K211460 · Diode Art Engineering Doing Business AS Air Relax · IRP · May 25, 2021 · Physical Medicine
Device Facts
Record ID
K211460
Device Name
Air Relax Plus Model AR-3.0
Applicant
Diode Art Engineering Doing Business AS Air Relax
Product Code
IRP · Physical Medicine
Decision Date
May 25, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Air Relax Plus Model AR-3.0 is intended for the temporary relief of minor muscle aches and for temporary increase in circulation to the treated areas in people who are in good health. The Air Relax Plus Model AR-3.0 simulates kneading and stroking of tissues by using an inflatable garment.
Device Story
Powered inflatable tube massager; consists of air compressor unit, control system, and inflatable garments (arms, legs, hips). Inputs: user-selected treatment time and pressure intensity (60-170 mmHg) via front panel buttons. Operation: compressor inflates garment segments in one of three sequences (Progressive, Sequential, Full) to simulate manual kneading/stroking. Used in clinics, hospitals, athlete training, and home environments by patients or clinicians. Output: mechanical compression of limbs/hips. Benefits: temporary muscle ache relief and increased circulation. Proprietary keyed connectors prevent misconnection of tubing/garments.
Clinical Evidence
No clinical testing was performed. Bench testing only, including control/indicator testing, performance verification, and compliance with AAMI/ANSI ES60601-1, IEC 60601-1-2, and IEC 60601-1-11.
Technological Characteristics
Powered inflatable tube massager; molded ABS enclosure; nylon with polyurethane laminate garments. Four inflatable segments. Pressure range 60-170 mmHg. Power: 100-240V, 50/60Hz. Type BF applied part. Proprietary keyed connectors. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11.
Indications for Use
Indicated for adults in good health for temporary relief of minor muscle aches and pains and temporary increase in circulation to treated areas.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
May 25, 2021
Diode Art Engineering doing business as Air Relax % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K211460
Trade/Device Name: Air Relax Plus Model AR-3.0 Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager Regulatory Class: Class II Product Code: IRP Dated: May 10, 2021 Received: May 11, 2021
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Vivek Pinto, PhD Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211460
Device Name Air Relax Plus Model AR-3.0
#### Indications for Use (Describe)
The Air Relax Plus Model AR-3.0 is intended for the temporary relief of minor muscle aches and for temporary increase in circulation to the treated areas in people who are in good health. The Air Relax Plus Model AR-3.0 simulates kneading and stroking of tissues by using an inflatable garment.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| | 510(k) Summary<br>K211460 |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | April 29, 2021 |
| Applicant | Diode Art Engineering doing business as Air Relax<br>9535 Brasher St<br>Pico Rivera, CA 90660<br>Tel – 1.323.285.4231 |
| Official Contact: | Beomjoon Lee, General Manager |
| Proprietary or Trade Name: | Air Relax Plus Model AR-3.0 |
| Common/Usual Name: | Powered Inflatable Tube Massager |
| Classification Name: | IRP - Massager Powered Inflatable Tube (CFR 890.5650 |
| Predicate Devices: | K191441 – Air Relax Model AR-1.0/AR-2.0 |
#### Device Description:
This submission is for the Diode Art Engineering Air Relax Plus Model AR-3.0. The Air Relax Plus Model AR-3.0 is a powered inflatable tube massager. It is intended to temporarily relieve minor muscle aches and pains, and to temporarily increase circulation to the treated areas. It simulates manual kneading and stroking of tissues by use of an inflatable pressure cuff. The device is to be used by people who are in good health.
The device is a Class II, type BF applied part that receives power from a separately approved external IEC 60601-1 compliant power supply or optional battery pack.
The Air Relax Plus Model AR-3.0 consists of an air compressor unit with a control system, an inflatable "garment" (arms, legs and hips), plastic air tubing with a proprietary connector for connecting the device to the garment A description of each of these components is provided below. The hip garment is also referred to as "shorts".
The user interface is a front panel display and buttons.
The Air Relax Plus Model AR-3.0 contains an air compressor with a system control that allows the user to adjust the amount of air coming from the air compressorand going to the individual segments of the inflatable garment.
There is no electrical contact with the user and the device does not transfer or detect energy to or from the user.
The user interface of the Air Relax Plus Model AR-3.0 provides for starting and stopping the massage treatment, allows for adjusting time and intensity (pressure) of the treatment. The device also provides a proprietary keyed connector to the tubing
{4}------------------------------------------------
which connects to the garment. The tubing connector to the garment is also proprietary. The proprietary connectors ensure that neither the device, tubing nor garment can be misconnected to any other device or garment.
Pressure selection is performed by pressing pressure button multiple times. There are four pressure levels with approximate pressure levels as below:
Level 1: 60 mmHg Level 2: 100 mmHg Level 3: 140 mmHg Level 4: 170 mmHg
There are three modes (Progressive, Sequential, Full massage P, S and F) that determine the inflation sequence.
#### Indications for Use:
The Air Relax Plus Model AR-3.0 is intended for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Air Relax Plus Model AR-3.0 simulates kneading and stroking of tissues by using an inflatable garment.
Patient Population: Adults
#### Environments of Use:
Clinics, hospital, athlete training, and home environments
{5}------------------------------------------------
| Model Name<br>510(k)<br>Number | Subject Device | Predicate Device | Comment |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Air Relax Plus Model AR-3.0<br>510(k) K211460 | Air Relax Model AR-1.0/AR-<br>2.0<br>510(k) K191441 | |
| Classification | Class II Device, IRP (21<br>CFR890.5650) | Class II Device, IRP (21<br>CFR890.5650) | Identical |
| Indications<br>for use | The Air Relax Plus Model AR-<br>3.0 is intended for the<br>temporary relief of minor<br>muscle aches and pains and for<br>temporary increase in<br>circulation to the treated areas<br>in people who are in good<br>health. The Air Relax Plus<br>Model AR-3.0 simulates<br>kneading and stroking of<br>tissues by using an inflatable<br>garment | The Air Relax /Compressible<br>limb Sleeve System (AR-1.0,<br>AR-2.0) is intended for the<br>temporary relief of minor muscle<br>aches and pains and for the<br>temporary increase in<br>circulation to the treated areas in<br>people who are in good health.<br>The Air Relax/ Compressible<br>Limb Sleeve System (Model:<br>AR-1.0, AR-2.0) simulates<br>kneading and stroking of tissues<br>by using an inflatable garment | Identical |
| OTC or<br>Prescription | OTC | OTC | Identical |
| Environment<br>of Use | Clinics, hospital, athlete<br>training, and home<br>environments | Clinics, hospital, athlete<br>training, and home<br>environments | Identical |
| Compliance<br>with<br>standards | ES 60601-1, IEC 60601-1-2,<br>IEC 60601-1-11 | ES 60601-1, IEC 60601-1-2,<br>IEC 60601-1-11 | Identical |
| Mode of<br>Operation | Sequential/Peristaltic | Sequential/Peristaltic | Identical |
| Power | 100~240V, 50/60Hz | 120 V, 60Hz | Although the<br>power rating of<br>the subject device<br>is different<br>both devices<br>comply with<br>ANSI/AAMI<br>ES60601-1, so the<br>difference does<br>not affect safety<br>and effectiveness. |
| Device<br>Pressure<br>range | 60 - 170 mmHg | 0-230 mmHg | Lower<br>range than<br>K191441 |
| Garments<br>material | Nylon with a Polyurethane<br>Laminate | Nylon with a Polyurethane<br>Laminate | Identical with<br>K191441 |
| Model Name | Subject Device | Predicate Device | Comment |
| 510(k) | Air Relax Plus Model AR-3.0 | Air Relax Model AR-1.0/AR- | |
| Number | 510(k) K211460 | 2.0<br>510(k) K191441 | |
| Leg Garment | Yes size in inches:<br>L 13.4 x 8.25 x 35.6<br>XL 15.4 x 8.25 x 43.1<br>XXL 15.4 x 8.25 x 47.0 | Yes size in inches:<br>L 13.4 x 8.25 x 35.6<br>XL 15.4 x 8.25 x 43.1<br>XXL 15.4 x 8.25 x 47.0 | Identical in size,<br>construction and<br>materials |
| Arm Garment | Yes | No | See reference<br>device comparison<br>below |
| Hip Garment | Yes | No | See reference<br>device comparison<br>below |
| Number of<br>Inflatable<br>garment<br>segments | 4 | 4 | Identical |
| Weight | 1.67kg (3.7 pounds) | 1.67kg (3.7 pounds) | Identical |
| Dimensions<br>(W x H x D) | 9.5" x 4.7" x 7.5" | 9.5" x 4.7" x 7.5" | Identical |
| Housing<br>Materials and<br>Constructions | Molded ABS enclosure | Molded ABS enclosure | Identical |
| Patient<br>contact | Non-conductive garments | Non-conductive garments | Identical |
| Safety<br>Features | Button at control unit allows<br>user to stop or pause<br>therapy session at any time. | Button at control unit allows<br>user to stop or pause<br>therapy session at any time. | Identical |
| Modes | 4 Modes :<br>"P" mode inflates<br>and deflates chambers from<br>bottom up, one at a time<br><br>"S" mode also inflates<br>chambers from bottom up, but<br>maintains pressure in lower<br>chambers as works its way to<br>top.<br><br>"F" mode is inflates all<br>chambers and maintained<br>pressure at same time and<br>release pressure all chambers.<br><br>At "Target" mode, user can<br>select specific chamber to<br>inflates | 4 Modes :<br>"A" mode inflates<br>and deflates chambers from<br>bottom up, one at a time<br><br>"B" mode also inflates<br>chambers from bottom up, but<br>maintains pressure in lower<br>chambers as works its way to<br>top.<br><br>"C" mode is inflates all<br>chambers and maintained<br>pressure at same time and<br>release pressure all chambers.<br><br>At "Manual" mode, user can<br>select specific chamber to<br>inflates | Identical to<br>predicate except<br>mode names differ |
| Model Name<br>510(k)<br>Number | Subject Device | Reference Device | Comment |
| Indications<br>for use | The Air Relax Plus<br>Model AR-<br>3.0 is intended for the<br>temporary relief of<br>minor muscle aches<br>and pains and for<br>temporary increase in<br>circulation to the<br>treated areas in<br>people who are in<br>good health. The Air<br>Relax Plus Model<br>AR-3.0 simulates<br>kneading and<br>stroking of tissues by<br>using an inflatable<br>garment | The Rapid Reboot<br>Compression Therapy<br>System is indicated for<br>the temporary relief of<br>minor muscle aches and<br>pains and for temporary<br>increase in circulation to<br>the treated areas in<br>people who are in good<br>health. The Rapid<br>Reboot Compression<br>Therapy System<br>simulates kneading and<br>stroking of tissues by<br>using an inflatable<br>garment. | Identical |
| Treatment<br>Time | User determines<br>Therapy time.<br>Choosefrom 15, 30 | User determines<br>Therapy time. Choose<br>from 10, 20, or 30<br>minute session time,<br>with option to add<br>additional 10 minutes<br>to<br>any therapy time. | Similar to K182668<br>except time is limited<br>to 30 minutes in<br>subject device |
| Arm<br>Garment | Yes<br>Long : 13.8" x 33.5"<br>Image: Arm Garment | Yes<br>Regular : 18" x 38"<br>Long : 18" x 44"<br>Image: Arm Garment | Similar in size and<br>construction. Materials<br>are identical to<br>K191441 |
| Hip Garment | Yes<br>Large : 24.5" x 31.5"<br>Image: Hip Garment | Yes<br>Regular : 26" x 32"<br>Large : 26" x 35"<br>Image: Hip Garment | Similar in size and<br>construction. Materials<br>are identical to<br>K191441 |
# Table of the Similarities and Differences of Predicate vs. Subject Device
{6}------------------------------------------------
{7}------------------------------------------------
# Comparison to reference device:
{8}------------------------------------------------
In the above detailed tables we have compared the Air Relax Plus Model AR-3.0 to the predicate for equivalence of:
### Indications -
The Air Relax Plus Model AR-3.0 is intended for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Air Relax Plus Model AR-3.0 simulates kneading and stroking of tissues by using an inflatable garment. These indications are identical to the predicate.
Prescriptive - The Air Relax Plus Model AR-3.0 is OTC as is the predicate.
Design, Technology and Principle of Operation - The Air Relax Plus Model AR-3.0 has equivalent design and features when compared to the predicate and has identical technology to the predicate.
Performance and Specifications - The Air Relax Plus Model AR-3.0 has equivalent specifications of performance when compared to the predicate.
Compliance with standards - The subject device declares compliance with ES 60601-1, IEC 60601-1-11 and IEC 60601-1-2 which is identical to the predicate.
### Materials -
The patient contacting materials of the Air Relax Plus Model AR-3.0 are the inflatable garments which are identical to the predicate device 510(K) K191441.
### Patient Population -
The Air Relax Plus Model AR-3.0 and predicates are indicated for adults.
### Environment of Use -
The Air Relax Plus Model AR-3.0 and predicates are for use in clinics, hospital, athlete training, and home environments.
### Differences -
There are no differences between the proposed device and the predicate device that raise any new safety and effectiveness concerns.
{9}------------------------------------------------
### Performance Testing
#### Bench:
The device has been tested to ensure that all requirements have been met, this includes:
- . Testing of all controls
- Testing of all indicators ●
- Testing of performance
The device has also been tested to the requirements of the following standards:
- AAMI / ANSI ES60601-1:2005 + A1: 2012 Medical electrical equipment -. part 1: general requirements for basic safety and essential performance
- IEC 60601-1-2: 2014 Collateral standard: Electromagnetic Disturbances -● Requirements and Tests
- IEC 60601-1-11: 2015, Collateral standard: requirements for medical ● electrical equipment and medical electrical systems used in the home healthcare environment
### Animal:
No animal testing was performed
### Clinical:
No clinical testing was performed
### Differences -
There are no differences between the proposed device and the predicate device that raise any new safety and effectiveness concerns.
### Substantial Equivalence Rationale
The Air Relax Plus Model AR-3.0 is viewed as substantially equivalent to the predicate device because:
Indications - are identical to the predicate
Prescriptive - The Air Relax Plus Model AR-3.0 and predicate are OTC.
{10}------------------------------------------------
Design, Technology and Principle of Operation - The Air Relax Plus Model AR-3.0 has equivalent design and features when compared to the predicate and has identical technology to the predicate.
Performance and Specifications – The Air Relax Plus Model AR-3.0 has equivalent specifications of performance when compared to the predicate.
Compliance with standards - The subject device declares compliance with ES 60601-1, IEC 60601-1-11 and IEC 60601-1-2 which is identical to the predicate.
Materials - The patient contacting materials of the Air Relax Plus Model AR-3.0 are the inflatable garments they are identical to the predicate.
Environment of Use - Clinics, hospital, athlete training, and home environments, not specified for predicate but predicate is OTC.
Features - The Air Relax Plus Model AR-3.0 has equivalent features when compared to the predicate.
#### Conclusion
The Air Relax Plus Model AR-3.0 is substantially equivalent to the predicate in indications for use, patient population, environment of use, technology characteristics, materials, specifications / performance and compliance with international standards. Minor differences as detailed in the substantial equivalence table above do not raise questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.