Urethral Warming Catheter Kit
K211149 · Varian Medical Systems, Inc. · GEH · Jan 5, 2022 · General, Plastic Surgery
Device Facts
| Record ID | K211149 |
| Device Name | Urethral Warming Catheter Kit |
| Applicant | Varian Medical Systems, Inc. |
| Product Code | GEH · General, Plastic Surgery |
| Decision Date | Jan 5, 2022 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4350 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Varian Urethral Warming Catheter Kit, a set of disposable components used in conjunction with a warmer system, sterile water or saline, and peristaltic pump, is intended to transfer heat to urethra during urological cryosurgery with Varian CRYOCARE Systems.
Device Story
The Urethral Warming Catheter Kit is a disposable, closed-circuit, double-lumen urethral balloon catheter used during urological cryosurgery. It functions by circulating sterile water or saline through the catheter to transfer heat to the urethral mucosa, preventing cold thermal injury. The kit includes the catheter, tubing, flow indicator, clamp, and luer connections. It is designed to be used with a third-party fluid warming system (e.g., 3M Ranger Model 245) and a roller-type peristaltic pump. The system is operated by clinical staff in a surgical setting. By maintaining urethral temperature, the device protects the urethra from the cold energy applied by Varian CRYOCARE cryosurgical systems, potentially reducing procedure-related complications.
Clinical Evidence
No clinical or animal studies were performed. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993 series), sterilization validation (ISO 11137, ISO 11737, ANSI/AAMI ST67), and usability testing (IEC 62366-1). Bench testing confirmed the system maintains proper fluid flow, remains leak-free, and effectively dissipates cold thermal energy.
Technological Characteristics
Double-lumen closed-circuit balloon catheter; polyurethane shaft with polyester balloon; 16-22 French size. Operates via external fluid warming system and roller-type peristaltic pump (nominal flow 500 mL/min). Sterilized via Gamma irradiation. Complies with ISO 10993 (biocompatibility), ISO 11137 (sterilization), and IEC 61010 (pump safety).
Indications for Use
Indicated for patients undergoing urological cryosurgery requiring urethral warming to protect urethral tissue.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
Predicate Devices
- Urethral Warming System (K963970)
Reference Devices
- CRYO-CS System (K153489)
- CRYO-38T (K201588)
Related Devices
- K963970 — ENDOCARE CRYOCARE CRYOSURGICAL SYSTEM URETHRAL WARMING SYSTEM · Endocare, Inc. · Feb 25, 1997
- K090978 — UC-CARE WARMING SYSTEM · UC-CARE , Ltd. · Aug 10, 2009
- K955011 — CANDELA HCS 2000 URETHERAL WARMER CATHETER · Candela Laser Corp. · Apr 3, 1996
- K965136 — UROSURGE UROTHERM FLUID WARMING SYSTEM · Princeton Regulatory Assoc. · Oct 16, 1998
- K961219 — PERCUTANEOUS INTRODUCER SYSTEM · Boston Scientific Corp · Jun 21, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
January 5, 2022
Varian Medical Systems, Inc. Peter Coronado Senior Director Regulatory Affairs 9825 Spectrum Drive, Building 2, Suite 250 Austin, Texas 78717
Re: K211149
Trade/Device Name: Urethral Warming Catheter Kit Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit and Accessories Regulatory Class: Class II Product Code: GEH Dated: December 7, 2021 Received: December 8, 2021
Dear Peter Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211149
Device Name Varian Urethral Warming Catheter Kit
Indications for Use (Describe)
The Varian Urethral Warming Catheter Kit, a set of disposable components used in conjunction with a warmer system, sterile water or saline, and peristaltic pump, is intended to transfer heat to urethra during urological cryosurgery with Varian CRYOCARE Systems.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "varian" in a simple, sans-serif font. The word is written in all lowercase letters and is black. The background is white, which makes the word stand out. The font is clean and modern, and the overall design is simple and effective.
Varian Medical
# PREMARKET NOTIFICATION
510(k) Summary
Urethral Warming Catheter Kit
The following information is provided as required by 21 CFR 807.92
#### l. Submitter's Information:
| Name and Address: | Varian Medical Systems<br>9825 Spectrum Drive Building 2, Suite 250<br>Austin, TX 78717 |
|-------------------|-----------------------------------------------------------------------------------------|
| Contact Name: | Mr. Peter J. Coronado, Senior Director Regulatory Affairs |
| Phone: | 650-424-6320 Fax: 650-646-9200 |
| E-mail: | submissions.support@varian.com |
| Date Prepared: | December 7, 2021 |
#### Device Information: II.
| Proprietary Name: | Urethral Warming Catheter Kit |
|----------------------|-----------------------------------|
| Common/ Usual Name: | Cryosurgical Unit and accessories |
| Classification Name: | Cryosurgical unit and accessories |
| Regulation Number: | 21 CFR 878.4350 |
| Product Code: | GEH |
#### Predicate Device: III.
Urethral Warming System (K963970)
#### IV. Device Description:
The Urethral Warming Catheter Kit (CRYO-77) warms a patient's urethra during the cryoablation procedure by transferring heat through a catheter inserted within the urethra. The catheter contains fluid that is warmed using dry heat from an electrical heating unit (Warming unit) and circulated in a closed circuit using a peristaltic pump (circulating pump).
The Urethral Warming Catheter Kit contains the following:
- 22 french urethral balloon catheter
- Tubing
- Flow indicator
- Tubing clamp ●
- IV bag decanter
- Two luer connections
- Tube placement markers for pump ●
- Tube retainer ●
and is used in conjunction with a compatible fluid warming system and peristaltic pump. The fluid warming system supplies a steady flow of warm, sterile water or saline to the urethral catheter which transfers heat to the urethra during urological cryosurgery. The water or saline is circulated by a peristaltic pump from the catheter to the I.V. bag in a closed circuit.
The Urethral Warming Catheter Kit is recommended to use with 3M™ Ranger™ Model 245 Fluid
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "varian" in a simple, sans-serif font. The word is written in all lowercase letters and is black. The background is white, which makes the word stand out. The font is modern and clean, and the overall design is simple and professional.
Varian Medical ctrum Drive ing 2 Suite 250
Warmer System and 3M™ Ranger™ Fluid Warming Set, and a peristaltic pump meeting the following technical specifications:
- Pump mechanism- Roller-type peristaltic pump ●
- Compatible Pump head- Masterflex L/S® Easy-Load for use with 13, 14, 16, 25, 17, 18 tubing (07518-60, 77200-60, 77201-60 or similar)
- Operational nominal flow rate- 500 mL/min
- Minimum Pump RPM range 0-300RPM
- . Maximum Continuous Pressure- 40PSIG
- . Rated voltages 100V-240V; 50/60Hz
- . IEC 61010 certified
The compatible fluid warming system and peristaltic pump are not owned or distributed by Varian. The system integration of Urethral Warming Catheter Kit with compatible accessories is the responsibility of end user.
{5}------------------------------------------------
### Varian Medical Systems. Inc. 9825 Spectrum Drive Building 2. Suite 250 Austin, TX 78717
#### V. Indications for Use Statement:
The Varian Urethral Warming Catheter Kit, a set of disposable components used in conjunction with a warmer system, sterile water or saline, and peristaltic pump, is intended to transfer heat to urethra during urological cryosurgery with Varian CRYOCARE Systems.
#### Comparison of Technological Characteristics with the Predicate Device: VI.
At a high level, the subject device is similar to the predicate based on the following characteristics:
- Both the subject and predicate device have same indication for use to transfer heat to urethra during cryosurgery and have same . urological clinical application
- The mechanism of action of urethral warming catheter is same in both subject and predicate device, the catheter in both devices . uses heat transference to maintain the temperature of urethral tissue.
At a high level, the subject device differs from the predicate as a result of the following characteristics:
- The predicate device is a complete system consisting of a fluid warming system, peristaltic pump, IV stand, urethral warming ● catheter and tubing set. The subject device is a stand-alone device consisting of urethral warming catheter and tubing.
- . Modification in indication for use of the subject device to include on-market available compatible accessories fluid warming system and peristaltic pump which deliver substantially equivalent tissue warming to that achieved by predicate device. The compatible accessories are not owned or distributed by Varian.
- . Modification in sterilization method of urethral warming catheter
- Modification in packaging of urethral warming catheter
- Removal of heat exchanger cartridge from Urethral Catheter Kit
- Change in temperature range of fluid warming system ●
- Change in flow range of peristaltic pump
The table below includes a high-level comparison of the predicate and subject devices.
| Feature and/ or<br>Specification | Predicate Device<br>Urethral Warming System (K963970) | Subject Device<br>Urethral Warming Catheter Kit | Comparison between<br>Subject and Predicate |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model Number | CRYO-60 | CRYO- 77 | NA |
| Indications for use | The Endocare Urethral Warming System is<br>intended to transfer heat to urethral tissue<br>during urological cryosurgery. | The Varian Urethral Warming Catheter Kit, a set of<br>disposable components used in conjunction with<br>a warmer system, sterile water or saline, and<br>peristaltic pump, is intended to transfer heat to | The indication and clinical<br>application remain the same.<br>The differences in wording are<br>due to the subject device's |
| Feature and/ or<br>Specification | Predicate Device<br>Urethral Warming System (K963970) | Subject Device<br>Urethral Warming Catheter Kit | Comparison between<br>Subject and Predicate |
| | | urethra during urological cryosurgery with Varian<br>CRYOCARE Systems. | compatibility with on-market<br>available fluid warming systems<br>and peristaltic pumps rather<br>than exclusive use with<br>Endocare Urethral Warming<br>System components - Alton<br>Dean FW-537 warming system<br>and adjustable RPM Rotary<br>Peristaltic pump. |
| Product Configuration | Fluid Warming System Peristaltic Pump IV Stand Urethral Catheter Kit containing<br>Urethral Warming Catheter, Tubing and<br>Heat Exchanger Cassette | Urethral Warming Catheter Tubing | Predicate device is a complete<br>system whereas the subject<br>device is a stand-alone device. |
| Compatibility with<br>Varian CRYOCARE<br>Systems | CRYO-CS System (K153489) CRYO-204-V and CRYO-2017-V<br>(K153489) | CRYO-CS System (K153489) CRYO-38T (K201588) CRYO-204-V and CRYO-207-V (K153489,<br>K201588) | Subject device will be<br>additionally compatible with<br>CRYO-38T |
| Fluid Warming System | Alton Dean FW-537 | 3M™ Ranger™ Model 245 Fluid Warmer<br>System 3M™ Ranger™ Fluid Warming Set | Subject device is compatible<br>with fluid warming system and<br>fluid warming set available on-<br>market. The compatible<br>accessories are not<br>manufactured and distributed<br>by Varian. |
| Temperature Range of<br>Fluid Warmer | 35-39°C | 41-43°C | Difference in the temperature<br>range of fluid warming system<br>due to on-market available<br>compatible accessories. |
| Pump | Adjustable RPM Rotary Peristaltic pump<br>have the following technical specifications, | Peristaltic Pump available on-market meeting the<br>following technical specifications, | Subject device is compatible<br>with peristaltic pump available<br>on-market. Peristaltic pump |
| Feature and/ or | Predicate Device | Subject Device | Comparison between |
| Specification | Urethral Warming System (K963970) | Urethral Warming Catheter Kit | Subject and Predicate |
| | Pump mechanism- Roller-type peristaltic pump Compatible Pump head- Masterflex L/S® Easy-Load for use with 13, 14, 16, 25, 17, 18 tubing (77200-60) | Pump mechanism- Roller-type peristaltic pump Compatible Pump head- Masterflex L/S® Easy-Load for use with 13, 14, 16, 25, 17, 18 tubing (07518-60, 77200-60, 77201-60 or similar) | should meet the mentioned technical specifications to be compatible with subject device and 3M™ Ranger™ Fluid Warmer System. The compatible accessories are not manufactured and distributed by Varian |
| Flow Range of<br>Peristaltic Pump | 140-1700 mL/min | 0-1140 mL/min | Difference in the flow range of peristaltic pump is mitigated by providing nominal operational flow rate |
| Operational Flow rate | Not specified in K963970 submission | Nominal 500 mL/min, recommended by 3M™ Ranger™ Fluid Warmer System | Operational flow rate of predicate device is not specified in K963970 submission; however, the subject device is tested to perform at operational flow rate of nominal 500 mL/min, the specified fixed flow rate prevents over/under fluid circulation. The test result showed the subject device and its on-market available components (warmer and pump) installed together as a system perform equivalent to predicate cleared device and transfer the heat to the urethra equally during cryoablation procedure. |
| Mechanism of Action | Heat transfer | Heat transfer | Same |
| Urethral Warming | Double lumen closed circuit catheter | Double lumen closed circuit catheter | Same |
| Feature and/ or<br>Specification | Predicate Device<br>Urethral Warming System (K963970) | Subject Device<br>Urethral Warming Catheter Kit | Comparison between<br>Subject and Predicate |
| Catheter Description | | | |
| Catheter Material | Polyurethane shaft with polyester balloon | Polyurethane shaft with polyester balloon | Same |
| Catheter Size | 16-22 French | 16-22 French | Same |
| Catheter Sterilization<br>Method | Ethylene Oxide | Gamma | Subject device uses Gamma<br>irradiation process |
| Packaging | Plastic film side and coated paper | Tray, Tyvek top, chipboard box | Subject uses Tray, Tyvek top,<br>chipboard box |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Varian Medical Systems, Inc. The logo consists of the word "varian" in a bold, sans-serif font on the left, and the words "Varian Medical Systems, Inc." stacked on top of the address "9825 Spectrum Drive, Building 2, Suite 250, Austin, TX 78717" on the right. A vertical line separates the company name from the address.
{7}------------------------------------------------
K211149
### Varian Medical Systems, Inc. varian 9825 Spectrum Drive Building 2, Suite 250 Austin, TX 78717
{8}------------------------------------------------
K211149
Image /page/8/Picture/1 description: The image shows the Varian Medical Systems, Inc. logo and address. The logo is on the left side of the image and the address is on the right side. The address is 9825 Spectrum Drive, Building 2, Suite 250, Austin, TX 78717. The logo is a stylized version of the word "varian".
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the word "varian" in a sans-serif font. The letters are black and the background is white. The letters are evenly spaced and the word is centered in the image.
Varian Medical
Systems, Inc.
9825 Spectrum Drive
Building 2, Suite 250
Austin, TX 78717
#### VII. Performance Data (Non-Clinical Testing)
Design Verification and Design Validation testing were completed to demonstrate that the Urethral Warming Catheter Kit performs as intended and meets its essential performance specifications.
- Catheter to include flow indicator to ensure fluid is not impeded and flowing correctly through closed circuit system.
- · Catheter is closed-circuit, free of fluid leaking or component separation
- Catheter is designed to dissipate cold thermal energy from the urethral mucosa
Bench testing included testing of the system, including:
- · Biocompatibility Testing in accordance with ISO 10993 series of standards, including ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-12
- Sterilization Validation in accordance with EN 556-1, ANSI/AAMI ST67, ISO 11137-1, ISO 11137-2, ISO 11737-1 and ISO 11737-2
- Usability Testing in accordance with IEC 62366-1 Annex C
No animal studies or clinical tests have been included in this pre-market submission.
#### VIII. Determination of Substantial Equivalence to the Predicate Device
A subset of technological characteristics and features of the subject device differs from the predicate device. These differences are all considered to be enhancements of the predicate, intended to configure urethral warming catheter as a stand-alone device compatible with onmarket available accessories.
The indications for use are substantially the same as the predicate device. Further, there are no changes in the principle of operation of the devices. The Verification and Validation demonstrates that the device is as safe and effective as the predicate. Varian therefore believes the data demonstrates that Varian Urethral Warming Catheter Kit is substantially equivalent to the predicate device, Endocare Urethral Warming System.
#### IX. Conclusion
The assessment following the outcomes observed in the performance testing determines that the Urethral Warming Catheter Kit conforms to the defined user needs and intended uses. Varian therefore considers the Urethral Warming Catheter Kit to be safe and effective and to perform at least as well as the predicate device.