Ultipor U55/U55N Breathing Circuit Filter and Heat and Moisture Exchanger
Applicant
Pall Corporation
Product Code
CAH · Anesthesiology
Decision Date
Nov 16, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5260
Device Class
Class 2
Indications for Use
The Pall® Ultipor® U55/U55N Filter is a single use bacterial/viral filter and heat and moisture exchanger (HME) for patient side or machine side installation in breathing systems. It is designed to reduce bacterial/viral transmissions between the patient, the equipment and the environment and to reduce the loss of patient heat and humidity. The Ultipor® U55/U55N breathing circuit filter has >99.999% bacterial and >99.995% viral efficiency. The filter is for single patient use for adult patients, and is intended for use within breathing systems in healthcare and home environments where ventilation is required and for a maximum duration of 24 hours.
Device Story
Disposable, bi-directional bacterial/viral filter and heat and moisture exchanger (HME); installed at patient or machine end of breathing circuits. Consists of pleated hydrophobic filter media within a molded plastic housing with tapered fittings. Operates via size exclusion (direct interception), inertial impaction, and diffusion to remove airborne/waterborne contaminants. Acts as HME by conserving heat/humidity from exhaled air for return during inspiration. Used in healthcare and home settings for adult patients on ventilators. Healthcare providers install the device; output is filtered, humidified, and warmed air delivered to the patient. Benefits include reduced transmission of pathogens between patient, equipment, and environment, and maintenance of patient airway moisture/heat.
Clinical Evidence
Bench testing only. Evaluated biocompatibility (ISO 10993), microbial retention (aerosol/liquid), filtration efficiency (PAO/NaCl), hydrostatic head resistance, heat/moisture exchange efficiency, pressure drop, leakage, connector compliance (ISO 5356-1), nebulization effects, and particulate/VOC emissions (ISO 18562). All tests met predefined acceptance criteria for unaged and aged (up to 5 years) samples.
Technological Characteristics
Disposable filter with pleated hydrophobic media in molded housing. Features 22mm/15mm conical fittings. Filtration mechanism: size exclusion, inertial impaction, diffusion. Connectivity: mechanical integration into breathing circuits. Non-sterile. Performance: >99.999% bacterial/viral efficiency, 100% waterborne microbial removal. Complies with ISO 9360-1 and ISO 5356-1.
Indications for Use
Indicated for adult patients requiring ventilation in healthcare and home environments. Used as a single-use bacterial/viral filter and heat and moisture exchanger (HME) in breathing systems for up to 24 hours.
Regulatory Classification
Identification
A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.
Predicate Devices
Intersurgical, Inc. Hydro-Guard Mini Breathing Filter and Heat and Moisture Exchanger (Model 1745) (K092409)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
November 16, 2021
Pall Corporation % Randy Prebula Partner Hogan Lovells U.S. LLP 555 Thirteenth Street NW Washington, District of Columbia 20004
Re: K211286
Trade/Device Name: Ultipor U55/U55N Breathing Circuit Filter and Heat and Moisture Exchanger Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Regulatory Class: Class II Product Code: CAH Dated: October 18, 2021 Received: October 18, 2021
Dear Randy Prebula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
#### K211286
Device Name
Ultipor® U55/U55N Breathing Circuit Bacterial Filter and Heat and Moisture Exchanger
Indications for Use (Describe)
The Pal® Ultipor® U55/U55N Filter is a single use bacterial/viral filter and heat and moisture exchanger (HME) for patient side or machine side installation in breathing systems. It is designed to reduce bacterial/viral transmissions between the patient, the equipment and the environment and to reduce the loss of patient heat and humidity. The Ultipor® U55/U55N breathing circuit filter has >99.999% bacterial and >99.995% viral efficiency.
The filter is for single patient use for adult patients, and is intended for use within breathing systems in healthcare and home environments where ventilation is required and for a maximum duration of 24 hours.
Type of Use (Select one or both, as applicable)
図 Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY - K211286
### Pall Corporation's Ultipor® U55/U55N Breathing Circuit Bacterial Filter and Heat and Moisture Exchanger
| Submitter: | Pall Corporation<br>25 Harbor Park Drive<br>Port Washington, NY 11050, USA |
|----------------------------|-----------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Brian Goetz<br>Senior Manager, Regulatory Affairs<br>Telephone Number: 516-801-9267<br>Fax Number: 516-801-9643 |
| Date Prepared: | November 2, 2021 |
| Trade Names: | Pall® Ultipor® U55/U55N Filter |
| Common Name: | Filter, Bacterial, Breathing-Circuit |
| Classification Regulation: | 21 C.F.R. § 868.5260, Breathing circuit bacterial filter |
| Regulatory Class: | Class II |
| FDA Product Code: | CAH |
| Predicate Device: | Intersurgical, Inc. Hydro-Guard Mini Breathing Filter and Heat<br>and Moisture Exchanger (Model 1745) (K092409) |
### Intended Use/Indications for Use
The Pall® Ultipor® U55/U55N Filter is a single use bacterial/viral filter and heat and moisture exchanger (HME) for patient side or machine side installation in breathing systems. It is designed to reduce bacterial/viral transmissions between the patient, the equipment and the environment and to reduce the loss of patient heat and humidity. The Ultipor® U55/U55N breathing circuit filter has
>99.999% bacterial and >99.995% viral efficiency.
The filter is for single patient use for adult patients, and is intended for use within breathing systems in healthcare and home environments where ventilation is required and for a maximum duration of 24 hours.
## Device Description
The Pall® Ultipor®U55/U55N Breathing Circuit Bacterial Filter and Heat and Moisture Exchanger is a disposable, highly efficient, bi-directional bacterial/viral filter for patient ventilation solutions and is comprised of the following components:
- . Filter housing - Is comprised of two molded halves, an inlet housing and an outlet housing that are joined and sealed together. The finished filter housing holds the pleated hydrophobic filter media and provides conical (tapered) fittings on opposing sides of the finished housing to provide connection to the conventional equipment used in the breathing circuit. One side of the finished housing has a coaxial conical fitting, with a conical fitting located on the opposing side.
{4}------------------------------------------------
- Hydrophobic filter media Provides airborne bacterial removal efficiency of >99.999%, . airborne viral removal efficiency of >99.995% and waterborne microbial contaminant removal efficiency of 100%. If the filter is used at the patient end, the filter media also acts as a heat and moisture exchanger (HME) by conserving a proportion of the heat and humidity present in the patient's exhaled air and returning it to the patient on the next inspiration.
# Summary of Technological Characteristics
The subject and predicate devices have the same fundamental design, consisting of the filter media and housing. In addition, the subject and predicate filters both operate based on size exclusion (direct interception filtration) as the primary filtration mechanism, supplemented by effects of inertial impaction, and diffusion impaction, to prevent certain particles from passing through the filter to the other side.
| | Subject Device | Predicate Device |
|----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Element | Pall® Ultipor® U55 and U55N<br>Breathing Circuit Bacterial Filter and<br>Heat and Moisture Exchanger | Intersurgical 1745 Hydro-Guard<br>Mini Breathing Filter and Heat<br>and Moisture Exchanger<br>(K092409) |
| Indications for Use | The Pall® Ultipor® U55/U55N Filter is a<br>single use bacterial/viral filter and heat and<br>moisture exchanger (HME) for patient side<br>or machine side installation in breathing<br>systems. It is designed to reduce<br>bacterial/viral transmissions between the<br>patient, the equipment and the environment<br>and to reduce the loss of patient heat and<br>humidity. The Ultipor® U55/U55N breathing<br>circuit filter has >99.999% bacterial and<br>>99.995% viral efficiency.<br>The filter is for single patient use for adult<br>patients, and is intended for use within<br>breathing systems in healthcare and home<br>environments where ventilation is required<br>and for a maximum duration of 24 hours. | The Intersurgical Hydro-Guard Mini<br>Filter and Heat and Moisture<br>Exchanger is for use at the patient<br>and equipment connections. It is<br>designed to reduce bacterial/viral<br>transmissions between the patient<br>and equipment and to reduce the<br>loss of patient heat and humidity.<br>The filter is for single patient use<br>for an adult target population, is<br>intended for use within breathing<br>systems in healthcare and home<br>environments where ventilation is<br>required and a maximum duration<br>of 24 hours. |
| Installation positions | Patient and machine ends | Patient and machine ends |
| Filtration efficiency<br>(%) | >99.999% bacterial; >99.995% viral | >99.999% |
| Filtration efficiency -<br>waterborne bacterial<br>removal (%) | 100 | Not available |
| Minimum tidal<br>volume (mL) | 150 (See Note 1 below) | >200 |
| Maximum tidal<br>volume (mL) | 1000 | 1000 |
| Pressure drop<br>(cmH2O) @ 60 L/min | 2.5 | 2.9 |
| Heat and moisture<br>exchanger loss H2O<br>@ 500 mL tidal<br>volume (mg/L) | 15 | 23 |
{5}------------------------------------------------
| Compressible volume<br>(dead space) (mL)<br>approx. | 50 | 63 |
|-----------------------------------------------------|-------------------|-------------------|
| Gas Leakage<br>(mL/min) | < 4 | <5 |
| Compliance (ml/kPa) | <1 | <1 |
| Connections patient<br>side (mm) | 22 male/15 female | 22 male/15 female |
| Connections<br>machine side (mm) | 22 female | 22 female/15 male |
| Dry weight (g) | 50 | 30 |
| Filter Life (hrs) | 24 | 24 |
NOTE: 1. Lowest tidal volume while providing adequate patient ventilation should be determined by the physician by clinical assessment, inter volume, ventilation settings and breathing circuit configuration.
## Performance Data
Testing was performed on the subject device or on representative filters reflecting the same fundamental design and materials, and the same manufacturing methods. The following non-clinical tests were performed to demonstrate the performance of the subject Ultipor® U55/U55N Filter and support the proposed indications for use:
| Methodology | Purpose | Acceptance Criteria | Results |
|---------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | Evaluate device's<br>biological safety<br>for the intended<br>use, in accordance<br>with ISO 10993-1<br>and FDA's<br>corresponding<br>guidance<br>document | • Cytotoxicity (L929 MEM<br>elution) per ISO 10993-5;<br>• Sensitization and<br>intracutaneous injection per<br>ISO 10993-10;<br>• Acute systemic toxicity per<br>ISO 10993-11 with both polar<br>and non-polar solvents (in<br>lieu of testing to ISO 18562-<br>4); and<br>• Material-mediated<br>pyrogenicity per ISO 10993-<br>11. | All results were acceptable. |
| Microbial<br>retention | Evaluate aerosol<br>bacterial and viral<br>removal | >99.999% effectiveness for<br>bacteria removal and<br>>99.995% for virus removal | Unaged and aged (5 years) filters<br>demonstrated bacterial<br>effectiveness of >99.999% and<br>viral effectiveness of >99.995%. |
| Microbial<br>retention | Evaluate liquid<br>bacterial removal | 100% retention | No bacteria were recovered from<br>the water placed on the machine<br>side following the challenge for<br>any unaged or aged (3 & 5 years)<br>filters. This testing also supports<br>filter media integrity and<br>hydrophobicity. |
| Poly-alpha-<br>olefin (PAO)<br>Removal1 | Evaluate filtration<br>efficiency | Penetration of ≤0.09%, | All unaged and 5-year aged filters<br>had a penetration of 0.09%. |
| Sodium<br>Chloride<br>Particulate<br>Removal | Evaluate filtration<br>efficiency | Pre- and post-conditioning<br>penetrations of <1% | Unaged and 5-year aged filters<br>had acceptable performance for<br>pre-conditioning and post-<br>conditioning penetration. |
| Methodology | Purpose | Acceptance Criteria | Results |
| Filter Media<br>Integrity with<br>50 cm<br>hydrostatic<br>head | Evaluate<br>hydrophobicity<br>(resistance to<br>penetration by a<br>measured column of<br>liquid) | No water breakthrough of the<br>filter media for 60 seconds<br>after application of a 50 cm<br>head of sterile water | No unaged or 3-year aged filter<br>showed water breakthrough. |
| Air Flow<br>Resistance<br>(Pressure<br>Drop) Pre- and<br>Post-<br>hydrostatic<br>head | Evaluate any<br>changes in the<br>filter's resistance<br>to air flow before<br>and after being<br>subjected to a 15<br>cm hydrostatic<br>head. | At 60 L/min, average<br>inspiratory/expiratory flow<br>resistance should be <2.5 cm<br>H2O pre-hydrostatic head and<br>≤6.0 cm H2O post-hydrostatic<br>head, and no evidence of water<br>breakthrough within 60 seconds<br>of the 15 cm hydrostatic head. | The pre-hydrostatic head and<br>post-hydrostatic maximum flow<br>resistance were passing across<br>all unaged, 3-year aged, and 5-<br>year aged filters. |
| Heat and<br>Moisture<br>Exchange<br>efficiency | Evaluate moisture<br>loss for tidal<br>volumes of 250<br>mL, 500 mL, 750<br>mL, and 1000 mL | <12 mg/L (10-12 mg H2O/L) at<br>250 ml tidal volume;<br><16 mg/L (13-16 mg H2O/L) at<br>500 ml tidal volume;<br><20 mg/L (19 mg H2O/L) at<br>750 ml tidal volume;<br><22 mg/L (19-22 mg H2O/L) at<br>1000 ml tidal volume. | The average moisture loss met<br>the acceptance criteria for the<br>specified tidal volumes for<br>unaged, and aged (1, 3, & 5<br>years filters. |
| | Evaluate pressure<br>drop pre and post<br>moisture loss<br>testing | The inspiratory/expiratory flow<br>resistance should be <2.5 cm<br>H2O at 60 L/min | For unaged as well as 1, 3 & 5<br>year aged filters, the maximum<br>inspiratory and expiratory flow<br>resistance were passing, showing<br>that the pressure drop does not<br>increase when conditioned for the<br>length of filter use or when<br>challenged with liquid. |
| | Evaluate leakage<br>(pressure decay)<br>under an applied<br>pressure pre and<br>post moisture loss<br>testing | ISO 9360-1 Section 6.4 does<br>not specify acceptance<br>criteria, only to record the<br>volume of air required;<br>however, Pall has specified an<br>internal acceptance criteria for<br>leakage of <25 mL/min at 7 ±<br>0.5 kPa. | The maximum leakage values for<br>unaged as well as for 1, 3 & 5<br>year aged filters met predefined<br>criteria both pre- and post-<br>moisture loss. |
| | Evaluate<br>compliance in<br>accordance with<br>ISO 9360 pre- and<br>post- moisture loss<br>testing | ISO 9360-1 Section 6.5 does<br>not specify an acceptance<br>criterion, only to record the<br>volume of air required. | There was no significant increase<br>in flow rate following filter<br>conditioning. All test values for<br>unaged and aged (1, 3, & 5<br>years) filters were acceptable. |
| Housing<br>Integrity (for<br>pressure<br>decay) | Evaluate filter<br>leakage at two<br>different pressures | Leakage <4 ml/min for unaged<br>filters at 70 mbar (7 kPa) and<br>150 mbar (15 kPa); leakage<br><25 ml/min at 70 mbar (7 kPa)<br>for aged filters. | Unaged as well as 1 & 3 year<br>aged filters had leakage values<br>meeting the predefined<br>acceptance criteria at the<br>evaluated pressures. |
| Connector<br>Compliance | Demonstrate inlet<br>and outlet ports<br>comply with ISO<br>5356-1 | All ports must comply with ISO<br>5356-1 | For both 15 mm and 22 mm<br>connectors, axial force and<br>rotation for unaged and 1 & 3<br>year aged filters were in<br>accordance with the standard. |
| Methodology | Purpose | Acceptance Criteria | Results |
| Nebulization | Evaluate changes<br>in air flow<br>resistance across<br>filter when used<br>with nebulized<br>drugs at patient<br>and machine ends<br>of breathing circuit | Air flow resistance <6 cm H2O<br>at 60 L/min airflow after 24<br>hours of nebulization | Unaged & 3 year aged filters:<br>Highest flow resistance values<br>were within specification for both<br>filter locations and for drugs<br>delivered by air-driven and<br>ultrasonic nebulizers as well as<br>drugs delivered by metered dose<br>inhalers, as applicable. |
| Particulate<br>Analysis | Assess if airborne<br>particulate is<br>emitted into the<br>gas stream in<br>accordance with<br>ISO 18562 | Particulate category PM2.5<br>and PM10 acceptable<br>concentrations: 12 µg/m3 and<br>150 µg/m3, respectively | The minimum, maximum, and<br>average particulate<br>concentrations are all below<br>acceptable limits |
| Volatile<br>Organic<br>Compounds<br>(VOCs)<br>Analysis | Assess if airborne<br>VOCs are emitted<br>from the filter into<br>the gas stream per<br>ISO 18562 | Threshold of Toxicological<br>Concern (TTC) of 360 µg/d for<br>the first 24 hours<br>and a limit of 120 µg/d for the<br>remaining 24 hours | Exposure of the individual VOCs<br>released are all below pre-defined<br>TTC |
| Shelf life<br>(stability /<br>maintained<br>performance) 1 | Evaluate shelf life | 3 years from the date of<br>manufacture based on the<br>stability testing | Stability of device confirmed from<br>performance testing conducted<br>on 3-year real-time aged and 5-<br>year accelerated aged devices. |
{6}------------------------------------------------
{7}------------------------------------------------
1 Testing also performed during filter manufacturing process.
² The subject filter is not supplied sterile and therefore does not have an expiry date based on validation of the integrity of a sterile barrier seal.
## Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, Intersurgical Inc.'s Model 1745 Hydro-Guard Mini Breathing Filter and Health and Moisture Exchanger (K092409).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.