K211637 · Daeju Meditech Engineering Co., Ltd. · GEX · Sep 3, 2021 · General, Plastic Surgery
Device Facts
Record ID
K211637
Device Name
AROMA GRAND
Applicant
Daeju Meditech Engineering Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 3, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AROMA GRAND is indicated for hair removal, permanent hair reduction and for the treatment of benign vascular and pigmented lesions. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regimen.
Device Story
Aroma Grand is a Class IV semiconductor diode laser surgical system (808nm) for hair removal and treatment of benign vascular/pigmented lesions. Device comprises a main unit and a laser hand-piece with a sapphire crystal optical guide. Physician/clinician operates device via footswitch to deliver laser energy to treatment area. Water cooling system manages thermal output. Device output parameters (fluence, pulse duration, spot size) are controlled by the user to target hair follicles or vascular/pigmented structures. Clinical benefit includes permanent hair reduction and lesion treatment. Device is intended for professional use in clinical settings.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included electrical safety (IEC 60601-1, IEC 60601-2-22), EMC (IEC 60601-1-2), laser safety (IEC 60825-1), and biocompatibility (ISO 10993-1). Accuracy testing verified energy output and spot size within tolerance.
Technological Characteristics
Diode laser (808nm); Class IV; 600W power; 1-10Hz frequency; 10-350ms pulse duration; 2.1cm2 spot size; water cooling; sapphire crystal optical guide. Complies with IEC 60601-1, IEC 60601-2-22, IEC 60825-1, and IEC 60601-1-2. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for hair removal, permanent hair reduction, and treatment of benign vascular and pigmented lesions in patients of all skin types (Fitzpatrick I-VI).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 3, 2021
Daeju Meditech Engineering Co., Ltd. % April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620
Re: K211637
Trade/Device Name: Aroma Grand Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: July 30, 2021 Received: August 5, 2021
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211637
Device Name AROMA GRAND
Indications for Use (Describe)
The AROMA GRAND is indicated for hair removal, permanent hair reduction and for the treatment of benign vascular and pigmented lesions.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regimen.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CER 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### Submitter
DAEJU MEDITECH ENGINEERING Co., Ltd. SEONGUN KIM #501-504, HausD Sejong Tower, 26, Seongsu-il-ro 10gil, Seongdong-gu Seoul, South Korea 04793 Email: globalsales@daejumedi.co.kr Phone: +82-2-2208-0905
# Device Information
- Trade Name: AROMA GRAND
- . Common Name: Powered laser surgical instrument
- Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
- Product Code: GEX
- Panel: General & Plastic Surgery
- Regulation Number: 21 CFR §878.4810
- Device Class: Class II
- Date prepared: 08/19/2021
# Predicate Device
Primary Predicate K151232, VIKINI DIODE LASER SYSTEM by ILOODA Company, Ltd.
### Indications for use
The AROMA GRAND is indicated for hair removal, permanent hair reduction and for the treatment of benign vascular and pigmented lesions.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regimen.
### Device Description
The subject device. AROMA GRAND is a surgical device, which is intended for hair removal, permanent hair reduction on all skin tvpes (Fitzpatrick skin type I-VI).
It utilizes a semiconductor diode as a laser source (808nm). The laser power is delivered to the treatment area via a laser hand-piece. The emission laser is activated by a footswitch.
### Official Correspondent Withus Group Inc. April Lee 106 Superior,
Irvine. CA USA 92620 Email: withus6664@gmail.com Phone: +1-909-274-9971
{4}------------------------------------------------
# Summary of Technological Characteristics
| ITEM | PROPOSED DEVICE | PREDICATE DEVICE | Remark |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | AROMA GRAND | VIKINI DIODE LASER<br>SYSTEM | |
| K number | K211637 | K151232 | |
| Product Code | GEX | GEX | Same |
| Regulation | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Indications for<br>Use | The AROMA GRAND is<br>indicated for hair removal,<br>permanent hair reduction and for<br>the treatment of benign vascular<br>and pigmented lesions.<br><br>Permanent hair reduction is<br>defined as the long-term, stable<br>reduction in the number of hairs<br>re-growing when measured at 6,<br>9, and 12 months after the<br>completion of a treatment<br>regimen. | The VIKINI DIODE LASER<br>SYSTEM is indicated for hair<br>removal, permanent hair<br>reduction and for the treatment of<br>benign vascular and pigmented<br>lesions.<br><br>Permanent hair reduction is<br>defined as the long-term, stable<br>reduction in the number of hairs<br>re-growing when measured at 6,<br>9, and 12 months after the<br>completion of a treatment<br>regimen. | Same |
| Configuration | Main Unit,<br>Hand-piece | Main Unit,<br>Hand-piece | Same |
| Principle of<br>Operation | Diode Laser | Diode Laser | Same |
| Laser Type | Diode Laser | Diode Laser | Same |
| Laser<br>Classification | Class IV | Class IV | Same |
| Laser<br>Wavelength | 808[nm] | 808[nm] | Same |
| Spot Size | 2.1[cm2] / 14*15[mm] | 1.32[cm2] / 12*11[mm] | Analysis 1 |
| Fluence | 100[J/cm²] | 120[J/cm²] | Analysis 2 |
| Max. Output | 100W/cm2 | 284W/cm2 | Analysis 3 |
| ITEM | PROPOSED DEVICE | PREDICATE DEVICE | Remark |
| Frequency | 1-10[Hz] | 1-10[Hz] | Same |
| Power | 600W | 600W | Same |
| Purse duration | 10-350[ms] | 5-625[ms] | Analysis 4 |
| Cooling system | Water cooling | Water cooling | Same |
| Optical Guide | Sapphire Crystal | Sapphire Crystal | Same |
| Power Supply | 100-240V, 50/60 Hz | 100-240V, 50/60 Hz | Same |
| Dimensions | 480*640*1,230[mm]<br>(L*W*H) | 380*400*840 [mm] | Analysis 5 |
| Weight | 67[kg] | 43[kg] | |
| Electrical Safety | Comply with<br>IEC 60601-1<br>IEC 60601-2-22<br>IEC 60825-1 | Comply with<br>IEC 60601-1<br>IEC 60601-2-22<br>IEC 60825-1 | Same |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Same |
{5}------------------------------------------------
{6}------------------------------------------------
### Analysis 1 - Spot size
The spot size of the proposed device is larger than the spot size of the predicate device. Fluency (energy density) is a key factor in laser therapy. The fluency of the proposed device is not much different from the fluency of the predicate device. Therefore, the difference on spot size is considered would not raise any issues in safety and effectiveness.
### Analysis 2-Fluence
The fluence for the proposed device is different from predicate device. However, this difference is very slight and performance testing has been conducted on the proposed device and the test result can comply with related standards requirement. Therefore, this difference does not raise issues of safety or effectiveness.
### Analysis 3-Max. Output
The Max. Output for the proposed device is different from predicate device. The value is calculated as Power*Pulse duration/Spot Size. Power is the same, but differences occur due to Pulse durations and Spot Size. First of all, check the analysis 1 and analysis 4. Therefore, this difference does not raise issues of safety or effectiveness.
### Analysis 4-Pulse duration
The Pulse duration is most important parameter how long time the energy will deliver to the patient s skin, it may affect the safety (too long action time) may burn patient`s skin. The Pulse duration of proposed device is 10-350ms, it's less than the Pulse duration of predicate device. That s means the proposed device could be consider as safety.
### Analysis 5-Dimension and weight
The dimension and weight for the proposed device is different from predicate device. However, the dimension and weight difference are just in physical specification and this difference will not raise any issues in safety and effectiveness.
{7}------------------------------------------------
### Non-clinical Testing Summary
The non-clinical testing results demonstrate that the subject device is as safe, as effective, and performs as well as the predicate device.
### Biocompatibility Testing
The biocompatibility evaluation for the subject device was conducted in accordance with the guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The biocompatible testing including cytotoxicity, Sensitization are conducted according to the ISO 10993-1: 2009 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
# Electrical Safety and electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on Diode Laser Hair Removal Machine. The device complies with the following standards
- AAMI/ANSI/ES 60601-1:2005(R) 2012 and A1:2012 Medical electrical equipment Part 1: ● General requirements for basic safety and essential performance.
- EN 60601-1-2:2015 Medical electrical equipment Part 1-2: General requirements for basic . safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
- IEC 60825-1:2014 Safety of laser products Part 1: Equipment classification and requirements. ●
# Particular Performance Testing
Performance testing was conducted on the device according to the following standard:
- . IEC 60601-2-22:2012 Medical electrical equipment - Part 2-22: Particular requirements for the safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
### Software Verification and Validation Testing
The software for this device was considered as a "Moderate" level of concern. Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
### Accuracy Testing
The accuracy test was conducted to verify that the energy output and spot size of the proposed laser system do not deviate the tolerance of the setting value of energy output or the fixed value of spot size. The non-clinical testing results demonstrate that the subject device is as safe, as effective, and performs as well as the predicate device.
### Clinical Testing
Not applicable.
### Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be as safe, as effective, and performs as well as the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.