrasound | Diagnostic ultrasound |
| Intended use | System is intended for use<br>by a qualified physician<br>for ultrasound evaluation | imaging and fluid flow<br>analysis |
| Indications for<br>Use | Abdominal, Obstetrics,<br>Gynaecology, Small Parts<br>(breast, thyroid, etc), | -Ophthalmic<br>-Fetal<br>-Abdominal |
| Criteria for<br>comparison | Subject device | Predicate device (K192107) |
| | Peripheral Vascular,<br>Urology. | -Intraoperative (Ab/Vasc)<br>-Pediatric<br>-Small organ<br>-Adult cephalic |
| | | -Trans<br>-rectal<br>-Trans<br>-vaginal<br>-Musculo-skel. (Conv.)<br>-Musculo-skel. (Superfic.)<br>-Urology<br>-Gynecology<br>-Cardiac adult<br>-Cardiac pediatric<br>-Peripheral vessel<br>-Carotid |
| Technological Characteristics | | -Needle guidance |
| Portability | Portable ultrasound system | Portable ultrasound system |
| Duration of Use | Limited (<24 hours) | Limited (<24 hours) |
| Transducer<br>Types | Convex Array<br>Linear Array<br>Phased Array<br>Intracavity | -Convex Array<br>-Linear Array<br>-Phased Array<br>-Intracavity |
| Frequency | 3.5-7.5MHz | 1-20MHz |
| Environment<br>of<br>Use | Hospital, clinic, and medical<br>office settings | Hospital, clinic, and medical<br>office settings |
| Power Source | Removable battery (Li-ion) | Removable battery (Li-ion) |
| Wireless<br>Capability | Communicates wirelessly<br>via Wi-Fi | Communicates wirelessly via<br>Wi-Fi and Bluetooth |
| Acoustic Output<br>Levels | ISPTA max=720 mw/cm²<br>MI max =1.9<br>MI display<br>TI display | Below Track 3 FDA limits in<br>accordance with Sept. 2008<br>ultrasound systems guidance<br>document |
| Imaging<br>Capacities | B mode<br>M mode<br>Color Doppler<br>Power Doppler<br>PWD and Combined (B+M;<br>B+CD; B+PD; B+PWD) | -B-mode<br>-M-mode<br>-Color Doppler<br>-Power Doppler<br>-PWD<br>-Combined (B+M; B+CD;<br>B+PD; B+PWD) |
| Criteria for<br>comparison | Subject device | Predicate device (K192107) |
| Patient<br>Population | For use in all patients | For use in all patients |
| Users | Healthcare professionals | trained healthcare<br>professionals. |
| Principle/<br>Method of<br>Operation | The basic principle is that<br>system transmits ultrasonic<br>energy into patient body and<br>implements post processing<br>of received echoes to<br>generate onscreen display of<br>anatomic structures and fluid<br>flow within the body.<br>The equipment to complete<br>the ultrasonic imaging<br>function; APP complete the<br>image display, measurement,<br>patient information and<br>image storage and other<br>applications. APP can be<br>deployed on Windows or<br>Android devices which<br>supported WIFI. There are<br>two free APPs to support<br>different systems. The<br>communication protocol<br>between APP and the system<br>is specially defined by<br>our-self, and other<br>manufacturers can not use<br>this protocol.The ultrasound<br>device and the APP software<br>transfer images and control<br>instructions via WIFI. | The Clarius Ultrasound<br>Scanner is a portable,<br>general-purpose,<br>software-controlled,<br>diagnostic ultrasound system<br>used to acquire and display<br>high-resolution, real-time<br>ultrasound data through an<br>off-the-shelf (OTS) iOS or<br>Android device. The Clarius<br>Ultrasound Scanner<br>comprises a series of wireless<br>transducers employing<br>Bluetooth and Wi-Fi-based<br>technology to communicate<br>with traditional tablet/smart<br>phone devices via direct<br>Wi-Fi. This allows the user to<br>export ultrasound images and<br>display them across a range<br>of portable personal devices. |
| Image<br>Unit | Display Windows environment (13.3<br>inches);<br>Android environment (9.6<br>inches) | Resolution (in pixels) of<br>960x640 (or 640x960) |
| Probe<br>Characteristics | -Convex Array<br>-Linear Array<br>-Phased Array<br>-Intracavity | -Convex Array<br>-Linear Array<br>-Phased Array<br>-Intracavity |
| Probe | Wireless | Wireless |
| Criteria for<br>comparison | Subject device | Predicate device (K192107) |
| Connection to<br>Display | | |
| Off-the-shelf<br>operating<br>system | Windows /Android | iOS/ Android |
| Software | Run as an app on<br>off-the-shelf mobile device | Run as an app on off-the-shelf<br>mobile device |
| Safety&Effectiveness | | |
| Patient<br>Contacting<br>Materials | Per ISO 10993-5 and ISO<br>10993-10 | Per ISO 10993-5 and ISO<br>10993-10 |
| Electrical Safety | Evaluated according to<br>IEC60601-1 | Evaluated according to<br>IEC60601-1 |
| EMC | Evaluated according to<br>IEC60601-1-2 | Evaluated according to<br>IEC60601-1-2 |
| Performance<br>Safety | Evaluated according to<br>IEC60601-2-37 | Evaluated according to<br>IEC60601-2-37 |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
GuangDong Youkey Medical Co., Ltd. Pocket Ultrasound System
#### Non-clinical tests
The pocket ultrasound system was subjected to extensive safety, performance, and verification testing. Extensive benchtop tests verified the performance of the device to meet the design requirements in an ideal non-clinical environment. All tests were performed to ensure that the device complies with industry and safety standards.
- Electrical safety testing per IEC 60601-1
- EMC testing per IEC 60601-1-2
- Biocompatibility testing per ISO 10993-5 and ISO 10993-10
- Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment per IEC 60601-2-37
- Acoustic Output Test per FDA Guide for Track 3
- Software verification per FDA Software Guidance
{9}------------------------------------------------
GuangDong Youkey Medical Co., Ltd. Pocket Ultrasound System
- Clinical accuracy performance per Guidance for Industry and FDA staff: Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers.
- Wireless performance and coexistence testing
#### Clinical tests
The subject of this premarket submission, Pocket ultrasound system did not require clinical studies to support substantial equivalence.
#### Conclusions
The non-clinical testing data support the safety of the device and the performance testing report demonstrate the subject device Pocket Ultrasound System perform as intended and is substantially equivalent to its predicate device with respect to safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.