K211147 · Path Medical GmbH · GWJ · Oct 14, 2021 · Neurology
Device Facts
Record ID
K211147
Device Name
ALGO 7i
Applicant
Path Medical GmbH
Product Code
GWJ · Neurology
Decision Date
Oct 14, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1900
Device Class
Class 2
Attributes
AI/ML, Pediatric
Indications for Use
ALGO 7i devices are designed for use in clinical environments, such as the well-baby nursery, neonatal intensive care unit (NICU), mother's bedside, audiology suite, outpatient clinic, or doctor's office as well as in community settings. The ALGO 7i is not intended for use in oxygen-rich environments.
Device Story
ALGO 7i is a portable, handheld newborn hearing screener. It uses AABR (Automated Auditory Brainstem Response) technology to measure evoked potentials. Input: click stimuli delivered to the cochlea via acoustic transducer assembly (ATA) cables; EEG signals recorded via electrodes on the head/neck. Operation: device firmware processes signals using a weighted-binary template-matching algorithm to separate ABR from background noise. Output: PASS/REFER/INCOMPLETE result displayed on a 3.5" touch screen. Used by audiologists, ENT doctors, nurses, and trained personnel in clinical/community settings. The device provides objective hearing status; results assist clinicians in identifying infants requiring further diagnostic evaluation. Benefits include non-invasive, automated screening with high statistical confidence, reducing the need for subjective interpretation.
Clinical Evidence
No new clinical performance data collected. Substantial equivalence is supported by bench testing and the use of licensed, validated AABR technology from the predicate ALGO 3i. The algorithm and ABR template (derived from 35 neonates) are identical to those used in previously cleared ALGO devices. Literature review confirms the ABR screening implementation matches current state-of-the-art.
Technological Characteristics
Handheld platform with 3.5" TFT-LC touch display. Uses AABR technology with click stimuli (35/40 dB nHL). Biocompatible materials (ISO 10993-1). Connectivity: USB for data transfer/PC management. Power: Rechargeable Li-ion battery. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-40, IEC 62304, IEC 62366, IEC 60645-7.
Indications for Use
Indicated for objective hearing status determination in newborns/infants from 34 weeks gestational age to 6 months old. Patients must be ready for hospital discharge and in a quiet or asleep state. Exclusions: infants on ventilators/incubators, CNS stimulants, ototoxic medications, or those with compromised skin or jaundice.
Regulatory Classification
Identification
An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
October 14, 2021
Path Medical GmbH Ing. Oswald Managing Director Landsberger Strasse 65 Germering, Bavaria 82110 Germany
Re: K211147
Trade/Device Name: ALGO 7i Regulation Number: 21 CFR 882.1900 Regulation Name: Evoked response auditory stimulator Regulatory Class: Class II Product Code: GWJ Dated: September 9, 2021 Received: September 13, 2021
Dear Ing. Oswald:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Data base located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including inf ormation about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211147
Device Name ALGO 7i
#### Indications for Use (Describe)
The ALGO® 71 Newborn Hearing Screener is a hand-held, portable hearing screener intended to objectively determine the hearing status of a newborn/infant from 34 weeks gestational age to 6 months old. Babies should be well enough for hospital discharge and should be asleep or in a quiet state at the time of screening.
Exclusion Criteria
The following criteria should be used to exclude an infant from screening with the Natus ALGO hearing screeners:
- = Infants not between 34 weeks gestational and 6 months of age
- Infants on ventilators or in incubators
- · Infants on CNS (central nervous system) stimulants
- Infants receiving ototoxic medications
- Infants with compromised skin or jaundice
ALGO 7i devices are intended for use by audiologists, ear-nose-throat (ENT) doctors, and other hearing health care professionals, nurses and audiologically trained personnel. Basic training with the device is sufficient for performing screening of patients in good health. It is not intended to be operated by lay users.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Path Medical. The logo consists of a red square with a white curved line running through it on the left side. To the right of the square is the word "PATH" in bold black letters, with the word "MEDICAL" in smaller black letters below it.
# 510(k) Summary
# SUBMISSION INFORMATION
| Date of preparation: | October 14, 2021 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter: | PATH MEDICAL GmbH<br>Landsberger Str. 65<br>82110 Germering<br>Germany<br>Phone: ++49-89-80076502<br>Fax: ++49-89-80076503 |
| Contact Person: | Dr. Johann Oswald, Director<br>Landsberger Str. 65<br>82110 Germering<br>Germany<br>Phone: ++49-89-80076502<br>Fax: ++49-89-80076503<br>Email: oswald@pathme.de |
# DEVICE INFORMATION
| Device Name: | ALGO 7i |
|------------------------------|--------------------------------------------------------------------|
| Device Trade Names: | ALGO 7i, A7i |
| Device Identification Codes: | 101049 |
| Common Name: | Evoked Response Auditory Stimulator |
| Classification Name: | Evoked Response Auditory Stimulator<br>FDA 21 CFR section 882.1900 |
# PREDICATE DEVICE
| ALGO 3i | 510(k) Number: K030823 |
|---------|------------------------|
|---------|------------------------|
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Path Medical. The logo consists of a red square with a white design inside, and the words "PATH MEDICAL" in black text to the right of the square. The word "PATH" is in a larger font than the word "MEDICAL".
### REFERENCE DEVICE
SENTIERO
510(k) Number: K133012
# DEVICE DESCRIPTION
ALGO 7i is an audiometric examination platform which consists of the ALGO 7i device with a touch screen display together with different accessories such as Multidata Cable, Docking Station, Patient Cable, ATA Cable. All connectors and transducers have a special mechanically coded plug-in order to ensure the correct connection to the device. All plugs of the transducers have a memory chip inside which stores the information about the respective transducer (including type of connector, calibration table). As a result, the ALGO 7i instrument can be connected flexibly to different ATA Cables while enabling the instrument to 'know' the calibration values and status of the connected Cable. This information is to guide the user (feedback via display) and help to ensure correct performance of the device. The ALGO 7i is designed as standalone examination platform and can be connected to a personal computer (PC) via USB using the Multidata Cable or the Docking Station for data review and management. The Device is portable and is meant to be mainly used as mobile device. Materials in contact with humans are selected to be biocompatible.
Additional features are direct printing of patient and measurement data on a label using label printer that can be connected via Multidata Cable or Docking Station.
The ALGO 7i offers hearing screening using AABR technology.
The measurement application is controlled from a self-contained firmware. The measurement flow is menu guided on a touch screen. Evaluation of test results is based on signal statistics. Besides that, result information is displayed as PASS/REFER/INCOMPLETE.
The ALGO 7i is designed to be used by trained personnel in a medical or home environment to examine hearing in infants from 34 weeks (gestational age) that are ready for discharge from the hospital up to 6 months old.
The following accessories are available to conduct a measurement:
- ATA (Acoustic Transducer Assembly) Cable: PATH ATA-S, ATA-L
- Equivalent to Natus ATA Cables, equivalent to the cables used with the predicate o ALGO 3i, similar to the PATH Ear Coupler Cables used with the reference Sentiero
- Electrode cable: PATH ABR-S. ABR-L
- shielded, passive cable to connect the instrument to electrodes o
These accessories can be connected to ALGO 7i using special color and mechanical coded plugs, which holds the information about the connected transducer / cable. By that, the firmware can
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Path Medical. The logo consists of a red square with a white design on the left and the words "PATH MEDICAL" in black on the right. The word "PATH" is in a larger font than the word "MEDICAL".
make use of this information and adapt the measurement procedure according to calibration values or provide information to the user via its display.
# INDICATIONS FOR USE
The ALGO® 71 Newborn Hearing Screener is a hand-held, portable hearing screener intended to objectively determine the hearing status of a newborn/infant from 34 weeks gestational age to 6 months old. Babies should be well enough for hospital discharge and should be asleep or in a quiet state at the time of screening.
Exclusion Criteria
The following criteria should be used to exclude an infant from screening with the Natus ALGO hearing screeners:
- = Infants not between 34 weeks gestational and 6 months of age
- Infants on ventilators or in incubators
- Infants on CNS (central nervous system) stimulants
- Infants receiving ototoxic medications
- Infants with compromised skin or jaundice
ALGO 71 devices are intended for use by audiologists, ear-nose-throat (ENT) doctors, and other hearing health care professionals, nurses, and audiologically trained personnel. Basic training with the device is sufficient for performing screening of patients in good health. It is not intended to be operated by lay users.
#### INTENDED USE
ALGO 7i devices are designed for use in clinical environments, such as the well-baby nursery, neonatal intensive care unit (NICU), mother's bedside, audiology suite, outpatient clinic, or doctor's office as well as in community settings. The ALGO 7i is not intended for use in oxygenrich environments.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Path Medical. The logo has a red square on the left side with a white curved line going through it. To the right of the square, the word "PATH" is written in large, bold, black letters, and below that, the word "MEDICAL" is written in smaller, bold, black letters.
# COMPARISON TO SIMILAR DEVICES
The ALGO 7i will be compared to the predicate device ALGO 3i in terms of intended purpose, physiological features, accessories and applied screening algorithm. For workflow, technology, applied standards and user interface a comparison will be drawn to the reference device Sentiero. COMPARISON TO THE PREDICATE DEVICE ALGO 3i:
| | ALGO 7i (subject) | ALGO 3i<br>(predicate) | Equivalency |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Intended purpose | | | |
| Intended Use | ALGO 7i devices are<br>designed for use in<br>clinical environments,<br>such as the well-baby<br>nursery, neonatal<br>intensive care unit<br>(NICU), mother's<br>bedside, audiology<br>suite, outpatient clinic,<br>or doctor's office as<br>well as in community<br>settings. The ALGO 7i<br>is not intended for use<br>in oxygen-rich<br>environments. | The ALGO 3i<br>screener is simple to<br>operate and does not<br>require special<br>technical skills or<br>interpretation of<br>results. Basic<br>training with the<br>equipment is<br>sufficient to learn<br>how to screen<br>infants correctly. A<br>typical screening can<br>be completed in 15<br>minutes or less in<br>any clinical<br>environment (i.e.<br>well-baby nursery,<br>NICU, mother's<br>bedside, audiology<br>suite, outpatient<br>clinic,<br>or doctor's office). | Same (different<br>wording,<br>description of<br>ALGO 7i is more<br>specific) |
| Indications for Use | The ALGO® 7i<br>Newborn Hearing<br>Screener is a hand-held,<br>portable hearing<br>screener intended to<br>objectively determine<br>the hearing status of a<br>newborn/infant from 34<br>weeks gestational age<br>to 6 months old. Babies<br>should be well enough<br>for hospital discharge<br>and should be asleep or | The ALGO 3i<br>Newborn Hearing<br>Screener is a<br>portable,<br>noninvasive device<br>for screening the<br>hearing of infants<br>between the ages of<br>34 weeks corrected<br>gestational age and<br>six months.<br>Screening can be<br>performed by trained<br>personnel in any | Same |
| | | | |
| | in a quiet state at the<br>time of screening. | clinical environment<br>(i.e. well-baby<br>nursery, NICU,<br>mother's bedside,<br>audiology suite,<br>outpatient clinic, or<br>doctor's office) | |
| | Exclusion Criteria<br>The following criteria<br>should be used to<br>exclude an infant from<br>screening with the<br>Natus ALGO hearing<br>screeners:<br>• Infants not between<br>34 weeks gestational<br>and 6 months of age<br>• Infants on ventilators<br>or in incubators<br>• Infants on CNS<br>(central nervous<br>system) stimulants<br>• Infants receiving<br>ototoxic medications<br>• Infants with<br>compromised skin or<br>jaundice | | |
| | ALGO 7i devices are<br>intended for use by<br>audiologists, ear-nose-<br>throat (ENT) doctors,<br>and other hearing<br>health care<br>professionals, nurses<br>and audiologically<br>trained personnel. Basic<br>training with the device<br>is sufficient for<br>performing screening<br>of patients in good<br>health. It is not<br>intended to be operated<br>by lay users. | | |
| Physiological features | | | |
| Biosignal | Evoked potential | Evoked potential | Same |
| Electrode position | Head and neck | Head and neck | Same |
| Stimulation target | Cochlea | Cochlea | Same |
| EEG recording<br>channels, number of<br>electrode contacts | 1,3 | 1,3 | Same |
| Accessories | | | |
| Transducer | ATA-Cable (different<br>plug) | ATA-Cable | Same |
| Electrode Cable | Patient Cable<br>(shielded, 3 clamps)<br>(different plug) | Patient Cable,<br>(shielded, 3<br>clamps) | Same |
| Electrodes | Natus Jelly Tabs | Natus Jelly Tabs | Same |
| Ear Coupler | Natus FlexiCoupler | Natus FlexiCoupler | Same |
| Implementation details | | | |
| Screening options | Each ear individually,<br>simultaneously or<br>sequentially | Each ear<br>individually,<br>simultaneously or<br>sequentially | Same |
| Stimulus | Click | Click | Same |
| Stimulus repetition<br>rate | 34/37 Hz | 34/37 Hz | Same |
| Stimulus level | 35, 40 dB nHL | 35, 40 dB nHL | Same |
| Noise cancellation<br>(acoustic & myogenic) | Ambient noise<br>microphone<br>(acoustic);<br>Dual channel<br>sampling (myogenic) | Ambient noise<br>microphone<br>(acoustic);<br>Dual channel<br>sampling<br>(myogenic) | Same |
| Applied algorithm | Natus AABR | Natus AABR | Same |
| Result interpretation | By the device | By the device | Same |
| Result representation | PASS/REFER/<br>INCOMPLETE | PASS/REFER/<br>INCOMPLETE | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for Path Medical. The logo consists of a red square with a white curved line running through it on the left side. To the right of the square is the word "PATH" in large, bold, black letters, and below that is the word "MEDICAL" in smaller, black letters.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for Path Medical. The logo consists of a red square with a white curved line running through it on the left side. To the right of the square is the word "PATH" in large, bold, black letters, with the word "MEDICAL" in smaller, black letters underneath.
#### Table 5. 1 Comparison to predicate device
The ALGO 7i will issue a PASS result when it collects sufficient data to establish with > 99% statistical confidence that an ABR signal is present and consistent with the template. This confidence level can be reached at a minimum of 1000 sweeps for 35dBnHL screening, and 2000 sweeps for 40dBnHL screening. The ALGO screener will continue to collect data up to 15,000 noise-weighted sweeps. If it has not established with > 99% statistical confidence that the ABR signal is present after 15,000 noise-weighted sweeps, it will issue a REFER result. Aborting a test will result in an INCOMPLETE result. The same procedure is implemented in the predicate ALGO 3i.
The ALGO 7i uses the same patented signal processing technology to separate the ABR from background noise and other brain activity as the predicate ALGO 3i. These responses are matched against a stored pattern called a "template", derived from the ABRs of normal-hearing infants. The ALGO 7i must detect the ABR with very high statistical confidence in order to issue a PASS result. This technology includes a patented dual-artifact rejection system to prevent non-ABR activity
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image is a logo for PATH Medical. The logo consists of a red square with a white, stylized "P" inside of it. To the right of the square, the word "PATH" is written in large, black letters, and below that, the word "MEDICAL" is written in smaller, black letters. The logo is simple and modern, and the use of red and black gives it a professional look.
from contributing toward a PASS result. This ensures a very high degree of accuracy of the PASS result issued by an ALGO 7i device. The template cannot be replaced or modified.
In conclusion, the ALGO 7 system is substantially equivalent to the predicate device ALGO 3i in terms of intended use, accessories used, operating principle and algorithm.
# COMPARISON TO REFERENCE DEVICE:
Design and working principles of ALGO 7i are very similar to Sentiero. The similarity can be attributed to the fact that the same engineers at PATH MEDICAL, who designed the Sentiero products, also designed the ALGO 7i. In order to expedite development, hardware properties (e.g. PCB design, housing, connectors) as well as firmware features (workflow, User Interface, development environment, compiler) were mirrored from the Sentiero Backbone as starting point of development). The ALGO 7i test method ABR is a subset of the test methods available on the Sentiero, which in addition to ABR also offers various different test methods such as OAE or TYMP as well as diagnostic modules. Due to the variety of available modules and their differences in operation principle, the comparison of the ALGO 7i to the reference device will be drawn to the Sentiero and its ABR module:
| | ALGO 7i (subject) | SENTIERO<br>(reference) | Equivalency |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Intended Purpose | | | |
| Patient population | Newborn/infant from 34<br>weeks gestational age to 6<br>months old | All ages | Same (Subset) |
| Indications for Use | Determination of the<br>hearing status of a<br>newborn/infant | Determination of the<br>hearing status | Same (Subset) |
| Intended User | Audiologists, ear-nose-<br>throat (ENT) doctors, and<br>other hearing health care<br>professionals, nurses and<br>audiologically trained<br>personnel. It is not<br>intended to be operated<br>by lay users. | Audiologists, ear-<br>nose-throat (ENT)<br>doctors, and other<br>hearing health care<br>professionals and<br>audiologically<br>trained technicians in<br>a medical<br>environment. It is<br>not intended to be<br>operated by lay<br>users. | Same |
| Medical principle | Determination of hearing<br>status by evoking and<br>recording potentials in the<br>auditory pathway. | Determination of<br>hearing status by<br>evoking and<br>recording potentials<br>in the auditory<br>pathway. | Same |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image is a logo for Path Medical. The logo consists of a red square on the left side with a white curved line running through it. To the right of the square is the word "PATH" in large, bold, black letters, with the word "MEDICAL" in smaller, black letters underneath.
| | Technical implementation | | |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Mode of operation | Handheld, standalone | Handheld, standalone | Same |
| User interface | Branded and reduced version of the Sentiero user interface | Sentiero user interface | Same (Subset) |
| Workflow | Operation via touch screen, device control via header/footer, online information | Operation via touch screen, device control via header/footer, online information | Same |
| Battery | Rechargeable | Rechargeable | Same |
| Connectors for accessories | Push-pull ODU-Plug | Push-pull ODU-Plug | Same |
| Display | 3.5" TFT-LC backlight touch-sensitive display | 3.5" TFT-LC backlight touch-sensitive display | Same |
| Power plug | Medical grade | Medical grade | Same |
| Data transfer | USB (via Multidata cable or docking station) | USB | Same |
| PC software | PC Software for data analysis and archiving. User management configurable. Can be used to load data from/ to the device. | PC Software for data analysis and archiving. User management configurable. Can be used to load data from/ to the device. | Same |
| Applied standards | All relevant Audiology, Electrical Safety and Biocompatibility standards | All relevant Audiology, Electrical Safety and Biocompatibility standards | Same (For a detailed list of applied standards see below) |
Table 5. 2 Comparison to reference device
The primary mechanical difference between ALGO 7i and Sentiero is the charging of the built-in rechargeable batteries. ALGO 7i can be connected to a docking station or a dedicated multidata cable for charging and data transfer. In contrast thereto, Sentiero must be connected to a custommade plug for connecting to the charger.
Due to the similarity in hardware and workflow, the ALGO 7i and Sentiero require compliance to the same standards. Both Devices show similar performance data and compliance to the following standards:
- EN 10993-1:2018 Biological evaluation of medical devices (relevant tests: Cytotoxicity, -Sensitization, Irritation)
- -ISO 15223-1:2012 Symbols, labeling, and information to be provided
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image is a logo for Path Medical. The logo consists of a red square with a white curved line running through it on the left side. To the right of the square, the word "PATH" is written in large, bold, black letters, and below that, the word "MEDICAL" is written in smaller, black letters.
- IEC 60601-1:2005/AMD1:2012 Medical Electrical Equipment: General requirements for safety and essential performance
- IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance
- IEC 60601-1-4:1996 Programmable electrical system -
- IEC 60601-1-6:2010 Usability -
- -IEC 60601-2-40:1998 and 2016 Safety of electromyographs and devices for evoked potentials
- -IEC 62304:2010 Software lifecycle
- EN 1041:2013 Information provided by the manufacturer -
- -IEC 60645-7:2009 Acoustically evoked potentials
- EN 60645-3:2007 Electroacoustics: Audiometric Equipment Part 3: Test signals of short duration
- -EN 62366:2008 + A1:2015 Application of usability engineering to medical devices
In conclusion, ALGO 71 is substantially equivalent to the reference device Sentiero with respect to technological characteristics and non-clinical performance data.
### Biocompatibility testing
The biocompatibility evaluation was conducted according to ISO 10993-1:2018. Following tests are considered applicable:
- Cytotoxicity ।
- -Sensitization
- Irritation -
The device and its accessories are classified as short-term contact and contact with skin. No issues were found during biocompatibility testing.
### Electrical safety and electromagnetic compatibility (EMC)
The ALGO 7i was tested according to and complies to following electrical safety and electromagnetic compatibility standards: IEC 60601-1:2005/AMD1:2012, IEC 60601-1-2:2014, IEC 60601-2-40:1998 and 2016.
### Software Verification and Validation Testing:
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
### Mechanical and Acoustic Testing:
- 트 Tensile strength of accessory cable
- I Flex life of accessory cable
- I Frequency content, polarity, timing, sound level, and repetition rate of the Click stimulus equivalent to the predicate ALGO 3i stimulus
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image contains the logo for Path Medical. On the left side of the logo is a red square with a white curved line design. To the right of the square is the text "PATH" in large, bold, black letters, with the word "MEDICAL" underneath in smaller, black letters.
- 트 Similar myogenic and acoustic noise detection and rejection, impedance detection of the ALGO 7i to the predicate ALGO 3i
- Maximum possible sound level will remain below a possibly threatening level even under I failure
- I Push, Drop, and Mould Stress Relief test
# Sensitivity and Specificity for PASS/REFER:
The algorithmic sensitivity of the ALGO is set to 99.9% for each ear using binomial statistics. Independent clinical studies published peer-reviewed clinical performance data showed a combined overall sensitivity of 98.4%. Specificity of the ALGO device ranged from 96% to 98% in these studies. Relevant studies are listed in the section Clinical Studies. Equivalence of the ALGO 7i to the predicate device in result detection has been proven in bench testing.
## Clinical Performance Data:
No clinical performance data was collected for the subject device, ALGO 7i. The core automated ABR (AABR) technology is licensed from the predicate device ALGO 3i's manufacturer, Natus Medical Incorporated. ALGO 71's substantial equivalence to predicate ALGO 3i was the main design goal. This was accomplished with the direct incorporation of this licensed technology, including the weighted-binary template-matching algorithm and associated ABR template used in ALGO 3i. Therefore, ALGO 7i device utilizes the exact same methods and parameters to evoke, record, process and detect ABR responses as implemented in the predicate ALGO 3i device. The equivalency of the presented stimulus and the use of the identical algorithm for interpretation of the recorded signal demonstrates substantial equivalence of the subject device to the predicate device in terms of clinical performance. As such, additional clinical performance data was not needed.
The equivalency of the acoustic stimuli and recording of the evoked potentials as well as the correct implementation of the template and algorithm was established with nonclinical performance data and code reviews during the verification phase. Additionally, the literature review in the clinical evaluation shows that the implementation of ABR screening on the ALGO 7i, which is identical to the implementation on the ALGO 3i, matches the current state of the art.
The ABR template used in the ALGO 7i is based on the morphology of normal hearing, nearthreshold, infant ABR waveforms, determined by superimposing the responses of 35 neonates to 35 dBnHL click stimuli. The data was collected at Massachusetts Eye and Ear Infirmary during the design and development of the original automated infant hearing screener using ABR. This technology was originally commercialized as ALGO I device, and details of the ALGO technology, including the template and its collection info was provided under 510(k) submission K852687. The same template and detection algorithm has been used in all the subsequent FDAcleared ALGO devices, including ALGO 2 (K936039), ALGO 3 (K013137), the predicate ALGO 3i (K030823), and ALGO 5 (K073665).
The template and AABR algorithm used in all ALGO devices were developed and validated in the following studies:
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image is a logo for Path Medical. The logo has a red square on the left side with a white curved line going through it. To the right of the square is the word "PATH" in black, with the word "MEDICAL" in black underneath it.
Peters, J. G. "An automated infant screener using advanced evoked response technology." The Hearing Journal 39 (1986): 25-30.
Herrmann, Barbara S., Aaron R. Thornton, and Janet M. Joseph. "Automated infant hearing screening using the ABR: development and validation." American Journal of Audiology 4.2 (1995): 6-14.
# SUBSTANTIAL EQUIVALENCE
The ALGO 7i's stimulus input and measuring algorithm are identical to the predicate ALGO 3i. Bench testing was performed for relevant audiological characteristics of the stimulus delivered to as well as interpretation of of the recorded potential (creating the patient a PASS/REFER/INCOMPLETE result). The bench tests included frequency, timing, polarity and sound level of the stimulus as well as noise resistance and the lowest potential measurable or detectable by the device. Additionally, biocompatibility, electrical safety, and EMC testing was conducted to demonstrate that the subject device is as safe and as effective as the predicate device. Although the hardware was updated with the state-of-the-art hardware (touch sensitive, colored display, PCB layout, robust Patient Cable, Smaller, color and mechanically coded plugs instead of D-SUB 9 plugs, more powerful, but smaller battery, lightweight material), bench testing demonstrated substantially equivalent performance between subject (ALGO 7i) and predicate device (ALGO 3i).
The ALGO 7 i hardware is based on the reference device (Sentiero). The ALGO 7i uses the same material and manufacturer for the housing, sockets, plugs, charger, display and PCB as the Sentiero device. The well-known Lilon battery technology as incorporated in the Sentiero was used in the ALGO 7i. Additionally, the user interface and PCB design of the ALGO 7i is based on the Sentiero.
The biocompatibility and electrical safety as well as EMC of the hardware was independently verified by external laboratories.
Main differences between the subject ALGO 7i and the predicate ALGO 3i are the hardware changes as described above and their implication on the handling of the device. The screen displays information in color, facilitating e.g. differentiation of ears by color coding to audiological conventions (right ear = red, left ear = blue). Also. the operating principle changed to maneuvering through the device by swiping and scrolling as current standard for e.g. smartphone use. This was enabled by implementing a touch- sensitive color display. The weight reduction makes the device easier to handle and carry, while color and mechanical coding minimizes the risk of using the wrong connector. Those features have been used in the reference device Sentiero for years without any adverse events reported.
Summarized, the main differences of the ALGO 7i compared to the predicate ALGO 3i represents a design which has been independently tested by third parties, bench tested and a long history of use for the same intended use. The differences between the subject ALGO 7i and predicate ALGO 3i do not raise any new questions concerning safety or effectiveness.
Nonclinical performance testing has shown that the ALGO 7i is substantially equivalent to the ALGO 3i.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image contains the logo for Path Medical. The logo consists of a red square on the left side with a white curved line running through it. To the right of the square is the text "PATH" in large, bold, black letters, with the word "MEDICAL" underneath in smaller, black letters. The background is white.
It can be concluded, that the ALGO 7i is substantially equivalent to the ALGO 3i in terms of measuring performance based on stimulus and algorithm and substantially equivalent to the reference Sentiero in terms of hardware and user interface.
# OVERALL CONCLUSION
The ALGO 7i shows similar safety, effectiveness and performance data as the predicate ALGO 3i. Through modernization of the hardware platform as well as the update of the applicable standards to their most recent versions and implementation of new relevant standards such as IEC 62366, the ALGO 7i is a state-of-the-art newborn hearing screener.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.