Small Bone Nailing System

K211439 · Skeletal Dynamics, Inc. · HTY · Sep 30, 2021 · Orthopedic

Device Facts

Record IDK211439
Device NameSmall Bone Nailing System
ApplicantSkeletal Dynamics, Inc.
Product CodeHTY · Orthopedic
Decision DateSep 30, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Skeletal Dynamics Small Bone Nailing System is indicated for the fixation of extra-articular fractures of the long bones of the hand including the metacarpals and the proximal phalanges.

Device Story

The Small Bone Nailing System is an orthopedic implant set used for the fixation of extra-articular fractures in the hand. The system consists of intramedullary nails and screws (2.1mm, 3.0mm, 4.0mm diameters) and transfixion screws. The device is used by surgeons in a clinical setting to provide rigid fixation within the medullary canal. The system utilizes locking compression screws at both proximal and distal ends to ensure implant/bone rotational stability. The device is provided non-sterile and requires sterilization at the user facility before implantation. By stabilizing fractures, the system facilitates bone healing and restores structural integrity to the affected metacarpals or phalanges.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including static and dynamic mechanical testing per ASTM F2193-20, ASTM F543-17, and ASTM F1264.

Technological Characteristics

System includes titanium alloy intramedullary nails/screws and Cobalt Chrome (CoCr) locking/transfixion screws. Stainless steel K-wires provided for provisional fixation only. Design features locking compression screws at both proximal and distal ends for rotational stability. Non-sterile; requires facility-based sterilization. Mechanical testing performed per ASTM F2193-20, ASTM F543-17, and ASTM F1264.

Indications for Use

Indicated for fixation of extra-articular fractures of the long bones of the hand, specifically metacarpals and proximal phalanges.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ September 30, 2021 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The Department of Health & Human Services logo is a stylized depiction of a human figure, while the FDA acronym and name are written in blue, with the acronym in a square and the name in a sans-serif font. Skeletal Dynamics, Inc. Ana Escagedo President 7300 N. Kendall Drive, Suite 400 Miami, Florida 33156 Re: K211439 Trade/Device Name: Small Bone Nailing System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HTY, HSB Dated: September 3, 2021 Received: September 7, 2021 Dear Ana Escagedo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Michael Owens Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known) K211439 Device Name | Small Bone Nailing System | |---------------------------| |---------------------------| Indications for Use (Describe) The Skeletal Dynamics Small Bone Nailing System is indicated for the fixation of extra-articular fractures of the long bones of the hand including the metacarpals and the proximal phalanges. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number {3}------------------------------------------------ # 510(k) SUMMARY Skeletal Dvnamics Inc.'s Small Bone Nailing System ### Submitter Skeletal Dynamics, Inc. 7300 N. Kendall Drive Suite 400 Miami, FL 33156 Phone: 305-596-7585 Facsimile: 305-596-7591 Contact Person: Ana M. Escagedo Date Prepared: May 3, 2021 #### Name and Classification | Name: | Small Bone Nailing System | |----------------------|-----------------------------------------------------------------------------------| | Common Name: | Pin, Fixation, Smooth | | | Rod Fixation, Intramedullary and Accessories | | Classification: | 21 CFR §888.3040 and §888.3020 | | Classification Name: | Smooth or Threaded Metallic Bone Fixation Fastener<br>Intramedullary Fixation Rod | | Regulatory Class: | Class II | | Product Code: | HTY, HSB | #### Predicate Devices Small Bone Fixation System, K033406 ### Device Description The Skeletal Dynamics Small Bone Nailing System consists of intramedullary nails and screws with diameters of 2.1mm, 3.0mm, and 4.0mm, and lengths from 25mm to 70mm to address proximal phalanx and metacarpal fractures. The intramedullary nails and screws are made of medical grade titanium allov. Cobalt Chrome 1.7mm transfixion screws are also provided in various lengths. Included in the system is specialized instrumentation. The System is provided non-sterile for sterilization in the user facility. The Small Bone Nailing System is comprised of: - Titanium alloy intramedullary nails and screws ● - CoCr locking screws and transfixion screws . - Stainless steel K-wires (for provisional fixation: not for implantation) . - System specific instrumentation . ### Indications for Use The Small Bone Nailing System is indicated for fixation of extra-articular fractures of the long bones of the hand including the metacarpals and the proximal phalanges. ### Summary of Technological Characteristics The substantial equivalence of the Small Bone Nailing System to the predicate device is demonstrated by similarities in intended use, indications for use, materials, design (fundamental scientific technology), performance, sterility and packaging and does not present any new issues of safety or effectiveness. Both the subject device and the predicate device are similar in that they offer fixation for extra {4}------------------------------------------------ articular fractures of the long bones of the hand, including the metacarpals and proximal phalanx and include locking features for implant/bone rotational stability. Differences between the subject and devices do not affect the safety and effectiveness of the subject device. These differences include: Fixation by means of two locking compression screws (one distal, one proximal) with the subject device; the predicate fixation is by way of a bent tip and one proximal locking device. The use of locking compression screws at both the proximal and distal ends with the subject device provides rigid fixation of the implant within the medullary canal. There is also a material difference between the subject device (made of medical grade titanium alloy) and the predicate device (made of stainless steel), however, both devices have the comparable mechanical properties. # Performance Testing Engineering analysis and mechanical testing demonstrated that the Skeletal Dynamics Small Bone Nailing System is equivalent to predicate devices currently marketed. Static and dynamic testing which established equivalency included: - . ASTM F2193-20, Standard Specifications and Test Methods for Components Used in the Surqical Fixation of the Spinal Skeletal System and ASTM-F543-17 Standard Specification and Test Methods for Metallic Medical Bone Screws. - ASTM F1264, Standard Specification and Test Methods for Intramedullary Fixation Devices. . Therefore, the subject device is as safe and effective as the legally marketed predicate device. # Conclusions The Skeletal Dynamics Small Bone Nailing System is substantially equivalent to the predicate device identified in this premarket notification.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...