Selebrek PTCA Balloon Dilatation Catheter

K211349 · Kossel Medtech (Suzhou) Co., Ltd. · LOX · Jan 7, 2022 · Cardiovascular

Device Facts

Record IDK211349
Device NameSelebrek PTCA Balloon Dilatation Catheter
ApplicantKossel Medtech (Suzhou) Co., Ltd.
Product CodeLOX · Cardiovascular
Decision DateJan 7, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5100
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Selebrek PTCA Balloon Dilatation Catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.

Device Story

Selebrek PTCA Balloon Dilatation Catheter is a sterile, single-use, rapid exchange (RX) intravascular device; used by physicians in clinical settings for percutaneous coronary intervention (PCI). Device consists of a proximal stainless steel shaft (PTFE coated), distal shaft (outer/inner tubes), and a semi-compliant Pebax balloon. Operates by inflating balloon with contrast medium via inflation lumen to dilate stenotic coronary arteries or bypass grafts; guidewire lumen facilitates advancement through stenosis. Radiopaque Platinum/Iridium marker band aids positioning. Output is mechanical dilatation of stenosis; clinical benefit is improved myocardial perfusion. Device does not include distal dye injection or pressure measurement lumens.

Clinical Evidence

Bench testing only. Performance data included dimensional verification, rated burst pressure, fatigue, compliance, inflation/deflation time, bond/tip pull strength, flexibility, torque, radiopacity, coating integrity, particulate evaluation, and shelf life. Biocompatibility leveraged from reference device (K182699) per ISO 10993-1, covering cytotoxicity, sensitization, reactivity, systemic toxicity, hemocompatibility, and pyrogenicity.

Technological Characteristics

Rapid exchange PTCA catheter; 142cm length. Materials: Pebax (balloon), stainless steel (proximal shaft), PTFE/hydrophilic coating. Semi-compliant balloon (1.5mm diameter, 6-30mm length); 14 atm rated burst pressure; 6 atm nominal pressure. Single radiopaque Platinum/Iridium marker band. 0.014" guidewire compatibility. Ethylene oxide sterilization.

Indications for Use

Indicated for balloon dilatation of stenotic coronary arteries or bypass grafts to improve myocardial perfusion in patients with coronary ischemia.

Regulatory Classification

Identification

Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.

Special Controls

*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b) *Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2) *Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". January 7, 2022 Kossel Medtech (Suzhou) Co., Ltd. Zane Wang Quality Director F2-F3, BLDG 6, No. 8, Jinfeng Road Suzhou, Jiangsu Province 215163 China Re: K211349 Trade/Device Name: Selebrek PTCA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II Product Code: LOX Dated: December 2, 2021 Received: December 8, 2021 Dear Zane Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {2}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K211349 Device Name Selebrek PTCA Balloon Dilatation Catheter Indications for Use (Describe) The Selebrek PTCA Balloon Dilatation Catheted for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/0 description: The image contains a logo for KOSSEL. The logo consists of a circular graphic on the left and the text "科塞尔医疗 KOSSEL" on the right. The circular graphic is made up of several curved lines that create a swirling effect. The text is in a sans-serif font and is a teal color. # 510(k) Summary – K211349 This 510(k) Summary is submitted in accordance with 21 CFR 807.92(c). | Submitter: | Kossel Medtech (Suzhou) Co., Ltd.<br>F2-F3, BLDG 6, No. 8 Jinfeng Road<br>Suzhou New District, Jiangsu Province, China | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Zane Wang<br>F3, BLDG 6, No. 8 Jinfeng Road<br>Suzhou New District, Jiangsu Province, China<br>Phone: +86 512 8717 4080-6003<br>Mail: wangzhigao@kosselmed.com | | Date Prepared: | December 30, 2021 | | Trade Name: | Selebrek PTCA Balloon Dilatation Catheter | | Common Name: | Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter | | Classification Name: | Catheters, transluminal coronary angioplasty, percutaneous (21 CFR 870.5100), Class II (special controls) | | Product Code: | LOX | | Predicate Device: | Sapphire II PRO Balloon Dilatation Catheter (K180921; cleared June 28, 2018) | #### Intended use/Indication for use: The Selebrek PTCA Balloon Dilatation Catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. #### Device Description: The Selebrek PTCA Balloon Dilatation Catheter is a rapid exchange (RX) PTCA Balloon Dilatation Catheter used for the purpose of dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The Selebrek is a sterile, single-use, intravascular medical device with a working length of 142cm. The proximal shaft is PTFE coated stainless steel tube, which allows for exceptional push ability and a smooth transition to the distal shaft, which is composed of an outer tube, an inner tube, and a balloon. A hydrophilic coating is applied to the distal section. The semicompliant balloons are available in diameters of 1.5mm and lengths from 6-30mm, and have a rated burst pressure of 14 atm. The proximal shaft of the catheter has two marker sections of 5mm length that aid in gauging dilatation catheter position relative to the guiding catheter located closest to the dilatation catheter adaptor is for femoral guiding catheters and the other marker is for brachial guiding catheters). The distal shaft of the catheter has an integrated shaft system. The shaft has a combination of single lumen and dual lumen tubing. One lumen is used for inflation of the balloon with contrast medium. The other lumen, in the distal shaft, permits the use of a quide wire to facilitate advancement of the dilatation catheter to and through the stenosis to be dilated. {5}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for KOSSEL. The logo consists of a circular graphic on the left and the text "科塞尔医疗 KOSSEL" on the right. The circular graphic is made up of several curved lines that form a spiral shape. The text is in Chinese and English, with the English text "KOSSEL" appearing below the Chinese text. The guidewire enters the catheter tip and advances coaxially out the distal Rx port, thereby allowing both coaxial quidance and rapid exchange of the catheter with a single standard length guidewire. One radiopaque Platinum/lridium marker band is located within the balloon segment to aid in positioning the balloon in the stenosis, and are designed to provide an expandable segment of known diameter and length at a specific pressure. The design of this dilatation catheter does not incorporate a lumen for distal dye injections and distal pressure measurements. ## Comparison with predicate device: Percutaneous Transluminal Coronary Angioplasty (PTCA) is the technological principle for both Selebrek PTCA Balloon Dilatation Catheter and the predicate device. PTCA is based on the use of Percutaneous Coronary Intervention (PCI) devices for the purpose of myocardial perfusion. The Selebrek PTCA Balloon Dilatation Catheter and predicate device have following same technological elements: - Operating principle - balloon dilatation of stenotic portion by pressurization of inflation medium - Intended User – for patients evidencing coronary ischemia for the purpose of improving myocardial perfusion - Fundamental catheter design – balloon, shaft, radiopaque marker, hub, coating - Shaft type – Rapid exchange - Sterilization - Ethylene oxide - Pressure - Nominal Pressure and Rated Burst Pressure There are following minor technological differences between Selebrek PTCA Balloon Dilatation Catheter and predicate device: - Combination of balloon diameter and balloon length - Distal shaft diameters - Catheter effective length - Crossing profile - Marker bands - Miniwrap folds Any differences that exist between the Selebrek PTCA Balloon Dilatation Catheter and the predicate device did not raise any new questions of safety or effectiveness. {6}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for KOSSEL, a medical company. The logo features a circular design with concentric rings on the left. To the right of the circular design, the company name "KOSSEL" is written in both Chinese and English. | | Subject Device | Predicate Device | | | |--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Comparison items | Selebrek PTCA Balloon Catheter | Orbus Neich Medical, Sapphire II PRO<br>Balloon Dilatation Catheter K180921 | Comments | | | Intended Use/Indications for<br>use | The Selebrek PTCA Balloon Dilatation<br>Catheter is indicated for balloon<br>dilatation of the stenotic portion of a<br>coronary artery or bypass graft stenosis<br>for the purpose of improving<br>myocardial perfusion. | The Sapphire® II PRO Balloon Dilatation<br>Catheter (Ø1.0-1.25mm configurations) is<br>indicated for:<br>● Balloon pre-dilatation of a stenotic portion<br>of a coronary artery or bypass graft stenosis<br>(≥70% stenosis) for the purpose of improving<br>myocardial perfusion.<br>The Sapphire® II PRO Balloon Dilatation<br>Catheter (Ø1.5-4.0mm configurations) is<br>indicated for:<br>● Balloon dilatation of the stenotic portion<br>of a coronary artery or bypass graft stenosis in<br>patients evidencing coronary ischemia for the<br>purpose of improving myocardial perfusion.<br>● Balloon dilatation of a coronary artery<br>occlusion for the treatment of acute myocardial<br>infarction.<br>The Sapphire® II PRO Balloon Dilatation<br>Catheter is also indicated for:<br>● percutaneous transluminal angioplasty in the<br>peripheral vasculature, including renal, femoral,<br>popliteal, infra-popliteal, tibial, and peroneal<br>arteries. | ● The intended use of predicate<br>device is larger than that of the subject<br>device, which not only for coronary use<br>but also support the peripheral use;<br>● The intended use of subject device<br>is identical to that of the predicate<br>device balloon size 1.5~4.0mm, while<br>the description of predicate device is<br>more detailed and specific.<br>● The Subject Device Selebrek PTCA<br>Balloon Catheter is intended use in<br>patients evidencing coronary ischemia<br>for improving myocardial perfusion.<br>The differences in description are not<br>critical to the intended therapeutic, or<br>surgical use of the device, so the<br>differences do not raise any new<br>questions of safety or effectiveness | | | | | Subject Device | Predicate Device | | | Comparison items | | Selebrek PTCA Balloon Catheter | Orbus Neich Medical, Sapphire II PRO<br>Balloon Dilatation Catheter K180921 | Comments | | Classification name | | Percutaneous Transluminal Coronary<br>Angioplasty (PTCA) Catheter<br>21 CFR 870.5100 (class II, special control) | Percutaneous Transluminal Coronary<br>Angioplasty (PTCA) Catheter<br>21 CFR 870.5100 (class II, special control) | Identical | | Product code | | LOX | LOX | Identical | | Chemical<br>composition | Balloon<br>material | Pebax | Pebax | Identical | | | Coating | PTFE / Hydrophilic | Hydrophilic (distal tip to guidewire exit<br>marker); Hydrophobic (guidewire lumen) | Equivalent | | Physical<br>characteristics<br>/Device<br>design | Balloon<br>Diameters<br>(mm) | 1.5 | 1.0 - 4.0 | Equivalent | | | Balloon<br>lengths (mm) | 6-30 | 5-30 | Equivalent | | | Balloon<br>compliance | Semi-Compliant | Semi-Compliant | Identical | | | Catheter Type | Rx | Rx | Identical | | | Marker bands | 1 marker | 1.0-1.5mm: 1 marker<br>1.5-4.0mm: 2 markers | Equivalent | | | Comparison items | Subject Device<br>Selebrek PTCA Balloon Catheter | Predicate Device<br>Orbus Neich Medical, Sapphire II PRO<br>Balloon Dilatation Catheter K180921 | Comments | | | Guide wire size<br>(inch) | 0.014" | 0.014" | Identical | | | Proximal shaft<br>diameter | 2.0 F | 2.0 F | Identical | | | Distal shaft<br>diameter | 2.36 F | 1.0-1.5mm: 2.36F<br>1.75-4.0mm: 2.36F – 2.7F | Equivalent | | | Crossing<br>Profile (inch) | 0.026"~0.043" | 3.0mm: 0.031" | Equivalent | | | Miniwrap folds | 3 folds | 1.0mm: 2 folds;<br>1.25-4.0mm: 3 folds | Equivalent | | | Catheter<br>length (cm) | 142 | 140 | Equivalent | | | Rated<br>burst<br>Pressure (atm) | 14 | 14 | Identical | | | Nominal<br>pressure (atm) | 6 | 6 | Identical | | | Sterilization Method | EO | EO | Identical | {7}------------------------------------------------ Image /page/6/Picture/0 description: The image contains a logo for KOSSEL, a medical company. The logo features a circular design with swirling lines on the left. To the right of the circular design, the company's name is written in both Chinese characters and English. {8}------------------------------------------------ Image /page/7/Picture/0 description: The image contains a logo with a circular design on the left and the text "科塞尔医疗 KOSSEL" on the right. The circular design consists of multiple curved lines arranged in a spiral pattern. The text is in Chinese and English, with "KOSSEL" written in capital letters below the Chinese characters. The logo appears to represent a medical or healthcare-related organization. {9}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for KOSSEL. The logo consists of a circular graphic on the left and the text "科塞尔医疗 KOSSEL" on the right. The circular graphic is made up of several concentric circles. The text is in a sans-serif font, with the Chinese characters "科塞尔医疗" appearing above the English word "KOSSEL". ### Non-clinical testing/Performance data: Non-clinical bench testing was performed on the subject device to determine substantial equivalence. The testing performed is as follows: - Dimensional verification - Balloon rated burst pressure - Balloon fatigue - Balloon compliance - Balloon inflation and deflation time - Catheter bond strength - Tip pull strength - Flexibility and kinkinq - Torque strength - Radiopacity - Coating integrity - Particulate evaluation - Shelf life testing The results met all acceptance criteria and ensure that the Balloon Dilatation Catheter design and construction are suitable for its intended use as recommended by the Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters (FDA; September 8, 2010). #### Biocompatibility: To demonstrate the biological safety of the body-contacting materials and substantial equivalence of the Selebrek PTCA Balloon Dilatation Catheter to the compared device Selethru PTCA Balloon Dilatation Catheter which had been cleared by FDA (K182699), the following biocompatibility testing, which was performed on the compared device (Selethru PTCA Balloon Dilatation Catheter) in accordance with "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"; Guidance for Industry and Food and Drug Administration Staff" were leveraged: Cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemocompatibility (hemolysis, complement activation, and in vivo thromboresistance), pyrogenicity were conducted. The results of the testing show that the subject device in this submission met all acceptance criteria and the subject device is biocompatible. #### Conclusion: The Selebrek PTCA Balloon Dilatation Catheter included in this notification is equivalent to the previously cleared predicate device, the Sapphire® II PRO PTCA Balloon Dilatation Catheter (K180921) in terms of intended use and technological characteristics. Any differences that exist between the Selebrek PTCA Balloon Dilatation Catheter and the predicate device did not raise any new questions of safety or effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...