K211431 · Olea Medical · QIH · Aug 2, 2021 · Radiology
Device Facts
Record ID
K211431
Device Name
breastscape v1.0
Applicant
Olea Medical
Product Code
QIH · Radiology
Decision Date
Aug 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
Indications for Use
breastscape V1.0 is an optional image processing software application that is intended for use on Olea Sphere 3.0 software package. It is intended to be used by trained breast imaging physicians and trained MRI technologists. breastscape V1.0 includes a software module (BreastApp) that supports the visualization, analysis, and reporting of lesions measurements and analysis. breastscape V1.0 supports the evaluation of dynamic MR data acquired during contrast administration and the calculation of parameters related to the uptake characteristics. breastscape V1.0 performs other user selected processing functions (such as image subtraction, multiplanar and oblique reformats, 3D renderings). The resulting information can be displayed in a variety of formats, including a parametric image overlaid onto the source image. breastscape V1.0 can also be used to provide measurements of the segmented tissue volumes (volumes of interest) based on uptake characteristics. These measurements include volume measurement, distances of volumes of interest to anatomical landmarks, 3D longest diameter and 2D long and short axis. breastscape V1.0 includes the option to add annotations based on the fifth edition of the American College of Radiology's Breast Imaging Reporting and Data System (BI-RADS®) Breast Imaging Atlas. breastscape V1.0 may be used as an image viewer of multi-modality digital images, including ultrasound and mammography. breastscape V1.0 is not intended for primary interpretation of digital mammography images. breastscape V1.0 includes a software module (BreastLoc) to assists users in planning MR guided breast interventional procedures. Using information from MR images, regarding userspecified target lesion and fiducial location coordinates, the software gives calculation of the targeted region of interest (such as suspected lesion) depth. When interpreted by a skilled physician, breastscape V1.0 provides information that may be used for screening, diagnosis, and interventional planning. Patient management decisions should not be based solely on the results of breastscape V1.0.
Device Story
breastscape V1.0 is an optional PACS software module operating within the Olea Sphere V3.0 environment. It processes DICOM-formatted MRI data to assist physicians and technologists in breast imaging. The BreastApp module performs semi-automatic segmentation of lesions, calculates uptake kinetics (wash-in, washout, time-to-peak), generates parametric maps, and computes volumetric/linear measurements. It also provides tools for multiplanar/oblique reformats and 3D rendering. The BreastLoc module assists in MR-guided interventional planning by calculating needle insertion depth and path based on user-defined target coordinates and fiducial locations. Output is displayed as parametric overlays or reports, including BI-RADS annotations. Used in hospitals and imaging centers, the device supports screening, diagnosis, and interventional planning. It does not provide automated diagnosis; clinical decisions must not rely solely on its output. It benefits patients by providing standardized, quantitative analysis of breast lesions and facilitating precise interventional procedure planning.
Clinical Evidence
Bench testing only. Validation included product risk assessment, software module verification, and software validation testing. Performance was compared against the predicate (MultiView V4.0.3.2) and reference (Olea Sphere V3.0) devices using anonymized patient cohorts. BreastLoc performance was evaluated against the predicate for procedural planning, while BreastApp metrics (kinetics, measurements, parametric maps) were validated against manual ground truth measurements.
Technological Characteristics
PACS software module; requires Olea Sphere V3.0 platform. DICOM-based image processing. Features: semi-automatic segmentation, kinetic curve computation, parametric mapping, 3D rendering, and interventional planning calculations. Software-based; no hardware components. Operates on standard clinical workstations.
Indications for Use
Indicated for trained breast imaging physicians and MRI technologists to assist in visualization, analysis, and reporting of breast MRI studies, including dynamic contrast-enhanced (DCE-MRI) data, and planning of MR-guided interventional procedures. Not intended for primary interpretation of digital mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
August 2, 2021
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Olea Medical % John J. Smith, M.D. J.D. Partner Hogan Lovells US LLP 555 Thirteenth Street NW WASHINGTON DC 20004
Re: K211431
Trade/Device Name: breastscape v1.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH, LLZ Dated: May 7, 2021 Received: May 7, 2021
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K211431
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below
510(k) Number (if known)
Device Name
e
#### breastscape V1.0
breastscape V1.0 is an optional image processing software application that is intended for use on Olea Sphere package. It is intended to be used by trained breast imaging physicians and trained MRI technologists.
breastscape V1.0 includes a software module (BreastApp) that supports the visualization, and reporting of lesions measurements and analysis. breastscape V1.0 supports the evaluation of data acquired during contrast administration and the calculation of parameters related to the uptake characteristics.
breastscape V1.0 performs other user selected processing function, multiplanar and oblique reformats, 3D renderings).
The resulting information can be displayed in a variety of formats, including a parametric image overlaid onto the source image.
breastscape V1.0 can also be used to provide measurements of the segmented tissue volumes of interest) based on uptake characteristics. These measurements include volume measurement, distances of volumes of interest to anatomical landmarks, 3D long and short axis.
breastscape V1,0 includes the option to add annotation of the American College of Radiology's Breast Imaging Reporting and Data System (BI-RADS®) Breast Imaging Atlas.
breastscape V1.0 may be used as an image viewer of multing ultrasound and mammography. breastscape V1.0 is not intended for primary interpretation of digital mammography images.
breastscape V1.0 includes a software module (BreastLoc) to assists users interventional procedures. Using information from MR images, regarding user-specified target lesion coordinates, the software gives calculation of the targeted region of interest (such as suspected lesion) depth.
When interpreted by a skilled physician, breatscape V1.0 provides information that may be used for screening, danning.
Patient management decisions should not be based solely on the results of breastscape V1.0.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY Olea Medical's breastscape V1.0 K211431
#### Submitter
Olea Medical 93 avenue des Sorbiers, ZI ATHELIA IV 13600, La Ciotat France Phone: +33 4 42 71 24 20 Facsimile: +33 4 42 71 24 27 Contact Person: Nathalie Palumbo
Date Prepared: August 2, 2021
Name of Device: breastscape V1.0
Common or Usual Name: PACS
Classification Name: Medical Image Management and Processing System
Regulatory Class: 21 CFR 892.2050
Product Code: QIH, LLZ
#### Predicate Device
MultiView V4.0.3.2,
- Manufacturer: Hologic, Inc. (K132316) .
- . Device Name: MultiView V4.0.3.2
- . Regulation Number: 21 CFR 892.2050
- Product Code: LLZ
### Reference Device
for Kinetics: Olea Sphere V3.0 (K152602)
- Manufacturer: Olea Medical
- Device Name: Olea Sphere V3.0 .
- . Regulation Number: 21 CFR 892.2050
- . Product Code: LLZ
### Device Description
breastscape V1.0 is an optional PACS software tool that is intended for use with the Olea Sphere V3.0 software package, cleared under K152602. The software accesses image series in DICOM format through Olea Sphere V3.0, which is a software package used to perform image viewing, processing and analysis of medical images.
{4}------------------------------------------------
breastscape V1.0 is made of two software modules: BreastApp and BreastLoc.
- 1. BreastApp
BreastApp is designed to assist in the visualization, analysis and reporting of Magnetic Resonance Imaging (MRI) breast studies. This module supports the evaluation of dynamic MR breast data acquired during contrast administration (DCE-MRI), and the calculation of parameters related to the lesion uptake characteristics. This module provides semi-automatic segmentation of volumes of interest, distance measurements and lesion volume measurements.
BreastApp provides the features below:
- Visualization of registered MR image series. It includes well-established to . standard image viewing, MIPs, reformats and 3D volume rendering.
- . Visualization of Mammography and Ultrasound image series for display purpose only.
- . Evaluation of dynamic MR breast data acquired during contrast administration. It includes:
- The computation of image subtractions (subtractions of each time point/phase image o of the dynamic series with the 1st time point (baseline) image to highlight tissue with contrast enhancement).
- The display of time intensity signal curves (kinetics curves) showing tissue contrast o enhancement evolution over time.
- The detection and display of the kinetics curve showing the worst kinetics behavior O (most important washout among pixels having peak enhancement superior at 50% enhancing threshold).
- The computation of semi-quantitative kinetics maps that are derived from the time o intensity signal curves and showing uptake characteristics (e.g., Time to Maximum contrast Enhancement, Wash in, Washout, etc.).
- . Automatic detection of breast morphological structures. It includes the automatic detection of nipple position, chest and skin border. The user can further adjust them if needed.
- . Semi-automatic lesion segmentation. It includes:
- Highlighting tissues showing significant contrast agent uptake based on an uptake O threshold.
- Semi-automatic segmentation of the suspected lesion identified by the user. The user O can further adjust the segmentation if needed or even manually segment the suspected lesion.
- Automatic computation of the suspected lesion 2D/3D diameter and lesion volume. o
{5}------------------------------------------------
- . Automatic computation of distances between the suspected lesion and the morphological structures (distances to nipple, chest and skin). The user can further adjust the distances if needed.
- . Reporting of user-selected findings and assessment through a dedicated breast report. It includes the option to add annotations based on the Fifth Edition of American College of Radiology's Breast Imaging Reporting and Data System (BI-RADS®) Breast Imaging Atlas. The software automatically reports the localization of the suspected lesion on a dedicated breast sector map. The position can be further adjusted by the user if needed.
- . Follow-up for multiple (more than two) studies from same patient. It includes tools to enhance the visualization and analysis of patient follow-up studies through the same layout.
- 2. BreastLoc
The BreastLoc module is designed to assist users in planning MR-guided breast interventional procedures. Based on user-specified target lesion and fiducial location coordinates, BreastLoc is used to compute and display the following features:
- . Needle insertion block position within the grid diagram;
- . Needle insertion point activation in the block;
- Depth of introducer, representing the graduation value where to put the depth stop on the introducer sheath;
- . Needle insertion path display on native images.
### Intended Use / Indications for Use
breastscape V1.0 is an optional image processing software application that is intended for use on Olea Sphere 3.0 software package. It is intended to be used by trained breast imaging physicians and trained MRI technologists.
breastscape V1.0 includes a software module (BreastApp) that supports the visualization, analysis, and reporting of lesions measurements and analysis. breastscape V1.0 supports the evaluation of dynamic MR data acquired during contrast administration and the calculation of parameters related to the uptake characteristics.
breastscape V1.0 performs other user selected processing functions (such as image subtraction, multiplanar and oblique reformats, 3D renderings).
The resulting information can be displayed in a variety of formats, including a parametric image overlaid onto the source image.
breastscape V1.0 can also be used to provide measurements of the segmented tissue volumes (volumes of interest) based on uptake characteristics. These measurements include volume measurement, distances of volumes of interest to anatomical landmarks, 3D longest diameter and 2D long and short axis.
breastscape V1.0 includes the option to add annotations based on the fifth edition of the American College of Radiology's Breast Imaging Reporting and Data System (BI-RADS®)
{6}------------------------------------------------
Breast Imaging Atlas.
breastscape V1.0 may be used as an image viewer of multi-modality digital images, including ultrasound and mammography. breastscape V1.0 is not intended for primary interpretation of digital mammography images.
breastscape V1.0 includes a software module (BreastLoc) to assists users in planning MR guided breast interventional procedures. Using information from MR images, regarding userspecified target lesion and fiducial location coordinates, the software gives calculation of the targeted region of interest (such as suspected lesion) depth.
When interpreted by a skilled physician, breastscape V1.0 provides information that may be used for screening, diagnosis, and interventional planning.
Patient management decisions should not be based solely on the results of breastscape V1.0.
Both breastscape V1.0 and MultiView V4.0.3.2 are user-defined software analysis tools used for the visualization, analysis and reporting of MRI studies. Both software tools provide assistance for the users in planning MR guided breast interventional procedures. Furthermore, both software devices support the same limitation of use regarding the display of lossy compressed mammographic images and digitized film screen images that must not be reviewed for primary image interpretations.
Similarly, both devices are for use in hospitals, imaging centers and radiologist reading practices by any trained professional who may require and are granted access to patient image, demographic, and report information. Importantly, neither software product is used for diagnosis. Patient management decisions should not be based solely on the results of either software. The intended use of the software is the same.
The only minor difference in the indications for use between the two devices is that breastscape V1.0 is a software application intended to be used on Olea Sphere V3.0 whereas MultiView V4.0.3.2 is a standalone software. breastscape V1.0 is only to be used with the cleared Olea Sphere V3.0 PACS (K152602), which is the reference device. Therefore, this minor difference does not impact the safety or effectiveness of the subject device.
# Summary of Technological Characteristics
breastscape V1.0 is an optional PACS software module intended for use with the cleared Olea Sphere V3.0 software package (K152602). breastscape V1.0 accesses MR image series in DICOM format through Olea Sphere V3.0, which is a software package for image viewing, processing and analysis of medical images. The application execution requires Olea Sphere V3.0 with its components (data and license management services) to be started and in operational use.
breastscape V1.0 allows the:
- . Visualization (standard imaqe viewing tools, MIPs, and reformats)
- Analysis (registration, subtractions, kinetic curves, parametric image maps, and 3D . volume rendering)
- . Reporting of user-selected findings and assessment
- Assistance for the users in planning MR guided breast interventional procedures. .
{7}------------------------------------------------
#### Breastscape V1.0 versus Multiview V4.0.3.2 .
As noted above, both the breastscape V1.0 and MultiView V4.0.3.2 are software analysis tools that are used for the visualization, analysis, and reporting of MRI studies. Both software tools provide assistance for the users in planning MR guided breast interventional procedures. Both software devices are used with combinations of biopsy devices (grids, blocks and MRI-guided vacuum-assisted breast biopsy systems).
Both the breastscape v1.0 and Multivew V4.0.3.2 have similar technological characteristics. The only minor difference is that MultiView V4.0.3.2 is intended to be used with one grid, the Sentinelle grid, and performs automatic detection of the grid and the fiducial. In contrast the breastscape V1.0 is intended to be used with four different grid models: generic, curved lateral access, curved medial access, and Sentinelle grids. Nevertheless, software validation testing demonstrates that the device operates as safely and effectively as its predicate device and does not raise different questions of safety and effectiveness.
As further confirmatory evidence, the company conducted additional validation testing with the subject breastscape V1.0 and predicate MultiView V4.0.3.2 device using anonymized images from a cohort of patients.
#### . Breastscape V1.0 versus Olea Sphere V3.0
As noted previously, the subject breastscape V1.0 will only be used with the company's cleared Olea Sphere V3.0 PACS viewer (K152602).
Additionally, the software architecture of the feature related to kinetics curve computation and parametric image maps of the breastscape V1.0 is identical to the architecture used by the reference device: Olea Sphere V3.0 manufactured by Olea Medical (K152602).
The results of performance evaluation support that the minor difference in the technological characteristics do not raise different questions of safety and effectiveness.
### Performance Data
Olea Medical has conducted extensive validation testing of the breastscape V1.0 as a PACS software module intended for use with the Olea Sphere V3.0 system, cleared under K152602. Internal verification and validation testing confirms that the product specifications are met, in support of the substantial equivalence of the intended use and technological characteristic as the reference devices.
breastscape V1.0, as an optional application of the Olea Sphere V3.0 software, has been validated to ensure that the system as a whole provides all the capabilities necessary to operate according to its intended use and in a manner substantially equivalent to the predicate device.
The main groups of tests performed include:
- Product risk assessment; ●
- Software modules verification tests; ●
- Software validation test.
{8}------------------------------------------------
Based on the clinical performance as documented in the pivotal clinical study, the breastscape v1.0 has a safety and effectiveness profile that is similar to the predicate device.
In addition to software performance testing, the company conducted additional validation testing to compare the results of breastscape V1.0 with the predicate and reference devices. MultiView™ MR Breast V4.0.3.2 (Hologic®, Bedford, MA, USA) was used as a comparison for BreastLoc™ to evaluate performance of the MR quided breast intervention procedural planning. For BreastApp™, direct manual measurements (ground truth) and Kinetics plugin, viewing tools, follow-up feature, breast dedicated report, mammography loading and visualization, measurements modifications within Olea Sphere V3.0, were used to evaluate the automatically calculated metrics and parametric maps, respectively.
## Conclusions
breastscape V1.0 is substantially equivalent to the predicate device: Multiview V4.0.3.2 and similar to the reference device: Olea Sphere V3.0. The breastscape V1.0 has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device.
In addition, the minor technological differences between the breastscape V1.0 and its predicate devices raise no new questions of safety or effectiveness. Performance data demonstrate that the breastscape V1.0 is as safe and effective as the Multiview V4.0.3.2 and Olea Sphere V3.0. Thus, the breastscape V1.0 is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.