MIS Ti-base abutment is a titanium base placed onto MIS dental implants to provide support for customized cement-retained or screw retained single or multiple-unit restorations. It is used with a digitally designed mesostructure. MIS Ti-base and the mesostructure make up a two- piece abutment used in conjunction with MIS dental implants, to be placed in the upper or lower jaw arches, in order to restore masticatory function. Narrow platform Ti-bases are indicated for use only in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws. MIS short implants are to be used only with straight abutments. Mesostructures for use with the MIS Ti-base abutment are to be made from inCoris ZI, IPS e.max® CAD Abutment or VITA ENAMIC® (IS), designed and manufactured using Sirona CEREC SW version 4.6.1 software. MIS Ti-base abutments are intended for use with the following MIS implants: C1 conical connection implant system, V3 conical connection implant system, SEVEN internal hex implant system, M4 internal hex implant system, Lance+ internal hex implant system.
Device Story
Device is a two-piece dental abutment system comprising a titanium base (Ti-base) and a CAD/CAM-milled ceramic mesostructure. Ti-base connects to MIS dental implants via prosthetic screw; mesostructure is cemented to Ti-base. Input: digital design parameters within Sirona CEREC SW 4.6.1 software. Output: customized abutment for single/multiple-unit restorations. Used in dental clinics by clinicians. Software enforces design constraints (max 20° angulation, min wall thickness) to ensure structural integrity. System restores masticatory function and aesthetics. Benefits include patient-specific restorative fit and compatibility with existing MIS implant platforms.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing, including ISO 14801 fatigue testing (2,000,000 cycles) and sterilization validation.
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F136). Components: Titanium base and prosthetic screw. Mesostructure materials: inCoris ZI, IPS e.max CAD, VITA ENAMIC (IS). Connection: Conical or internal hex. Sterilization: Non-sterile, user-performed moist steam sterilization. Software: CEREC SW 4.6.1 with locked design library.
Indications for Use
Indicated for partially or fully edentulous patients requiring single or multiple-unit cement or screw-retained restorations in the upper or lower jaw. Narrow platform bases restricted to mandibular central/lateral incisors and maxillary lateral incisors. Compatible with specific MIS implant systems (C1, V3, SEVEN, M4, Lance+).
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
July 28, 2021
Dentsply Sirona Rebecca Sporer Regulatory Affairs Specialist 221 West Philadelphia Street, Suite 60W York, Pennsylvania 17401
# Re: K211225
Trade/Device Name: MIS Ti-base abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: June 28, 2021 Received: June 30, 2021
# Dear Rebecca Sporer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Andrew Steen Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name MIS Ti-base abutment
#### Indications for Use (Describe)
MIS Ti-base abutment is a titanium base placed onto MIS dental implants to provide support for customized cementretained or screw retained single or multiple-unt restorations. It is used with a digitally designed mesostructure. MIS Tibase and the mesostructure make up a two- piece abutment used in conjunction with MIS dental in the upper or lower jaw arches, in order to restore masticatory function.
Narrow platform Ti-bases are indicated for use only in the mandbular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws.
MIS short implants are to be used only with straight abutments. Mesostructures for use with the MIS Ti-base abutment are to be made from inCoris ZI, IPS e.max® CAD Abutment or VITA ENAMIC® (IS), designed and manufactured using Sirona CEREC SW version 4.6.1 software.
MIS Ti-base abutments are intended for use with the following MIS implants:
- · C1 conical connection implant system
- · V3 conical connection implant system
- SEVEN internal hex implant system
- M4 internal hex implant system
- Lance+ internal hex implant system
Type of Use (Select one or both, as applicable)
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
Image /page/3/Picture/1 description: The image contains the logo for Dentsply Sirona. On the left side of the logo is a gray abstract shape. To the right of the shape is the text "Dentsply" on the top line and "Sirona" on the bottom line. The text is in a sans-serif font.
# 510(k) SUMMARY for MIS Ti-base abutment (K211225)
#### 1. Submitter:
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
| Contact Person: | Rebecca Sporer |
|-------------------|-----------------------------------|
| Telephone Number: | 717-849-4793 |
| Fax Number: | 717-849-4343 |
| Email: | rebecca.sporer@dentsplysirona.com |
Date Prepared: June 28, 2021
#### 2. Device Name:
- Trade/Proprietary Name: MIS Ti-base abutment ●
- . Common/Usual Name: Dental abutment
- Classification Name: Endosseous dental implant abutment
- Regulation Number: 872.3630
- . Product Code: NHA
- Device Class: Class II
- . Classification Panel: Dental Devices Panel
#### 3. Predicate Device(s):
Primary predicate device:
- . MIS Ti-base abutment cleared under K191152. Reference devices:
- IPS e.max® CAD Abutment Solutions clearedunder K191382
- . VITA ENAMIC® Implant Solutions (IS) cleared underK153645
- Multilink Hybrid Abutment Cement cleared under K130436
- . Sirona Dental CAD/CAM System cleared under K181520
{4}------------------------------------------------
## 4. Device Description:
The subject MIS Ti-base abutments are endosseous dental implant abutments intended to be connected to MIS dental implants and used to support CAD/CAM customized cement-retained or screw retained single or multiple-unit restorations.
MIS Ti-base abutments consist of a titanium base and a prosthetic screw, both made of TI-6AI-4V ELI complying with ASTM F136. The prosthetic screw tightens the finished CAD/CAM abutment to the dental implant.
MIS Ti-base abutments are the bottom-half/base of a two-piece custom ceramic-titanium abutment consisting of a ceramic coping/mesostructure and a titanium base.
The top-half custom ceramic coping/mesostructure or crown is intended to be fabricated from Sirona inCoris ZI zirconium oxide ceramic block, IPS e.max® CAD ceramic block, or from IPS e.max® CAD ceramic block or VITA ENAMIC® (IS) ceramic block and designed and milled using Sirona chairside Dental CAD/CAM System, with software version: CEREC SW version 4.6.1. The mesostructure design will be subject to the Sirona system controls, such as: A maximum angulation of 20° and minimum wall thickness of 0.5mm for inCoris ZI and e.max materials and 0.8mm for VITA ENAMIC material.
lt is not permitted to reduce the Ti-base's diameter, shorten the Ti-base or modify its implant-abutment connection and emergence profile in any way.
The subject, pre-fabricated titanium base abutment is designed with interface compatibility to specific MIS dental implant systems. The subject MIS Ti-base abutments are MIS connection and internal hex connection Ti-base abutments, and their connection is compatible with MIS conical connection C1 and V3 implants, and MIS SEVEN, M4 and Lance+ internal hex implants, which are not subject to this submission and were previouslycleared.
## 5. Indications for Use:
MIS Ti-base abutment is a titanium base placed onto MIS dental implants to provide support for customized cement-retained or screw retained single or multiple-unit restorations. It is used with a digitally designed mesostructure. MIS Ti-base and the mesostructure make up a two- piece abutment used in conjunction with MIS dental implants, to be placed in the upper or lower jaw arches, in order to restore masticatory function.
Narrow platform Ti-bases are indicated for use only in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws.
MIS short implants are to be used only with straight abutments. Mesostructures for use with the MIS Tibase abutment are to be made from inCoris ZI, IPS e.max® CAD Abutment or VITA ENAMIC® (IS), designed and manufactured using Sirona CEREC SW version 4.6.1 software.
{5}------------------------------------------------
MIS Ti-base abutments are intended for use with the following MIS implants:
- C1 conical connection implant system
- . V3 conical connection implant system
- SEVEN internal hex implant system
- M4 internal hex implant system
- . Lance+ internal hex implant system
## 6. Substantial Equivalence Discussion:
The subject MIS Ti-base abutments have the same intended use as the predicate MIS Ti-base abutments cleared under K191152. They are intended for use in partially or fully edentulous mandibles in support of single or multiple-unit cement retained restorations, in conjunction with endosseous dental implants to restore the function and aesthetics in the oral cavity.
The difference between the subject and predicate Ti-base is the material from which the CAD/CAM mesostructure is to be fabricated. The predicate MIS Ti-base abutments are indicated for use with an inCoris mesostructure designed and milled with Sirona Dental CAD/CAM System, while the subject MIS Ti-base abutments are indicated for use with an IPS e.max CAD or VIRA ENAMIC (IS) mesostructure. Both subject and predicate MIS Ti-base abutments are indicated for use with the same MIS dental implants.
The principal of operation of the predicate and subject MIS Ti-base abutments is the digitally designed mesostructure is cemented to the Ti-base to make up a two-piece abutment. This two-piece abutment is attached to the implant by a prosthetic screw.
{6}------------------------------------------------
## Table 1 – Comparison of MIS Ti-base abutment Characteristics
| Trade Name | MIS Ti-base abutments | MIS Ti-base abutments |
|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Subject | K191152 |
| Manufacturer | MIS Implants Technologies Ltd. | MIS Implants Technologies Ltd. |
| Device Class | Class II | Class II |
| Product Code(s) | NHA | NHA |
| Regulation<br>Description | Endosseous dental implant abutment | Endosseous dental implant abutment |
| Regulation Number | 872.3630 | 872.3630 |
| Indications for use: | MIS Ti-base abutment is a titanium base placed onto MIS dental<br>implants to provide support for customized cement-retained or<br>screw retained single or multiple-unit restorations.<br>It is used with a digitally designed mesostructure. MIS Ti-base and<br>the mesostructure make up a two-piece abutment used in<br>conjunction with MIS dental implants, to be placed in the upper or<br>lower jaw arches, in order to restore masticatory function.<br>Narrow platform Ti-bases are indicated for use only in the<br>mandibular central, lateral incisor and maxillary lateral incisor<br>regions of partially edentulous jaws.<br>MIS short implants are to be used only with straight abutments.<br>Mesostructures for use with the MIS Ti-base abutment are to be<br>made from inCoris ZI, IPS e.max® CAD Abutment or VITA ENAMIC®<br>(IS), designed and manufactured using Sirona CEREC SW version<br>4.6.1 Software.<br>MIS Ti-base abutments are intended for use with the following MIS<br>implants:<br>C1 conical connection implant system, V3 conical connection<br>implant system, SEVEN internal hex implant system, M4 internal<br>hex implant system and Lance+ internal hex implant system. | MIS Ti-base abutment is a titanium base placed onto MISdental<br>implants to provide support for customized cement-retained or<br>screw retained single or multiple-unit restorations.<br>It is used with a digitally designed mesostructure. MIS Ti-base and<br>the mesostructure make up a two-piece abutment used in<br>conjunction with MIS dental implants, to be placed in the upper or<br>lower jaw arches, in order to restore masticatory function.<br>Narrow platform Ti-bases are indicated for use only in the<br>mandibular central, lateral incisor and maxillary lateral incisor<br>regions of partially edentulous jaws.<br>MIS short implants are to be used only with straight abutments.<br>Mesostructures for use with the MIS Ti-base abutment are to be<br>made from inCoris ZI, designed and manufactured using Sirona<br>CEREC SW version 4.6.1 Software.<br>MIS Ti-base abutments are intended for use with the following MIS<br>implants:<br>C1 conical connection implant system, V3 conical connection<br>implant system, SEVEN internal hex implant system, M4 internal<br>hex implant system and Lance+ internal hex implant system. |
| Ti-base Material | Abutment and Abutment screw Ti 6Al 4V ELI per ASTM | Abutment and Abutment screw Ti-6Al-4V per ASTM F136 |
| Implant to<br>abutment<br>Connection | Conical connection Internal hex connection | Conical connection Internal hex connection |
| Restoration | Single-unit Multiple-unit | Single-unit Multiple-unit |
| Trade Name | MIS Ti-base abutments | MIS Ti-base abutments |
| Prosthesis<br>Attachment | Cement-retained<br>Screw-retained<br>● | Cement-retained<br>Screw-retained<br>● |
| Platforms | NP 3.3<br>●<br>SP 3.75, 3.9, 4.2, 4.3, 5.0<br>●<br>WP 5.0, 6.0<br>● | NP 3.3<br>●<br>SP 3.75, 3.9, 4.2, 4.3, 5.0<br>●<br>WP 5.0, 6.0<br>● |
| Gingiva Height | NP: 0.5, 1.5 mm<br>●<br>SP: 0.5, 1.5, 3.0 mm<br>●<br>WP: 0.5, 1.5, 3.0 mm<br>● | NP: 0.5, 1.5 mm<br>●<br>SP: 0.5, 1.5, 3.0 mm<br>●<br>WP: 0.5, 1.5, 3.0 mm<br>● |
| Post Height | Minimum 4.0 mm | Minimum 4.0 mm |
| Compatible<br>implant system | MIS C1 Conical Connection<br>●<br>MIS V3 Conical Connection<br>●<br>MIS SEVEN Internal Hex<br>●<br>MIS M4 Internal Hex<br>●<br>MIS Lance+ Internal Hex<br>● | MIS C1 Conical Connection<br>●<br>MIS V3 Conical Connection<br>●<br>MIS SEVEN Internal Hex<br>●<br>MIS M4 Internal Hex<br>●<br>MIS Lance+ Internal Hex<br>● |
| Compatible<br>implant diameter<br>(mm) | NP:<br>●<br>C1/V3/ SEVEN/M4/Lance+: 3.3<br>SP:<br>●<br>C1/ SEVEN/M4/Lance+: 3.75, 4.2<br>V3: 3.9, 4.3, 5.0<br>WP:<br>●<br>C1: 5.0<br>SEVEN/M4/Lance+: 5.0, 6.0 | NP:<br>●<br>C1/V3/ SEVEN/M4/Lance+: 3.3<br>SP:<br>●<br>C1/ SEVEN/M4/Lance+: 3.75, 4.2<br>V3: 3.9, 4.3, 5.0<br>WP:<br>●<br>C1: 5.0<br>SEVEN/M4/Lance+: 5.0, 6.0 |
| Finished CAD/CAM<br>Abutment<br>Angulation | 0° - 20° | 0° - 20° |
| Two-piece<br>Abutment<br>Mesostructure<br>Material | inCoris ZI<br>●<br>IPS e.max<br>●<br>VITA ENAMIC (IS)<br>● | InCoris ZI zirconium oxide |
| Sterilization<br>Method | Product provided non-sterile | Product provided non-sterile |
| Sterilization by<br>end user | Moist steam sterilization | Moist steam sterilization |
{7}------------------------------------------------
{8}------------------------------------------------
## 7. Non-Clinical Performance Data:
As part of demonstrating the substantial equivalence of MIS Ti-base abutments to the predicate device (K191152) listed in this 510(k) submission, MIS Implants Technologies completed a number of non-clinical performance tests:
- . Biocompatibility - The subject device is manufactured using identical manufacturing methods, in the same manufacturing facility, and using the same raw material as the previously cleared predicate device, K191152. The subject device has the same patient contact duration and type as the identified predicate device. For these reasons, biocompatibility testing was not required to support substantial equivalence of the subject device. Consideration of biocompatibility requirements was made with reference to Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
- . Biocompatibility (two-piece abutments mesostructure) – The final two-piece CAD/CAM abutments which are constructed utilizing the subject MIS Ti-base components are completed with CAM milled mesostructures composed of the IPS e.max or VITA ENAMIC (IS) material. The IPS e.max material is cleared under K191382 the VITA ENAMIC (IS) material is cleared under K153645. No modifications to the materials are included in this premarket notification. Therefore, no new biocompatibility data relating to the IPS e.max or VITA ENAMIC (IS) material was included in this submission in support of substantial equivalence.
- . Fatigue Testing - Mechanical testing of MIS Ti-base abutments in accordance to ISO 14801:2016 was conducted. The worst case abutments chosen for the tests were the narrowest abutments from both narrow and standard platforms, with a 20° mesostructure which is the maximum angulation possible in the Sirona Dental CAD/CAM System. Both IPS e.max and VITA ENAMIC (IS) mesostructures were tested. The test articles were able to withstand 2,000,000 cycles without failure at a substantially equivalent load to the cited predicates. The fatigue test conducted on the standard platform worst case abutment supports the wide platform abutments as the SP is a worst case in terms of diameter and wall thickness, and both SP and WP abutments are made of the same material. The results of fatigue testing support substantialequivalence. Fatigue testing was conducted with reference to the Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments.
- . Sterilization Testing - MIS Ti-base abutments are supplied non-sterile and intended to be steam sterilized by the user.
- O For inCoris ZI mesostructure: the steam sterilization parameters were validated according to ANSI/AAMI/ISO 17665- 1:2006 and ANSI/AAMI/ISO 17665-2:2009 for two methods: gravity displacement steam sterilization and pre vacuum steam sterilization validation of the recommended sterilization process for the subject devices was conducted with reference to Guidance for Industry and FDA Staff: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling.
{9}------------------------------------------------
- For IPS e.max and VITA ENAMIC (IS) mesostructures: the recommended steam sterilization o parameters are according to the parameters that were validated in the predicate K191382 and K153645 respectively.
- Reverse engineering analysis was not conducted in support of substantial equivalence due to the fact that the subject MIS Ti-base Abutments are proposed for compatibility only with MIS dental implant systems. Dentsply Sirona is the sponsor/applicant of this premarket notification as the Owner/Operator of MIS Implants Technologies, Ltd.
- . Software verification and validation testing was provided for the subject abutment design library to demonstrate use with the "chairside" CAD/CAM software, CEREC SW version 4.6.1. Software verification and validation testing was conducted to demonstrate that the restrictions prevent design of the mesostructure component outside of design limitations. In addition, the encrypted abutment design library was validated to demonstrate that the established design limitations are locked and cannot be modified within the abutment design library.
## 8. Clinical Performance Data:
There was no human clinical data included in support of substantial equivalence. The non-clinical testing detailed in this submission support the substantial equivalence of the device.
#### 9. Summary:
The comparison between the subject device and the predicate devices has shown that the indications for use, principles of operation, technological characteristics and materials are the same. The results of non- clinical performance testing support a conclusion that the subject device is at least equivalent to the predicate devices with respect to dynamic fatigue performance.
#### 10. Conclusion Regarding Substantial Equivalence:
The subject MIS Ti-base abutments have the same intended use, incorporate the same fundamental technology, and have similar indications for use as the predicate (K191152). Test data to verify the performance of MIS Ti- base abutments has been provided including: dynamic fatigue and sterilization validation and the results of this testing, combined with the design and intended use comparison with the predicate device, support substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.