Steinmann Pins and Kirschner Wires

K211508 · Stryker GmbH · JDW · Oct 5, 2021 · Orthopedic

Device Facts

Record IDK211508
Device NameSteinmann Pins and Kirschner Wires
ApplicantStryker GmbH
Product CodeJDW · Orthopedic
Decision DateOct 5, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Steinmann Pins are intended to be used in conjunction with a compatible Stryker external supporting frame for stabilization of open and/or unstable fractures and where soft tissue injury may preclude the use of other fracture treatments such as IM rodding, casting and other means of internal fixation. The Kirschner Wires are intended for use in skeletal traction for alignment and reduction of long bone fractures, and as guide wires in hip pinning, and as fracture fixation devices in certain other small bone fractures. • For use as guide wires in hip pinning procedures, • For use in aligning and reducing long bone fractures, • For use in securing temporary stabilization of bone fractures such as olecranon fractures; tibia plateau fractures; hand and foot bone fractures; humeral, radial and ulnar fractures. • For use with cerclage wire/cable in treating greater trochanter fractures.

Device Story

Steinmann Pins and Kirschner Wires are metallic bone fixation fasteners; used by orthopedic surgeons in clinical settings. Pins/wires inserted into bone to provide skeletal traction, fracture alignment, reduction, or temporary stabilization. Steinmann Pins used with external supporting frames for open/unstable fractures; Kirschner Wires used as guide wires for hip pinning or fixation for small bone fractures. Devices feature varying diameters, lengths, and tip geometries (trocar or diamond); some are threaded. Devices provide mechanical support to bone segments; aid in guiding cannulated instruments or evaluating screw trajectories. Single-use, sterile devices; facilitate fracture healing by maintaining bone position.

Clinical Evidence

No clinical data provided. Substantial equivalence demonstrated via bench testing and comparison to predicate devices. Packaging validated per ISO 11607-1 and ISO 11607-2; bench testing performed per ASTM standards.

Technological Characteristics

Material: Stainless steel (1.4441). Design: Smooth or threaded pins/wires with trocar or diamond tip geometries. Dimensions: Variable diameters and lengths. Sterilization: Radiation. Standards: ISO 5838-1 (implants for surgery), ISO 11607-1/2 (packaging).

Indications for Use

Indicated for patients requiring skeletal traction, fracture alignment, reduction, or temporary stabilization of long bone and small bone fractures (e.g., olecranon, tibia plateau, hand, foot, humeral, radial, ulnar, greater trochanter). Also indicated for use as guide wires in hip pinning procedures and with external fixation frames for open/unstable fractures where other internal fixation methods are precluded.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 5, 2021 Stryker GmbH Keith Neligan Senior Staff Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430 Re: K211508 Trade/Device Name: Steinmann Pins and Kirschner Wires Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HTY, JDW Dated: August 18, 2021 Received: August 24, 2021 Dear Keith Neligan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Brent L. Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K211508 Device Name Steinmann Pins Indications for Use (Describe) The Steinmann Pins are intended to be used in conjunction with a compatible Stryker external supporting frame for stabilization of open and/or unstable fractures and where soft tissue injury may preclude the use of other fracture treatments such as IM rodding, casting and other means of internal fixation. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K211508 Device Name Kirschner Wires ## Indications for Use (Describe) The Kirschner Wires are intended for use in skeletal traction for alignment and reduction of long bone fractures, and as guide wires in hip pinning, and as fracture fixation devices in certain other small bone fractures. · For use as guide wires in hip pinning procedures, · For use in aligning and reducing long bone fractures, · For use in securing temporary stabilization of bone fractures such as olecranon fractures; tibia plateau fractures; hand and foot bone fractures; humeral, radial and ulnar fractures. · For use with cerclage wire/cable in treating greater trochanter fractures. Type of Use (Select one or both, as applicable) | <div> <input checked="true" type="checkbox"/> <span>Prescription Use (Part 21 CFR 201 Subpart D)</span> </div> | |----------------------------------------------------------------------------------------------------------------| | <div> <input type="checkbox"/> <span>Over-The-Counter Use (Part 21 CFR 201 Subpart C)</span> </div> | > Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary | Proprietary Names: | Steinmann Pins<br>Kirschner Wires | |---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Pin, Fixation, Threaded; Pin, Fixation, Smooth | | Regulation Description: | Smooth or threaded metallic bone fixation fastener | | Regulation Number: | 21 CFR 888.3040 | | Classification Product Code: | JDW, HTY | | Device Class: | II | | Submitter: | Stryker GMBH<br>Bohnackerweg 1<br>2545 Selzach, Switzerland | | Contact Person: | Keith Neligan<br>Senior Staff Regulatory Affairs Specialist<br>IDA Business & Technology Park, Carrigtwohill,<br>Co. Cork, Ireland T45 HX08<br>Phone: +00 353 87 7995645 | | Date: | August 09, 2021 | | Primary Predicate Device | Apex Fixation Pins (K001886) | | Additional Predicate Device | Osteo Kirschner Wires (K971962) | | Device Description: | The Steinmann Pins and Kirschner Wires are designed for multiple purposes<br>which allow the use in skeletal traction for alignment and reduction of bone<br>fractures, function as fracture fixation devices in certain fractures, preliminary<br>fixation for implants (for example plates) and for evaluation of screw<br>trajectories or guidance for cannulated instruments.<br><br>The Steinmann Pins and Kirschner Wires are available in different dimensions<br>and provide several design features. The dimensions vary in diameter and<br>length. The design features are the tip geometry which can be a three-sided<br>trocar tip or a two-sided diamond tip on either one end or both ends of the pin.<br>Additionally, some pins are threaded.<br><br>The pins and wires are made from stainless steel (1.4441). | | Indications for Use: | The Steinmann Pins are intended to be used in conjunction with a compatible<br>Stryker external supporting frame for immobilization of open and/or unstable<br>fractures and where soft tissue injury may preclude the use of other fracture<br>treatments such as IM rodding, casting and other means of internal fixation. | | | The Kirschner Wires are intended for use in skeletal traction for alignment and<br>reduction of long bone fractures, and as guide wires in hip pinning, and as<br>fracture fixation devices in certain other small bone fractures.<br>• For use as guide wires in hip pinning procedures,<br>• For use in aligning and reducing long bone fractures,<br>• For use in securing temporary stabilization of bone fractures such as<br>olecranon fractures; patella fractures; tibia plateau fractures; small hand and<br>foot bone fractures; humeral, radial and ulnar fractures.<br>• For use with cerclage wire/cable in treating greater trochanter fractures. | | Summary of Technologies: | Steinmann Pins:<br>The intended use of the subject devices (Steinmann Pins) are similar to those<br>detailed in the predicate device (Apex Fixation Pins). There is no change in the<br>fundamental scientific technology shared by both the subject device and<br>predicate device. | | | Kirschner Wires<br>The intended use of the subject devices (Kirschner Wires) are similar to those<br>detailed in the predicate device (Osteo Kirschner Wires). There is no change in the<br>fundamental scientific technology shared by both the subject device and<br>predicate device. | | Performance Data (Nonclinical): | Non-Clinical Performance and Conclusions:<br>The Steinmann Pins and Kirschner wires are single use instruments, sterilized<br>by means of radiation; they are sterile stainless steel wires with diameter and<br>lengths comparable to those of the predicate devices. The design is<br>substantially equivalent to that of the predicate devices; the indications for use<br>are included in those of the predicate devices; the material of the Steinmann<br>Pins and Kirschner wire are the same as that of the predicate device. Moreover,<br>Steinmann Pins and Kirschner wires conform to the international standard ISO<br>5838-1. Testing, therefore, is not needed to demonstrate that the subject devices<br>are substantially equivalent to other legally marketed Steinmann Pins and<br>Kirschner wires. | | | Packaging tests were performed according ISO 11607-1 and ISO 11607-2. All<br>bench tests performed in accordance with ASTM standards. | | | Clinical Performance and Conclusions:<br>Clinical data and conclusions were not needed for these devices to demonstrate<br>substantial equivalence to the predicate devices. | | Conclusion: | The subject devices have the same intended use and similar indications for use<br>as the predicate devices. The subject devices use the same operating principle,<br>incorporate the same basic design and labeling and are manufactured and<br>sterilized using the same materials and processes as the predicate devices.<br>The performance data and analyses demonstrate that:<br>• any differences do not raise new questions of safety and effectiveness as<br>established with performance testing; and<br>• the subject devices are at least as safe and effective as the legally marketed<br>predicate devices | {5}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...