← Product Code [MQB](/productcode/MQB) · K211108

# Prudent 1717, Prudent 1417, Prudent 1212 (K211108)

_Pixxgen Corporation · MQB · Jun 4, 2021 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K211108

## Device Facts

- **Applicant:** Pixxgen Corporation
- **Product Code:** [MQB](/productcode/MQB.md)
- **Decision Date:** Jun 4, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Indicated for use in general radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding fluoroscopic, angiographic, and mammographic applications.

## Device Story

Digital radiography system featuring integrated flat panel digital detector (FPD); replaces conventional film/screen systems. Captures X-ray signals to produce digital radiographic images; supports filmless environment and PACS integration. Used in clinical settings by healthcare professionals. Operates via wired (Gigabit Ethernet) or wireless (IEEE 802.11ac) connectivity; powered by AC line or rechargeable lithium battery. Output is a DICOM image for diagnostic review by radiologists. Benefits include improved cost-effectiveness and digital image quality. Device models (Prudent 1717, 1417, 1212) offer varying pixel sizes and resolutions.

## Clinical Evidence

Clinical image evaluation performed per FDA guidance. Images were reviewed by a Board Certified Radiologist and determined to be of excellent diagnostic quality. No formal clinical trial with statistical endpoints was required; substantial equivalence supported by bench testing, MTF/DQE measurements, and clinical image assessment.

## Technological Characteristics

Flat panel digital detector (FPD) with CsI or GOS scintillator. 16-bit A/D conversion. Connectivity: Gigabit Ethernet (1000BASE-T) and IEEE 802.11ac. Power: AC line or rechargeable lithium battery. Compliance: IEC 60601-1 (Electrical Safety), IEC 60601-1-2 (EMC), IEC 62133 (Battery Safety). Software: Revision 5.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- PIXX 1717, PIXX 1212 ([K182533](/device/K182533.md))
- CareView 3600RF ([K202995](/device/K202995.md))
- CareView 1800Cw ([K201932](/device/K201932.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/10 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health and Human Services emblem. To the right of the emblem is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue font. The word "FDA" is in a larger font size than the rest of the text.

PIXXGEN Corporation % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114

Re: K211108

Trade/Device Name: Prudent 1717, Prudent 1417, Prudent 1212 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: April 14, 2021 Received: April 15, 2021

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

June 4, 2021

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K211108

Device Name Prudent 1717, Prudent 1417, Prudent 1212

#### Indications for Use (Describe)

Indicated for use in general radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding fluoroscopic, angiographic applications.

Type of Use (Select one or both, as applicable)

| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------|

| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------|
|-----------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary 510(k) Number K211108 PIXXGEN Corporation 5F, SMART BAY, 123, Beolmal-ro, Dongan-gu, Anyang-si, Gyeonggi-do, 431804, Korea Tel +82.70.4846.8888 Date Prepared: April 29, 2021 Contact: Young Kim, President

### 1. Identification of the Device:

Proprietary-Trade Names: Prudent 1717, Prudent 1417, Prudent 1212 Classification Name: Solid State X-Ray Imager (Flat Panel/Digital Imager), Common/Usual Name: Digital X-Ray Receptor Panel Device Class/Regulation Number: Class II per regulation 892.1680 Product Code: MQB

- 2. Equivalent legally marketed device: K182533 PIXX 1717; PIXX 1212; Digital Diagnostic X-Ray Receptor Panels Classification Name: Solid State X-Ray Imager (Flat Panel/Digital Imager) MQB,
Common/Usual Name: Digital X-Ray Receptor Panel

Device Class/Regulation Number: Class II per regulation 892.1680

Product Code: MQB

Alternate Predicate Device 1: (for MTF/DQE comparison) K202995, CareRay Digital Medical Technology Co., Ltd

Proprietary-Trade Name: CareView 3600RF, CareRay Digital Medical Technology Co., Ltd.

Classification Name: Solid State X-Ray Imager (Flat Panel/Digital Imager),

Common/Usual Name: Digital X-Ray Receptor Panel

Device Class/Regulation Number: Class II per regulation 892.1680

Product Code: MQB

Alternate Predicate Device 2: (for MTF/DQE comparison) K201932, CareRay Digital Medical Technology Co., Ltd

Proprietary-Trade Name: CareView 1800Cw, CareRay Digital Medical Technology Co., Ltd.

Classification Name: Solid State X-Ray Imager (Flat Panel/Digital Imager),

Common/Usual Name: Digital X-Ray Receptor Panel

Device Class/Regulation Number: Class II per regulation 892.1680 Product Code: MQB

- 3. Indications for Use (intended use) Indicated for use in general radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding fluoroscopic, angiographic, and mammographic applications.
### 4. Description of the Device:

The Prudent 1717, Prudent 1417, Prudent 1212 are digital radiography systems, featuring an integrated flat panel digital detector (FPD). It is designed to perform digital radiographic examinations as a replacement for conventional film. This integrated platform provides the benefits of PACS with the advantages of digital radiography for a filmless environment and improves cost effectiveness. The major functions and principle of operation of the updated panels are the same as our previous panel retaining the Wi-Fi wireless features and rechargeable battery operation. The Prudent 1717 is

{4}------------------------------------------------

available in 3 pixel sizes: 100/140/168 um whereas the Prudent 1417, Prudent 1212 are available in two pixel sizes: 100/140 µm. The available resolutions vary according to the comparison table below. All of the models are Wi-Fi wired) and rechargeable battery (or AC line) operated. The device employs the same software as cleared in the predicate with only minor changes made.

- 5. Safety and Effectiveness, comparison to predicate device. The results of clinical image inspection, bench, and test laboratory results indicates that the new device is as safe and effective as the predicate device. Clinical images collected demonstrate equal or better image quality as compared to our predicate.

|                                | K182533, PIXX 1717, PIXX 1417, PIXX 1212                                                                                                                                                                                                                       | Prudent 1717, Prudent 1417, Prudent 1212                                                                                                                                       |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                   | Indicated for use in general radiographic<br>images of human anatomy. It is intended to<br>replace radiographic film/screen systems in<br>all general-purpose diagnostic procedures,<br>excluding fluoroscopic, angiographic, and<br>mammographic applications | UNCHANGED                                                                                                                                                                      |
| Configuration                  | This submission is for the Digital Panel and<br>Software only, no generator or stand<br>provided.                                                                                                                                                              | UNCHANGED                                                                                                                                                                      |
| Pixel Pitch                    | 140 μm                                                                                                                                                                                                                                                         | Prudent 1717: 100/140/168 μm (3 available<br>pixel sizes)<br>Prudent 1417: 100/140 μm (2 available pixel<br>sizes)<br>Prudent 1212:<br>100/140 μm (2 available<br>pixel sizes) |
| Limiting<br>Resolution         | 3 lp/mm                                                                                                                                                                                                                                                        | 5.0 lp/mm, 3.6 lp/mm 3.0 lp/mm<br>(Equal or better)                                                                                                                            |
| A/D Conversion                 | 16 bits                                                                                                                                                                                                                                                        | SAME                                                                                                                                                                           |
| Active Area                    | 17 x 17 inch<br>14 x 17 inch<br>12 x 12 inch                                                                                                                                                                                                                   | SAME                                                                                                                                                                           |
| Dimensions(mm)/<br>Weights(Kg) | 460(W)x461(L)x15(H)/3.0Kg<br>385(W) x 460(L) x 15(H)/ 2.8Kg<br>308.5(W) x 319.5(L) x 15(H)/ 1.9Kg                                                                                                                                                              | 460(W)X460(L)X15(H)/ 3.9Kg<br>382(W)X460(L)X15(H)/ 3.3Kg<br>372(W)X315(L)X15(H)/ 2.0Kg<br>Comparable sizes                                                                     |
| Pixels                         | 3,072 x 3,072<br>2,560 x 3,072<br>2,048 x 2,048                                                                                                                                                                                                                | 1717: 3072x3072/2560x2560/<br>4302x4302<br>1417: 2500x3052 / 3534x4302<br>1212: 2048x2048 / 2864x2864<br>Similar resolutions                                                   |
| Software                       | Outputs a DICOM image                                                                                                                                                                                                                                          | SAME                                                                                                                                                                           |
| DICOM                          | Yes                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                            |
| Scintillator                   | Csl or GOS                                                                                                                                                                                                                                                     | SAME                                                                                                                                                                           |
|                                | K182533, PIXX 1717, PIXX 1417, PIXX 1212                                                                                                                                                                                                                       | Prudent 1717, Prudent 1417, Prudent 1212                                                                                                                                       |
| Interface                      | Wired: Gigabit Ethernet<br>(1000BASE-T)<br>Wireless:IEEE802.11ac, backward compatible                                                                                                                                                                          | SAME                                                                                                                                                                           |
| Power source                   | AC Line and/or  Rechargeable Lithium<br>Battery;  5 hours/300 images                                                                                                                                                                                           | AC Line and/or Rechargeable Lithium<br>Battery;<br>6~8 hours/480~600 images                                                                                                    |
| Standards                      | Electrical Safety per IEC 60601-1:2012 and<br>EMC per IEC 60601-1-22007+AC:2010 as well<br>as IEEE802.11ac. Meets FCC requirements<br>plus IEC 62133 Battery safety.                                                                                           | SAME                                                                                                                                                                           |

### 6. Substantial Equivalence Chart

{5}------------------------------------------------

|          | Alternate predicates                           | Prudent 1717, Prudent 1417, Prudent 1212                                                                   |
|----------|------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| DQE(CSI) | K201932 Alternate predicate At 2 lp/mm 45%     | At 2 lp/mm   60 % / 44% / 47%<br>(@100/140/168 um respectively)<br>Same or better than alternate predicate |
| MTF(CSI) | K201932 Alternate predicate at 1 lp/mm:<br>35% | At 1 lp/mm 70%,   53%,   55%<br>(@100/140/168 um respectively)<br>Better than alternate predicate          |
| DOE(GOS) | K202995 Alternate predicate at 1 lp/mm 20%     | At 1 lp/mm 36%, 27%, 30%<br>(@100/140/168 um respectively)<br>Better than alternate predicate              |
| MTF(GOS) | K202995 Alternate predicate at 1 lp/mm:  50%   | At 1 lp/mm 56%, 55%, 54%<br>(@100/140/168 um respectively)<br>Better than alternate predicate              |

- 7. Summary of Bench Testing Conducted: IEC Standards were employed for: Electrical Safety and Electromagnetic Compatibility, and Battery Safety Tests. Standards met: Electrical safety per: IEC/UL 60601-1, Medical Electrical Equipment – Part 1: General Requirements for Safety. (General) Electromagnetic Compatibility per IEC 60601-1-2, Collateral Standard: Electromagnetic compatibility Requirements and tests. IEC 62133 Edition 2.0 2012-12 Secondary cells and batteries containing alkaline or other non-acid
electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications

Risk Analysis was conducted in accordance with ISO 14971:2012 and EN 62304. The software remains essentially the same as in K182533 but moved from Revision 4 to Revision 5. A Software Validation Report for Revision 5 was produced.

MTF and DQE measurements were made in accordance with Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices Guidance for Industry and Food and Drug Administration Staff Document issued on: September 1, 2016

Battery life testing was conducted to confirm that the Prudent panels run for 6-8 hours/480-600 images.

{6}------------------------------------------------

The panels were evaluated for Usability according to IEC 62366-1:2015 Medical devices — Part 1: Application of usability engineering to medical devices. The evaluation concluded that the intended user can safely use the device in the intended environment without use error.

Cybersecurity precautionary labeling was added per the FDA guidance: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, Guidance for Industry and Food and Drug Administration Staff, Document Issued on: October 2, 2014.

- 8. Summary of Clinical Testing: Clinical images obtained in accordance with Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices 1 Guidance for Industry and Food and Drug Administration Staff Document issued on: September 1, 2016. The images were evaluated by a Board Certified Radiologist and found to be of excellent diagnostic quality.
- 9. Conclusion: After analyzing bench, clinical image, and external laboratory testing to applicable standards, it is the conclusion of Pixxgen that the Prudent 1417, Prudent 1417, Prudent 1212 Digital Diagnostic X-Ray Receptor Panels are as safe and effective as the predicate device, have few technological differences, and has the identical indications for use, thus rendering them substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K211108](https://fda.innolitics.com/device/K211108)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
