← Product Code [KWQ](/productcode/KWQ) · K211718

# Venus Cervical Plate System (K211718)

_Artfx Medical, LLC · KWQ · Jun 28, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K211718

## Device Facts

- **Applicant:** Artfx Medical, LLC
- **Product Code:** [KWQ](/productcode/KWQ.md)
- **Decision Date:** Jun 28, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Venus Cervical Plate System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications: -Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) -Trauma (including fractures) -Tumors -Deformities or curvatures (including kyphosis, lordosis, or scoliosis) -Pseudarthrosis -Failed previous fusion -Decompression of the spinal cord following total or partial cervical vertebrectomy -Spondylolisthesis -Spinal stenosis.

## Device Story

Venus Cervical Plate System provides temporary anterior spinal stabilization during fusion; indicated for cervical levels C2-T1. System comprises titanium alloy (Ti6Al4V-ELI) plates and fixed-angle, self-tapping bone screws. Plates feature integrated blocking mechanisms to prevent screw back-out. Used by surgeons in clinical settings to fixate vertebrae; implants remain static post-surgery. Benefits include structural support during bone healing. Device is non-sterile, single-use.

## Clinical Evidence

Bench testing only. No clinical data provided. Mechanical performance verified via ASTM F543-17 (pull-out, driving torque, torsion, self-tapping) and ASTM F1717-18 (fatigue, static compression, static tensile, static torsion) testing.

## Technological Characteristics

Material: Titanium alloy (Ti6Al4V-ELI) per ASTM F136. Components: Plates (17-90mm lengths) and bone screws (3.5-4.5mm diameters, 12-22mm lengths). Sensing/Actuation: Mechanical fixation via fixed-angle screws and anti-back-out blocking mechanism. Energy: None. Connectivity: None. Sterilization: Non-sterile (user-sterilized).

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Stryker Spine NuLock Anterior Cervical Plating System ([K083562](/device/K083562.md))

## Reference Devices

- Mikron Spinal Fixation System ([K171497](/device/K171497.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 28, 2022

ArtFx Medical LLC Ozgen Ozfidan CEO 50 Laura Street North, 25th Floor Jacksonville, Florida 32202

Re: K211718

Trade/Device Name: Venus Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWO Dated: June 10, 2022 Received: June 14, 2022

Dear Ozgen Ozfidan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K211718

Device Name VENUS CERVICAL PLATE SYSTEM

The Venus Cervical Plate System is intervertebral screw fixation of the cervical spine at levels C2-T 1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:

-Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)

-Trauma (including fractures)

-Tumors

-Deformities or curvatures (including kyphosis, lordosis, or scoliosis)

-Pseudarthrosis

-Failed previous fusion

-Decompression of the spinal cord following total or partial cervical vertebrectomy

-Spondylolisthesis

-Spinal stenosis.

Type of Use (Select one or both, as applicable)

| X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

Contact Person/Applicant: NAME: OZGEN OZFIDAN COMPANY NAME: ARTFX MEDICAL LLC ADDRESS: 50 LAURA ST N 25™ FLOOR, JAKCSONVILLE, FL, 32202 USA PHONE: +1 917 445 2085 FAX: -EMAIL: ozgen@artfxmed.com ESTABLISHMENT REGISTRATION: 3017435639 DATE : 10.06.2022 DEVICE IDENTIFICATION: TRADE NAME: VENUS CERVICAL PLATE SYSTEM DEVICE COMMON NAME:ANTERIOR CERVICAL PALTE CLASSIFICATION NAME: APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY REGULATION NUMBER: 888.3060 REVIEW PANEL: ORTHOPEDIC PRODUCT CODE: KWQ REASON FOR SUBMISSION: NEW DEVICE FDA GUIDANCE DOCUMENTS:

## INDICATIONS FOR USE:

The Venus Cervical Plate System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:

-Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)

-Trauma (including fractures)

{4}------------------------------------------------

## -Tumors

- -Deformities or curvatures (including kyphosis, lordosis, or scoliosis)
- -Pseudarthrosis
- -Failed previous fusion
- -Decompression of the spinal cord following total or partial cervical vertebrectomy
- -Spondylolisthesis
- -Spinal stenosis.

## DEVICE DESCRIPTION:

The Venus Cervical Plate System is supplied non-sterile, single use and are fabricated from titanium alloy (Ti6Al4V-ELI) that conforms to ASTM F136. Various sizes of these components are available. The Venus Cervical Plate System configurations ranging in lengths from 17 mm to the

one-level plates, 27 mm to 40 mm for the two-level plates, 45 mm for the three-level plates, , and 70 mm to 90 mm for the four -level plates. All of the plate levels incorporate blocking mechanism to aid in prevention of bone screw back-out. The bone screws are provided with fixed angles available in self- tapping design. The fixed bone screws are inserted at a defined angle. The bone screws are offered in 3.5 mm, 4.0 mm and 4.50 mm diameters in lengths of 12 mm – 22 mm. The implants (bone screws and cervical plates) are provided as single-use, non-sterile devices manufactured from implantable grade titanium alloy (TI6Al4V-ELI).

The Venus Cervical Plate System is intended for use as an aid in cervical spinal fusion .

The Venus Cervical Plate System is supplied non-sterile, single use and are fabricated from titanium alloy (Ti6Al4V-ELI) that conforms to ASTM F136. Various sizes of these components are available.

## NON-CLINICAL TESTING:

Non-clinical testing including Pull-out Test (ASTM F543-17), Driving Torque Test (ASTM F543-17) Torsion Test (ASTM F543-17), Self-Tapping Test (ASTM F543-17), Fatigue Test (ASTM F1717-18) Static Compression Test (ASTM F1717-18), Static Tensile Test (ASTM F1717-18), and Static Torsion Test (ASTM F1717-18) were conducted. The results showed that the performance of the proposed devices is substantially equivalent.

{5}------------------------------------------------

## PREDICATE DEVICE:

TRADE/ DEVICE NAME: STRYKER SPINE NULOCK ANTERIOR CERVICAL PLATING SYSTEM REGULATION NUMBER: 21 CFR 888.3060 REGULATION NAME: APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY REGULATORY CLASS: Class II PRODUCT CODE: KWQ 510(K) NUMBER: K083562

## Technological Characteristics, Comparison to Predicate Device

The Venus Cervical Plate System has similar technological characteristics as the predicate devices, including the materials, design, function, range of sizes, manufacturing processes, surgical techniques, and intended use. The minor differences in design and sizing options do not present new issues of safety and effectiveness.

REFERENCE DEVICE: TRADE/ DEVICE NAME: Mikron Spinal Fixation System REGULATION NUMBER: 21 CFR 888.3070 REGULATION NAME: THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM REGULATORY CLASS: Class II PRODUCT CODE: NKB 510(K) NUMBER: K171497

The 510(K) number of the cleared device is K171497 for "Polyaxial screws and rods" under Mikron Makine ,it is refence device for biological evaluation of this submission i

The chemical characterization for raw material and end product of "MSFX – MİKRON SPINAL FIXATION POLYAXIAL SCREW " as per ISO 10993-18 in biological evaluation report in Annex IX.

Also we mentioned these steps in "ANNEX X Manufacturing Flows for VENUS CERVICAL PLATE SYSTEM " and "Annex XIII - Manufacturing Flow for Screw".

{6}------------------------------------------------

## CONCLUSION:

The intended use and material oh the subject anterior cervical plate are identical to this of the predicate anterior cervical plates.The predicate STRYKER SPINE NULOCK ANTERIOR CERVICAL PLATING SYSTEM consists of various size plates are implanted and remain static.The subject system introduces various length plates that has antiback out mechanism as predicate device.

Non-clinical testing including Pull-out Test,Driving Torque Test ,Torsion Test ,Self-Tapping Test , Fatigue Test , Static Compression Test , Static Tensile Test and Static Torsion Test were conducted. The results showed that the performance of the proposed devices is substantially equivalent.

---

**Source:** [https://fda.innolitics.com/device/K211718](https://fda.innolitics.com/device/K211718)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
