K211502 · Stryker Endoscopy · MBI · Aug 4, 2021 · Orthopedic
Device Facts
Record ID
K211502
Device Name
Alpha Vent PEEK Suture Anchor
Applicant
Stryker Endoscopy
Product Code
MBI · Orthopedic
Decision Date
Aug 4, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AlphaVent PEEK Suture Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures: Shoulder: Rotator Cuff Repair, Bankart Repair, Siceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. Foot and Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair and Bunionectomy. Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis and Quadriceps Tendon Repair. Secondary or adjunct fixation for ACL/PCL reconstruction or Repair. Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis Repair. Hip: Capsular Repair, Acetabular Labral Repair.
Device Story
Bone anchor system for soft tissue-to-bone fixation; screw-in PEEK anchor design; pre-loaded with UHMWPE suture; includes inserter for pilot hole placement; used by surgeons in clinical settings; provides mechanical fixation via interference fit; benefits patient by securing soft tissue during orthopedic repairs; sterile, single-use.
Clinical Evidence
Bench testing only. Evaluated ultimate tensile strength (UTS) and insertion performance. Results demonstrated pull-out strength similar or superior to predicate. Bacterial endotoxin testing met acceptance criteria.
Technological Characteristics
Materials: PEEK anchor, UHMWPE suture. Design: Cannulated, screw-in anchor with integrated soft eyelet. Sizes: 4.75mm, 5.5mm, 6.5mm. Sterilization: Ethylene Oxide (EO), SAL 10^-6. Biocompatibility: ISO 10993-1 compliant.
Indications for Use
Indicated for soft tissue-to-bone fixation in shoulder, foot/ankle, knee, hand/wrist, elbow, and hip orthopedic procedures.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 4, 2021
Stryker Endoscopy Christie Samsa Principal Regulatory Affairs Specialist 5900 Optical Ct. San Jose, California 95138
Re: K211502
Trade/Device Name: Alpha Vent PEEK Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: May 13, 2021 Received: May 14, 2021
Dear Christie Samsa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211502
#### Device Name AlphaVent PEEK Suture Anchors
#### Indications for Use (Describe)
The AlphaVent PEEK Suture Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:
Shoulder: Rotator Cuff Repair, Bankart Repair, Siceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
Foot and Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair and Bunionectomy.
Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis and Quadriceps Tendon Repair. Secondary or adjunct fixation for ACL/PCL reconstruction or Repair.
Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction,
Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis Repair.
Hip: Capsular Repair, Acetabular Labral Repair.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span style="font-size:12pt"> <b> <span style="font-family:Symbol">☒</span> </b> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:12pt"> <b> <span style="font-family:Symbol">☐</span> </b> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Stryker logo in black, with the text "K211502" above it. The logo is a stylized version of the company name, with bold, sans-serif letters. The text "K211502" is in a smaller font size and is centered above the logo. The logo and text are set against a white background.
# 510(k) Summary
## Submitter:
| Applicant | Stryker Endoscopy<br>5900 Optical Court<br>San Jose, CA 95138 |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Christie Samsa<br>Principal Regulatory Affairs Specialist<br>Phone: (978) 500-1303<br>Facsimile: (408) 754-2598<br>Email: christie.samsa@stryker.com |
| Date Prepared | May 13, 2021 |
#### Subject Device:
| Name of Device | AlphaVent PEEK Suture Anchor |
|----------------------|-----------------------------------------------------------------|
| Common or Usual Name | Suture, Fastener, Fixation, Nondegradable, Soft Tissue |
| Classification Name | Fastener, Fixation, Nondegradable, Soft Tissue, 21 CFR 888.3040 |
| Regulatory Class | Class II |
| Product Code | MBI |
### Predicate Device:
| Name of Device | Arthrex Swivelock Anchors, K191226 |
|----------------|------------------------------------|
|----------------|------------------------------------|
Note: The predicate device has not been subject to a design-related recall.
### Device Description:
The AlphaVent PEEK Suture Anchor (herein referred to as the proposed device(s)) are bone anchors with a screw-in design. Each anchor is vented and is provided pre-loaded with one or more working sutures with or without needles. The working sutures run through the cannulated anchor body and are attached to the distal end of the anchor by an integrated suture loop, referred to as a "soft eyelet." The anchor with working sutures is provided pre-assembled on an inserter, which enables insertion of the anchor into bone after creation of a pilot hole. The anchors are offered in a poly-ether-ether-keytone (PEEK) material with the integrated suture eyelet offered in non-absorbable USP braided ultra-high molecular weight polyethylene (UHMWPE) high strength suture. The devices are provided sterile and are packaged in sterile barrier systems (SBS) that include one anchor pre-loaded with suture on an inserter.
{4}------------------------------------------------
### Indications for Use:
The AlphaVent PEEK Suture Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
Foot and Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair/Tendon Repair and Bunionectomy.
Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis and Quadriceps Tendon Repair. Secondary or adjunct fixation for ACL/PCL reconstruction or Repair.
Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction.
Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis Repair.
Hip: Capsular Repair, Acetabular Labral Repair.
{5}------------------------------------------------
| | Subject Device<br>(This Submission) | Predicate Device<br>(K191226) |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Feature | AlphaVent PEEK Suture Anchor | Arthrex SwiveLock<br>Suture Anchor |
| Intended Use | Fixation of soft tissue to bone | Same as subject device |
| Indications for<br>Use | The AlphaVent PEEK Suture Anchors are intended<br>for fixation of suture (soft tissue) to bone in the<br>shoulder, foot/ankle, knee, hand/wrist, elbow, and<br>hip in the following procedures:<br><br>Shoulder: Rotator Cuff Repair, Bankart Repair,<br>SLAP Lesion Repair, Biceps Tenodesis, Acromio-<br>Clavicular Separation Repair, Deltoid Repair,<br>Capsular Shift or Capsulolabral Reconstruction.<br><br>Foot and Ankle: Lateral Stabilization, Medial<br>Stabilization, Achilles Tendon Repair, Hallux<br>Valgus Reconstruction, Midfoot Reconstruction,<br>Metatarsal Ligament Repair/Tendon Repair and<br>Bunionectomy.<br><br>Knee: Anterior Cruciate Ligament Repair, Medial<br>Collateral Ligament Repair, Lateral Collateral<br>Ligament Repair, Patellar Tendon Repair, Posterior<br>Oblique Ligament Repair, Illiotibial Band Tenodesis<br>and Quadriceps Tendon Repair. Secondary or<br>adjunct fixation for ACL/PCL reconstruction or<br>Repair.<br><br>Hand/Wrist: Scapholunate Ligament Reconstruction,<br>Ulnar or Radial Collateral Ligament Reconstruction.<br><br>Elbow: Biceps Tendon Reattachment, Tennis Elbow<br>Repair, Ulnar or Radial Collateral Ligament<br>Reconstruction, Lateral Epicondylitis Repair.<br><br>Hip: Capsular Repair, Acetabular Labral Repair. | Same as subject device |
| Operational<br>Principle | Fixation of suture to bone achieved by interference<br>fit with screw-in anchor, suture fixated by eyelet | Fixation of suture to bone<br>achieved by interference<br>fit with sutures and screw-<br>in anchor |
| Key Patient-<br>Contacting<br>Materials | Material: Poly-ether-ether ketone (PEEK),<br>UHMWPE | Same as subject device |
| Anchor System<br>Design | Screw in anchor with eyelet<br>Inserter<br>Sutures for fixation | Similar to subject device |
| Eyelet Design | Integrated Closed eyelet | Separate Closed eyelet |
| Feature | Subject Device<br>(This Submission) | Predicate Device<br>(K191226) |
| | AlphaVent PEEK Suture Anchor | Arthrex SwiveLock<br>Suture Anchor |
| Anchor<br>Diameter | 4.75mm<br>5.5mm<br>6.5mm | 3.5mm - 9mm |
| Sterilization<br>Method, SAL | Ethylene Oxide (EO),<br>SAL: 10-6 | Similar to subject device |
| Packaging | Double Barrier Blister or Pouch<br>Single Use Only | Similar to subject device |
| Biocompatibility | Device conforms to ISO 10993-1 for patient contact<br>nature and duration | Similar to subject device |
# Comparison of Technological Characteristics with the Predicate Device:
{6}------------------------------------------------
### Performance Testing:
Non-clinical benchtop testing was performed to evaluate the performance characteristics of the Stryker AlphaVent PEEK Suture Anchor, including ultimate tensile strength (UTS) and insertion testing. The proposed devices demonstrated similar or increased pull-out strength to the predicate devices, and no new issues of safety and effectiveness were identified. The shelf life of the device is 2 years. Additionally, bacterial endotoxin testing was performed and results of testing met the acceptance criteria.
#### Conclusions:
The AlphaVent PEEK Suture Anchor is substantially equivalent in design, intended use, principles of operation, technological characteristics and safety features to the predicate device. There are no new issues of safety and/or effectiveness introduced by the AlphaVent PEEK Suture Anchor when used as instructed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.