K211652 · Apyx Medical Corporation (Formerly Bovie Medical Corporation · GEI · May 25, 2022 · General, Plastic Surgery
Device Facts
Record ID
K211652
Device Name
Renuvion Dermal Handpiece, Renuvion Dermal System
Applicant
Apyx Medical Corporation (Formerly Bovie Medical Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
May 25, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K211652 · May 25, 2022
Renuvion Dermal Handpiece, Renuvion Dermal System
Apyx Medical Corporation (Formerly Bovie Medical Corporation
Published clinical literature (Kilmer et al., 2007)
The sponsor used published clinical literature to establish a comparative safety profile for the predicate device (NeoGen PSR System) to support the substantial equivalence of the subject device.
Adverse event comparison; Literature review; Predicate device safety profile
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Kilmer et al., 2007 (Lasers Med Sci); Pilot study
Patients undergoing full facial rejuvenation procedures; Sample Size: 24
Not applicable for this study
Adverse events
Indications for Use
The Renuvion® Dermal System is an electrosurgical device for dermatological procedures for the treatment of moderate to severe wrinkles and rhytides, limited to patients with Fitzpatrick skin types I, II or III. The treatment is achieved through controlled heating of the outer layers of the skin so that part or all of the epidermis becomes non-viable and there is controlled thermal modification to the underlying dermis. The Renuvion® Dermal Handpiece is only compatible with Apyx Medical Electrosurgical Generators.
Device Story
Renuvion Dermal Handpiece is a sterile, single-use monopolar electrosurgical device used with Apyx Helium Plasma Generators. It delivers radiofrequency energy and helium plasma to skin tissue. Operated by trained medical staff in clinics or surgery centers, the device uses a tip to deliver a precise helium plasma beam, requiring a patient grounding pad to complete the circuit. Activation occurs via a handpiece button or foot pedal. The device heats the epidermis to render it non-viable and provides controlled thermal modification to the underlying dermis. This process aims to reduce facial wrinkles and rhytides. Clinical benefits include improved skin appearance with a safety profile comparable to existing nitrogen plasma devices.
Clinical Evidence
Prospective, multicenter, single-arm clinical study (G170151) of 55 subjects. Primary endpoint: proportion of subjects with at least 1-point improvement on Fitzpatrick Wrinkle and Elastosis Scale (FWS) at 90 days. Results: 92.7% achieved at least 1-point improvement. No serious adverse events related to device/procedure. Common AEs: edema, erythema, post-inflammatory hyperpigmentation. Performance supported by bench testing and two porcine animal studies comparing thermal damage zones.
Technological Characteristics
Monopolar electrosurgical handpiece; helium plasma and radiofrequency energy source; 4.0kVpeak rating; single-use sterile (EO); compatible with Apyx generators (APYX-200P, APYX-RS3). Complies with ANSI/AAMI/IEC 60601-1, 60601-2-2, and 62366. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for patients with Fitzpatrick skin types I, II, or III seeking treatment for moderate to severe facial wrinkles and rhytides.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 25, 2022
Apyx Medical Corporation (formerly Bovie Medical Corporation Priscilla Herpai Global Regulatory Director 5115 Ulmerton Road Clearwater, Florida 33760-4004
Re: K211652
Trade/Device Name: Renuvion Dermal Handpiece, Renuvion Dermal System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: May 24, 2022 Received: May 25, 2022
Dear Priscilla Herpai:
We have reviewed your Section 510/k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A ) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K211652
Device Name
Renuvion® Dermal Handpiece (APYX-044-DERM)
Indications for Use (Describe)
The Renuvion® Dermal System is an electrosurgical device for dermatological procedures for the treatment of moderate to severe wrinkles and rhytides, limited to patients with Fitzpatrick skin types I, II or III. The treatment is achieved through controlled heating of the outer layers of the skin so that part or all of the epidermis becomes non-viable and there is controlled thermal modification to the underlying dermis.
The Renuvion® Dermal Handpiece is only compatible with Apyx Medical Electrosurgical Generators.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the Apyx Medical logo. The logo is orange and blue. The word "Apyx" is in orange, and the word "Medical" is in blue.
### 510(k) Summary
### K211652
| 1. | General Information | |
|----|-------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| | Submitted by: | Apyx Medical Corporation<br>5115 Ulmerton Road<br>Clearwater, Florida 33760 -4004<br>United States of America |
| | Establishment Registration #: | 3007593903 |
| | Submitter FAX Number: | (727) 322-4465 |
| | Contact Person: | Mrs. Priscilla Herpai, Global Regulatory Director<br>Phone: (727) 803-8512<br>Email: priscilla.herpai@apyxmedical.com |
| | Date Prepared: | May 27, 2021 |
| | Trade Names (Model Numbers): | Renuvion® Dermal Handpiece<br>(APYX-044-DERM) |
| | Common Name: | Electrosurgical Handpiece |
| | Classification: | Class II per 21CFR 878.4400 - Electrosurgical<br>Cutting and Coagulation Device and Accessories<br>Product Code GEI |
| | Predicate Devices: | Predicate Device<br>NeoGen PSR System, Energist Ltd (K132754) -<br>(formerly Portrait PSR System, Rhytec Inc.) |
### 2. Indications for Use
The Renuvion® Dermal System is an electrosurgical system for dermatological procedures for the treatment of moderate to severe wrinkles and rhytides, limited to patients with Fitzpatrick Skin Types I, II or III. The treatment is achieved through controlled heating of the outer layers of the skin so that part or all of the epidermis becomes non-viable and there is controlled thermal modification to the underlying dermis. The Renuvion® Dermal Handpiece is only compatible with Apyx Medical Electrosurgical Generators.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Aryx Medical. The logo is orange and blue. The word "Aryx" is in orange, and the word "Medical" is in blue.
### K211652
### 3. Device Description
The Renuvion® Dermal Handpiece is a sterile single use electrosurgical (monopolar) device intended to be used in conjunction with a compatible Apyx Helium Plasma Generator (APYX-200P, APYX-RS3 - cleared under 510(k) K170188 & K192867) for the delivery of radiofrequency energy and/or helium plasma for the delivery of helium plasma for the noninvasive treatment of facial wrinkles and rhytides, in patients with Fitzpatrick skin types I, II or III. The Renuvion® Dermal Handpiece is pictured below.
The major components of the Renuvion® Dermal Handpiece are pictured in Figure 1.
#### Figure 1. Renuvion Dermal Handpiece
Image /page/4/Figure/9 description: The image shows the Renuvion Dermal Handpiece and its components. The handpiece has a tip, shaft, activation button, and handle. There are also standard and periorbital spacers, a plug connector, and a cable.
The Renuvion Dermal handpiece is part of the Renuvion System composed of the compatible Apyx generator, Foot switch, helium source, and patient grounding pad. The components of the system are pictured in Figure 2.
The handpiece is activated by pressing the activation button or by pressing the foot pedal on the footswitch accessory
The tip of the Renuvion® Dermal Handpiece is used to deliver thermal energy to tissue through a precise helium plasma beam. The device connects to the electrosurgical generator which is also connected to a helium gas tank and regulator. The device requires the use of a grounding pad plugged into the generator to complete the circuit. The device is offered only in one size, 44mm, catalog number APYX-044-DERM. The handpiece is activated by pressing the activation button on the handpiece or by pressing the purple foot pedal on the foot control accessory. The device is rated at 4.0kVpeak.
Figure 2. Renuvion Dermal System.
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image is a logo for Apyx Medical. The logo features the word "Apyx" in orange with a small, dark blue triangle in place of the "A". To the right of the word "Apyx" is the trademark symbol. Below the word "Apyx" is the word "MEDICAL" in dark blue. There is a thin, dark blue line below the word "MEDICAL".
510(k) Summary
K211652
Image /page/5/Figure/5 description: The image shows a medical device setup with several components labeled. A red helium tank is connected to a gas regulator via a blue tube. The gas regulator is connected to a generator, which also has a handpiece and cable attached. A foot switch and grounding pad are also part of the setup.
### 4. Performance Data
Performance testing was conducted in accordance with FDA's Guidance for Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery originally issued on August 15, 2016 and updated on March 9, 2020
#### a. Bench Testing
Performance testing was conducted to verify that the Renuvion® Dermal Handpiece and associated accessory (Renuvion® Dermal Spacer) met performance specifications. All acceptance criteria were met, and the device met all its performance and product requirements. The testing performed is summarized in the following table:
| Test | Objective | Result |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| Mechanical<br>Verification &<br>Functionality | Verify Mechanical functionality of the Renuvion®<br>Dermal Handpiece and associated accessories,<br>Renuvion® Dermal Spacers. | Mechanical<br>functionality<br>requirements met |
| System<br>Compatibility<br>Testing | Verify the device, when connected to the<br>generators (with all the components and<br>accessories working together as a system), are<br>working as intended and are compatible with all<br>system components. | System<br>compatibility<br>requirements met |
| Thermal Tissue<br>Effect | Compare the tissue thermal effects of the subject<br>device (Renuvion® Dermal Handpiece) with the<br>reference device (Renuvion®/J-Plasma Precise<br>Open Handpiece) using 4 different tissue types. | Thermal tissue<br>effect<br>requirements met |
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for Apyx Medical. The logo features the word "Apyx" in orange, with a blue triangle in place of the letter "A". Below the word "Apyx" is the word "MEDICAL" in blue. The letters are sans-serif and the logo has a clean, modern look.
### K211652
| Test | Objective | Result |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Inspectional<br>Verification | Document parameters that can be verified<br>through inspection for the Renuvion® Dermal<br>Handpiece. | Inspectional<br>requirements met |
| Packaging<br>Validation | Evaluate simulated (and/or real-time) aging and<br>associated package integrity and shelf-life claims.<br>Assess the potential for damage to the device<br>(e.g., drop tests of the instrument in its<br>packaging). | Packaging<br>requirements met |
#### b. Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing was conducted on the system (subject device, Renuvion® Dermal Handpiece (HP), and the compatible Electrosurgical Generator, APYX-200P and APYX-RS3) to ensure compliance with the following recognized standards and all results were passing:
ANSI/AAMI/IEC ES60601-1:2005/(R)2012 and A1:2012, Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance
ANSI/AAMI/IEC 60601-2-2:2017 Medical electrical equipment - Part 2-2: Requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surqical accessories
IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and Tests.
ANSI/AAMI/IEC 62366:2015 Medical devices - Part 1: Application of usability engineering to medical devices
#### c. Biocompatibility Testing
The biocompatibility evaluation for the Renuvion® Dermal Handpiece was conducted in accordance with the June 16, 2016, FDA Guidance, "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and the AAMI/ANSI/ISO 10993-1:2018 and AAMI/ANSI/ISO 10993-1:2009 standards, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process". The Renuvion® Dermal Handpiece (subject device) is considered a surface contacting device with contact to breached or compromised surfaces. Therefore, the following tests were conducted and passed in accordance with the respective ISO 10993 series recognized consensus standard:
- Cytotoxicity ■
- Systemic toxicity
- Material Mediated Pyrogen/Pyrogenicity Sensitization ■
- 트 Irritation
■
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image shows the logo for Apyx Medical. The logo is orange and blue, with the word "Apyx" in orange and a blue triangle in the letter "A". Below the word "Apyx" is the word "MEDICAL" in blue. The logo is simple and modern.
### K211652
### 5. Pre-Clinical & Animal Studies
Two pre-clinical animal studies were conducted to establish the equivalence between the subject device and the predicate device, NeoGen® PSR System by Energist (formerly the Portrait PSR System by Rhytec). These pre-clinical studies were designed to provide a measurement (using histology under magnification) of the thermal damage zone (i.e., coagulation necrosis) on live porcine skin tissue at their minimum and maximum recommended power settings for dermatological wrinkle reduction procedures for moderate to severe wrinkles and rhytides.
These two (2) pre-clinical studies demonstrated that the depth of thermal effect for the Renuvion® Dermal System (subject Device) was less than or comparable to the depth of thermal effect for the predicate NeoGen PSR system in a porcine model. The results of these pre-clinical studies support the safety and substantial equivalence of the Renuvion® Dermal System for dermatological wrinkle reduction procedures in comparison to the predicate nitrogen plasma technology (NeoGen PSR System, K132754).
#### 6. Human Clinical Trials
The clinical performance of the subject device (Renuvion® Dermal Handpiece) is supported by the results of a Clinical IDE study performed (G170151) using the subject device. This clinical study was a prospective, multicenter, single arm clinical study evaluating the use of the Renuvion® Dermal System for dermatological wrinkle reduction conducted at 3 investigational centers in the United States as an evaluator-blinded prospective study of 55 study subjects who were seeking a procedure for the purpose of improving facial appearance by reducing facial wrinkles and rhytides. Each study subject received one single-pass fullface treatment with the Renuvion® Dermal System.
Primary Effectiveness Endpoint: The primary effectiveness endpoint was the proportion of subjects with at least one-point improvement from Baseline in Fitzpatrick Wrinkle and Elastosis Scale (FWS) at 90-days post-procedure as determined by 2 out of 3 blinded Independent Photographic Reviewers (IPR) assessment of photographs.
Primary Safety Endpoint: The primary safety endpoint was the evaluation of adverse events up to the 3-month visit following the procedure.
#### a. Effectiveness Results
Of 55 subjects treated. 92.7% (51/55) were scored to have at least a one-point improvement, 25.5% were scored to have at least a two- point improvement, and 5.5% were scored to have at least a three-point improvement on the Fitzpatrick Wrinkle and Elastosis Scale (FWS) at 90 days as determined by Independent Photographic Reviewers (IPRs).
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image shows the logo for Apyx Medical. The logo has the word "Apyx" in orange with a blue triangle in place of the "A". Below the word "Apyx" is the word "MEDICAL" in blue. The TM symbol is in the upper right corner of the logo.
### K211652
#### b. Safety Results
The study reported no serious adverse events (SAEs) related to the study device or procedure, and adverse events (AEs) were in line with the expectations of dermatological wrinkle reduction procedures involving the application of controlled heat to the epidermis to selectively render it non-viable with subsequent thermal modification of the underlying dermis to prompt a healing response. The most common events reported were, edema, erythema, and post-inflammatory hyperpigmentation.
The rate of AEs for the subject device is consistent with the expected range that has been observed for the predicate device, as reported in the literature. Based on the clinical performance as documented in the clinical study (G170151), the Renuvion® Dermal System was found to have a safety and effectiveness profile that is consistent with other legally marketed dermatological wrinkle reduction technologies such as the predicate device.
Table 2 is a direct comparison of adverse events for the Renuvion Dermal Handbiece (subject device) compared to the Rhytec/Energist (predicate device) established in literature.
| Adverse Event | Renuvion<br>Dermal System<br>(n=55) | Rhytec/Energist Predicate<br>(n=24)1 |
|----------------------------------------|-------------------------------------|--------------------------------------|
| Crusting | 1.8% | 100%1 |
| Edema | 34.5% | 100%1 |
| Erythema | 16.4% | 100%1 |
| Milia/Acne | 12.7% | Not Reported |
| Pruritus/Itching | 9.1% | Not Reported |
| Hypertrophic Scar | 3.6% | 4.2%1 |
| Post-inflammatory<br>Hyperpigmentation | 14.5% | "A number of patients" at D901 |
| Telangiectasia | 0% | Not Reported |
Table 2. Adverse Event Profile Comparison for Subject and Predicate Device
{9}------------------------------------------------
Image /page/9/Picture/2 description: The image shows the logo for Aryx Medical. The logo features the word "Aryx" in orange, with a blue triangle replacing the "A". Below the word "Aryx" is the word "MEDICAL" in red. The logo is simple and modern, and the colors are bright and eye-catching.
### K211652
| Discoloration/Hypopigmentation<br>n | 0% | 0%1 |
|--------------------------------------------------------|------|--------------|
| Delayed Healing | 7.3% | Not Reported |
| Pain<br>D0 Post-Op VAS Score<br>(0=lowest, 10=highest) | 4.3 | 4.31 |
| Lid Retraction | 0% | Not Reported |
Kilmer S, Semchyshyn N, Shah G, Fitzpatrick R. A pilot study on the use of a plasma skin regeneration device (Portrait PSR3) in full facial rejuvenation procedures. Lasers Med Sci. 2007 Jun;22(2):101-9 doi: 10.1007/s10103-006-0431-9. Epub 2007 Mar 7. PMID: 17342383.
#### c. Human Clinical Trial Outcome Summary
The clinical study demonstrated both safety and effectiveness, as endpoints related to effectiveness were met and there were no serious adverse events reported related to the study device or procedure. The risks and AE profile are comparable between the subject device and what is reported in published literature for the predicate device as long as the recommended treatment parameters are followed. The effectiveness outcomes are as good or better for the subject device as compared to the predicate device.
- ❖ The totality of the data shows benefit to all subjects via the reduction of facial wrinkles and rhytides in subjects with Fitzpatrick Skin Type I, II, and III. The primary effectiveness endpoint for this study was achieved at 92.7%.
- * The primary safety endpoint identified the type and frequency of adverse events, and no new or different risks or adverse events were identified. There were no serious adverse events reported as related to the study device or the study procedure. The adverse events were comparable to those reported for the predicate device and fully ablative dermatological wrinkle reduction procedures in general using other FDA-cleared energybased modalities as long as the recommended treatment parameters are followed.
- * These data support the proposed Indications for Use of the product and substantial equivalence to the predicate device (Energist NeoGen Nitrogen Plasma K132754) for a 510(k) submission.
Based on the totality of data (safety variables and success criteria), there is sufficient assurance that safety and effectiveness has been demonstrated and clinically meaningful benefits have been observed in a majority of subjects. The safety and effectiveness conclusions drawn from the clinical studies demonstrate that there is statistically significant data to support the safety, effectiveness, and performance of the Renuvion® Dermal System for the treatment of moderate to severe facial wrinkles and rhytides in patients with Fitzpatrick Skin Types I, II or III and the benefits of Renuvion outweigh the risks. The subject device clinical study results, when compared to the predicate device safety and effectiveness data as stated in literature also supports substantial equivalence of the Renuvion Dermal Handpiece.
{10}------------------------------------------------
Image /page/10/Picture/2 description: The image shows the logo for Aryx Medical. The logo is primarily orange, with the word "Aryx" in a stylized font. A small blue triangle is incorporated into the "A" of "Aryx." Below "Aryx" is the word "MEDICAL" in blue, in a sans-serif font.
### K211652
### 7. Substantial Equivalence
The predicate device was selected because the subject device's intended use is a subset of the intended use for the primary predicate device (electrosurgical device for dermatologic treatment of wrinkles and rhytides - the treatment is achieved through controlled heating of the outer layers of the skin so that part or all the epidermis becomes non-viable and there is controlled thermal modification to the underlying dermis). The reference device was selected due to having the same principle of operation, mechanism of action, sterilization methods and technology used for device performance as the subject device, other than the few design modifications described in the submission. A summary of the subject device and predicate device comparison is provided below.
| | Subject Device | Predicate Device |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature/<br>Characteristic | Renuvion® Dermal Handpiece | NeoGen PSR System,<br>K132754 |
| Indications for<br>Use | The Renuvion® Dermal system is an<br>electrosurgical device for<br>dermatological procedures for the<br>treatment of moderate to severe<br>wrinkles and rhytides, limited to<br>patients with Fitzpatrick skin types I,<br>II or III. The treatment is achieved<br>through controlled heating of the<br>outer layers of the skin so that part or<br>all the epidermis becomes non-viable<br>and there is controlled thermal<br>modification to the underlying dermis.<br>The Renuvion® Dermal Handpiece is<br>only compatible with Apyx Medical<br>Electrosurgical Generators | The NeoGen PSR system is an<br>electrosurgical device for the<br>treatment of dermatologic<br>conditions including acne scars,<br>actinic keratosis, superficial skin<br>lesions and the treatment of<br>wrinkles and rhytides. The<br>treatment is achieved through<br>controlled heating of the outer<br>layers of the skin so that part or<br>all the epidermis becomes non-<br>viable and there is controlled<br>thermal modification to the<br>underlying dermis. |
| Software | Renuvion® Dermal handpiece does<br>not contain software; Renuvion<br>generator contains firmware | NeoGen PSR handpiece does<br>not contain software; NeoGen<br>generator contains software |
| Sterility | Single Use disposable Sterile (EO)<br>Handpiece | Handpiece clean but non-sterile |
| Use<br>Environment | Doctor's Office, Surgery Center, or<br>similar environment | Doctor's Office/ Clinics or<br>hospitals |
| Clinical<br>Application | Dermatologic on intact & broken skin;<br>Dermal procedures (facial) | Dermatologic on intact & broken<br>skin; Dermal procedures (facial &<br>non-facial) |
| Prescription or<br>OTC | Prescription Use Only | Prescription Use Only |
| Energy Source | Electrosurgical generator | Electrosurgical generator |
| Energy Type | Inert Gas Plasma (helium) | Inert Gas Plasma (nitrogen) |
| Waveform | Radiofrequency waveform | Radiofrequency waveform |
{11}------------------------------------------------
Image /page/11/Picture/2 description: The image shows the logo for Apyx Medical. The logo is orange and blue. The word "MEDICAL" is written in blue below the logo.
### K211652
| | Subject Device | Predicate Device |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature/<br>Characteristic | Renuvion® Dermal Handpiece | NeoGen PSR System,<br>K132754 |
| Device<br>Activation | Foot Pedal Activation or<br>button hand activation | Foot Pedal Activation |
| User Interface | Generator buttons and displays | Generator buttons and displays |
| Limitations on<br>device use | Device to be used by trained medical<br>staff only, in a Doctor's Office,<br>Surgery Center or similar<br>environment | Device to be used by trained<br>medical staff only, in a Doctor's<br>Office, Surgery Center or similar<br>environment |
| Mode of Action | The treatment is achieved through<br>controlled heating of the outer layers<br>of the skin so that part or all the<br>epidermis becomes non-viable and<br>there is controlled thermal<br>modification to the underlying dermis | The treatment is achieved<br>through controlled heating of the<br>outer layers of the skin so that<br>part or all of the epidermis<br>becomes non-viable and there is<br>controlled thermal modification to<br>the underlying dermis |
| Power<br>delivered to<br>patient | 3.2 - 8.55 W | 1 – 10 W |
| Procedures &<br>Anatomical<br>site | Dermatological procedures | Dermatological procedures |
| Plasma<br>Settings | Controls on the generator allows the<br>user to vary the power delivered to<br>the patient | Controls on the generator allows<br>the user to vary the power<br>delivered to the patient |
| Compatibility | Compatible only with specific<br>generator | Compatible only with specific<br>generator |
| Connector | Compatible only with specific<br>generator | Compatible only with specific<br>generator |
### 8. Conclusion
Clinical testing of the Renuvion® Dermal System (subject device) demonstrated that the device performed as intended with a favorable safety profile. The primary effectiveness endpoint FWS improvement score, investigator FWS score, and patient assessment of results yielded comparable or more improved outcomes than the predicate device. The safety profile of the subject device is comparable to the predicate device; no new or different risks were identified, and adverse event rates were comparable to the predicate device rates when used following recommended treatment settings. The Renuvion® Dermal Handpiece safety and performance has been further confirmed by the results of the performance bench and animal testing. Technologically, there are no differences between the subject and predicate device that would raise new or different questions regarding safety or effectiveness; the maximum depths of thermal effect measured for the Renuvion® Dermal Handpiece are less than or statistically equivalent to the predicate device as measured by histopathology. In conclusion,
{12}------------------------------------------------
Image /page/12/Picture/2 description: The image shows the logo for Apyx Medical. The logo is orange, with a blue triangle in the lower left corner of the "A". The word "MEDICAL" is in blue and is located below the main logo. There is a trademark symbol in the upper right corner of the logo.
# K211652
the totality of the data (clinical, pre-clinical animal data, performance bench testing, and benefit-risk assessment) demonstrate that the Renuvion® Dermal Handpiece is substantially equivalent to the predicate device (NeoGen PSR System, Energist Ltd.- K132754) for the requested indication for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.