Disposable Medical Synthetic Nitrile Examination Gloves (Non-Sterile)
Applicant
Changzhou Universal Medical Equipment Co., Ltd.
Product Code
LAZ · Clinical Toxicology
Decision Date
Sep 23, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3280
Device Class
Class 1
Indications for Use
Disposable Medical Synthetic Nitrile Examination Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Device Story
Disposable synthetic nitrile and vinyl compound examination gloves; non-sterile; blue color; powder-free. Worn by healthcare personnel or patients to provide a barrier against contamination between the examiner and the patient's body, fluids, waste, or environment. Single-use device. Provides physical protection during medical examinations.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ASTM D6319-19 for dimensions, physical properties (tensile strength/elongation), freedom from holes (AQL 2.5), and residual powder. Biocompatibility testing performed per ISO 10993-10 (skin irritation/sensitization) and ISO 10993-5 (in vitro cytotoxicity).
Technological Characteristics
Material: Nitrile and vinyl compound. Form factor: Disposable, non-sterile, blue, powder-free gloves. Sizes: Small, Medium, Large. Standards: ASTM D6319-19 (physical dimensions, tensile strength, elongation, residual powder, freedom from holes). Biocompatibility: ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for use as a disposable medical device worn on the examiner's hand to prevent contamination between patient and examiner. Suitable for over-the-counter use.
Regulatory Classification
Identification
A clinical toxicology control material is a device intended to provide an estimation of the precision of a device test system and to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects. This generic type of device includes various single, and multi-analyte control materials.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 23, 2021
Changzhou Universal Medical Equipment Co. Ltd Johnson Liu Consultant CNMED Consultant 31 Archer St Upper MT Gravatt, QLD 4122 Australia
Re: K211667
Trade/Device Name: Disposable Medical Synthetic Nitrile Examination Gloves (Non-Sterile) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: August 16, 2021 Received: August 24, 2021
Dear Johnson Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K211667
Device Name
Disposable Medical Synthetic Nitrile Examination Gloves (Non-Sterile)
Indications for Use (Describe)
Disposable Medical Synthetic Nitrile Examination Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo with a circular teal border. Inside the border are five white stars evenly spaced around the top half of the circle. In the center of the circle are two Chinese characters stacked vertically. Below the characters is the text "Lian Shou" in a sans-serif font, also in teal. In the upper right corner of the image is the registered trademark symbol.
# 510(K) Summary K211667
#### DISPOSABLE MEDICAL SYNTHETIC NITRILE EXAMINATION GLOVES, NON-STERILE
Preparation Date: Aug 16th, 2021
#### 1. SUBMITTER
Company Name: Changzhou Universal Medical Equipment Co.,Ltd Company Address: No.6,Xinxi Road, Xinbei District Changzhou, CN 213000 Contact Person: Johnson Liu Telephone Number: +614-0158-9995 Email: Johnson@cnmed.com.au
#### 2. NAME OF THE DEVICE
Trade Name / Proprietary Name: Disposable Synthetic Nitrile Examination Gloves (Non Sterile) Device Name: Disposable Medical Synthetic Nitrile Examination Gloves (Non-Sterile) Device Classification Name: Patient Examination Gloves Device Class: Class I Device Classification Number: 21 CFR 880.6250 Product Code: LZA
# 3. IDENTIFICATION OF THE LEGALLY MARKETED PREDICATE DEVICE
Primary Predicate Device: K102593 Applicant: Zibo Yingbo Medical Products Co., Ltd Device Name: Synthetic Nitrile Patient Examination Gloves - Powder free, Blue Colour Device Classification Name: Patient Examination Gloves Device Classification Number: 21 CFR 880.6250 Device Class: Class I Product Code: LZA Review Panel: General Hospital
### 4. DEVICE DESCRIPTION
The subject device in this 510(k) Notification is Disposable Medical Synthetic Nitrile Examination Gloves. Non-sterile. The subject device is a patient examination glove made from nittile and vinyl compound, blue color, powder free and non-sterile (Per 21 CFR 880.6250, Class I). The device meets the specifications in ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.
# 5. INDICATIONS FOR USE OF THE DEVICE
Disposable Medical Synthetic Nitrile Examination Gloves, Non-sterile is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
{4}------------------------------------------------
# 6. TECHNOLOGICAL CHARACTERISTIC COMPARISON FOR THE PROPOSED AND PREDICATE DEVICES
| CHARACTERISTICS | DEVICE PERFORMANCE | | PRIMARY PREDICATE | Result |
|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| | SUBJECT | | | |
| 510(k) Number | K211667 | | K102593 | - |
| Device Name | Disposable Medical Synthetic Nitrile Examination<br>Gloves Powder Free, Non sterile | | Synthetic Nitrile Patient Examination Gloves -<br>Powder free, Blue Color | - |
| Product Code | LZA | | LZA | Same |
| Indications for Use | Disposable Nitrile Examination Gloves Non-sterile<br>(Non sterile) is disposable devices intended for<br>medical purpose that are worn on the examiner's hand<br>to prevent contamination between patient and<br>examiner. The device is for over the-counter use. | | A patient examination glove is a disposable<br>device intended for medical purposes that is<br>worn on the hand of healthcare and similar<br>personnel to prevent contamination between<br>healthcare personal and the patient's body,<br>fluids, waste or environment | Same |
| Materials of Use | Nitrile and Vinyl compound | | Synthetic Nitrile | Similar |
| Color | Blue | | Blue | Same |
| Size<br>(ASTM D6319-19) | Small, Medium, Large, | | Small, Medium, Large,<br>Extra Large | Same |
| Sterilization | Non-sterile | | Non-sterile | Same |
| Usage | Single usage | | Single usage | Same |
| Dimensions | Width:<br>80mm ±10mm (for small size)<br>95mm ±10mm (for medium size)<br>110 mm ±10mm (for large size)<br>Length:<br>220mm (Minimum) (for small size)<br>230 mm (Minimum)(for medium, large size) | | Comply with ASTM D6319-05 | Same |
| Physical Properties | Before Aging<br>Tensile Strength<br>Min 14 Mpa<br>Ultimate Elongation Min 500%<br>After Aging<br>Tensile Strength<br>Min 14 Mpa<br>Ultimate Elongation Min 400%<br>(ASTM D6319-19) | | Before Aging<br>Tensile Strength<br>Min 14 Mpa<br>Ultimate Elongation Min 500%<br>After Aging<br>Tensile Strength<br>Min 14 Mpa<br>Ultimate Elongation Min 400%<br>(ASTM D6319-05) | Same |
| Thickness | Palm min 0.05 mm<br>Finger min 0.05 mm<br>(ASTM D6319-19) | Palm min 0.05 mm<br>Finger min 0.05 mm<br>(ASTM D6319-05) | Same | |
| Powder Free | <2 mg/glove (ASTM D6319-19) | <2 mg/glove (ASTM D6319-05) | Same | |
| Freedom from Holes (Water<br>Tight -1000 ml) AQL2.5 | Passed<br>ASTM D6319-19 (Cross Reference D5151 | Passed<br>ASTM D6319-05 (Cross Reference D5151) | Same | |
| Biocompatibility -SKIN<br>SENSITIZATION - ISO<br>10993-10: 2010 (E) | Under the conditions of the study not a sensitizer | Under the conditions of the study not a sensitizer | Same | |
| Biocompatibility -SKIN<br>IRRITATION - ISO10993-<br>10: 2010 (E) | Under the conditions of study not an irritant | Under the conditions of study not an irritant | Same | |
| Biocompatibility - <i>IN VITRO</i><br>CYTOTOXICITY – ISO<br>10993-5: 2009(E) | Under the conditions of the cytotoxicity study, testing<br>results showed mild systemic toxicity and cytotoxic<br>level was 2. | No Testing Result available | Different | |
| Manufacturer(s) | Changzhou Universal Medical Equipment Co.,Ltd | Zibo Yingbo Medical Products Co., Ltd | -- | |
{5}------------------------------------------------
To sum up, the subject device and predicate devices are equivalent in terms of intended use, design, safety and performance.
{6}------------------------------------------------
# 7. NON-CLINICAL TESTING SUMMARY
## PERFORMANCE DATA
| Test Method | Purpose | Acceptance Criteria | Result |
|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| ASMT D6319-19<br>Standard Specification<br>for Nitrile Examination<br>Gloves for Medical<br>Application - Physical<br>Dimensions Test | To determine the<br>width, length, and<br>thickness of the<br>gloves | Width:<br>80 mm (±10mm) (for small size)<br>95 mm (±10mm) (for medium size)<br>110 mm (±10mm) (for large size)<br>Length:<br>220 mm (Minimum)(for small size)<br>230 mm (Minimum)(for medium, large<br>size)<br>Thickness: (for all sizes)<br>Finger -0.05 mm (Minimum)<br>Palm -0.05mm (Minimum) | Passed |
| ASMT D6319-19<br>Standard Specification<br>for Nitrile Examination<br>Gloves for Medical<br>Application - Physical<br>Requirements Test | To determine the<br>tensile strength and<br>ultimate elongation<br>before and after<br>acceleration aging | Before Acceleration Aging:<br>Tensile Strength (MPa): 14 (Minimum)<br>Ultimate Elongation (%): 500 (Minimum)<br>After Acceleration Aging:<br>Tensile Strength (MPa): 14 (Minimum)<br>Ultimate Elongation (%): 400(Minimum) | Passed |
| ASTM D6319-19<br>(ASTM D5151-11)<br>Standard Test Method<br>for Detection of Holes in<br>Medical Gloves | To determine the<br>holes in the gloves | AQL 2.5 | Passed |
| ASMT D6319-19<br>(ASTM D6124-11)<br>Standard Test Method<br>for Residual Powder on<br>Medical Gloves | To determine the<br>residual powder in<br>the gloves | < 2.0 mg/glove | 0.6 mg/glove, Passed |
{7}------------------------------------------------
#### BIO-COMPATIBILITY DATA
| Test Method | Purpose | Acceptance<br>Criteria | Result |
|----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-10 Biological evaluation<br>of medical devices — Part 10: Tests<br>for skin irritation and skin<br>sensitization | To determine the potential of the<br>material under test to produce skin<br>irritation in rabbits | Under the condition of<br>study not an irritant | Under the<br>condition of study<br>not an irritant |
| ISO 10993-10 Biological evaluation<br>of medical devices — Part 10: Tests<br>for skin irritation and skin<br>sensitization | To determine the skin sensitization<br>potential of the material both in<br>terms of induction and elicitation in<br>guinea pigs. | Under the conditions of<br>the study not a sensitizer. | Under the condition<br>of the study not a<br>sensitizer. |
| ISO 10993-5 Biological evaluation of<br>medical devices — Part 5: Tests for in<br>vitro cytotoxicity | To evaluate the in vitro cytotoxic<br>potential of the test item (both inner<br>and outer surface) Extracts in L-929<br>mouse fibroblasts cells using elution<br>method | No more than grade 2<br>cytotoxicity at 100%<br>extract concentration | Under the conditions<br>of the cytotoxicity<br>study, the extract of<br>the test article<br>showed mild<br>cytotoxicity to L-929<br>cells and the<br>cytotoxic level was<br>2. |
## 8. CLINICAL TESTING SUMMARY
Not applicable - Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
### 9. CONCLUSION
The conclusions drawn from the non-clinical test demonstrate that the subject device in 510(k) submission, the Disposable Medical Synthetic Nitrile Examination Gloves Non sterile is as safe, as effective, and performs as well as or better than the legally marketed predicate device K102593- Synthetic Nitrile Patient Examination Gloves – Powder free, Blue.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.