3N Contact Lenses Adjunct Cleaner is intended for use as an adjunct in deproteinization and storage for soft hydrophilic contact lenses with only either OPTI-FREE EXPRESS Multi-Purpose Disinfecting Solution or Bausch & Lomb ReNu MultiPlus Multi-Purpose Solution.
Device Story
3N Contact Lenses Adjunct Cleaner (Model TN083) is an electrical device for home use to deproteinize and store soft hydrophilic contact lenses. It utilizes protein electrophoresis; negatively charged proteins in the lens are attracted to a positive electrode (anodic cleaning probe) within the device's cleaning chamber. The device consists of a power base and a partitioned fluid reservoir. Users place lenses in the reservoir with approved multi-purpose solution; the device performs an automated cleaning cycle and shuts down automatically. The process removes protein from the lens surface and air holes. The device is powered by a built-in lithium battery. It functions as both a cleaner and a storage case, providing a physical cleaning mechanism that supplements the chemical action of the contact lens solution.
Clinical Evidence
No clinical data. Bench testing only. Deproteinization efficiency testing showed >92.5% protein removal rate for specific soft contact lens materials (Ocufilcon D, Etafilcon A). Service life testing confirmed performance after 151 cycles (3-month shelf life). Physical compatibility testing confirmed lens parameter stability per ISO 18369-2:2017. Biocompatibility testing included cytotoxicity (MTT), sensitization (Guinea Pig Maximization), and ocular irritation.
Technological Characteristics
Materials: Medical grade ABS. Principle: Protein electrophoresis (electrical). Power: Built-in lithium battery. Form factor: Power base with partitioned fluid reservoir. Connectivity: None. Sterilization: Non-sterile. Software: Minor level of concern. Standards: ISO 10993-1, ISO 10993-11, IEC 61010-1, IEC 61326-1, ISO 18369-2.
Indications for Use
Indicated for deproteinization and storage of soft hydrophilic contact lenses in conjunction with specific multi-purpose disinfecting solutions (OPTI-FREE EXPRESS or Bausch & Lomb ReNu MultiPlus).
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 13, 2022
Suzhou 3N Biological Technology Co., Ltd Wang Xiaofang Quality and Regulatory Division Manager NO.218 Xinghu Street Suzhou, Jiangsu 215000 China
Re: K211361
Trade/Device Name: 3N Contact Lenses Adjunct Cleaner Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LYL Dated: December 10, 2021 Received: December 10, 2021
Dear Wang Xiaofang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211361
Device Name 3N Contact Lenses Adjunct Cleaner
Indications for Use (Describe)
3N Contact Lenses Adjunct Cleaner is intended for use as an adjunct in deproteinization and storage for soft hydrophilic contact lenses with only either OPTI-FREE EXPRESS Multi-Purpose Disinfecting Solution or Bausch & Lomb ReNu MultiPlus Multi-Purpose Solution.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image contains a logo with the number 3 in a stylized font, followed by the text "3N科技" and "3N TECH". The number 3 is designed with blue geometric shapes, and the text is in a simple, sans-serif font. The logo appears to represent a technology company named 3N Tech.
## 510(K) Summary
This 510(k) Summary is provided per the requirements of section 21 CFR 807.92.
The assigned 510(k) number is K211361.
Date Summary Prepared: 1/11/2022.
### 1. Submitter
| Submitter's Name: | Suzhou 3N Biological Technology Co., Ltd |
|-------------------|------------------------------------------------------------------|
| Contact person: | Wang Xiaofang<br>Head of Quality and Regulatory |
| Address: | NO.218, Xinghu Street, BioBay, SIP Suzhou, Jiangsu, PRC<br>China |
| Telephone: | 0512-62757627/+8617317293095 |
| Email: | xiaofang.wang@3nbio.com |
#### 2. Device
| Trade Name | 3N Contact Lenses Adjunct Cleaner |
|------------------------|-----------------------------------|
| Common Name | Contact Lenses Adjunct Cleaner |
| Regulation Description | Contact lens care products |
| Regulation number | 21 CFR 886.5928 |
| Device Class | Class II |
| Product Code | LYL |
#### 3. Predicate Device
RIM-WIDE COMPANIES, INC. LENSERVER AUTOMATIC CONTACT LENS CLEANING ACCESSORY, K982115 (RIM-WIDE COMPANIES, INC.) FDA cleared on 01/13/1999.
#### 4. Description of Proposed Device:
The proposed device, 3N Contact Lenses Adjunct Cleaner, Model TN083 is an electrical cleaner which is intended for use as an adjunct in deproteinization and storage for soft hydrophilic contact lenses using approved contact lens solutions.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for 3N Tech. The logo consists of a stylized blue number 3 on the left, followed by the text "3N" in a simple sans-serif font. Below "3N" is the text "3N TECH" in a smaller font. The logo is clean and modern, with a focus on the number 3 and the company's name.
The 3N Contact Lenses Adjunct Cleaner adopt the principle of protein electrophoresis, take advantage of the characteristics that negatively charged protein in certain environment and swarm towards positive pole in electric field, then the tear protein will be removed from the surface and air holes of contact lens and absorbed by anodic cleaning probe to achieve the goal of cleaning contact lenses.
The cleaning components of the 3N Contact Lenses Adjunct Cleaner (Model: TN083) consist of a cleaning unit (Power Base) and a partitioned fluid reservoir functioning as the cleaning chamber.
The cleaning chamber is detachably connected to the cleaning unit (power base). In the cleaning chamber, there are two independent cleaning tanks; and there is a cleaning probe on each side of the cleaning tank, namely electrode. In the cleaning unit, the circuit board is fixedly provided with electrode connectors electrically connected to the circuit board. The end of the electrode connector can be stretched and elastically connected with the bottom of the electrode.
#### 5. Indications for Use
3N Contact Lenses Adjunct Cleaner is intended for use as an adjunct in deproteinization and storage for soft hydrophilic contact lenses with only either OPTI-FREE EXPRESS Multi-Purpose Disinfecting Solution or Bausch & Lomb ReNu Multi-Purpose Solution.
#### Comparison of Technological Characteristics with the Predicate Devices 6.
The following table compares the device to the predicate device with basic technological characteristics.
Both subject and predicate devices are electrical contact lens cleaning device and used in conjunction with approved contact lens solutions.
3N Contact Lenses Adjunct Cleaner, TN083 has been compared to LENSERVER Automatic Contact Lens Cleaning Accessory as a predicate device for substantial equivalence. A table comparing the two devices is provided as follows:
| Features | Proposed Device<br>(K211361) | Predicate Device<br>(K982115) | Comparison |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Product code | LYL | LYL | Same |
| Regulation No. | 21 CFR 886.5928 | 21 CFR 886.5928 | Same |
| Device Class | Class II | Class II | Same |
| Indications for Use | 3N Contact Lenses Adjunct<br>Cleaner is intended for use as an | The LENSERVER Automatic<br>Contact Lens Cleaning Accessory | Similar<br>Note 1 |
| | adjunct in deproteinization and<br>storage for soft hydrophilic<br>contact lenses with only either<br>OPTI-FREE EXPRESS Multi-<br>Purpose Disinfecting Solution or<br>Bausch & Lomb ReNu MultiPlus<br>Multi-Purpose Solution. | is intended for use as an adjunct<br>in cleaning both soft hydrophilic<br>and rigid gas permeable contact<br>lenses using approved contact<br>lens cleaning solutions.<br><br>Examples of approved cleaning<br>solutions indicated for use with<br>the LENSERVER Automatic for<br>soft hydrophilic lenses are the<br>Allergan Soft Mate Consept-1<br>Cleaning and Disinfecting<br>Solution, Allergan Soft Mate<br>Cdrisept-2 Neutralizing and<br>Rinsing Solution, or Spray. The<br>LENSERVER Automatic is<br>indicated for cleaning rigid gas<br>permeable lenses in conjunction<br>with approved solutions such as<br>Allergan Gas Permeable Daily<br>Cleaner and Allergan<br>ComfortCare Gas Permeable<br>Wetting and Soaking Solution. | |
| Technical Principle | Protein electrophoresis | Ultrasonic Cleaning | Different<br>Note 2 |
| Mechanism of<br>Action | Electrical | Electro-mechanical | Different<br>Note 3 |
| Configuration | Cleaning Unit(Power Base)<br>Cleaning chamber | Motorized cleaning unit<br>Cleaning beaker | Similar<br>Note 4 |
| Shutdown mode | Automatic timing shutdown | Automatic timing shutdown | Same |
| Power | Built-in lithium battery | Dry Battery | Different<br>Note 5 |
| Material | Medical grade ABS | Medical grade<br>polymethylpentene | Similar<br>Note 6 |
| Sterile | Non-sterile | Non-sterile | Same |
| Single use | No | No | Same |
| Environment of<br>Use | home | home | Same |
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image contains a logo for 3N Tech. The logo consists of a blue stylized number 3 on the left, followed by the text "3N 科技" in Chinese characters and "3N TECH" in English. The number 3 is made up of three connected lines, with a small square above the top line.
The following technological differences exist between the subject and predicate devices:
Note 1
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image contains a logo with the number 3 in a stylized font, along with the text "3N科技" and "3N TECH". The number 3 is in blue and is made up of two parts, a darker blue and a lighter blue. The text "3N科技" is in black and is written in a Chinese font. The text "3N TECH" is in black and is written in a sans-serif font.
The intended use of proposed device and predicate device are similar. The proposed device has storage function. It is expected that the proposed device has more storage functions than the predicate device, just like a contact lens case.
It will not affect the main function and intended use of the device, and indications for use of proposed device is clearly indicated in user manual and label. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.
### Note 2 and Note 4
The technical principle and configuration of proposed device and predicate device are different, but it will not affect the main function and the intended use of the device. They are both electrical means. And they are both compliance with electrical safety Standard. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.
### Note 3
The mechanism of action of proposed device and predicate device are different, but it will not affect the main function and the intended use of the device. They are both meet the related standards and requirements of physical compatibility. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.
### Note 5
The battery type of proposed device and predicate device are different. They are both compliance with electrical safety standard, it will not raise any safety or effectiveness issue.
### Note 6
The main materials of the proposed device and predicate device are different. They are both compliance with medical requirements and standard ISO 10993-1, 10993-11. It will not raise any safety or effectiveness issue.
#### 7. Non-Clinical Data
The following performance data were provided in support of the substantial equivalence determination.
## 7.1 Biocompatibility testing
The biocompatibility evaluation for the 3N Contact Lenses Adjunct Cleaner is conducted in accordance with the International Standard ISO10993-1 " Biological Evaluation of Medical Devices -Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The worst case of the device is considered mucosal membranes contacting for duration of less than 24 hours. And the testing included the following tests:
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for 3N Tech. The logo consists of a stylized number 3 in blue, with a small square in the upper right corner. To the right of the number 3 is the text "3N科技" in Chinese characters, followed by the text "3N TECH" in English.
- Cytotoxicity Study using MTT Method
- Guinea Pig Maximization Test
- Ocular Irritation test
# 7.2 Software Verification and Validation
Software documentation, including verification, was provided following FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software with a minor level of concern.
# 7.3 Safety and EMC
Testing was performed to verify the basic safety and essential performance of the 3N Contact Lenses Adjunct Cleaner, TN 083. The following tests were performed: IEC 61010-1:2010+A1:2016 Safety requirements for electrical equipment for measurement, control, and laboratory use-Part1: General requirements IEC 61326-1:2013 Electrical equipment for measurement, control and laboratory use-EMC
requirements- Part 1: General requirements
## 7.4 Performance Testing- Bench
The following tests were performed to demonstrate that the proposed 3N contact lenses adjunct cleaner met the applicable design and performance requirements and support a determination of substantial equivalence. Where applicable, testing was done per applicable ISO and other international standards.
### ● Deproteinization efficiency testing and Service life testing
Testing was performed to verify the Protein removal effect of the 3N contact lenses adjunct cleaner. This verification test selected lenses of four kinds of materials from Johnson and Cooper Optics. After soaking in artificial tears, Ocufilcon D and Etafilcon A soft contact lenses had a large amount of protein adsorption. The protein cleaning rate was more than 92.5% with 3N contact lenses adjunct cleaner combined with care solution. However, Comfilcon A and Senofilcon A soft contact lenses have very little protein adsorption, so it is impossible to calculate the clearance effect. The protein removal rate of Ocufilcon D soft contact lens was 95% after 3N contact lenses adjunct cleaner was recycled for 0, 151 times. So the shelf life of the 3N contact lenses adjunct cleaner is 3 months.
### ● Physical compatibility testing
The testing was considerably streamlined just assessing 3 low modulus high water SCLs (>50wt% water, for example 3 commercial lenses from any of these SCL materials, hioxifilcon A, nelfilcon A or nesofilcon A, or other high water materials) with the 3N device.
Deviation of the parameters of contact lenses before and requirements after cleaned by 3N Contact
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image contains a logo with the number 3 on the left side, and the text "3N科技" and "3N TECH" on the right side. The number 3 is stylized with a blue color scheme. The text "3N科技" is in Chinese characters, and the text "3N TECH" is in English. The logo is likely for a technology company.
Lenses Adjunct Cleaner complies with the relevant requirements in ISO 18369-2:2017.
- Cleaning Validation was performed to demonstrate functionality is not affected after maximum number of cleaning cycles. The device was cleaned following the procedure defined in the User's Manual.
## 7.5 Performance Testing - Animal
This submission does not include any animal performance testing. It was determined that no such testing was required to demonstrate substantial equivalence.
## 7.6 Performance Testing - Clinical
This submission does not include any clinical performance testing. It was determined that no such testing was required to demonstrate substantial equivalence.
#### 8. Conclusion
From the bench test conducted on the proposed device provided in above, the test result showed that the proposed device can meet the requirements of related standards. Therefore, it can be determined that the proposed devices are Substantially Equivalent (SE) to the predicate device, K982115.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.