K211649 · Baxter Healthcare Corporation · PWH · Nov 23, 2021 · General Hospital
Device Facts
Record ID
K211649
Device Name
PCEA Syringe Set
Applicant
Baxter Healthcare Corporation
Product Code
PWH · General Hospital
Decision Date
Nov 23, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Intended Use
For the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device.
Device Story
PCEA Syringe Set is a single-use, sterile, non-pyrogenic disposable device; facilitates administration of anesthetics/analgesics from a syringe to a patient's epidural space via an epidural access device. Components include non-DEHP PVC tubing/bushing, female/male NRFit locks/caps, on-off clamp, and anti-siphon valve. Anti-siphon valve prevents inadvertent free flow and backflow by requiring specific opening pressure. Used in clinical settings; operated by healthcare professionals. Output is the controlled delivery of medication to the epidural space. Benefits include reduced risk of free flow and secure connection via ISO 80369-6 compliant NRFit connectors.
Clinical Evidence
Bench testing only. Performance evaluated via ISO 80369-6 (connector testing), ISO 8536-9 (tensile strength/leakage), ISO 8536-12 (counter flow/blocking/opening pressure), ISO 8536-14 (clamp shut-off), USP <788> (particulate matter), and biocompatibility per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, genotoxicity). All tests met acceptance criteria.
Technological Characteristics
Materials: PVC (tubing/bushing), Copolyester (NRFit locks), Acrylic (caps), Polycarbonate/Silicone (anti-siphon valve). Energy: None (mechanical). Connectivity: None. Sterilization: Gamma irradiation (SAL 10^-6). Dimensions: 216 cm length, 1.1 mm ID, 2.5 mm OD. Standards: ISO 80369-6 (connectors), ISO 8536-9/12/14 (performance), ISO 10993-1 (biocompatibility), USP <788> (particulates).
Indications for Use
Indicated for the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
CADD Yellow Extension Sets with NRFit™ connector (K162219)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 23, 2021
Baxter Healthcare Corporation Jeffrey Thompson Manager, Regulatory Affairs 25212 West Illinois Route 120 Round Lake, Illinois 60073
Re: K211649
Trade/Device Name: PCEA Syringe Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: PWH Dated: October 20, 2021 Received: October 25, 2021
Dear Jeffrey Thompson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211649
Device Name PCEA Syringe Set
Indications for Use (Describe)
For the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif and slightly italicized. The word is the main focus of the image and is positioned horizontally.
### K211649
#### 510(k) Summary
#### OWNER:
Baxter Healthcare Corporation 25212 W. IL Route 120 Round Lake, Illinois 60073
# CONTACT PERSON:
Jeffrey E. Thompson Manager, Regulatory Affairs 25212 W. IL Route 120 Round Lake, Illinois 60073 Telephone: (224) 270 3806 Fax: (224) 270 4119
Date Preparation: November 23, 2021
## SUBJECT DEVICE:
Trade/Device Name: PCEA Syringe Set Classification Panel: 80 General Hospital Regulation Number: 21 CFR 880.5440 Common Name: Administrations Sets with Neuraxial Connectors Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: PWH
| Code # | Device Description |
|--------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2P3334 | PCEA Syringe Set with Anti-Siphon Valve, 85" (216 cm),<br>Vol 2.2 mL<br>Image: [device]<br>1: Female NRFit Cap<br>2: Female NRFit Lock<br>3: Anti-Siphon Valve<br>4: On-Off Clamp<br>5: Male NRFit Lock<br>6: Male NRFit Cap<br>7: Non-DEHP Tube Bushing<br>8,9: Non-DEHP Tube |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is slightly italicized, giving the word a sense of movement. The word is the logo for the Baxter International Inc. company.
## PREDICATE DEVICE:
| Device | Company | Predicate 510(k) | Clearance<br>Date |
|-----------------------------------------------------|--------------------------|------------------|-------------------|
| CADD Yellow Extension Sets<br>with NRFit™ connector | Smiths Medical ASD, Inc. | K162219 | July 20, 2017 |
### Table 2: Predicate Device
### REASON FOR SUBMISSION:
The basis for this premarket notification is the intent to market a patient controlled epidural analgesia (PCEA) syringe set. The proposed device in this submission is a single-use, disposable device, intended for the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device.
### DESCRIPTION OF THE DEVICE:
The proposed device is a patient controlled epidural analgesia (PCEA) syringe set. It is a single use disposable device intended for the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device. It is a non-pyrogenic, sterile device that can be directly attached to a syringe.
The PCEA syringe set consists of non-DEHP PVC tubing/bushing, female NRFit cap, female NRFit lock, anti-siphon valve, on-off clamp, male NRFit lock, and male NRFit cap. It is used to administer anesthetics and/or analgesics from a syringe to the patient's epidural space through an epidural access device.
The anti-siphon valve component adds specific functionalities to facilitate the administration of anesthetics and/or analgesics to the patient's epidural access device, as described below:
- Anti-siphon valve: reduces the risk of any inadvertent free flow of solution to the patient. The higher opening pressure does not allow administration of flow into the patient's epidural access device, unless the stipulated pressure is applied. Another safety feature is that the valve does not allow any backflow. This prevention of backflow is achieved through the valve's ability to withstand a counterflow pressure.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif and slightly italicized. The word is horizontally oriented and takes up most of the frame.
## INDICATIONS FOR USE:
For the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device.
## TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:
The proposed device is substantially equivalent to the predicate device, previously cleared under 510(k) premarket notification K162219 on July 20, 2017. The intended use and function of the proposed device is equivalent to the predicate device.
Table 3 is a device comparison table outlining the differences between the predicate and proposed devices.
| Feature | Predicate<br>K162219 | Subject Device<br>K211649 | Assessment of Differences |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | CADD Yellow<br>Extension Sets with<br>NRFitTM connectors<br>are designed for use<br>only with CADD<br>Yellow Medication<br>Cassette Reservoirs<br>with NRFitTM<br>connectors for the<br>delivery of regional<br>anesthetics or<br>narcotics. | For the<br>administration of<br>anesthetics and/or<br>analgesics from a<br>syringe into the<br>patient's epidural<br>space through an<br>epidural access<br>device. | Both the proposed and<br>predicate device are intended<br>to deliver similar<br>medications to the patient<br>through NRFit connections.<br>The specified fluid<br>containers for the proposed<br>and predicate devices are<br>different; however, the<br>connections to both fluid<br>containers claim compliance<br>to the same ISO 80369-6<br>standard ("NRFit"). Design<br>control activities have been<br>conducted and have<br>confirmed that the different<br>technological characteristics<br>of the proposed device does<br>not raise different questions<br>of safety and effectiveness. |
| Sterility | Sterile; Ethylene<br>Oxide | Sterile; Gamma<br>Irradiation | Both the proposed and<br>predicate devices have been<br>validated to a sterility<br>assurance level (SAL) of ≤<br>10-6 using ISO 11137 for the<br>proposed device and ISO<br>11135 for the predicate<br>device. The sterilization<br>modalities for the proposed<br>and predicate devices are |
| Feature | Predicate<br>K162219 | Subject Device<br>K211649 | Assessment of Differences |
| | | | different (gamma irradiation<br>and ethylene oxide,<br>respectively); however, the<br>validated SAL requirement<br>for both devices aligns with<br>the same ANSI/AAMI ST67<br>standard (Sections 4.2.1 and<br>4.2.1.1), and the<br>technological differences<br>between the sterilization<br>modalities do not raise<br>questions of safety and<br>effectiveness. |
| Non-Pyrogenic | Yes | Same | N/A |
| Single Use | Yes | Same | N/A |
| Length | 90" | 85" (216 cm) | The proposed device length<br>is nominally shorter than the<br>predicate device length.<br>Design control activities<br>have been conducted and<br>have confirmed that the<br>different technological<br>characteristics of the<br>proposed device does not<br>raise different questions of<br>safety and effectiveness.<br>Lengths were calculated<br>using nominal values from<br>component and tubing part<br>specifications per Baxter<br>procedure. |
| Tubing Inner / Outer<br>Diameter | Information not<br>available | Inner Diameter - 1.1<br>mm<br>Outer Diameter - 2.5<br>mm | The information is not<br>available for the predicate<br>device. Design control<br>activities have been<br>conducted on the proposed<br>device and have confirmed<br>that if there are different<br>technological characteristics<br>to the predicate device, they<br>would not raise different<br>questions of safety and<br>effectiveness.<br>Tensile strength and leakage<br>were evaluated per ISO<br>8536-9. |
| Feature | Predicate<br>K162219 | Subject Device<br>K211649 | Assessment of Differences |
| Priming Volume | 2.3 mL | 2.2 mL | The proposed device priming<br>volume is nominally less<br>than the predicate device<br>priming volume. Design<br>control activities have been<br>conducted and have<br>confirmed that the different<br>technological characteristics<br>of the proposed device does<br>not raise different questions<br>of safety and effectiveness.<br>Priming volume was<br>determined per Baxter test<br>method. |
| Fluid Path Components/Materials | | | |
| Anti-Siphon Valve | Polycarbonate<br>(Housing)<br>Silicone<br>(Membrane) | Same | N/A |
| Tubing | Polyvinyl Chloride | Same | N/A |
| Tube Bushing | Not Applicable | Polyvinyl Chloride | The predicate device does<br>not have tube bushings.<br>Design control activities<br>have been conducted and<br>have confirmed that the<br>different technological<br>characteristics of the<br>proposed device does not<br>raise different questions of<br>safety and effectiveness.<br>Biocompatibility<br>assessments were conducted<br>per ISO-10993-1. |
| Female NRFit Lock | Not Applicable | Copolyester | The predicate device does<br>not have a female NRFit |
| Feature | Predicate<br>K162219 | Subject Device<br>K211649 | Assessment of Differences |
| | | | lock. Design control<br>activities have been<br>conducted and have<br>confirmed that the different<br>technological characteristics<br>of the proposed device does<br>not raise different questions<br>of safety and effectiveness.<br>Biocompatibility<br>assessments were conducted<br>per ISO-10993-1.<br>Functional performance was<br>evaluated per ISO 80369-6. |
| Male NRFit Lock | Acrylonitrile<br>Butadiene Styrene | Copolyester | The predicate device does<br>not have the same type of<br>material formulation.<br>Design control activities<br>have been conducted and<br>have confirmed that the<br>different technological<br> |
| Male NRFit Cap | Linear Low Density<br>Polyethylene | Acrylic | The predicate device does<br>not have the same type of<br>material formulation.<br>Design control activities<br>have been conducted and<br>have confirmed that the<br>different technological<br>characteristics of the<br>proposed device does not<br>raise different questions of<br>safety and effectiveness.<br>Biocompatibility<br>assessments were conducted<br>per ISO-10993-1. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be italicized, with the letters slightly slanted to the right. The word is the logo for the Baxter International Inc. company.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly slanted to the right, giving it a dynamic appearance.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be italicized, with the letters slightly slanted to the right. The word is the logo for the Baxter International Inc. company.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "Baxter" in a bold, italicized, sans-serif font. The color of the text is a dark blue. The word is the logo for the Baxter International Inc. company.
## DISCUSSION OF NONCLINICAL TESTS:
Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses. All test results meet their acceptance criteria and support that the proposed device is appropriately designed for its intended use.
## Performance Data:
The following bench tests (Table 4) were conducted to evaluate the functional performance of the proposed device:
| Test | Acceptance Criteria |
|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 80369-6 Tests on Male NRFit (Neuraxial Fit) Lock<br>Connector | ISO 80369-6: 2016, Clause 5 (as applicable),<br>ISO 80369-6:2016, Clause 6.1,<br>ISO 80369-6:2016, Clause 6.2,<br>ISO 80369-6:2016, Clause 6.3,<br>ISO 80369-6:2016, Clause 6.4,<br>ISO 80369-6:2016, Clause 6.5,<br>ISO 80369-6:2016, Clause 6.6 |
| ISO 80369-6 Tests on Female NRFit (Neuraxial Fit)<br>Lock Connector | ISO 80369-6: 2016, Clause 5 (as applicable),<br>ISO 80369-6:2016, Clause 6.1,<br>ISO 80369-6:2016, Clause 6.2,<br>ISO 80369-6:2016, Clause 6.3,<br>ISO 80369-6:2016, Clause 6.4,<br>ISO 80369-6:2016, Clause 6.5,<br>ISO 80369-6:2016, Clause 6.6 |
| Tensile Strength Test | ISO 8536-9:2015, Clause 5.3 |
| Leak Test | ISO 8536-9:2015, Clause A.4 |
| Counter Flow Test | ISO 8536-12:2007+A1:2013, Clause A.4 |
| Blocking Performance Test | ISO 8536-12:2007+A1:2013, Clause A.6 |
| Opening Pressure Test | ISO 8536-12:2007+A1:2013, Clause A.7.1<br>and per Baxter Test Method |
| Particulate Matter Test | USP Chapter <788> |
| Clamp Activation Force Test | Per Baxter Test Method |
| Clamp Shut-Off Test | ISO 8536-14:2018, Clause A.1 |
| Non-DEHP Claim Verification | Per Baxter Test Method (as tested in<br>K161808) |
# Table 4: Performance Data
All tests met the acceptance criteria.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly slanted to the right, giving it a dynamic appearance. The font is sans-serif and has a modern look.
## Biocompatibility:
Biocompatibility assessments were conducted based on ISO-10993-1. Biological Evaluation of Medical Devices for prolonged contact duration, external communicating device, tissue/bone/dentin, and FDA-2013-D-0350 Guidance for Industry & FDA Staff, Use of ISO 10993-1, "Biological evaluation and medical devices - Part 1: Evaluation and testing within a risk management process, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing", as recommended in the Intravascular Administration Sets guidance, "Guidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)]". Biocompatibility assessments include the following assays:
- Cytotoxicity ISO 10993-5
- Sensitization ISO 10993-10
- Intracutaneous (Irritation) Reactivity ISO 10993-10
- Acute Systemic Toxicity ISO 10993-11
- . Material Mediated Pyrogen ISO 10993-11
- Genotoxicity ISO 10993-3
- Subacute Systemic Toxicity ISO 10993-11
Based upon the results the device met the designated ISO 10993-1 categorization and is biocompatible and appropriate for the intended use.
### Sterility:
The proposed device is sterilized with gamma radiation. The product is in the bioburden (sub) category "General Sets Labeled 'Sterile''. The Minimum Sterilizing Dose (MSD) required to provide a 106 Sterility Assurance Level (SAL) for this (sub) category was established and validated at the manufacturing facility as described in ANSI/AAMI/ISO 11137-2, "Sterilization of health care products - Radiation-Part 2: Establishing the sterilization dose." The dose setting method used includes, but is not limited to, Method 1 or VDmax. Generally, the MSDs are between 14.2 - 25.0 kGy. The continued validity of the MSD for this (sub) category is confirmed via periodic dose audit studies. In addition, routine periodic pre-sterilization bioburden testing is performed for each (sub) category. The endotoxin limit is 2.15 EU/device per USP <161>.
### Shelf Life:
Baxter has performed aging testing to support a shelf-life claim of 3 (three) years.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif and slightly italicized. The word is presented in a clean and professional manner.
## CONCLUSION:
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The non-clinical data demonstrate that the subject device is substantially equivalent and performs comparably to the predicate device that is legally marketed for the same intended use.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.