The Advantage-C PEEK Cervical Interbody Fusion Device is a cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disease (DDD) with / without radicular symptoms at one level from C2 - T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The Advantage-C PEEK Cervical Interbody Fusion Device is intended for use with supplemental fixation systems and with autogenous bone graft.. The Advantage-C PEEK Cervical Interbody Fusion Device is supplied sterile and is intended for one-time use.
Device Story
Advantage-C PEEK Cervical Interbody Fusion Device is a sterile, single-use implant for anterior cervical discectomy and fusion (ACDF) procedures. Device consists of polyetheretherketone (PEEK) cage with Tantalum markers for radiographic visualization. Implanted via anterior approach to provide supplemental stabilization of cervical spinal column. Used in conjunction with supplemental spinal fixation systems and autogenous bone graft to facilitate fusion. Available in various footprints, heights, and lordotic angles to accommodate patient anatomy. Surgeon selects appropriate size based on patient needs. Device acts as a spacer to maintain disc height and promote fusion between vertebral bodies.
Clinical Evidence
No clinical studies were performed. Safety and efficacy were established through non-clinical bench testing, including subsidence testing (ASTM F2267-04), monotonic and dynamic compressive/torsional strength (ASTM F2077-11), and wear particle analysis (ASTM F1877-16).
Technological Characteristics
Materials: PEEK (ASTM F2026) and Tantalum (ASTM F560). Form factor: Interbody cage with graft window and expulsion-resistant features. Dimensions: Heights 6-12mm, Widths 11-20mm, Lengths 11-15mm; Parallel (0°) and Lordotic (7°) configurations. Sterilization: Gamma radiation per ISO/AAMI/ANSI 11137-1:2006. Biocompatibility: ISO 10993-1:2018.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) with or without radicular symptoms at one level from C2-T1, confirmed by history and radiography, following six weeks of failed non-operative treatment. Intended for use with supplemental fixation and autogenous bone graft.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Zimmer-Biomet Vista®-S PEEK Cervical Cage System (K111983)
Zimmer-Biomet Vista®-S PEEK Cervical Cage System (K133784)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
August 24, 2021
Intelivation, LLC Jack Griffis Regulatory Affairs 70 Gruber Lane Saint Simons Island, Georgia 31522
Re: K211501
Trade/Device Name: Advantage-C PEEK Cervical Interbody Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: July 23, 2021 Received: July 26, 2021
Dear Jack Griffis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K211501
#### Device Name
Advantage-C PEEK Cervical Interbody Fusion Device
#### Indications for Use (Describe)
The Advantage-C PEEK Cervical Interbody Fusion Device is a cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disease (DDD) with / without radicular symptoms at one level from C2 - T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The Advantage-C PEEK Cervical Interbody Fusion Device is intended for use with supplemental fixation systems and with autogenous bone graft.. The Advantage-C PEEK Cervical Interbody Fusion Device is supplied sterile and is intended for one-time use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Intelivation Technologies. The logo features a lightbulb with a brain inside of it on the left side. To the right of the lightbulb is the word "INTELIVATION" in a stylized font. Below that is the word "TECHNOLOGIES" in a smaller font.
# 510(k) Summary
510(k) Number: K211501
July 23″d, 2021 Date Submitted:
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
- A. Submitter: Intelivation, LLC 70 Gruber Lane Saint Simons Island, GA 31522 912-434-2780
- B. Company Contact: Jack Griffis Regulatory Affairs (404) 583-6889 (direct) jackgriffisiii@gmail.com
- C. Device Information: Trade Name: Advantage-C PEEK Cervical Interbody Fusion Device Intervertebral Fusion Device, Cervical Common Name:
- D. Classification: Intervertebral Fusion Device 21 CFR §888.3080 (ODP) Class II
- E. Predicate Device:
Zimmer-Biomet Vista®-S PEEK Cervical Cage System, K111983 (Primary) Zimmer-Biomet Vista®-S PEEK Cervical Cage System, K133784
- F. Reference Predicate Device: Zavation Interbody Fusion System (IBF), K162206
- G. Physical Description:
The proposed Intelivation Advantage-C Cervical Interbody Fusion Device is a sterile, single use implant grade polyetheretherketone (PEEK; ASTM F2026) device and Tantalum (ASTM F560), available in varied footprints and heights, designed for supplemental stabilization of the cervical spinal column in anterior cervical discectomy and fusion (ACDF) procedures. The device is intended for spinal fusion procedures at one level in skeletally mature patients with degenerative disc disease (defined as pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. The implant should be used with supplemental spinal fixation systems and with autogenous bone graft.
- H. Indications for Use: The Advantage-C Cervical Interbody Fusion Device is a cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with / without radicular symptoms at one level from C2 – T1. DDD is
{4}------------------------------------------------
defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The Advantage-C Cervical Interbody Fusion Device is intended for use with supplemental fixation systems and with autogenous bone graft. The Advantage-C Cervical Interbody Fusion Device is implanted via an anterior approach.
- l. Comparison to Predicate Device:
The Intelivation Advantage-C PEEK Cervical Interbody Fusion Device is substantially equivalent with respect to intended use and technological characteristics to the Zimmer Vista-S, cleared under K111983 & 133784 and the Zavation Interbody Fusion (IBF) System cleared under K162206. These devices all operate under the same principle of operation and with a similar form factor (PEEK cage with graft window and superior and inferior features to resist expulsion). The devices differ in geometry only, and that the proposed Advantage-C device is provided packaged and pre-sterilized via gamma irradiation. Intelivation asserts that any differences between the Advantage-C and the predicate devices do not affect safety or efficacy.
| Parameter | Advantage-C<br>(Intelivation Proposed) | Zimmer Vista-S<br>(Predicate; K111983 & K133784) | Zavation IBF System<br>(Reference Predicate, K162206) |
|-----------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Manufacturer | Intelivation LLC | Zimmer | Zavation |
| 510(k) Number | Unassigned | K111983 & K133784 | K162206 |
| Product Code | ODP | ODP | ODP |
| Material | Magnolia® PEEK<br>Tantalum Markers | Optima® PEEK<br>Tantalum Markers | Magnolia® PEEK<br>Tantalum Markers |
| Sizes | Heights: 6mm - 12mm<br>Widths: 11mm - 20mm<br>Lengths: 11mm - 15mm<br>Parallel (0°) and Lordotic (7°) | Heights: 4mm - 12mm<br>Widths: 11mm - 20mm<br>Lengths: 11mm - 15mm<br>Parallel (0°) and Lordotic (7°) | Heights: 5mm - 12mm<br>Widths: 14mm - 16mm<br>Lengths: 12mm - 14mm<br>Parallel (0°) and Lordotic (6° & 10°) |
| Supplied Sterile? | Yes | No | Yes |
| Sterilization<br>method | Gamma radiation | N/A | Gamma radiation |
| Intended for<br>single use? | Yes | Yes | Yes |
| Packaging<br>Configuration | Tyvek-poly pouch with an outer<br>shelf carton | Poly bag | Tyvek-poly pouch with an outer<br>shelf carton |
| Shelf Life | 12 months | >12 months | >12 months |
| | | | Table 1. Table of Substantial Equivalence |
|--|--|--|-------------------------------------------|
|--|--|--|-------------------------------------------|
In addition, the proposed device was subjected to the following performance tests to support the assertion of substantial equivalence:
- J. Summary of Non-Clinical Tests:
Product characterization using known standards and/or clinically relevant acceptance criteria was performed on the proposed device. A summary of this testing is provided in Table 2 below.
Image /page/4/Picture/9 description: The image shows the logo for Intelivation Technologies. The logo features a lightbulb with a brain inside of it on the left side. To the right of the lightbulb is the company name, "INTELIVATION", in a gradient color from green to blue. Below the company name is the word "TECHNOLOGIES" in a smaller font size.
{5}------------------------------------------------
# Table 2. Non-Clinical Testing Information
| Requirement |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subsidence Testing:<br>◦ Subsidence characterization as defined in ASTM F2267-04 |
| Monotonic Compressive and Torsional Strength:<br>◦ Yield and/or ultimate force and moment characterization as defined in ASTM F2077-11 - PASS |
| Dynamic Compressive and Torsional Strength:<br>◦ Maximum runout ( $5\times10^6$ cycles) force or overall fatigue resistance as defined in ASTM F2077-11 - PASS |
| Compressive and Torsional Wear Particle Analysis:<br>◦ Wear particle size, shape, and distribution within the reported range of PEEK wear particles as defined in ASTM F1877-16 - PASS |
- K. Biocompatibility Testing:
Biocompatibility of the predicate device has been established in accordance with ISO 10993-1:2018 – Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process to demonstrate that the device is safe for permanent contact (>30 days) implantation. Refer to MAF #2735.
- L. Sterilization:
The method employed to ensure a sterility assurance level of 10th of the proposed device is provided in Table 3. The sterilization process is identical to the reference predicate device.
| Test | Test Method Summary | Results |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Sterilization<br>validation | ISO/AAMI/ANSI 11137-1:2006 Sterilization of health care products. Radiation. Requirements<br>for development, validation and routine control of a sterilization process for medical devices | Pass |
| LAL | ANSI/AAMI ST72:2019, USP<161>, USP<85>, EP 2.6.14 and JP 4.01 | Pass |
# Table 3. Sterilization Information
#### M. Clinical Studies:
No human studies were necessary to prove the safety and efficacy of the device.
- N. Conclusion:
The subject device demonstrated equivalent mechanical performance to the cited predicate under the same test conditions.
Jack Griffis Regulatory Affairs (404) 583-6889 (direct) jackgriffisiii@gmail.com
Image /page/5/Picture/15 description: The image shows the logo for Intelivation Technologies. The logo features a lightbulb with a brain inside of it on the left side. To the right of the lightbulb is the company name, "INTELIVATION" in a thin, sans-serif font, with the word "TECHNOLOGIES" in a smaller font below it.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.