Compressible Limb and Circulation Therapy System, Model POWER-Q2300
Applicant
Wonjin Mulsan Co., Ltd.
Product Code
IRP · Physical Medicine
Decision Date
Sep 14, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
POWER-Q2300 is intended for the temporary relief of minor muscle aches and pains, and for temporary increase in blood circulation to the treated areas in people who are in good health. POWER-Q2300 simulates kneading of tissues by using an inflatable garment (cuff).
Device Story
POWER-Q2300 is an AC-powered, inflatable tube massager used in clinics, hospitals, athlete training, and home environments. The system consists of a main control unit, power cord, hoses, and 4-chamber inflatable cuffs (leg, arm, hip, half-leg). The user selects from four compression modes (A, B, C, D) and treatment times (15 or 30 minutes). The main unit sequentially inflates and deflates the cuff chambers to simulate tissue kneading and stroking. Pressure ranges from 20 to 200 mmHg. The device is intended for over-the-counter use by healthy individuals to relieve minor muscle aches and improve circulation. It is designed to be used over clothing to avoid skin irritation or sensitization.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by bench testing, including pressure accuracy, seam strength, and fail mode testing, alongside compliance with electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and usability (IEC 60601-1-6, IEC 62366-1) standards.
Technological Characteristics
AC-powered (110-120VAC, 50/60Hz) pneumatic compression system. Features 4-chamber inflatable cuffs connected via single or double hoses. Pressure range: 20-200 mmHg. Four operation modes with adjustable (0/5/10/30 sec) or fixed (30 sec) rest intervals. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, and IEC 62366-1. Software developed per IEC 62304.
Indications for Use
Indicated for healthy individuals for temporary relief of minor muscle aches/pains and temporary increase in blood circulation to treated areas (leg, arm, hip, half-leg).
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym and name are in blue, with the acronym in a larger font size than the name. The name is written as "U.S. FOOD & DRUG ADMINISTRATION".
September 14, 2021
Wonjin Mulsan Co., Ltd. HaYong Jung QMR 89. Geomdan-ro. Seo-gu Incheon, 22653 Republic of Korea
Re: K211283
Trade/Device Name: Compressible Limb and Circulation Therapy System, Model POWER-02300 Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: August 24, 2021 Received: August 31, 2021
Dear HaYong Jung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vivek Pinto, PhD Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211283
#### Device Name
Compressible Limb and Circulation Therapy System, Model POWER-Q2300
Indications for Use (Describe)
POWER-Q2300 is intended for the temporary relief of minor muscle aches and pains, and for temporary increase in blood circulation to the treated areas in people who are in good health. POWER-Q2300 simulates kneading of tissues by using an inflatable garment (cuff).
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
[as required by 21 CFR 807.02]
| Date Prepared: | Sep. 8, 2021 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Wonjin Mulsan Co., Ltd.<br>89, Geomdan-ro, Seo-gu, Incheon, 22653, Republic of Korea<br>Tel: 82-32-816-0552, Fax: 82-32-816-0557<br>Establishment Registration Number: 3006797972<br>Contact Person: HaYong Jung<br>Email: wonjin@wonjininc.com |
| Trade Name: | Compressible Limb and Circulation Therapy System<br>Model POWER-Q2300 |
| Common Name: | Powered Inflatable Tube Massager |
| Classification Name: | Massager, Powered Inflatable Tube |
| Regulation Number: | 890.5650 |
| Product Code: | IRP |
| Classification: | Class II |
| Predicate Device: | Rapid Reboot Compression Therapy System which was cleared for<br>marketing under K182668 |
## Device Description
POWER-Q2300 is comprised of a main body, a power cord, cuffs, and hoses for connecting the device to the cuffs. The device is AC-powered. There are four cuffs to apply to different body areas, such as leg, arm, hip, and half-leg. The cuffs have 4-chambers. The cuff inflates and deflates sequentially to apply the pressure on the target body areas which are controlled by the main body.
Double hose is to connect main body and leg cuff. Single hose is to connect main body and arm cuff (or hip cuff or half-leg cuff). Extension zipper is for bigger leg. Blocked jack is to block an air outlet that will not be used. The colors of the cuffs and extension zipper are available in pink or gray.
## Indications for Use
POWER-Q2300 is intended for the temporary relief of minor muscle aches and for temporary increase in blood circulation to the treated areas in people who are in good health. POWER-Q2300 simulates kneading and stroking of tissues by using an inflatable garment (cuff).
| Item | Description |
|----------------|----------------------------------------------------------------------------------------------|
| Operation Mode | 4 modes such as Mode A, Mode B, Mode C, and Mode D |
| Pressure Range | 20 mmHg to 200 mmHg |
| Operation Time | 15 and 30 minutes |
| Rest Time | Adjustable at interval of 0/5/10/30 sec at modes A and B<br>Fixed at 30 sec at modes C and D |
## Technologic Characteristics
{4}------------------------------------------------
## Non-Clinical Testing
The device has been tested and met the requirements of the following standards:
IEC 60601-1:2005+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance ANSVAAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R) 2012 and A2:2010/(R)2012
IEC 60601-1-6 Edition 3.1 2013-10 Medical electrical equipment - Part 1-6: General requirements for safety and essential performance - Collateral Standard: Usability
IEC 60601-1-11 Edition 2.0 2015-01 Medical electrical equipment - Part 1-11: General requirements for safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
IEC 62366-1 Edition 1.0 2015-02 Medical device – Part 1: Application of usability engineering to medical devices
IEC 62304 Edition 1.1 2015-06 Medical device software - Software life cycle process
Performance testing for pressure accuracy, seam strength, and fail mode testing has been performed.
The biocompatibility of the materials has not been verified by the FDA and contact of the cuffs/ accessories to direct skin may lead to skin irritation, skin sensitization and/or cytotoxicity. The device is intended to be used only over clothing.
## Clinical Testing
No clinical testing was performed.
## Conclusions
Wonjin Mulsan Co., Ltd. believes that all applicable items of information specified in this submission have been supplied in full. The enclosed non-clinical data demonstrate that the subject device described herein raises no new questions concerning safety or effectiveness and may therefore be properly considered by the Agency as substantially equivalent to the predicate device that has previously been legally distributed in interstate commerce in the United States.
{5}------------------------------------------------
# Comparison with Predicate Devices
| Item | Subject Device | Predicate Device | Differences Discussion |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | POWER-Q2300 | Rapid Reboot Compression Therapy System | NA |
| Manufacturer | Wonjin Mulsan Co., Ltd. | Rapid Reboot Recovery Products, LLC | NA |
| Classification | Class II Device, IRP (21 CFR 890.5650) | Class II Device, IRP (21 CFR 890.5650) | NA |
| 510(k) number | | K182668 | NA |
| Indications for use | POWER-Q2300 is intended for the temporary relief of minor muscle aches and pains, and for temporary increase in blood circulation to the treated areas in people who are in good health. POWER-Q2300 simulates kneading and stroking of tissues by using an inflatable garment (cuff). | The Rapid Reboot Compression Therapy System is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Rapid Reboot Compression Therapy System simulates kneading and stroking of tissues by using an inflatable garment. | Identical |
| Prescriptive or OTC | OTC | OTC | Identical |
| Environment of Use | Clinics, hospital, athlete training, and home environments | Clinics, hospital, athlete training, and home environments | Identical |
| Power Source(s) | 110-120VAC, 50/60Hz | 110VAC, 60Hz | Minor difference |
| Weight | 3.1 kg | 2.63 kg (5.8 pounds) | Minor difference |
| Dimensions (W x H x D) | 290 x 260 x 172 mm | 25.4 x 16.51 x 12.7 cm (10" x 6.5" x 5") | Minor difference |
| Device pressure range | 20-200 mmHg | 0-200 mmHg | Pressure range of subject device is smaller than that of predicate device. |
| Treatment Time | User can select operation time from 15 and 30 minutes. | User determines therapy time. Choose from 10, 20, or 30 minute session time, with option | Minor difference |
| | | to add additional 10 minutes to<br>any therapy time. | |
| Number of<br>inflatable<br>appliance<br>segments | 4 | 4 | Identical |
| Anatomical<br>site | Leg (including of foot, calf,<br>knee, upper leg)<br>Hip (including of upper leg,<br>glutes, hips, lower lack)<br>Arm (including of entire arm,<br>shoulder)<br>Half-Leg (including of foot,<br>calf, knee) | Leg (including of foot, calf,<br>knee, upper leg)<br>Hip (including of upper leg,<br>glutes, hips, lower lack)<br>Arm (including of entire arm,<br>shoulder, upper chest and back) | Identical except<br>half-leg |
| Mode of<br>compression | 4 modes (Mode A, Mode B,<br>Mode C and Mode D) | 2 modes (Sequential and<br>Peristaltic) | Refer to the<br>visual<br>description<br>below. |
| Mode of<br>compression<br>(visual<br>description) | Image: Mode A | Image: Mode A | Mode A of<br>POWER-Q2300<br>is similar to<br>Mode A of the<br>predicate. |
| | Image: Mode B | Image: Mode A | Mode B of<br>POWER-Q2300<br>is identical to<br>Mode A of the<br>predicate. |
| | Image: Mode C | Image: Mode B | Mode C of<br>POWER-Q2300<br>is identical to<br>Mode B of the<br>predicate. |
| | Image: Mode D | Image: Mode A<br>Image: Mode B | Mode D of<br>POWER-Q2300<br>is a combination<br>of Mode B and<br>Mode C, and is<br>identical to a<br>combination of<br>Mode A and<br>Mode B of the<br>predicate. |
| Appearance<br>and size of<br>Cuffs | Leg | Leg | Minor<br>difference |
| | Image: Leg cuffs | Image: Leg cuffs<br>X-Short: 14" x 41"<br>Short: 14" x 43"<br>Medium: 14" x 45"<br>Long: 14" x 48"<br>X-Long: 14" x 52"<br>Image: Leg cuff | Minor difference |
| | Image: Leg cuff with dimensions<br>310<br>940 | | |
| | Medium: 310x940mm<br>Large: 390x990 | | |
| | Hip<br>Image: Hip cuff | Hip<br>Image: Hip cuff<br>Regular:<br>26" x 32"<br>Large: 26"<br>35"<br>Image: Hip cuff | Minor<br>difference |
| | Image: Hip cuff with dimensions<br>530<br>720 | | |
| | 530x720mm | | |
| | Arm<br>Image: Arm cuff | Arm<br>Image: Arm cuffs<br>Regular: 18" x 38"<br>Long: 18" x 44"<br>Image: Arm cuff | Minor<br>difference |
| | Image: Arm cuff with dimensions<br>255<br>870 | | |
| | Medium: 255x870mm<br>Large: 330x895mm | | |
| | Half-Leg | None | Minor<br>difference |
| Other<br>accessories | Image: Half-Leg<br>Image: Half-Leg with dimensions 300, 700<br>300x700mm<br>Extension zipper<br>Image: Extension zipper with dimensions 120, 900<br>Medium: 120x900mm<br>Large: 150x950mm<br>Double hose<br>Image: Double hose<br>Length: 1,430mm<br>Single hose<br>Image: Single hose<br>Length: 1,320mm<br>Blocked jack<br>Image: Blocked jack with dimensions 47.1, 13.9, 53.8<br>53.8x47.1x13.9mm | "Not publicly available in the<br>official 510K Summary" | Minor<br>difference |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.