← Product Code [NKB](/productcode/NKB) · K211323

# ASTRA Spine System (K211323)

_Spinecraft, LLC · NKB · Jun 3, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K211323

## Device Facts

- **Applicant:** Spinecraft, LLC
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Jun 3, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The ASTRA Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic. lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra: degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudo-arthrosis). The ASTRA Spine System is also a sacral/iliac screw fixation system of the indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis). When used in a percutaneous, posterior approach with AVANT Spine MIS instrumentation, the ASTRA Spine System is intended for non-cervical pedicle fixation for the following indications: degenerative disc discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudo-arthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine. When used for posterior non-cervical pedicle screw fixation in pediatric patients, the ASTRA Spine System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the ASTRA Spine System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are intended to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.

## Device Story

ASTRA Spine System; top-loading, posterior spinal fixation system. Components: rods, cannulated/non-cannulated monoaxial/uniplanar/polyaxial screws, hooks, iliac/rod/cross connectors. Used by surgeons in clinical settings for spinal stabilization/fusion. System provides mechanical immobilization of spinal segments. No changes to function or design from predicate. Pediatric indications added for clarity.

## Clinical Evidence

No clinical studies were performed in support of this submission.

## Technological Characteristics

Materials: CoCr28Mo6 (ASTM F1537, ISO 5832-12) and Ti-6Al-4V-Eli (ASTM F136). Mechanical system; no energy source or software.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- ASTRA Spine System ([K150417](/device/K150417.md))
- CD Horizon Spinal System ([K202771](/device/K202771.md))
- EXPEDIUM Spine System, VIPER and VIPER 2 Systems ([K160904](/device/K160904.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

June 3, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

SpineCraft, LLC Ami Akallal-Asaad Vice President of Regulatory Affairs & Quality Assurance 777 Oakmont Lane, Suite 200 Westmont. Illinois 60559

Re: K211323

Trade/Device Name: ASTRA Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWP, KWQ Dated: April 27, 2021 Received: April 30, 2021

Dear Ami Akallal-Asaad:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K211323

Device Name ASTRA Spine System

## Indications for Use (Describe)

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

The ASTRA Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic. lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra: degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudo-arthrosis).

The ASTRA Spine System is also a sacral/iliac screw fixation system of the indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).

When used in a percutaneous, posterior approach with AVANT Spine MIS instrumentation, the ASTRA Spine System is intended for non-cervical pedicle fixation for the following indications: degenerative disc discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudo-arthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the ASTRA Spine System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the ASTRA Spine System is intended to treat pediatric patients diagnosed with the following conditions:

spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are intended to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

# 510(k) Summary

# SpineCraft ASTRA Spine System

In accordance with 21 CFR 807.92 of the Federal Code of Requlations the following 510(k) summary is submitted for the ASTRA Spine System.

Date Prepared: April 27, 2021

| 1. Submitter:          | Contact Person:                           |
|------------------------|-------------------------------------------|
| SpineCraft, LLC        | Ami Akallal-Asaad                         |
| 777 Oakmont Lane       | Vice President of Regulatory Affairs & QA |
| Westmont, IL 60559 USA | SpineCraft, LLC                           |
| Tel: 1 630-920-7300    | a.asaad@spinecraft.com                    |
| Fax: 1 630-920-7310    |                                           |

| 2.   Trade name:     | ASTRA Spine System                                          |
|----------------------|-------------------------------------------------------------|
| Common Name:         | Pedicle screw system                                        |
| Classification Name: | Thoracolumbosacral pedicle screw system per 21 CFR 888.3070 |
| Product Codes:       | NKB, KWP, KWQ                                               |
| Classification:      | Class II                                                    |

- 3. Predicate or legally marketed devices which are substantially equivalent:
#### Primary

- . ASTRA Spine System (K150417) / SpineCraft
Additional

- CD Horizon Spinal System (K202771) / Medtronic .
- . EXPEDIUM Spine System, VIPER and VIPER 2 Systems (K160904) / DePuy Synthes Spine
- 4. Description of the device:

The subject ASTRA Spine System is top loading, multiple components, posterior system which consists of rods, cannulated and non-cannulated monoaxial, uniplanar and polyaxial screws, hooks, iliac connectors, rod connectors, and cross connectors. Most of the components are available in a variety of sizes to match the patient's anatomy more closely.

The subject ASTRA Spine System includes the pediatric use in the indications and other labeling updates to provide more clarity to the device insert.

#### Materials:

CoCr28Mo6 per ASTM F1537 and ISO 5832-12 Ti-6Al-4V-Eli per ASTM F136

#### Function:

No changes have been made to the function of the ASTRA Spine System.

- 5. Summary of Similarities and Differences in Technological Characteristics and Performance:

{4}------------------------------------------------

The design and material characteristics of the subject device remain unchanged from the currently marketed predicate devices.

#### 6. Indications for Use:

The ASTRA Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudo-arthrosis).

The ASTRA Spine System is also a sacral/iliac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).

When used in a percutaneous, posterior approach with AVANT Spine MIS instrumentation, the ASTRA Spine System is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudo-arthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the ASTRA Spine System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the ASTRA Spine System is intended to treat pediatric patients diagnosed with the following conditions: spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

#### 7. Non-clinical Test Summary:

Non-clinical tests were not performed in support of this submission.

#### Clinical Test Summary: 8.

No clinical studies were performed in support of this submission.

#### 9. Conclusion:

Based on the supporting evidence provided, we believe that the subject ASTRA Spine System is substantially equivalent to the below predicate devices in terms of indications for use, design, material, and function.

#### Primary (Predicate1)

- ASTRA Spine System (K150417) / SpineCraft ●
#### Additional

- Predicate 2: CD Horizon Spinal System (K202771) / Medtronic .
- Predicate 3: EXPEDIUM Spine System, VIPER 2 Systems (K160904) / DePuy . Synthes Spine.

---

**Source:** [https://fda.innolitics.com/device/K211323](https://fda.innolitics.com/device/K211323)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
