← Product Code [JAK](/productcode/JAK) · K211591

# NAEOTOM Alpha, Scan&GO (K211591)

_Siemens Medical Solutions USA, Inc. · JAK · Sep 30, 2021 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K211591

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [JAK](/productcode/JAK.md)
- **Decision Date:** Sep 30, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. This CT system can be used for low dose lung cancer screening in high risk populations * * As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

## Device Story

NAEOTOM Alpha is a dual-source CT scanner utilizing photon-counting detector technology (Cadmium-Telluride) to generate cross-sectional images. It processes x-ray transmission data to produce DICOM images for diagnosis, treatment planning, and radiation therapy. Operated by trained staff in clinical settings, it features an optional 'Scan&GO' mobile application on tablet hardware for in-room scan planning and control. The system employs iterative reconstruction (Quantum Iterative Reconstruction) and spectral imaging capabilities. Output images assist physicians in clinical decision-making, including calcium scoring and motion-artifact-reduced cardiac imaging. Benefits include improved image quality, potential for lower radiation dose, and enhanced workflow efficiency via mobile control.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including phantom evaluations, image quality analysis, and functional verification of hardware/software modifications. Performance metrics for new features (e.g., Quantum Iterative Reconstruction, CARE keV) were assessed via phantom measurements comparing results to predicate device performance.

## Technological Characteristics

Dual-source CT scanner; QuantaMax photon-counting detectors (CdTe); Vectron X-ray tubes; 2x 120 kW generator; 82 cm gantry bore; 0.2 mm slice width. Connectivity: DICOM, networked. Software: SOMARIS/10 syngo CT VA40 (Windows-based). Cybersecurity: IT hardening. Standards: IEC 60601-2-44, IEC 60601-1-2, NEMA PS 3.1-3.20. Sterilization: N/A (non-invasive).

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- SOMATOM Force ([K190578](/device/K190578.md))
- SOMATOM X.cite, Scan&GO ([K200524](/device/K200524.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

September 30, 2021

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized depiction of human figures. To the right of it is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Siemens Medical Solutions USA, Inc. % Tabitha Estes Regulatory Affairs Specialist 810 Innovation Drive KNOXVILLE TN 37932

Re: K211591

Trade/Device Name: NAEOTOM Alpha, Scan&GO Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: August 19, 2021 Received: August 20, 2021

Dear Tabitha Estes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

2

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

, for

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

### K211591

Device Name NAEOTOM Alpha Scan&GO

#### Indications for Use (Describe) NAEOTOM Alpha

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning.

This CT system can be used for low dose lung cancer screening in high risk populations *

* As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

## Scan&GO

This in-room scan application is a planning and information system the necessary functions required for planning and controlling scans of supported SIEMENS CT scanners. It allows users to work in close proximity to the scanner.

The in-room scan application runs on standard information technology hardware and software, utilizing the standard information technology operating systems and user interface. Communication and data exchange are done using special protocols.

Type of Use (Select one or both, as applicable)

| <div> <span style="font-size:10pt"> <span style="font-family:Wingdings">☒</span> </span>     Prescription Use (Part 21 CFR 801 Subpart D)    </div> | <div> <span style="font-size:10pt"> <span style="font-family:Wingdings">□</span> </span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </div> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K211591

# 510(K) SUMMARY FOR NAEOTOM ALPHA CT SCANNER AND SCAN&GO Submitted by:

Siemens Medical Solutions USA, Inc. 810 Innovation Drive Knoxville, TN 37932 Date Prepared: February 21, 2021

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

#### I. Submitter

Siemens Medical Solutions USA, Inc. 810 Innovation Drive Knoxville, TN 37932 Establishment Registration Number 1034973

#### Importer/Distributor

Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Establishment Registration Number 2240869

#### Location of Manufacturing Site (1)

Siemens Healthcare GmbH Siemensstr. 1 D-91301 Forchheim, Germany Establishment Registration Number 3004977335

Location of Manufacturing Site (2) SIEMENS SHANGHAI, MEDICAL EQUIPMENT LTD 278 Zhou Zhu Rd Shanghai, CHINA, 201318 Establishment Registration Number: 3003202425

Note: Description in this submission use the short company name Siemens. It covers both manufacturing locations and names as listed above. Brand name on all products is Siemens Healthineers.

## Contact Person:

Tabitha Estes Requlatory Affairs Siemens Medical Solutions USA, Inc. (865) 804-4553 (work cell) (865) 218-3019 FAX

{4}------------------------------------------------

# II. Device Name and Classification

| Product Name:                  | NAEOTOM Alpha                               |
|--------------------------------|---------------------------------------------|
| Trade Name:                    | NAEOTOM Alpha                               |
| Classification Name:           | Computed Tomography X-ray System            |
| Classification Panel:          | Radiology                                   |
| Regulation Number:             | 21 CFR §892.1750                            |
| Device Class:                  | Class II                                    |
| Product Code:                  | JAK                                         |
| Product Name:                  | Scan&GO                                     |
| Propriety Trade Name:          | Scan&GO                                     |
| Classification Name:           | Computed Tomography X-ray System            |
| Secondary Classification Name: | Picture Archiving and Communications System |
| Classification Panel:          | Radiology                                   |
| CFR Section:                   | 21 CFR §892.1750                            |
| Secondary CFR Section:         | 21 CFR §892.2050                            |
| Device Class:                  | Class II                                    |
| Product Code:                  | JAK                                         |
| Secondary Product Code:        | LLZ                                         |

## III. Predicate Device

Primary Predicate Device: SOMATOM Force Trade Name:

| 510(k) Number:        | K190578                                                                            |
|-----------------------|------------------------------------------------------------------------------------|
| Clearance Date:       | June 27, 2019                                                                      |
| Classification Name:  | Computed Tomography X-ray System                                                   |
| Classification Panel: | Radiology                                                                          |
| Regulation Number:    | 21 CFR §892.1750                                                                   |
| Device Class:         | Class II                                                                           |
| Product Code:         | JAK                                                                                |
| Recall Information:   | All predicate device recalls have been considered in the subject device<br>design. |

#### Secondary Predicate Device:

Trade Name:

SOMATOM X.cite, Scan&GO

| 510(k) Number:        | K200524                                                                            |
|-----------------------|------------------------------------------------------------------------------------|
| Clearance Date:       | April 01, 2020                                                                     |
| Classification Name:  | Computed Tomography X-ray System                                                   |
| Classification Panel: | Radiology                                                                          |
| Regulation Number:    | 21 CFR §892.1750                                                                   |
| Device Class:         | Class II                                                                           |
| Product Code:         | JAK                                                                                |
| Recall Information:   | All predicate device recalls have been considered in the subject device<br>design. |

Note: K200524 was a bundle submission with various Siemens CT Scanner Systems, including SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim, SOMATOM go.Open Pro, SOMATOM X.cite and Scan&GO software.

In this submission, the primary predicate device SOMATOM Force and the secondary predicate device SOMATOM X.cite including Scan&GO are being used, to demonstrate substantial equivalence of technological characteristics.

{5}------------------------------------------------

# IV. Device Description

Siemens intends to market a new CT scanner system NAEOTOM Alpha supporting software version, SOMARIS/10 syngo CT VA40 with mobile workflow options.

## Dual Source CT Scanner System:

- NAEOTOM Alpha
- Scan&GO Mobile Medical Application (optional mobile workflow component) .

The subject device NAEOTOM Alpha with SOMARIS/10 syngo CT VA40 is a dual-source Computed Tomography (CT) x-ray system featuring two detectors based on new photon counting technology. The CT scanner system algorithm is designed to allow image reconstruction by using photon counting data generated by the subject device. The reconstruction results are comparable with the predicate devices, but support with improved technological characteristics.

The NAEOTOM Alpha with Software SOMARIS/10 synqo CT VA40 produces CT images in DICOM format, which can be used by trained staff for post-processing applications commercially distributed by Siemens and other vendors as an aid in diagnosis, treatment preparation and therapy planning support (including, but not limited to, Brachytherapy, Particle including Proton Therapy, External Beam Radiation Therapy, Surgery). The computer system delivered with the CT scanner is able to run optional post processing applications.

The Scan&GO mobile workflow is an optional planning and information software designed to perform the necessary functions required for planning and controlling of the NAEOTOM Alpha. Scan&GO can be operated on a Siemens provided various tablet hardware that meets certain minimum technical requirements.

NOTE: Scan&GO does not support storage of images. Additionally, Scan&GO cannot trigger a scan or radiation release.

The software version for the NAEOTOM Alpha, syngo CT VA40 (SOMARIS/10 syngo CT VA40), is a command-based program used for patient management, data manaqement, X-ray scan control, image reconstruction, and image archive/evaluation.

The software platform provides a software plugin interface that allows for the use of specific commercially available post processing software algorithms in an unmodified form from the cleared stand-alone post processing version.

New software version syngo CT VA40 (SOMARIS/10 syngo CT VA40) is a modified software version based on syngo CT VA30A (SOMARIS/10 syngo CT VA30) which was cleared for the secondary predicate device and supports the same plugin interfaces for the subject device Scan&GO mobile workflow and integration of post-processing tasks as the secondary predicate device Scan&GO cleared in (K200524).

## V. Indications for Use

## NAEOTOM Alpha:

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning.

This CT system can be used for low dose lung cancer screening in high risk populations *

*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

{6}------------------------------------------------

## Scan&GO:

This in-room scan application is a planning and information system designed to perform the necessary functions required for planning and controlling scans of supported SIEMENS CT scanners. It allows users to work in close proximity to the scanner.

The in-room scan application runs on standard information technology hardware and software, utilizing the standard information technology operating systems and user interface. Communication and data exchange are done using special protocols.

## VI. Comparison of Technological Characteristics with the Predicate Device

The NAEOTOM Alpha scanner provides comparable technological characteristics in terms of materials, energy source, and control mechanisms when compared to the primary predicate device SQMATOM Force. The software and hardware components of this scanner have been modified or improved in comparison to the predicate devices to support enhanced device functionality and a to introduce a new detector technology.

The subject device NAEOTOM Alpha is building on the dual-source CT system configuration of the primary predicate device SOMATOM Force.

NAEOTOM Alpha features comparable technological characteristics of a Siemens dual-source CT system, such as fast scan speed and temporal resolution down to 66ms. It uses the same general geometric dimensions of the measurement field (such as detector z coverage and scan-field-of-view) in comparison to the primary predicate device SOMATOM Force CT scanner system.

As a dual source CT scanner system, the primary predicate device SOMATOM Force features two conventional energy-integrating detectors made of scintillator ceramic (product marketing name "Ultra Fast Ceramic", abbreviated UFC, based on scintillator ceramic). The subject device NAEOTOM Alpha features two photon counting detectors built from Cadmium-Telluride (CdTe).

Supported by the subject device, SOMARIS/10 syngo CT VA40 software version is a further development of the SOMARIS/10 syngo CT VA30 software version which is cleared in K200524. It reuses all unmodified software features of the legacy software syngo CT VA30 as described below in the section "Unmodified Features used in the Subject Devices".

As with the secondary predicate device SOMATOM X.cite, the subject device NAEOTOM Alpha is supported by an optional mobile workflow, Scan&GO has been modified to add new iPad hardware and operating software that supports the application software installation.

Software version SOMARIS/10 synqo CT VA40 is designed to reuse hardware independent extended functionalities and GO technologies provided by Siemens cleared software applications. The intended use and fundamental scientific technology for the NAEOTOM Alpha remains unchanged from the cleared predicate devices.

At a high level, the subject device NAEOTOM Alpha and the primary predicate device SOMATOM Force or the secondary predicate device SOMATOM X.cite are based on the same subset of technological elements:

- Scanner Principle- Whole body X-Ray Computed Tomography Scanner .
- System Acquisition Continuously rotating tube detector system ●
- Iterative Reconstruction Support of various iterative reconstruction principles ●
- Workplaces – Support of workplaces that include reconstruction and image evaluation software
- Patient table ●
- Patient table foot switch for movement
- Tin filtration technology ●
- Vectron X-ray Tube
- . Power Generator
- Mobile Medical application Software functionality (Scan&GO) .
- Mobile workflow (Tablet) .
- Support 3D Camera operation for fast patient positioning workflow

{7}------------------------------------------------

- Scanner display and control functionality ●
- Remote Scan Control ●
- Long scan range

### The following technological differences exist between the subject device NAEOTOM Alpha and the primary predicate device SOMATOM Force or the secondary predicate device SOMATOM X.cite:

- Software version SOMARIS/10 syngo CT VA40
- Support updated cybersecurity features
- Additional options for Inline and GO technologies ●
- Iterative Reconstruction Methods .
- . QuantaMax Detector based on Quantum Technology

The following technological differences exist between the subject device Scan&GO mobile application software and the secondary predicate device Scan&GO mobile application software cleared in K200524:

- Software version SOMARIS/10 syngo CT VA40
- iPad hardware to support mobile workflow options .

The NAEOTOM Alpha will support the following modifications/further developments in comparison to the predicate devices as listed in the tables below.

The configuration table and comparison table use the following terms to describe various technological characteristics in comparison to the predicate device information:

| Table 01: Overview term definition |                                                                                                                                                                                                                                                                                                              |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Term                               | Definition                                                                                                                                                                                                                                                                                                   |
| N/A                                | The feature is not supported for the subject device                                                                                                                                                                                                                                                          |
| New                                | The feature is newly supported for Siemens CT Scanners and the subject device                                                                                                                                                                                                                                |
| Modified                           | This feature is modified from the predicate devices                                                                                                                                                                                                                                                          |
| Unmodified                         | This feature remains unchanged from the predicate device                                                                                                                                                                                                                                                     |
| enabled                            | This feature is currently supported by other cleared Siemens CT systems or cleared<br>Siemens stand-alone software applications. This feature will be supported for the subject<br>device with software version SOMARIS/10 syngo CT VA40 and is substantially equivalent<br>compared to the cleared version. |

#### New/modified hardware features:

Table 02: Overview of hardware modifications of NAEOTOM Alpha

|    |                                                           | Subject Device |
|----|-----------------------------------------------------------|----------------|
|    | Hardware properties supported by SOMARIS/10 syngo CT VA40 | NAEOTOM Alpha  |
| 01 | Detector - QuantaMax                                      | new            |
| 02 | Tablet hardware for Scan&GO - iPad                        | modified       |
| 03 | Patient Table Configuration                               | modified       |

## New software version SOMARIS/10 syngo CT VA40 with enabled, modified and new software features:

Table 03: Overview of software modifications of NAEOTOM Alpha

|    |                                                                | Subject Device |
|----|----------------------------------------------------------------|----------------|
|    | Software properties supported by SOMARIS/10 syngo CT VA40      | NAEOTOM Alpha  |
| 01 | Precision Matrix (10242) (large image matrices)                | enabled        |
| 02 | Quantum Iterative Reconstruction                               | new            |
| 03 | Always Dual Energy                                             | new            |
| 04 | Calcium scoring with keV images                                | modified       |
| 05 | Imaging - Cardio BestPhase                                     | enabled        |
| 06 | CARE keV                                                       | new            |
| 07 | Quantum Pure Lumen                                             | new            |
| 08 | Cardiac CT imaging - Motion artifact reduced ECG-gated imaging | modified       |
| 09 | Motion artifact reduced non-gated imaging                      | modified       |

{8}------------------------------------------------

|    |                                                           | Subject Device |
|----|-----------------------------------------------------------|----------------|
|    | Software properties supported by SOMARIS/10 syngo CT VA40 | NAEOTOM Alpha  |
| 10 | myExam Console and myExam Satellite                       | modified       |
| 11 | Recon&GO Inline Results - DE SPP                          | modified       |
| 12 | Recon&GO Spectral Recon                                   | modified       |

A tabular summary of the comparable hardware and software properties between the subject device NAEOTOM Alpha with software version syngo CT VA40 and predicate devices are listed in Table 04 and Table 05 below (only the changes from the predicate devices are highlighted in gray in the sections below sections).

| Table 04: NAEOTOM Alpha comparable hardware properties (modified) |                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                          |
|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Hardware<br>properties                                            | subject device<br>NAEOTOM Alpha<br>Dual Source                                                                                                                                                                                                   | primary predicate device<br>SOMATOM Force<br>Dual Source                                                                                                                                                                                                                                                                                                                            | secondary predicate device<br>SOMATOM X.cite<br>Single Source                                                                                                                                                                                                                                                                                                            |
| Scanner                                                           | whole body X-ray computed<br>tomography scanner                                                                                                                                                                                                  | whole body X-ray computed<br>tomography scanner                                                                                                                                                                                                                                                                                                                                     | whole body X-ray computed<br>tomography scanner                                                                                                                                                                                                                                                                                                                          |
| Generator<br>max. power                                           | 2x 120 kW                                                                                                                                                                                                                                        | 2x 120 kW                                                                                                                                                                                                                                                                                                                                                                           | 105 kW                                                                                                                                                                                                                                                                                                                                                                   |
| Detector<br>technology                                            | Direct Conversion with<br>"Quantum Technology"                                                                                                                                                                                                   | Energy-integrating                                                                                                                                                                                                                                                                                                                                                                  | Energy-integrating                                                                                                                                                                                                                                                                                                                                                       |
| Detector<br>volume coverage                                       | 2 x 57.6 mm                                                                                                                                                                                                                                      | 2 x 57.6 mm                                                                                                                                                                                                                                                                                                                                                                         | 38.4 mm                                                                                                                                                                                                                                                                                                                                                                  |
| Detector<br>physical rows                                         | 2 x 288                                                                                                                                                                                                                                          | 2 x 96                                                                                                                                                                                                                                                                                                                                                                              | 64                                                                                                                                                                                                                                                                                                                                                                       |
| Detector<br>slice width                                           | 0.2 mm                                                                                                                                                                                                                                           | 0.6 mm                                                                                                                                                                                                                                                                                                                                                                              | 0.6 mm                                                                                                                                                                                                                                                                                                                                                                   |
| Detector<br>channel No.                                           | 2752 (A system)<br>1984 (B system)                                                                                                                                                                                                               | 920 (A system)<br>640 (B system)                                                                                                                                                                                                                                                                                                                                                    | 840                                                                                                                                                                                                                                                                                                                                                                      |
| X-ray Tube                                                        | Vectron                                                                                                                                                                                                                                          | Vectron                                                                                                                                                                                                                                                                                                                                                                             | Vectron                                                                                                                                                                                                                                                                                                                                                                  |
| Tube<br>kV steps                                                  | 90 kV, 100 kV,<br>120 kV, 140 kV                                                                                                                                                                                                                 | (in 10kV steps)<br>70 kV to 150kV                                                                                                                                                                                                                                                                                                                                                   | (in 10kV steps)<br>70 kV to 150kV                                                                                                                                                                                                                                                                                                                                        |
| Tube<br>max. current                                              | 2 x 1300 mA                                                                                                                                                                                                                                      | 2 x 1300 mA                                                                                                                                                                                                                                                                                                                                                                         | 1200 mA                                                                                                                                                                                                                                                                                                                                                                  |
| Tube<br>tube focus                                                | $0.4×0.5$ mm / $0.6x0.7$ mm /<br>$0.8x1.1$ mm<br>(for both tubes)                                                                                                                                                                                | $0.4×0.5$ mm / $0.6×0.7$ mm /<br>$0.8x1.1$ mm<br>(for both tubes)                                                                                                                                                                                                                                                                                                                   | $0.6x0.7$ mm / $0.8x1.1$ mm                                                                                                                                                                                                                                                                                                                                              |
| Tube<br>heat capacity                                             | higher than 30 MHU                                                                                                                                                                                                                               | higher than 30 MHU                                                                                                                                                                                                                                                                                                                                                                  | higher than 30 MHU                                                                                                                                                                                                                                                                                                                                                       |
| Tube<br>Collimator                                                | adaptive dose shied<br>supported                                                                                                                                                                                                                 | adaptive dose shied<br>supported                                                                                                                                                                                                                                                                                                                                                    | adaptive dose shied<br>supported                                                                                                                                                                                                                                                                                                                                         |
| Gantry<br>bore size                                               | 82 cm                                                                                                                                                                                                                                            | 78 cm                                                                                                                                                                                                                                                                                                                                                                               | 82 cm                                                                                                                                                                                                                                                                                                                                                                    |
| Gantry<br>Scan FoV                                                | 50 cm                                                                                                                                                                                                                                            | 50 cm                                                                                                                                                                                                                                                                                                                                                                               | 50 cm                                                                                                                                                                                                                                                                                                                                                                    |
| Gantry<br>rotation time (sec)                                     | 0.25s; 0.5s; 1.0s;                                                                                                                                                                                                                               | 0.25s, 0.285s, 0.33s, 0.5s,<br>1.0s                                                                                                                                                                                                                                                                                                                                                 | 0.3s, 0.5s, 1.0s                                                                                                                                                                                                                                                                                                                                                         |
| Gantry<br>Tilt (degree)                                           | N/A                                                                                                                                                                                                                                              | N/A                                                                                                                                                                                                                                                                                                                                                                                 | +/- 25                                                                                                                                                                                                                                                                                                                                                                   |
| Patient Table<br>type                                             | Vario 2.D:<br>2000 mm,<br>Vitus:<br>2000 mm                                                                                                                                                                                                      | PHS5, MPT4:<br>1600 mm or 2000 mm                                                                                                                                                                                                                                                                                                                                                   | Vario RT: 1600 mm,<br>Vario 2: 2000 mm                                                                                                                                                                                                                                                                                                                                   |
| Max. Scan length<br>Topogram                                      | Vario 2.D:<br>2080 mm,<br>Vitus:<br>2080 mm                                                                                                                                                                                                      | 1600 mm, 2000 mm                                                                                                                                                                                                                                                                                                                                                                    | 1680 mm, 2080 mm                                                                                                                                                                                                                                                                                                                                                         |
| Max. Scan length<br>Image acquisition                             | Vario 2.D:<br>2000 mm,<br>Vitus:<br>2000 mm                                                                                                                                                                                                      | 1600 mm, 2000 mm                                                                                                                                                                                                                                                                                                                                                                    | 1600 mm, 2000 mm                                                                                                                                                                                                                                                                                                                                                         |
| Spectral filtration<br>Option*                                    | Tin Filter for both tubes                                                                                                                                                                                                                        | Tin Filter for both tubes                                                                                                                                                                                                                                                                                                                                                           | Combined Split Filter / Tin<br>Filter supported, plus extra<br>Tin Filter                                                                                                                                                                                                                                                                                                |
| 3D Camera<br>for patient positioning                              | option for patient positioning<br>with 3D Camera                                                                                                                                                                                                 | option for patient positioning<br>with 3D Camera                                                                                                                                                                                                                                                                                                                                    | option for patient positioning<br>with 3D Camera                                                                                                                                                                                                                                                                                                                         |
| Properties<br>software                                            | subject device<br>NAEOTOM Alpha<br>(syngo CT VA40 )                                                                                                                                                                                              | primary predicate device<br>SOMATOM Force<br>(K190578) (syngo CT VB20)                                                                                                                                                                                                                                                                                                              | secondary predicate device<br>SOMATOM X.cite<br>(K200524)                                                                                                                                                                                                                                                                                                                |
| Operating<br>System                                               | Windows based<br>SOMARIS/10<br>syngo CT VA40                                                                                                                                                                                                     | Windows based<br>SOMARIS/7<br>syngo CT VB20                                                                                                                                                                                                                                                                                                                                         | Windows based<br>SOMARIS/10<br>syngo CT VA30 A                                                                                                                                                                                                                                                                                                                           |
| Acquisition<br>Workplace                                          | syngo Acquisition Workplace (AWP)<br>syngo Viewing, syngo Filming<br>and syngo Archiving &<br>Networking<br>2nd Acquisition Workplace<br>supported with myExam<br>Satellite                                                                      | syngo Acquisition Workplace (AWP)<br>syngo Viewing, syngo Filming<br>and syngo Archiving &<br>Networking<br>2nd Acquisition Workplace:<br>RRWP                                                                                                                                                                                                                                      | syngo Acquisition Workplace (AWP)<br>syngo Viewing, syngo Filming<br>and syngo Archiving &<br>Networking<br>No 2nd Acquisition Workplace<br>supported                                                                                                                                                                                                                    |
| IRS                                                               | Image Reconstruction from<br>photon counting data                                                                                                                                                                                                | Image Reconstruction from<br>classic Siemens dual source<br>CT scanner                                                                                                                                                                                                                                                                                                              | Image Reconstruction from<br>classic Siemens single source<br>CT scanner                                                                                                                                                                                                                                                                                                 |
| Detector                                                          | QuantaMax detector firmware<br>supported                                                                                                                                                                                                         | Stellar detector firmware<br>supported                                                                                                                                                                                                                                                                                                                                              | Stellar detector firmware<br>supported                                                                                                                                                                                                                                                                                                                                   |
| Teamplay                                                          | Support teamplay Protocols<br>Support of:                                                                                                                                                                                                        | Support teamplay Protocols<br>Support of:                                                                                                                                                                                                                                                                                                                                           | Support teamplay Protocols<br>Support of:                                                                                                                                                                                                                                                                                                                                |
| Protocols                                                         | Protocol supporting<br>contrast bolus-triggered<br>data acquisition Contrast media protocols Pediatric Protocols Flex Dose Profile Turbo Flash Spiral Dual Energy acquisition                                                                    | Protocols for Radiation<br>Therapy Planning Protocol supporting<br>contrast bolus-triggered<br>data acquisition Contrast protocols – CARE<br>Contrast III Pediatric Protocols Flex Dose Profile Turbo Flash Spiral Dual Energy acquisition Adaptive 4D Spiral Protocols that allow<br>scanning with support of<br>an 3rd party respiratory<br>gating system (ANZAI,<br>Varian RGSC) | Protocols for Radiation<br>Therapy Planning support<br>patient marking Protocol supporting<br>contrast bolus-triggered<br>data acquisition Contrast media protocols Pediatric Protocols Flex Dose Profile TwinBeam DE TwinSpiral DE Flex 4D Spiral Protocols that allow<br>scanning with support of<br>an 3rd party respiratory<br>gating system (ANZAI,<br>Varian RGSC) |
| Advanced<br>Reconstruction                                        | Recon&GO –<br>- Spectral Recon (Dual Energy<br>Reconstruction from photon<br>counting data) / including<br>Virtual Unenhanced,<br>Monoenergetic plus<br>- Inline Results DE SPP<br>(Spectral Post-Processing with<br>photon counting image data) | Advanced reconstruction tools<br>supported:<br>The syngo acquisition<br>workplace provides, image<br>reconstruction, and routine<br>postprocessing. Various<br>advanced reconstruction<br>features supported by the CT<br>Scanner, e.g., FAST and CARE<br>applications.                                                                                                             | Advanced reconstruction tools<br>supported via Recon&GO –<br>- Spectral Recon (Dual Energy<br>Reconstruction)<br>- Inline Results DE SPP<br>(Spectral Post-Processing) /<br>including<br>Virtual Unenhanced,<br>Monoenergetic plus<br>- Inline Results DE Ranges<br>(Parallel/Radial) / Inline DE                                                                        |
| Post-<br>Processing                                               | Software Plugin functions<br>enabled via software interface<br>Recon&GO - Inline Results<br>various methods of cleared<br>software applications                                                                                                  | syngo.via - Wide Range of<br>individual applications,<br>syngo.via is a software solution<br>intended to be used for<br>viewing, manipulation,<br>communication, and storage of<br>medical images. It can be used<br>as a standalone device or                                                                                                                                      | Software Plugin functions<br>enabled via software interface<br>Recon&GO - Inline Results<br>various methods of cleared<br>software applications                                                                                                                                                                                                                          |
|                                                                   | subject device                                                                                                                                                                                                                                   | primary predicate device                                                                                                                                                                                                                                                                                                                                                            | secondary predicate device                                                                                                                                                                                                                                                                                                                                               |
| Properties<br>software                                            | NAEOTOM Alpha<br>(syngo CT VA40)                                                                                                                                                                                                                 | SOMATOM Force<br>(K190578) (syngo CT VB20)                                                                                                                                                                                                                                                                                                                                          | SOMATOM X.cite<br>(K200524)                                                                                                                                                                                                                                                                                                                                              |
|                                                                   |                                                                                                                                                                                                                                                  | cleared and unmodified syngo based software options.                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                          |
| Cybersecurity                                                     | IT Hardening                                                                                                                                                                                                                                     | IT Hardening                                                                                                                                                                                                                                                                                                                                                                        | IT Hardening                                                                                                                                                                                                                                                                                                                                                             |
| Standard<br>Technologies                                          | FAST Features CARE Features GO technology CARE keV                                                                                                                                                                                               | FAST Features CARE Features CARE kV                                                                                                                                                                                                                                                                                                                                                 | FAST Features CARE Features GO technology CARE kV                                                                                                                                                                                                                                                                                                                        |
| Iterative<br>Reconstruction<br>Methods                            | Quantum Iterative<br>Reconstruction                                                                                                                                                                                                              | iMAR, SAFIRE, ADMIRE                                                                                                                                                                                                                                                                                                                                                                | iMAR, ADMIRE                                                                                                                                                                                                                                                                                                                                                             |
| Precision<br>Matrix                                               | Precision Matrix resolution<br>support image matrix sizes of<br>512x512, 768x768 pixels and<br>1024x1024 pixel (auto mode<br>supported)                                                                                                          | Precision Matrix resolution<br>support image matrix sizes of<br>512x512, 768x768 pixels and<br>1024x1024 pixel (auto mode<br>supported)                                                                                                                                                                                                                                             | Matrix resolution 512x512<br>(auto mode not supported)                                                                                                                                                                                                                                                                                                                   |
| Ca Scoring                                                        | Photon counting technology<br>offers monoenergetic images<br>which can be used as a base<br>for calcium scoring<br>independent from tube voltage<br>kV and beam filtration settings.                                                             | Classic SOMATOM Force<br>technology offers images<br>which can be used for calcium<br>scoring, independent from<br>tube voltage kV and beam<br>filtration settings based on a<br>dedicated single energy image<br>reconstruction method.                                                                                                                                            | Classic SOMATOM X.cite<br>technology offers images<br>which can be used for calcium<br>scoring, independent from<br>tube voltage kV and beam<br>filtration settings based on a<br>dedicated single energy image<br>reconstruction method.                                                                                                                                |

Table 04: NAEOTOM Alpha comparable hardware properties (modified)

{9}------------------------------------------------

Table 05: NAEOTOM Alpha comparable software properties

{10}------------------------------------------------

Any differences in technological characteristics do not raise different questions of safety and effectiveness. Siemens believes that the subject device is substantially equivalent to the predicate devices. Testing and validation are completed. Test results show that the subject device, the NAEOTOM Alpha, is comparable to the predicate devices in terms of technological characteristics and safety and effectiveness and therefore is substantially equivalent to the primary predicate SOMATOM Force and secondary predicate device SOMATOM X.cite.

# VII. Performance Data

## Non Clinical Testing

Non-clinical testing, (integration and functional) including phantom tests were conducted for the NAEOTOM Alpha during product development. The modifications described in this Premarket Notification were supported with verification and validation testing.

The general purpose of each tests is to verify and validate the functionality of the subject device modifications.

Testing will cover all related subsystems that contribute to the device modifications. Test levels are defined. For each test level several test activities are performed. The test specification and acceptance criteria are related to the corresponding requirements. Various test activities are performed to specific modifications on different test levels to ensure safe and effective integration in the system. Three test levels are defined:

System Validation test:

- Acceptance test (workflow and user manual test) .
- Legal and Regulatory test ●

System Verification test:

- System Integration Test (functional) ●
- Functionality verification ●
- Image Quality (IQ) Evaluation ●

{11}------------------------------------------------

Tests are conducted for all software components developed in product development and for the complete product itself. Several activities are considered for this process, including creation of test specifications that relate to software/hardware requirements including tests to address risk mitigations that are identified, documented and traced by hazard keys.

Additional evaluation tests are performed as bench tests to support the new device or device modification on Non-Clinical Performance Testing as listed in table 07 below.

| #  | Feature                                                        | Non-Clinical Performance Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|----|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 01 | Quantum Iterative Reconstruction                               | Quantum Iterative Reconstruction - QIR is routinely used in all reconstructions on the subject device NAEOTOM Alpha.<br>As with the primary predicate device SOMATOM Force supporting ADMIRE, the subject device supports iterative reconstruction principle.<br>NAEOTOM Alpha supports Quantum Iterative Reconstruction with image data generated with Quantum Technology (QuantaMax Detector) and optimized for photon counting data.<br>Iterative Beam Hardening Correction - IBHC is an iterative Beam Hardening Correction algorithm optimized for the subject device NAEOTOM Alpha.<br>Performance testing was done to support all technological characteristics, that are associated with this new technology.                                                                                                                                                                                                         |
| 02 | Detector - QuantaMax                                           | The subject device NAEOTOM Alpha supports two "QuantaMax" detectors for CT data acquisition. The QuantaMax detectors are built on the new "Quantum Technology" which is Siemens Healthineers' implementation of photon counting technology for CT detectors.<br>The performed tests include the measurement and analysis of raw data for single photon counting, the nonexistence of classical noise and the ability to measure multiple energies simultaneously.<br>Furthermore, the performance evaluation provides an in-depth evaluation of NAETOM Alpha Image Quality for general CT imaging, based on phantom evaluation of Typical Modes, compared to the predicate device SOMATOM Force. It also includes parameters for supporting the suitability of the subject device for low dose lung cancer screening.                                                                                                         |
| 05 | Cardiac CT imaging - Motion artifact reduced ECG-gated imaging | As with the primary predicate device the subject device supports ECG-gated imaging based on ECG-controlled dose modulation and triggering derived from continuous monitoring of the ECG. With same technological principle as cleared with the primary predicate device, the subject device supports with an unmodified algorithm, the desired prediction of the target start- and stop phase in retrospective spiral and in sequence mode. The prediction of the start phase in ECG triggered high pitch scanning is modified to improve the prediction over a wider range of heart rates.<br>The associated test evidence contains a basic test scenario and the results of measurements based on an andromorphic motion heart phantom to support the clinical approach based on technical phantom measurements, which are to some extend clinically realistic.                                                             |
| 06 | CARE keV                                                       | CARE kV is supported by the primary predicate device SOMATOM Force. The subject device NAEOTOM Alpha based on Quantum Technology supports CARE keV. CARE keV takes the dose efficiency of Monoenergetic reconstructions into account, by automatically adjusting the appropriate kV and effective mAs settings to optimize the applied dose while the image quality is maintained. The test procedure includes phantom measurements with clinically relevant phantom diameters and contrast materials to support the contrast, noise, and radiation dose related CARE keV information.                                                                                                                                                                                                                                                                                                                                        |
| 07 | Quantum Pure Lumen                                             | The subject device NAEOTOM Alpha supports with its new feature "VCR - Pure Lumen" a modified algorithm, which allows to reconstruc…

---

**Source:** [https://fda.innolitics.com/device/K211591](https://fda.innolitics.com/device/K211591)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
