K211531 · Guilin Woodpecker Medical Instrument Co., Ltd. · EKX · Dec 17, 2021 · Dental
Device Facts
Record ID
K211531
Device Name
Cordless Prophy System, Model: i-Polish
Applicant
Guilin Woodpecker Medical Instrument Co., Ltd.
Product Code
EKX · Dental
Decision Date
Dec 17, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic, Pediatric
Indications for Use
i-Polish is a cordless prophylaxis handpiece equipped with control buttons and wireless foot control for use with disposable prophylaxis angles in hygiene operatory to perform cleaning and polishing procedures on teeth surface and fillings.
Device Story
Cordless, battery-powered dental handpiece; used by dental professionals in hygiene operatory for cleaning/polishing teeth and fillings. System includes handpiece, removable/autoclavable outer sheath, AC-powered charging station, wireless foot control, and AC adapter. Operates via two modes: foot control (variable speed 500-4000 rpm) or handpiece-integrated centralized control button (six constant speeds: 500, 1000, 1500, 2000, 3000, 4000 rpm). Handpiece requires use with FDA-cleared disposable prophy angles and disposable sleeves. Provides rotary motion to prophy angles; features auto-off after 5 minutes of inactivity. Benefits include improved ergonomics and portability for dental hygiene procedures.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), wireless coexistence (ANSI/IEEE C63.27), battery safety (IEC 62133-2), and biocompatibility (ISO 10993-5, ISO 10993-10).
Indicated for dental professionals to perform cleaning and polishing procedures on teeth surfaces and fillings for patients of all ages in a professional dental operatory.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
Predicate Devices
Young INFINITY Cordless Handpiece System (K171377)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 17, 2021
Guilin Woodpecker Medical Instrument Co., Ltd. % Yoyo Chen Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town, Nanshan District Shenzhen, Guangdong 518100 China
Re: K211531
Trade/Device Name: Cordless Prophy System, Model: i-Polish Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EKX Dated: November 17, 2021 Received: November 29, 2021
Dear Yoyo Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K211531
Device Name Cordless Prophy System, Model: i-Polish
Indications for Use (Describe)
i-Polish is a cordless prophylaxis handpiece equipped with control buttons and wireless foot control for use with disposable prophylaxis angles in hygiene operatory to perform cleaning and polishing procedures on teeth surface and fillings.
| Type of Use (Select one or both, as applicable) | <table border="0"><tr><td><table border="0"><tr><td>☑ Remediation Under 21 CFR 1271 Subpart D</td></tr></table></td><td><table border="0"><tr><td>☐ Use That Is Subject to 21 CFR 1271 Subpart C</td></tr></table></td></tr></table> | <table border="0"><tr><td>☑ Remediation Under 21 CFR 1271 Subpart D</td></tr></table> | ☑ Remediation Under 21 CFR 1271 Subpart D | <table border="0"><tr><td>☐ Use That Is Subject to 21 CFR 1271 Subpart C</td></tr></table> | ☐ Use That Is Subject to 21 CFR 1271 Subpart C |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|------------------------------------------------|--------------------------------------------------------------------------------------------|------------------------------------------------|
| <table border="0"><tr><td>☑ Remediation Under 21 CFR 1271 Subpart D</td></tr></table> | ☑ Remediation Under 21 CFR 1271 Subpart D | <table border="0"><tr><td>☐ Use That Is Subject to 21 CFR 1271 Subpart C</td></tr></table> | ☐ Use That Is Subject to 21 CFR 1271 Subpart C | | |
| ☑ Remediation Under 21 CFR 1271 Subpart D | | | | | |
| ☐ Use That Is Subject to 21 CFR 1271 Subpart C | | | | | |
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(K) information is submitted As Required by requirements of SMDA and 21 CFR §807.92.
### 1. Administrative Information
| Submission Date | November 17, 2021 |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information | Guilin Woodpecker Medical Instrument Co., Ltd.<br>Address:<br>Information Industrial Park, Guilin National High-Tech Zone,<br>Guilin, Guangxi, 541004, P.R. China<br>Contact person: Ning Jiakang (宁加康)<br>TEL: +86-773-2350532<br>FAX:+86-773-2350532<br>E-Mail: zmnbg03@glwoodpecker.com |
| Submission<br>Correspondent | Shenzhen Joyantech Consulting Co., Ltd.<br>1713A, 17th Floor, Block A, Zhongguan Times Square, Liuxian<br>Avenue, Xili Town, Nanshan District, Shenzhen, Guangdong<br>Province, China.<br>Contact person: Ms. Yoyo Chen<br>E-Mail: yoyo@cefda.com; field@cefda.com<br>Image: Logo |
| Establishment<br>registration number | 3005581016 |
# 2. Device Information
| Type of 510(k)<br>Submission: | Traditional |
|-------------------------------|-------------------------------------|
| 510 (k) number | K211531 |
| Device Name: | Cordless Prophy System |
| Model: | i-Polish |
| Classification Name: | Handpiece, Direct Drive, Ac-Powered |
| Review Panel: | Dental |
| Device Class: | 1 |
| Regulation Number: | 872.4200 |
| Product Code: | EKX |
{4}------------------------------------------------
### 3. Primary Predicate Device
| Manufacturer | Young Dental Manufacturing Co 1, LLC |
|------------------------|------------------------------------------|
| Device name | Young INFINITY Cordless Handpiece System |
| 510(K) Number: | K171377 |
| Regulation Description | Dental handpiece and accessories |
| Product Code | EKX |
### 4. Reference Device
| Manufacturer | Parkell Products Inc. |
|------------------------|----------------------------------|
| Device name | Low-Speed Prophy Handpiece |
| 510(K) Number: | K983413 |
| Regulation Description | Dental handpiece and accessories |
| Product Code | EKX |
## 5. Device Description
The Cordless Prophy system is a cordless handpiece which is intended for use by dental professionals for cleaning and polishing teeth. This is a general hygiene procedure that is performed on people of all ages in a professional dental operatory.
The Cordless Prophy System is comprised of a cordless, battery-powered handpiece, a removable and autoclavable outer sheath, an AC powered battery charging station, a battery-powered wireless foot control and an AC adapter. Accessories to the Cordless Prophy System include three style of Disposable Prophy Angle (DPA), which is cleared as Class I, Product code ESG, under premarket notification K030603.
Additionally, the handpiece of Cordless Prophy System must be used with a Disposable Sleeve, which is cleared as Class II, Product Code PEM, under premarket notification K151123.
The handpiece features a removable outer sheath that is to be cleaned and steam sterilized prior to first use and after each patient use.
The i-Polish has two speed control mode for operation:
- . Foot control mode: One is using it with the wireless foot control, where the amount of vertical actuation on the wireless foot control correlates to the speed of the handpieces supplied to the DPA, the corresponding variable speed range of the DPA is controlled and adjusted through varying pressure on the foot control. The adjustable range is 500 rpm to 4000 rpm;
- . Handpiece control mode: And the other uses a Centralized control button located on the handpieces for six constant speed level, 500 rpm, 1500 rpm, 1500 rpm, 2000 rpm,
{5}------------------------------------------------
3000 rpm and 4000 rpm.
### 6. Intended use/Indication for use
i-Polish is a cordless prophylaxis handpiece equipped with control buttons and wireless foot control for use with disposable prophylaxis angles in hygiene operatory to perform cleaning and polishing procedures on teeth surface and fillings.
## 7. Comparison with Predicate Device
| Items | Subject Device<br>(K211531) | Predicate Device<br>(K171377) | Conclusion |
|------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Product Code | EKX | EKX | Same |
| Class | 1 | 1 | Same |
| Regulation number | 872.4200 | 872.4200 | Same |
| Intended use/Indication<br>for use | i-Polish is a cordless prophylaxis<br>handpiece equipped with control<br>buttons and wireless foot control<br>for use with disposable<br>prophylaxis angles in hygiene<br>operatory to perform cleaning and<br>polishing procedures on teeth<br>surface and fillings. | Battery driven electrical drive<br>unit with wireless foot<br>controller for use with<br>disposable prophylaxis angles<br>in hygiene operatory to<br>perform cleaning and polishing<br>of tooth surfaces and fillings. | Same |
| Use | Rx Only | Rx Only | Same |
| Handpiece geometry | Cylindrical shape with reverse<br>radius geometry to aid in device<br>handling. Tapered, swiveled<br>nosecone area. | Cylindrical shape with reverse<br>radius geometry to aid in<br>device handling. Tapered,<br>swiveled nosecone area. | Same |
| Handpiece power | Lithium-ion Battery capable of<br>being recharged multiple<br>times by inclusion of an AC/DC<br>power supply. | Lithium-ion Battery capable of<br>being recharged multiple<br>times by inclusion of an<br>AC/DC power supply. | Same |
| Foot control power | Foot control contains a Lithium-<br>ion battery capable of being<br>recharged multiple times by<br>inclusion of<br>AC/DC power supply. | Foot control contains a<br>Lithium-ion battery capable of<br>being recharged multiple times<br>by inclusion of<br>AC/DC power supply. | Same |
| Charge time | Handpiece: Approximately 2.5<br>hours<br>Foot control: Approximately 2<br>hours | Handpiece: Approximately 2<br>hours<br>Foot control: Approximately 3<br>hours | Different<br>(Note 1) |
| Handpiece Dimension | 27.6mm Dia × 192mm | 25mm Dia × 156mm | Different<br>(Note 2) |
| Items | Subject Device (K211531) | Predicate Device (K171377) | Conclusion |
| Prophy Angle Fit | The device could be fit with<br>Disposable prophy angle which<br>cleared under premarket notification<br>K030603. | Similar to most corded<br>handpieces on the market<br>today, our device will have a<br>Doriot style nose which allows<br>most prophy angles to be<br>used on the device. | Different<br>(Note 3) |
| Nose Cone<br>(Also called as Outer<br>Sheath in the subject<br>device) | Outer sheath will swivel, the<br>Outer sheath will also be<br>removable and autoclavable for<br>infection control. | Nose cone will swivel. The<br>nosecone will also be<br>removable and autoclavable<br>for infection control. | Same |
| Infection<br>Control/Sterilization<br>Method | Outer sheath is to be to cleaned<br>and sterilized prior to first use and<br>after each patient. And the<br>handpiece is to be covered with<br>an FDA cleared Disposable<br>Sleeve which is cleared as Class<br>II, Product Code PEM, under<br>premarket notification K151123. | Remove nosecone and<br>sterilize via autoclave.<br>The nosecose is to be cleaned<br>and sterilized prior to first use<br>and after each patient.<br>Handpiece is to be covered<br>with an FDA cleared<br>Disposable Sleeve | Same |
| Lubrication Method | Lubricant Free Motor. Do not use<br>Lubrication | Lubricant Free Motor. Do not<br>use Lubrication | Same |
| User Interface on<br>Handpiece | That handpiece will have a<br>Centralized control button, which<br>will enable connection to the foot<br>control for activation. | That handpiece will have a<br>power button, which will enable<br>connection to the foot pedal for<br>activation. | Same |
| Auto-off | The foot control and handpiece will<br>automatically shut down if the<br>standby time exceeds 5 minutes.<br>The user would then have to press<br>the Centralized control button to<br>activate the handpiece again. | Handpiece will enter a standby<br>mode if idle for more than 4<br>minutes. The user would then<br>have to press the POWER<br>button to activate the<br>handpiece again. | Similar |
| Mode of Operation | Rotary | Rotary | Same |
| Speed Control | i-Polish has two speed control<br>mode for operation.<br><br><b>Handpiece control mode:</b><br>Speed is controlled and adjusted<br>through press the Centralized<br>control button to achieve six<br>constant speed level, 500 rpm,<br>1000 rpm, 1500 rpm, 2000 rpm,<br>3000 rpm and 4000 rpm. | Speed is controlled and<br>adjusted through varying<br>pressure on the foot pedal.<br>The motor itself has a limit of<br>3000RPM (±10%). | Different<br>(Note 4) |
| Items | Subject Device<br>(K211531) | Predicate Device<br>(K171377) | Conclusion |
| | Foot control mode:<br>Speed is controlled and adjusted<br>through varying pressure on the<br>foot control. The adjustable range<br>is 500 rpm to 4000 rpm. | | |
| Speed Range (±10%) | 500~4000 RPM | 500-3000 RPM | Different<br>(Note 4) |
| Maximum Torque<br>(±10%) | 1.2Ncm | 1Ncm | Similar |
| Operating environment | Ambient temperature: +5°C ~<br>+40°C<br>Relative humidity: 30% ~ 75%<br>Atmospheric pressure: 70kPa<br>~106kPa | Ambient temperature: +10°C ~<br>+35°C;<br>Relative humidity: 15% ~ 80% | Different<br>(Note 5) |
| Transport and Storage<br>Condition | Ambient temperature: -20°C ~<br>+55°C<br>Relative humidity: 10% ~ 93%<br>Atmospheric pressure: 70kPa<br>~106kPa | Ambient temperature: -20°C ~<br>+60°C;<br>Relative humidity: 8% ~ 80% | Different<br>(Note 5) |
| Compliance Standards | IEC 60601-1;<br>IEC 60601-1-2;<br>ISO 10993-5.<br>ISO 10993-10.<br>ISO 14457: | IEC 60601-1;<br>IEC 60601-1-2;<br>ISO 10993-5.<br>ISO 10993-10.<br>ISO 14457: | Same |
{6}------------------------------------------------
{7}------------------------------------------------
### Note 1: Charge time
Although the charge time of the subject device is different with predicate device, but the battery is complied with the IEC62133-2:2017 standard, Otherwise, the subject device has been demonstrated to comply with the requirements of electrical safety IEC 60601-1:2015. This difference will not raise any new safety and effectiveness issues.
### Note 2: Handpiece Dimension
The handpiece dimension has a little bit difference with predicate device, but both products allow for similar interaction with the user. Otherwise, the dimension of subject device has been demonstrated to comply with the requirements of ISO14457-2017standard. This difference will not raise any new safety and effectiveness issues.
### Note 3: Prophy Angle Fit
The subject device only could be fit with Disposable prophy angle which cleared under premarket notification K030603. The prophy angle fit does not impact the user experience during cleaning and polishing procedures. This difference will not raise any new safety and effectiveness issues.
{8}------------------------------------------------
#### Note 4: Speed Control, Speed Range
Although the speed of subject device could be controlled and adjusted through press the Centralized control button and varying pressure on the foot pedal. But both devices deliver similar torque and speed profiles.
The low-end speed limit(500RPM) is the same. The top speed (4000RPM) is much higher than the predicate device, but the top speed is still lower than a reference device K983413 (the claimed top speed is able to operate up to 5000 RPM). In addition, a comparison test carried out between subject device and predicate device to demonstrate the difference on the top speed will not arise new safety and effectiveness issues.
Otherwise, the subject device has been demonstrated to comply with the requirements of IEC 60601-1, ISO 80601-2-60, and IEC60601-1-2 standard requirement. The difference will not raise any new safety and effectiveness issues.
#### Note 5: Operating environment and Transport and Storage Condition
The difference will not raise new safety or effectiveness issue, because, the subject device has tested to conform with the IEC 60601-1 standard.
### 8. Non-Clinical Test Summary
### 8.1. Electromagnetic Compatibility and Electrical Safety Test
The subject device has passed safety testing in according to following standards.
- IEC 60601-1:2005+AMD 1: 2012 Medical electrical equipment Part 1: General 1) requirements for basic safety and essential performance
- 2) IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 80601-2-60 Edition 2.0 2019-06 Medical electrical equipment Part 2-60: 3) Particular requirements for the basic safety and essential performance of dental equipment
- 4) ISO 14457: 2017 Dentistry - Handpieces and motors
- 5) ANSI/IEEE C63.27:2017 American National Standard for Evaluation of Wireless Coexistence
- 6) FCC Rules and Regulations, Part 15, Subpart C
- 7) The rechargeable lithium battery has passed the IEC 62133-2:2017 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary lithium cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems.
### 8.2. Biocompatibility Test
The subject device has passed safety testing in according to following standards.
- 1) ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro
{9}------------------------------------------------
cytotoxicity
- 2) ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
## 9. Software Validation
Software documentation consistent with moderate level of concern is submitted in this 510(k). System validation testing presented in this 510(k) demonstrates that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels
## 10. Clinical Data
Substantial equivalence does not depend on the clinical test data.
# 11. Conclusion
The subject device is substantially equivalent to the primary predicate device (K171377). This conclusion is based upon comparison on indication for use, technological characteristics, and applicable safety standards. Any difference in the technological characteristics does not raise any new issues or concerns of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.