BASHIR Endovascular Catheter (Ref. 7201), BASHIR Plus 10 Endovascular Catheter (Ref. 7210), BASHIR Plus 20 Endovascular Catheter (Ref. 7220), BASHIR Plus 30 Endovascular Catheter (Ref. 7230), BASHIR Plus 40 Endovascular Catheter (Ref. 7240)
K211061 · Thrombolex, Inc. · QEY · Jun 10, 2021 · Cardiovascular
Device Facts
Record ID
K211061
Device Name
BASHIR Endovascular Catheter (Ref. 7201), BASHIR Plus 10 Endovascular Catheter (Ref. 7210), BASHIR Plus 20 Endovascular Catheter (Ref. 7220), BASHIR Plus 30 Endovascular Catheter (Ref. 7230), BASHIR Plus 40 Endovascular Catheter (Ref. 7240)
Applicant
Thrombolex, Inc.
Product Code
QEY · Cardiovascular
Decision Date
Jun 10, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K211061 · Jun 10, 2021
BASHIR Endovascular Catheter (Ref. 7201), BASHIR Plus 10 Endovascular Catheter (Ref. 7210), BASHIR Plus 20 Endovascular Catheter (Ref. 7220), BASHIR Plus 30 Endovascular Catheter (Ref. 7230), BASHIR Plus 40 Endovascular Catheter (Ref. 7240)
Thrombolex, Inc.
Sequential case information from 15 clinical institutions
Retrospective analysis of 45 patients treated with the BASHIR Endovascular Catheter and BASHIR Plus Endovascular Catheters to support the safety and effectiveness of the device for the treatment of venous thrombus.
Venous thrombus; Real-world use; Retrospective case series
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Sequential case information from Real-World use; Retrospective case series
Patients with venous thrombus (primarily iliofemoral thrombus); Sample Size: 45; Number of Sites: 15
Not applicable for this study
Complete or partial resolution of venous thrombus; incidence of bleeding or other adverse events
Indications for Use
The BASHIR™ Endovascular Catheter and the BASHIR™ Plus Endovascular Catheters (BASHIR™+10, BASHIR™+20, BASHIR™+30, BASHIR™+40) are intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature, enabling the restoration of blood flow in patients with venous thrombus.
Device Story
Endovascular catheters designed for controlled, selective infusion of physician-specified fluids (e.g., thrombolytics) into peripheral vasculature; used to restore blood flow in patients with venous thrombus. Device consists of infusion basket and shaft; delivered via sheath over guidewire under fluoroscopic guidance. Infusion basket expands within vessel to fragment clot and create channel; multiple laser-drilled holes in infusion limbs/shaft allow fluid delivery. BASHIR Plus models include second shaft port for additional infusion coverage (10-40 cm). Operated by interventional specialists in clinical settings. Mechanical device; no electrical components. Output is physical delivery of therapeutic fluids to target site; clinical benefit is restoration of blood flow via thrombus resolution.
Clinical Evidence
Real-world clinical data from 15 institutions; 45 patients with venous thrombus (primarily iliofemoral) treated with BASHIR/BASHIR Plus catheters. Results showed complete or partial resolution of venous thrombus with no reported bleeding or adverse events.
Technological Characteristics
Continuous flush catheter; mechanical design; no electrical components. Features infusion basket and shaft with laser-drilled holes for fluid delivery. Delivered via sheath over guidewire under fluoroscopic guidance. Materials are standard vascular catheter materials.
Indications for Use
Indicated for patients with venous thrombus requiring controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature to restore blood flow.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
February 3, 2022
Thrombolex, Inc. % Diane Horwitz Regulatory Consultant Eminence Clinical Research Inc. 5 Lake Como Ct. Greenville. South Carolina 29609
### Re: K211061
Trade/Device Name: BASHIR Endovascular Catheter, Reference No. 7201 and BASHIR Plus Endovascular Catheter, Reference No. 7210, 7220, 7230, 7240 (BASHIR+10, BASHIR+20, BASHIR+30, BASHIR+40) Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: OEY, KRA
Dear Diane Horwitz:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated June 10, 2021. Specifically, FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell. OHT2: Office of Cardiovascular Devices. (301) 796-6075. Gregory. Oconnell@FDA.HHS.gov.
Sincerely,
Gregory W. Digitally signed by
Gregory W. Gregory W. O'connell -S O'connell -S Date: 2022.02.03
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 10, 2021
Thrombolex, Inc. % Diane Horwitz Regulatory Consultant Eminence Clinical Research Inc. 5 Lake Como Ct. Greenville, South Carolina 29609
### Re: K211061
Trade/Device Name: BASHIR Endovascular Catheter (Ref. 7201), BASHIR Plus 10 Endovascular Catheter (Ref. 7210), BASHIR Plus 20 Endovascular Catheter (Ref. 7220), BASHIR Plus 30 Endovascular Catheter (Ref. 7230), BASHIR Plus 40 Endovascular Catheter (Ref. 7240) Regulation Number: 21 CFR 870.1210
Regulation Name: Continuous flush catheter Regulatory Class: Class II Product Code: KRA Dated: April 8, 2021 Received: April 9, 2021
Dear Diane Horwitz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Gregory W. Gregory W. O'connell -S O'connell -S Pate: 2021.06.10
12:43:23 -04'00'
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
#### 510(k) Number: K211061
#### Device Name
BASHIR™ Endovascular Catheter and BASHIR™ Plus Endovascular Catheters (BASHIR™+20, BASHIR™+30, BASHIR™+40)
#### Indications for Use (Describe)
The BASHIR™ Endovascular Catheter and the BASHIR™ Plus Endovascular Catheters (BASHIR™+10, BASHIR™+20, BASHIR™+30, BASHIR™+40) are intended for the controlled and selective infusion of physicianspecified fluids, including thrombolytics, into the peripheral vasculature, enabling the restoration of blood flow in patients with venous thrombus.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K211061
Image /page/4/Picture/1 description: The image contains the logo for THROMBOLEX. The logo is red and has a stylized X at the end. Below the logo, the text "INNOVATIVE ENDOVASCULAR CATHETERS" is written in a smaller font.
### 510(k) Summary
#### 1. GENERAL INFORMATION
#### Submitter and 510(k) Owner 1.1
Thrombolex, Inc. 75 Britain Dr. New Britain PA 18901
#### Official Correspondent 1.2
Diane Horwitz, Ph.D. 5 Lake Como Ct. Greenville, SC 29609 Telephone: 703.307.2921 Email: dhorwitz@ecr-inc.com
#### 1.3 Date of Preparation
April 8, 2021
## 2. NAME OF THE DEVICE
#### 2.1.1 Trade/Proprietary Name
BASHIR™ Endovascular Catheter, Reference No. 7201 BASHIR™ Plus Endovascular Catheter, Reference No. 7210, 7220, 7230, 7240 (BASHIRTM +10, BASHIR™+20, BASHIR™+30, BASHIR™+40)
#### 2.1.2 Common/Usual Name
Continuous Flush Catheter
#### 2.1.3 Classification Information
| Classification Name: | Continuous Flush Catheter |
|----------------------------|---------------------------------|
| Classification Regulation: | 21 CFR 870.1210 |
| Class: | II |
| Product Code: | KRA – Continuous Flush Catheter |
| Panel: | Cardiovascular |
## 3. PREDICATE DEVICE
The predicate devices are the BASHIR™ Endovascular Catheter, K183290 and the BASHIR™ Plus Endovascular Catheters, K193071.
1
{5}------------------------------------------------
# 4. DESCRIPTION OF THE DEVICE
The BASHIR™ Endovascular Catheter and the BASHIR™ Plus Endovascular Catheter are continuous flush catheters designed to enable delivery of physician-specified fluids, including thrombolytics, into the peripheral circulation. There is one model of the BASHIR™ Endovascular Catheter (Ref. No. 7201) and four models of the Bashir™ Plus Endovascular Catheter, (Ref. No. 7210, 7220, 7230, 7240). The design of the catheters allow for delivery and positioning of the catheter at the target peripheral location via a sheath, over a guidewire using fluoroscopic guidance. The infusion can be initiated into a distal infusion basket via the Basket Port either in the non-expanded position (for vessels with smaller diameters), or by controlled expansion of the infusion basket to a desired dimension inside the vessel by the interventional specialist under fluoroscopic guidance. When the infusion basket is expanded into a clot, the clot is fragmented and a channel created, providing an opportunity for restoration of endogenous blood flow. The physician-specified solution is delivered via multiple laser-drilled holes through the infusion limbs of the catheters after placement.
The BASHIR™ Plus Endovascular Catheters have a second Shaft Port, for delivery of infusion solutions into the shaft lumen adjacent to the infusion basket where multiple laserdrilled holes extend over 10 cm (BASHIR+10), 20 cm (BASHIR+20), 30 cm (BASHIR+30) or 40 cm (BASHIR+40) of shaft length.
After the completion of solution delivery, the infusion limbs are collapsed to a closed position and the BASHIR™ Endovascular Catheter and/or BASHIR™ Plus Endovascular Catheter are retracted into the sheath and removed.
The catheters are comprised of common materials used in vascular catheters are mechanical and contain no electrical components.
## 5. INTENDED USE / INDICATION FOR USE
The BASHIR™ Endovascular Catheter and the BASHIR™ Plus Endovascular Catheters (BASHIR™+10, BASHIR™+20, BASHIR™+30, BASHIR™+40) are intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature, enabling the restoration of blood flow in patients with venous thrombus.
## 6. COMPARISON OF THE INTENDED USE WITH THE PREDICATE DEVICE
The Intended Use statement is identical to that of the predicate devices with the exception of adding a specific indication for use, as an example of the types of patients that can be treated with the BASHIR™ Endovascular Catheter and the BASHIR™ Plus Endovascular Catheters. The modification, addition of the wording the restoration of blood flow in patients with venous thrombus," is an example of the patient population to be treated.
{6}------------------------------------------------
## TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE
The BASHIR™ Endovascular Catheter and the BASHIR™ Plus Endovascular Catheters are lentical in design, operational and technological characteristics and supports that no new safety concerns are being raised by change in the Intended Use/Indications for Use statement and thus raises no new issues of safety or effectiveness.
# PERFORMANCE TESTING
The BASHIR™ Endovascular Catheter and the BASHIR™ Plus Endovascular Catheters were previously subjected to a comprehensive test program that included animal and bench testing, biocompatibility, sterilization and packaging testing. Clinical data for the treatment of venous thrombus, already included in the intended use of the device, support the additional wording in the intended use statement as summarized below:
Tlinical Evidence: Sequential case information from Real-World use of the BASHIR™ indovascular Catheter and the range of BASHIR™ Plus Endovascular Catheters in 15 ifferent institutions provides evidence for expanding the labeling to include the indication for use in patients with venous thrombus. The devices were used to successfully treat 45 patients with venous thrombus (primarily iliofemoral thrombus) with either complete or partial resolution of venous thrombus, with no bleeding or other adverse events.
# 9. CONCLUSIONS
n summary, the Real-World clinical information presented in this 510(k) submission upports the additional indications for use wording for the BASHIR™ Endovascular Catheter and the BASHIR™ Plus Endovascular Catheters to restore blood flow in patients with venous thrombus, and raises no new issues of safety or effectiveness, leading to the conclusion of substantial equivalence to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.