SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900, S40 Sterilant Concentrate
K211607 · STERIS Corporation · MED · Sep 7, 2021 · General Hospital
Device Facts
Record ID
K211607
Device Name
SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900, S40 Sterilant Concentrate
Applicant
STERIS Corporation
Product Code
MED · General Hospital
Decision Date
Sep 7, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6885
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities. The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use. The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
Device Story
Automated, microprocessor-controlled liquid chemical sterilization system; processes heat-sensitive, immersible medical devices (flexible/rigid endoscopes, microsurgical instruments). Uses S40 Sterilant Concentrate (peracetic acid) diluted with water; 6-minute exposure at 45.5–60°C. System includes processor, interchangeable trays/containers, and Quick Connects for internal lumen access. Post-sterilization, load rinsed with extensively treated water (pre-filtered, UV-irradiated, 0.1-micron filtered). Used in healthcare facilities; operated by trained staff. Provides printed cycle documentation; monitors critical parameters (temperature, contact time, sterilant concentration, filter integrity). Ensures device sterility for immediate use; critical devices not used immediately must be reprocessed. Benefits include rapid sterilization of heat-sensitive instruments without damage.
Clinical Evidence
Bench testing only. Efficacy validated via AOAC Official Methods (966.04, 955.17, 955.14, 955.15, 964.02) and EPA virucidal testing (DIS/TSS-7). Simulated-use testing demonstrated >6-log reduction of Geobacillus stearothermophilus spores. Material compatibility confirmed through 300 cycles for flexible endoscopes and 150 cycles for rigid devices.
Technological Characteristics
Microprocessor-controlled liquid chemical sterilizer. Uses peracetic acid (S40 Sterilant Concentrate). Features: interchangeable processing trays, Quick Connects, UV water treatment, 0.1-micron (or 0.2-micron) bacterial retentive filtration. Standardized, unalterable cycle parameters. Connectivity: USB for data download, barcode scanner, touchscreen interface, web-based data management. Sterilization: 6-minute exposure at 45.5–60°C.
Indications for Use
Indicated for liquid chemical sterilization of cleaned, immersible, reusable critical and semi-critical heat-sensitive medical devices and accessories in healthcare facilities.
Regulatory Classification
Identification
A liquid chemical sterilant/high level disinfectant is a germicide that is intended for use as the terminal step in processing critical and semicritical medical devices prior to patient use. Critical devices make contact with normally sterile tissue or body spaces during use. Semicritical devices make contact during use with mucous membranes or nonintact skin.
Special Controls
*Classification.* Class II (special controls). Guidance on the Content and Format of Premarket Notification (510(k)) Submissions for Liquid Chemical Sterilants/High Level Disinfectants, and user information and training.
Predicate Devices
SYSTEM 1E Liquid Chemical Sterilant Processing System (K210737)
SYSTEM 1endo Liquid Chemical Sterilant Processing System, Model P6800 (K210737)
SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900 (K210737)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 7, 2021
STERIS Corporation Jennifer Nalepka Lead Regulatory Affairs Specialist 5960 Heisley Road Mentor, Ohio 44060
Re: K211607
Trade/Device Name: SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900, S40 Sterilant Concentrate Regulation Number: 21 CFR 880.6885 Regulation Name: Liquid Chemical Sterilants/High Level Disinfectants Regulatory Class: Class II Product Code: MED Dated: September 1, 2021 Received: September 2, 2021
Dear Jennifer Nalepka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K211607
### Device Name
SYSTEM 1E Liquid Chemical Sterilant Processing System
Indications for Use (Describe)
The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities.
The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use.
The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
* The extensive treatment of EPA potable water consists of:
- 1. Pre-filtration through two pre-filters:
- · Pre-filter A is a gross depth filter that removes approximately 2.5 micron or larger particles/contaminants.
- · Pre-filter B is a surface filter that removes particles/contaminants > 0.1 micron.
- 2. UV Irradiation:
- · During transit through the UV water treatment chamber, a UV dose sufficient to achieve a > or equal to 6-log reduction of MS2 virus is delivered to the water.
- 3.0.1 micron filtration:
- · The water prepared by pre-filtration is filtered through redundant, 0.1-micron (absolute rated) membranes to remove bacteria. fungi and protozoa > 0.1 micron.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K211607
Device Name
SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800
Indications for Use (Describe)
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical medical devices and their accessories in healthcare facilities.
The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution(> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span></span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K211607
Device Name
SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900
Indications for Use (Describe)
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable sensitive medical devices and their accessories in healthcare facilities.
The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the STERIS logo. The word "STERIS" is written in a serif font in all capital letters. Below the word is a graphic of several horizontal wavy lines in blue, resembling water waves. The registered trademark symbol is located to the right of the word STERIS.
# 510(k) Summary For SYSTEM 1E Liquid Chemical Sterilant Processing System
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
Jennifer Nalepka Contact: Lead Regulatory Affairs Specialist Tel: 440-392-7458 Fax: 440-357-9198
Summary Date: September 1, 2021
Premarket Notification Number: K211607
{6}------------------------------------------------
#### 1. Device Name
| Trade Name: | SYSTEM 1E Liquid Chemical Sterilant Processing<br>System |
|------------------------|-------------------------------------------------------------------------|
| Device Classification: | Class II |
| Common/usual Name: | Liquid Chemical Sterilizer |
| Classification Name: | Sterilant, Medical devices, Liquid Chemical<br>Sterilants/Disinfectants |
| Classification Number: | 21 CFR 880.6885 |
| Product Code: | MED |
#### 2. Predicate Device
SYSTEM 1E Liquid Chemical Sterilant Processing System, K210737
#### 3. Description of Device
The SYSTEM 1E Liquid Chemical Sterilant Processing System is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible, heat sensitive, flexible and rigid endoscopes and their accessories, and microsurgical instruments. The system consists of the SYSTEM 1E Processor and the S40 Sterilant Concentrate, interchangeable processing trays/containers and Quick Connects. The current submission is provided to describe modifications for:
- . Obsolescence and replacement of interface board and connector
- Obsolescence and replacement of silicone tubing ●
- S40 Sterilant Concentrate design modification ●
The SYSTEM 1E Processor is an automated, self-contained device which creates and maintains the conditions necessary for liquid chemical sterilization in 6 minutes. Following processing, the liquid chemically sterilized articles are rinsed with extensively treated water produced by passing EPA potable tap water through pre-filters, an ultraviolet light treatment subsystem, and then through two 0.1micron filter membranes. The processor, which is computer controlled and continually monitored, provides printed documentation of each cycle.
S40 Sterilant Concentrate is a single use chemical sterilant concentrate developed for use in the SYSTEM 1E Processor. The active ingredient in S40 Sterilant Concentrate, peracetic acid, is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.
{7}------------------------------------------------
The interchangeable processing trays/containers are made to accommodate a variety of instrument types, models and procedure specific sets. Each container is designed to maintain instruments in appropriate position while specific Quick Connects for the SYSTEM 1E Processor, if required, facilitate delivery of the liquid chemical sterilant use dilution and rinse water to internal channels.
#### 4. Indications for Use
The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semicritical heat-sensitive medical devices in healthcare facilities.
The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use.
The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
- * The extensive treatment of EPA potable water consists of:
- 1. Pre-filtration through two pre-filters:
- Pre-filter A is a gross depth filter that removes approximately 2.5 . micron or larger particles/contaminants.
- Pre-filter B is a surface filter that removes particles/contaminants > . 0.1 micron.
- 2. UV Irradiation:
- During transit through the UV water treatment chamber, a UV dose . sufficient to achieve a > 6-log reduction of MS2 virus is delivered to the water.
- 3. 0.1 micron filtration:
- The water prepared by pre-filtration and UV irradiation is filtered . through redundant, 0.1-micron (absolute rated) membranes to remove bacteria, fungi and protozoa > 0.1 micron.
#### ട. Technological Characteristic Comparison Table
The SYSTEM 1E Liquid Chemical Sterilant Processing System is the same as the predicate device; the specific modifications described in this submission are for a firmware update and obsolescence to components.
{8}------------------------------------------------
| Table 1. Processor Device Comparison Table | | | |
|-------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Feature | Proposed<br>SYSTEM 1E Processor | Predicate (K210737)<br>SYSTEM 1E Processor | Comparison |
| Indications for<br>Use | The SYSTEM 1E Liquid<br>Chemical Sterilant Processing<br>System is intended for liquid<br>chemical sterilization of cleaned,<br>immersible, and reusable critical<br>and semi-critical heat-sensitive<br>medical devices in healthcare<br>facilities. | The SYSTEM 1E Liquid<br>Chemical Sterilant Processing<br>System is intended for liquid<br>chemical sterilization of cleaned,<br>immersible, and reusable critical<br>and semi-critical heat-sensitive<br>medical devices in healthcare<br>facilities. | Identical |
| | The SYSTEM 1E Processor<br>dilutes the S40 Sterilant<br>Concentrate to its use dilution<br>(>1820 mg/L peracetic acid),<br>liquid chemically sterilizes the<br>load during a controlled 6-<br>minute exposure at 45.5 to 60°C,<br>and rinses the load with<br>extensively treated* potable<br>water. After completion of a<br>cycle, critical devices should be<br>used immediately; semi-critical<br>devices should be used<br>immediately or may be handled<br>and stored in a manner similar to<br>that of high level disinfected<br>endoscopes. Critical devices not<br>used immediately should be<br>processed again before use. | The SYSTEM 1E Processor<br>dilutes the S40 Sterilant<br>Concentrate to its use dilution<br>(>1820 mg/L peracetic acid),<br>liquid chemically sterilizes the<br>load during a controlled 6-<br>minute exposure at 45.5 to 60°C,<br>and rinses the load with<br>extensively treated* potable<br>water. After completion of a<br>cycle, critical devices should be<br>used immediately; semi-critical<br>devices should be used<br>immediately or may be handled<br>and stored in a manner similar to<br>that of high level disinfected<br>endoscopes. Critical devices not<br>used immediately should be<br>processed again before use. | |
| | The SYSTEM 1E Processor uses<br>only S40 Sterilant Concentrate to<br>liquid chemically sterilize<br>medical devices. | The SYSTEM 1E Processor uses<br>only S40 Sterilant Concentrate to<br>liquid chemically sterilize<br>medical devices. | |
| | * The extensive treatment of<br>EPA potable water consists of:<br>1. Pre-filtration through two<br>pre-filters:<br>• Pre-filter A is a gross<br>depth filter that removes<br>approximately 2.5 micron<br>or larger<br>particles/contaminants.<br>• Pre-filter B is a surface<br>filter that removes<br>particles/contaminants ><br>0.1 micron. | * The extensive treatment of<br>EPA potable water consists of:<br>1. Pre-filtration through two pre-<br>filters:<br>• Pre-filter A is a gross depth<br>filter that removes<br>approximately 2.5 micron<br>or larger<br>particles/contaminants.<br>• Pre-filter B is a surface<br>filter that removes<br>particles/contaminants > 0.1 micron. | |
| Feature | Proposed<br>SYSTEM 1E Processor | Predicate (K210737)<br>SYSTEM 1E Processor | Comparison |
| | • During transit through the<br>UV water treatment<br>chamber, a UV dose<br>sufficient to achieve a ≥<br>6-log reduction of MS2<br>virus is delivered to the<br>water.<br>3. 0.1 micron filtration:<br>• The water prepared by<br>pre-filtration and UV<br>irradiation is filtered<br>through redundant, 0.1-<br>micron (absolute rated) | • During transit through the<br>UV water treatment<br>chamber, a UV dose<br>sufficient to achieve a ≥ 6-<br>log reduction of MS2 virus<br>is delivered to the water.<br>3. 0.1 micron filtration:<br>• The water prepared by pre-<br>filtration and UV irradiation<br>is filtered through<br>redundant, 0.1-micron<br>(absolute rated) membranes<br>to remove bacteria, fungi | |
| | membranes to remove<br>bacteria, fungi and<br>protozoa > 0.1 micron. | and protozoa > 0.1 micron. | |
| Operating<br>Principles/<br>Technology | A microprocessor controlled unit<br>with interchangeable processing<br>trays/containers. The processor<br>lid opens to reveal the<br>processing chamber in which the<br>load is placed. Devices with<br>internal lumens are interfaced<br>with the processor using<br>connectors. Sterilant<br>Concentrate is placed in a<br>specialized compartment and<br>when the processor fills with<br>water, it creates the sterilant use<br>dilution from the single use<br>sterilant cup. The processor<br>monitors and controls the use<br>dilution temperature and contact<br>time. The processor<br>automatically rinses the load<br>with extensively treated water to<br>remove sterilant residuals. | A microprocessor controlled unit<br>with interchangeable processing<br>trays/containers. The processor<br>lid opens to reveal the<br>processing chamber in which the<br>load is placed. Devices with<br>internal lumens are interfaced<br>with the processor using<br>connectors. Sterilant<br>Concentrate is placed in a<br>specialized compartment and<br>when the processor fills with<br>water, it creates the sterilant use<br>dilution from the single use<br>sterilant cup. The processor<br>monitors and controls the use<br>dilution temperature and contact<br>time. The processor<br>automatically rinses the load<br>with extensively treated water to<br>remove sterilant residuals. | Identical |
| Process<br>Parameters | Standardized cycle parameters<br>cannot be altered by operator.<br>The critical process parameters<br>are:<br>• Contact Time<br>• Use Dilution Temperature<br>• Peracetic acid concentration<br>• Bacterial retentive water<br>• filter integrity<br>• UV irradiation | Standardized cycle parameters<br>cannot be altered by operator.<br>The critical process parameters<br>are:<br>• Contact Time<br>• Use Dilution Temperature<br>• Peracetic acid concentration<br>• Bacterial retentive water<br>• filter integrity<br>• UV irradiation | Identical |
| Process<br>Monitors | • Cycle Printout documents<br>successful cycle completion | • Cycle Printout documents<br>successful cycle completion | Identical |
| Feature | Proposed<br>SYSTEM 1E Processor | Predicate (K210737)<br>SYSTEM 1E Processor | Comparison |
| | or identifies fault if cycle aborts Alarms if thermocouples indicate temperature out of specification Alarms if pressure switch indicates that high pressure pump is not operating Alarms if conductivity probe indicated conductivity specification not met Alarms if pressure transducer indicates circulation pressure is out of specification during Diagnostic cycle Alarms if pressure transducer indicates 0.1-micron water filter failed integrity test during liquid chemical sterilant processing and Diagnostic cycles. Alarms if UV monitor indicates UV intensity out of specification | or identifies fault if cycle aborts Alarms if thermocouples indicate temperature out of specification Alarms if pressure switch indicates that high pressure pump is not operating Alarms if conductivity probe indicated conductivity specification not met Alarms if pressure transducer indicates circulation pressure is out of specification during Diagnostic cycle Alarms if pressure transducer indicates 0.1-micron water filter failed integrity test during liquid chemical sterilant processing and Diagnostic cycles. Alarms if UV monitor indicates UV intensity out of specification | |
| Design<br>Features | Microprocessor controlled unalterable and standardized liquid chemical sterilant processing and Diagnostic cycles Intended for use with S40 Sterilant Concentrate Processor provides dual 0.1 micron filtered, UV treated water for liquid chemical sterilant processing and rinsing Automated dilution and delivery of sterilant Make up air for processor during drain sequences is filtered through a 0.2-micron membrane air filter. | Microprocessor controlled unalterable and standardized liquid chemical sterilant processing and Diagnostic cycles Intended for use with S40 Sterilant Concentrate Processor provides dual 0.1 micron filtered, UV treated water for liquid chemical sterilant processing and rinsing Automated dilution and delivery of sterilant Make up air for processor during drain sequences is filtered through a 0.2-micron membrane air filter. | Identical |
| Processing Cycle | | | Comparison |
| Incoming water<br>temperature | ≥ 43°C | ≥ 43°C | Identical |
| Feature | Proposed<br>SYSTEM 1E Processor | Predicate (K210737)<br>SYSTEM 1E Processor | Comparison |
| Temperature to<br>start exposure<br>phase | $\geq$ 46°C | $\geq$ 46°C | Identical |
| Temperature<br>alarm point<br>during the<br>exposure phase | <45.5 or >60°C | <45.5 or >60°C | Identical |
| Temperature<br>range during a<br>typical Liquid<br>Chemical<br>Sterilant<br>Processing<br>Cycle | 46 - 55°C | 46 - 55°C | Identical |
| Exposure<br>Time | 6 minutes | 6 minutes | Identical |
| Rinse water<br>preparation | Hot potable tap water is:<br>• pre-filtered<br>• flowed through a UV Light<br>treatment chamber to<br>achieve $\geq$ a 6-log reduction<br>of virus<br>• Filtered through redundant<br>0.1-micron filter<br>membranes | Hot potable tap water is:<br>• pre-filtered<br>• flowed through a UV Light<br>treatment chamber to<br>achieve $\geq$ a 6-log reduction<br>of virus<br>• Filtered through redundant<br>0.1-micron filter<br>membranes | Identical |
| Number of<br>rinses | 2 | 2 | Identical |
| Air Purge | Aids in removing excess water<br>from instrument lumens after<br>rinsing | Aids in removing excess water<br>from instrument lumens after<br>rinsing | Identical |
| Water Filter<br>Integrity Test | Conducted at the end of every<br>liquid chemical sterilant<br>processing cycle and during the<br>Diagnostic cycle | Conducted at the end of every<br>liquid chemical sterilant<br>processing cycle and during the<br>Diagnostic cycle | Identical |
| Approximate<br>Cycle time | 25 minutes | 25 minutes | Identical |
| Diagnostic<br>Cycle | Performs 15 tests on processor's<br>systems confirming proper<br>function (same tests as predicate<br>device except for an added UV<br>monitor test). Recommended to<br>perform every 24 hours. After a<br>failed Diagnostic cycle, a liquid<br>chemical sterilant processing<br>cycle cannot be performed until<br>the problem is rectified and a<br>successful Diagnostic cycle has<br>been completed. | Performs 15 tests on processor's<br>systems confirming proper<br>function (same tests as predicate<br>device except for an added UV<br>monitor test). Recommended to<br>perform every 24 hours. After a<br>failed Diagnostic cycle, a liquid<br>chemical sterilant processing<br>cycle cannot be performed until<br>the problem is rectified and a<br>successful Diagnostic cycle has<br>been completed. | Identical |
| Interchangeable Components | | | |
| Feature | Proposed<br>SYSTEM 1E Processor | Predicate (K210737)<br>SYSTEM 1E Processor | Comparison |
| Processing<br>Tray /<br>Containers | Uses interchangeable processing<br>trays/containers<br>• Universal Flexible Processing<br>Tray (C1160E)<br>• General Processing<br>Container/Tray (C1200)<br>• Directed Flow Processing<br>Container/Tray (C1220)<br>• Flexible Endoscope<br>Processing Container / Tray<br>(C1140)<br>• Ultrasound Processing Tray<br>(C3000XL) | Uses interchangeable processing<br>trays/containers<br>• Universal Flexible Processing<br>Tray (C1160E)<br>• General Processing<br>Container/Tray (C1200)<br>• Directed Flow Processing<br>Container/Tray (C1220)<br>• Flexible Endoscope<br>Processing Container / Tray<br>(C1140)<br>• Ultrasound Processing Tray<br>(C3000XL) | Identical |
| | Accessories | | |
| Sterilant<br>Concentrate | Uses S40 Sterilant Concentrate | Uses S40 Sterilant Concentrate | Identical |
| Quick<br>Connects | Uses Quick Connects to adapt<br>instrument lumens to the<br>Tray/Container ports | Uses Quick Connects to adapt<br>instrument lumens to the<br>Tray/Container ports | Identical |
| Chemical<br>Indicator | VERIFY Chemical Indicator for<br>the S40 Sterilant Concentrate | VERIFY Chemical Indicator for<br>the S40 Sterilant Concentrate | Identical |
| Spore Test<br>Strip | VERIFY Spore Test Strip for<br>S40 Sterilant Concentrate | VERIFY Spore Test Strip for<br>S40 Sterilant Concentrate | Identical |
| Operator<br>Maintenance<br>Requirements | Periodic replacement of printer<br>tape, water filters and air filter | Periodic replacement of printer<br>tape, water filters and air filter | Identical |
Table 1. Processor Device Comparison Table
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
# Table 2. S40 Sterilant Concentrate Device Comparison Table
| Feature | Proposed Device<br>S40 Sterilant Concentrate | Predicate Device<br>S40 Sterilant Concentrate<br>(K210737) | Comparison |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for<br>Use | The SYSTEM 1E Processor uses<br>only S40 Sterilant Concentrate to<br>liquid chemically sterilize<br>medical devices. | The SYSTEM 1E Processor uses<br>only S40 Sterilant Concentrate to<br>liquid chemically sterilize<br>medical devices. | Identical |
| Germicidal<br>claim | Liquid Chemical Sterilant | Liquid Chemical Sterilant | Identical |
| Germicide<br>Exposure<br>Time (min) for<br>intended use | 6 | 6 | Identical |
| Use<br>Temperature | 45.5-60°C – allowable<br>46-55°C - typical<br>Potency and simulated use<br>evaluations conducted at <43°C | 45.5-60°C – allowable<br>46-55°C - typical<br>Potency and simulated use<br>evaluations conducted at <43°C | Identical |
| Reuse | Single use | Single use | Identical |
| Feature | Proposed Device<br>S40 Sterilant Concentrate | Predicate Device<br>S40 Sterilant Concentrate<br>(K210737) | Comparison |
| Human<br>Factors | Dispensed ready to use.<br>Container is opened and diluted<br>by the processor, thus limiting<br>user exposure to the active<br>ingredient…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.