Rotarex Atherectomy System

K211738 · C.R. Bard, Inc. · MCW · Sep 14, 2021 · Cardiovascular

Device Facts

Record IDK211738
Device NameRotarex Atherectomy System
ApplicantC.R. Bard, Inc.
Product CodeMCW · Cardiovascular
Decision DateSep 14, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4875
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K211738 · Sep 14, 2021Rotarex Atherectomy SystemC.R. Bard, Inc.Medical literature (clinical studies/cohorts)The sponsor conducted a literature review of real-world clinical studies to evaluate the safety and efficacy of the Rotarex™ Atherectomy System in treating in-stent restenosis (ISR), iliac artery occlusions, and bypass graft debulking compared to other standard-of-care treatments (PTA, thrombectomy devices, and open surgery).Literature review; In-stent restenosis; Peripheral artery disease; Retrospective clinical outcomes

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Literature review of Rotarex™ Atherectomy System clinical outcomes; Systematic literature review of clinical studiesPatients with in-stent restenosis, iliac artery occlusions, and bypass graft occlusionsPercutaneous transluminal angioplasty (PTA), Ekos, Indigo, and open surgerySafety, technical success, and clinical outcome

Indications for Use

The Rotarex™ Atherectomy System is intended for use as an atherectomy device and to break up and remove thrombus from native peripheral arteries or peripheral arteries fitted with stents, stent grafts or natificial bypasses.

Device Story

Rotarex™ Atherectomy System is a percutaneous, over-the-wire device for removing atheromatic plaque and thrombi. System includes a single-use catheter, guidewire, collecting bag, and a drive system (control unit, motor, foot switch). Catheter features a rotating head with abrading facets and an internal helix; driven by a motor at 40,000–60,000 RPM. Mechanical rotation creates aspiration, drawing material into the catheter head for removal. Used in clinical settings by physicians. Output is the physical removal of occlusive material from peripheral arteries, improving vessel patency. Benefits include debulking of occlusions in native arteries, stents, stent grafts, and bypasses, potentially reducing need for open surgery.

Clinical Evidence

No new clinical trials were conducted. Substantial equivalence supported by a review of existing medical literature regarding the Rotarex™ Atherectomy System's use in treating in-stent restenosis (ISR), iliac artery occlusions, and bypass graft debulking. Literature compared outcomes against PTA, other thrombectomy devices (Ekos, Indigo), and open surgery. Results supported the benefit-risk profile for the expanded indications.

Technological Characteristics

Catheter components: flexible outer covering, rotating head (two metal cylinders with side openings), rotating helix. Materials: latex and phthalate-free. Guidewire: Nitinol core, PTFE coating, hydrophilic distal coating. Energy: electromechanical (motor-driven). Sterilization: ethylene oxide. Connectivity: standalone drive system. Sizes: 6F and 8F diameters; 85–135 cm lengths.

Indications for Use

Indicated for patients requiring atherectomy or thrombectomy in native peripheral arteries or peripheral arteries fitted with stents, stent grafts, or native/artificial bypasses. Contraindicated in patients unsuitable for atherectomy/thrombectomy, or for use in cardiopulmonary, coronary, carotid, cerebral, renal, or venous vasculature, or in undersized vessels.

Regulatory Classification

Identification

An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 14, 2021 C.R. Bard, Inc. Aaron Conovaloff Regulatory Affairs Manager 1625 West 3rd St Tempe, Arizona 85281 Re: K211738 Trade/Device Name: Rotarex™ Atherectomy System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW, DQX Dated: August 17, 2021 Received: August 18, 2021 Dear Aaron Conovaloff: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K211738 Device Name Rotarex™ Atherectomy System Indications for Use (Describe) The Rotarex™ Atherectomy System is intended for use as an atherectomy device and to break up and remove thrombus from native peripheral arteries or peripheral arteries fitted with stents, stent grafts or natificial bypasses. Contraindications: - · In patients not suitable for atherectomy/thrombectomy - · In the cardiopulmonary, coronary, carotid, cerebral and renal vasculature - · In vessels that are undersized for the particular device used - In the venous vasculature Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### Rotarex™ Atherectomy System #### 510(k) Summary 21 CFR 807.92 As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (1)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows: #### Submitter Information: | Applicant: | Bard Peripheral Vascular, Inc<br>1625 West 3rd Street<br>Tempe, Arizona 85281 | |------------|-------------------------------------------------------------------------------| | Phone: | 602-830-5453 | | Fax: | 321-949-0436 | | Contact: | Aaron Conovaloff, Regulatory Affairs Manager | | Date: | June 4, 2021 | ### Subiect Device Name: | Device Trade Name: | Rotarex™ Atherectomy System | |-----------------------|------------------------------| | Common or Usual Name: | Intraluminal artery stripper | | Product Code: | MCW, DQX | | Classification: | Class II | | Review Panel: | Cardiovascular | | Regulation Number: | 21 CFR 870.4875 | #### Predicate Device: - Straub Endovascular System (K172315; cleared April 12, 2018) . #### Reference Device: - B-Laser Atherectomy System (K181642; cleared October 5, 2018) . #### Device Description: The Rotarex™ Atherectomy System is made up of a single use Rotarex™ Atherectomy Catheter Set and the Drive System, consisting of the control unit, motor and foot switch. For more detail on the Drive System, please see the manual provided with the Drive System. Read and understand all instructions prior to attempting any atherectomy procedure with the Rotarex™ Atherectomy System or any of its components. The Rotarex™ Atherectomy Catheter Set is composed of multiple components, including the Rotarex™ Atherectomy Catheter, guidewire, collecting bag, and sterile drape. Rotarex™ Atherectomy Catheters are over-the-wire, single use, percutaneous devices for the removal of {4}------------------------------------------------ atheromatic plaque and thrombi in native arteries fitted with stents, stent grafts or native or artificial bypasses. The catheters are latex and phthalate free, and consist of a flexible outer covering, a rotating head, and a rotating helix which runs the length of the catheter. A lumen for the passage of the supplied guidewire runs the entire length of the helix and through the head of the catheter. The catheter head is made up of two overlying metal cylinders, with two side openings. The outer cylinder is connected to the rotating helix, and the inner cylinder to the catheter shaft. The helix and the catheter head rotate at approximately 40,000-60,000 rpm depending on the model, by means of a gear box in the catheter housing and a motor contained within the catheter handle driven by the Drive System. The rotating outer cylinder is fitted with abrading facets at its foremost tip. The Rotarex™ Atherectomy System Guidewires are 0.018" and are included with each Rotarex™ Atherectomy Catheter Set. These guidewires consist of a Nitinol core and a PTFE coating with a 9.5cm hydrophilic distal coating and a 4 cm angled flex tip. All components within the Rotarex™ Atherectomy Catheter Sets are supplied sterile for single use only. The method of sterilization is ethylene oxide. | Catheter<br>Size | Minimum<br>Vessel<br>Diameter | Catheter<br>External<br>Diameter | Nominal<br>Rotation<br>(RPM) | Maximum<br>Aspiration<br>(ml/min) | Guidewire<br>Size &<br>Length | |------------------|-------------------------------|----------------------------------|------------------------------|-----------------------------------|-------------------------------| | 6F<br>110 cm | 3 mm | 2.0 mm | 60,000 | 45 | 0.018"<br>270 cm | | 6F<br>135 cm | 3 mm | 2.0 mm | 60,000 | 45 | 0.018"<br>320 cm | | 8F<br>85 cm | 5 mm | 2.7 mm | 40,000 | 75 | 0.018"<br>220 cm | | 8F<br>110 cm | 5 mm | 2.7 mm | 40,000 | 75 | 0.018"<br>270 cm | # Indications for Use of Device: The Rotarex™ Atherectomy System is intended for use as an atherectomy device and to break up and remove thrombus from native peripheral arteries or peripheral arteries fitted with stents, stent grafts or native or artificial bypasses. # Contraindications: - In patients not suitable for atherectomy/thrombectomy - In the cardiopulmonary, coronary, carotid, cerebral and renal vasculature ● - In vessels that are undersized for the particular device used - . In the venous vasculature # Comparison to Predicate Devices: The subject Rotarex™ Atherectomy System has the following similarities to the predicate Rotarex® S catheter (a component of the Straub Endovascular System: K172315, cleared April 12, 2018): - . Intended use - . Target population/conditions of use (anatomical location of use, how device interacts with {5}------------------------------------------------ other devices, interaction with patient) - Principles of operation/mechanism of action ● - . Fundamental scientific technology - Sterilization method ● - . Sterility assurance level The subject Rotarex™ Atherectomy System has the following similarities to the reference B-Laser Atherectomy System (K181642, cleared October 5, 2018): - Intended use . - . Indications for use - . Target population/conditions of use (anatomical location of use, how device interacts with other devices, interaction with patient) - . Sterilization method The subject Rotarex™ Atherectomy System incorporates the design of the Rotarex® S catheter predicate with the in-stent restenosis indication of the reference B-Laser Atherectomy System device. Note that the design, materials, and methods of manufacture of the subject device and the Rotarex® S catheter predicate are identical; the only differences are the expansion of the indications for use to include use in peripheral arteries fitted with stents or stent grafts and in native or artificial bypasses, and to remove several of the contraindications in the IFU. The removed contraindications are not critical to the intended use of this device. Some of these contraindications are being removed completely, either in light of data to support their removal or because the corresponding risk information is already communicated in other areas of the IFU. Other contraindications are being restated as warnings have been determined to be a more appropriate way to present the corresponding risk information. Because the corresponding risk information is still presented appropriately in the IFU, removal of these contraindications does not affect the safety and effectiveness of the device when used as labeled. ### Performance Data: To demonstrate substantial equivalence of the subject device to the predicate device, its technological characteristics and performance criteria were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Assessment procedures, the following in vitro tests were performed on the subject device: - Simulated use of atherectomy device in a stent . - Heat generation . - . Embolization analysis #### Clinical Literature A significant body of medical literature confirms the frequent usage of the Rotarex™ Atherectomy System for the treatment of ISR as well as for the debulking of occlusions in illac arteries and bypass grafts. Literature was reviewed to examine the safety and efficacy of the Rotarex™ Atherectomy System compared to percutaneous transluminal angioplasty (PTA) when used for ISR treatment. In addition, the safety, technical success, and clinical outcome for iliac arteries was addressed by description of the results from all-comer Rotarex™ Atherectomy System studies. The debulking of bypass grafts using the Rotarex™ Atherectomy System was compared with thrombectorny devices, Ekos (Boston Scientific) and Indigo (Penumbra), and with open surgery. The results of the literature review were used to determine the acceptability of the benefit-risk profile of Rotarex™ Atherectorny Systems used forthe treatment of ISR and debulking of occlusions in iliac-arteries and bypass-grafts. {6}------------------------------------------------ ### Conclusions: The results from both in vitro studies and clinical literature demonstrate that the technological characteristics and performance criteria of the subject Rotarex™ Atherectomy System are substantially equivalent to the predicate device, and that it can perform in a manner equivalent to devices currently on the market for the same intended use. The subject device, the Rotarex™ Atherectomy System, met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. Rotarex™ Atherectomy System is substantially equivalent to the legally marketed predicate device, the Rotarex® S catheter.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...