Retrospective clinical studies were conducted to validate the performance of the CereFlow™ V1.2 software, specifically assessing its ability to generate accurate quantitative perfusion parameters.
Validation of performance of the CereFlow™ V1.2 meeting the acceptance criteria
Indications for Use
CereFlow™ V1.2 is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" workstation and can be used to perform image viewing, processing, image collage and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities. CereFlow™ V1.2 provides visualization and automatic semi-quantitative quantification of Arterial Spin Labeling (ASL) data acquired from Magnetic Resonance Imaging (MRI) scanners at 1.5T and 3.0T. CereFlow™ V1.2 is able to process single-delay and multi-delay ASL data to quantify multiple perfusion parameters (Cerebral Blood Flow/CBF, Arterial Transit Time/ATT), and exports the results of ASL image analysis as DICOM images and in a report.
Device Story
CereFlow™ V1.2 is a standalone medical image management and processing software; operates on standard Windows 10 64-bit PC workstations. Inputs consist of raw ASL DICOM images imported via CD-R. Software performs image pre-processing (file validation, motion correction) and applies a one-compartment perfusion model and nonlinear iterative curve-fitting to calculate Cerebral Blood Flow (CBF) and Arterial Transit Time (ATT). Outputs include processed perfusion maps, quality control parameters, and reports exported as DICOM images or PDF. Used in clinical settings by trained professionals; provides quantitative data to assist clinicians in evaluating patient perfusion status. Does not replace clinical judgment. Benefits include automated, standardized quantification of ASL perfusion data to support diagnostic workflows.
Clinical Evidence
Retrospective clinical studies conducted to validate performance against acceptance criteria. Bench testing included system-level functional testing (DICOM import, case management, quantification, report generation), digital phantom validation, and instrumental validation studies.
Technological Characteristics
Software-only; runs on standard Windows 10 64-bit PC. Processes DICOM-compliant ASL MRI data (1.5T/3.0T). Implements one-compartment perfusion model and nonlinear iterative curve-fitting. Standalone deployment; no network/PACS interface. Compliant with DICOM 3.1, ISO 14971:2019, and IEC 62304:2006/AMD 1:2015.
Indications for Use
Indicated for trained professionals, including physicians and medical technicians, to perform visualization and automatic semi-quantification of ASL MRI data (1.5T and 3.0T) to generate perfusion parameters (CBF, ATT) for clinical analysis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 31, 2023
Translational MRI, LLC % Andrew Wu General Manager, Software Consultant Rook Quality Systems Taiwan Branch Office 802, No. 502, Sec 2, Ren'ai Road, Linkou District New Taipei City, 24 TAIWAN
Re: K211059
Trade/Device Name: CereFlow™ V1.2 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: February 15, 2023 Received: February 27, 2023
Dear Andrew Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211059
Device Name CereFlow™ V1.2
#### Indications for Use (Describe)
CereFlow™ V1.2 is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelt" workstation and can be used to perform image viewing, processing, image collage and analysis of medical images are acquired through DICOM compliant imaging devices and modalities.
CereFlow™ V1.2 provides visualization and automatic semi-quantification of Arterial Spin Labeling (ASL) data acquired from Magnetic Resonance Imaging (MRI) scanners at 1.5T and 3.0T.
CereFlow™ V1.2 is able to process single-delay and multi-delay ASL data to generate multiple perfusion parameters (Cerebral Blood Flow/CBF, Arterial Transit Time/ATT), and exports the results of ASL image analysis as DICOM images and in a report.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Date Prepared
Feb 13th, 2023
# Manufacturer and 510(k) Owner
Translational MRI, LLC 3200 S Sepulveda Blvd Suite K30 Los Angeles, CA 90034
| Official Contact: | Zhihai Ricky Xiang |
|-------------------|--------------------------|
| Telephone | +1-530-760-9006 |
| Email | zhihaixiang@transmri.com |
# Representative/Consultant
| Andrew Wu, General Manager Principal Software Consultant | | |
|---------------------------------------------------------------------------------------|----------------------|--|
| Rook Quality Systems Taiwan Branch | | |
| Office 802, No. 502, Sec 2, Ren'ai Road, Linkou District, New Taipei City, 244, R.O.C | | |
| Telephone: | +886-912-258980 | |
| Email: | andrew.wu@rookqs.com | |
#### Device Information
| Trade/Proprietary Name | CereFlow™ V1.2 |
|------------------------|------------------------------------------------|
| Classification Name | Medical image management and processing system |
| Regulation Number | 21 CFR 892.2050 |
| Product Code(s) | LLZ |
| Classification | Class II |
| Review Panel | Radiology Device |
| Use | Prescription |
{4}------------------------------------------------
#### Indications for Use
CereFlow™ V1.2 is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" workstation and can be used to perform image viewing, processing, image collage and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities.
CereFlow™ V1.2 provides visualization and automatic semi-quantification of Arterial Spin Labeling (ASL) data acquired from Magnetic Resonance Imaging (MRI) scanners at 1.5T and 3.0T.
CereFlow™ V1.2 is able to process single-delay and multi-delay ASL data to quantify multiple perfusion parameters (Cerebral Blood Flow/CBF, Arterial Transit Time/ATT), and exports the results of ASL image analysis as DICOM images and in a report.
#### Device Description
CereFlow™ V1.2 is a medical viewing, analysis and processing, Medical Image Management and Processing System software implemented in Java and Java tools, compliant with the DICOM standard and running on standard PC (Windows 10 64-bit). CereFlow™ V1.2 is intended for visualization and automatic semi-quantitative quantification of arterial spin labeling (ASL) data acquired from MRI scanners at 1.5T and 3.0T. CereFlow™ V1.2 is able to process single-delay and multi-delay ASL data to generate multiple perfusion parameters (Cerebral Blood Flow/CBF, Arterial Transit Time/ATT), and exports the results of ASL image analysis as DICOM images and in a report.
The software consists of three modules, namely the interface module, and the database module. The interface module is responsible for interacting with users and helping them to manage cases easily. The report module is responsible for processing raw DICOM images and calculating parameters to generate ATT/CBF images which can be exported as DICOM images and in a report. Image pre-processing includes sorting and checking files to ensure designated protocols are used, having a complete set of reference (MO) image and label/control images, and implementing motion correction. The calculation of CBF and ATT parameters is implemented using the standard one-compartment perfusion model recommended by the consensus ASL white paper and standard nonlinear iterative curve-fitting approach for multi-delay ASL data. For single-delay ASL data, CBF is calculated according to the standard one-compartment perfusion model recommended by the ASL white paper.
The database module is responsible for managing cases including storing, query and deleting case data. The software applies an arrange of quality control parameters to ensure the quality of acquired ASL data and processed results. These quality control parameters can also be exported as a report.
The entire workflow is processed locally, and the report module shall only allow raw ASL imaging data importing from DICOM CD-R only. CereFlow™ V1.2 is not interface with DICOM servers, MR scanners, or PACS over wired network. It is a standalone software application designed to run on a dedicated PC workstation.
{5}------------------------------------------------
CereFlow™ V1.2 is for prescription use. Information provided by CereFlow™ V1.2 is not intended in any way to eliminate, replace, or substitute for, in whole or in part, the healthcare provider's judgment and analysis of the patient's condition.
# Technological Characteristics
The CereFlow™ V1.2 is a software-only device that runs on a workstation computer that meets the minimum requirements.
The CereFlow™ V1.2 receives DICOM datasets acquired from MRI scanners at 1.5T and 3.0T and provides visualization and automatic semi-quantification of ASL data to the physicians. The user can import single-delay and multi-delay ASL data, and export the results of ASL image analysis as DICOM images and in a PDF report.
CereFlow™ V1.2 does not contact the patient, nor does it control any life sustaining devices. Information provided by the CereFlow™ V1.2 is not intended in any way to eliminate, replace, or substitute for, in whole or in part, the healthcare provider's judgment and analysis of the patient's condition.
# Predicate Device Identification
The CereFlow™ V1.2 is substantially equivalent to the Olea Sphere V3.0 (K152602).
The CereFlow™ V1.2 has the same intended use and similar indications, technological characteristics, and principles of operation to the device cleared in K152602. The differences in technological characteristics between the subject device and the predicate device do not raise new issues of safety and effectiveness. Therefore, K152602 is listed as the predicate device for the CereFlow™ V1.2.
#### Equivalence to Predicate Device
Translational MRI, LLC submits the following information to demonstrate that the CereFlow™ V1.2 is substantially equivalent to the following legally marketed predicate device:
| 510(k) Number | Predicate Device<br>Name/Manufacturer | Primary Predicate |
|---------------|---------------------------------------|-------------------|
| K152602 | Olea Sphere V3.0 | Yes |
Translational MRI, LLC intends to use the reference device listed below for validation purpose. Translational MRI, LLC intends to compare the performance (i.e. quantification accuracy of CBF maps) of CereFlow™ V1.2 to the performance of GE 3D ASL at each post-labeling delay using the ASL data acquired via GE scanner.
| 510(k) Number | Predicate Device<br>Name/Manufacturer | Reference Device |
|---------------|---------------------------------------|------------------|
| K092925 | 3D ASL/GE Healthcare | Yes |
{6}------------------------------------------------
The subject device has the same intended use and similar technological characteristics (e.g. software for visualization and quantification of perfusion parameters of human brain acquired from MR scanners of high magnetic field strength (1.5T, 3.0T) using designated ASL protocol) to the device cleared in K152602. The differences in technological characteristics between the subject device and predicate device are explained below.
- In the predicate device, the MR image acquired by protocols of conventional MRI, ASL, and Diffusion Weighted Imaging (DWI), whereas the subject device receives MR images only based on ASL protocol acquired from commercial MRI scanners at 1.5T and 3.0T. Hence, this technological difference does not raise new issues of safety and effectiveness as compared to the predicate device.
- . In the predicate device, the image process algorithm can only process single-delay ASL data, whereas the subject device is able to process single-delay and multi-delay ASL data to generate multiple perfusion parameters (Cerebral Blood Flow/CBF, and Arterial Transit Time/ATT). Additional verification tests were carried out to demonstrate that the subject device yields accurate quantitative perfusion parameters using multi-delay ASL data. Hence, this technological difference does not raise new issues of safety and effectiveness as compared to the predicate device.
- In the predicate device, the software outputs the perfusion map and allows user to customize the segmentation masks based on user defined areas, whereas the subject device only outputs multi-parametric maps of the whole brain region. Hence this technological difference does not raise new issues of safety and effectiveness as compared to the predicate device
- . The predicate device can be operational on Windows and Linux, while the subject device can be operational on Windows only. Hence, this technological difference does not raise new issues of safety and effectiveness as compared to the predicate device.
- . The predicate device can acquire images from multiple modalities or network using DICOM protocol, while the subject device can only import DICOM images via DICOM CD-R. Hence, this technological difference does not raise new issues of safety and effectiveness as compared to the predicate device.
The subject device does not have the following functionalities whereas the predicate has:
- . Permeability post-processing (the calculation of parameters related to leakage of injected contrast material from intravascular to extracellular space)
- . Relaxometry post-processing (the calculation of parameters related to the MR longitudinal and transversal relaxation time and rate)
- . Metabolic post-processing (the calculation of parameters related to the fat signal fraction based on a MR technique using opposed-phase imaging)
Translational MRI, LLC believes that the CereFlow™ V1.2 described in this notification and for use under the conditions of the proposed labeling is substantially equivalent to the legally marketed predicate device (K152602) based on the information summarized in the following Table 1 – Substantial Equivalence Summary.
{7}------------------------------------------------
| | Subject Device | Primary Predicate Device |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | CereFlow™ V1.2 | Olea Sphere V3.0 |
| Manufacturer | Translational MRI, LLC | Olea Medical |
| 510(k) Number | K211059 | K152602 |
| Class | II | II |
| Device Classification Name | Medical image management<br>and processing system | Medical image management<br>and processing system |
| Regulation Number | 892.2050 | 892.2050 |
| Product Code | LLZ | LLZ |
| Indications for Use | CereFlow™ V1.2 is an image<br>processing software package to<br>be used by trained professionals<br>including but not limited to<br>physicians and medical<br>technicians. The software runs<br>on a standard "off-the-shelf"<br>workstation and can be used to<br>perform image viewing,<br>processing, image collage and<br>analysis of medical images. Data<br>and images are acquired<br>through DICOM compliant<br>imaging devices and modalities.<br>CereFlow™ V1.2 provides<br>visualization and automatic<br>semi-quantitative quantification<br>of Arterial Spin Labeling (ASL)<br>data acquired from Magnetic<br>Resonance Imaging (MRI)<br>scanners at 1.5T and 3.0T. It is<br>recommended the minimal SNR<br>of perfusion images in white<br>matter should be equal to or<br>larger than 1. | Olea Sphere V3.0 is an image<br>processing software package to<br>be used by trained professionals<br>including but not limited to<br>physicians and medical<br>technicians. The software runs<br>on a standard "off-the-shelf"<br>workstation and can be used to<br>perform image viewing,<br>processing, image collage and<br>analysis of medical images. Data<br>and images are acquired<br>through DICOM compliant<br>imaging devices and modalities.<br>Olea SphereV3.0 provides both<br>viewing and analysis capabilities<br>of functional and dynamic<br>imaging datasets acquired with<br>MRI or other relevant<br>modalities, including a Diffusion<br>Weighted Imaging (DWI) / Fiber<br>Tracking Module and a Dynamic<br>Analysis Module (e.g. dynamic<br>exogenous or endogenous<br>contrast enhanced imaging data<br>for MRI and CT). The DWI<br>Module is used to visualize local<br>water diffusion properties from<br>the analysis of Diffusion<br>Weighted MRI data. The Fiber<br>Tracking feature utilizes the<br>directional dependency of the<br>diffusion to display the white<br>matter structure in the brain or<br>more generally the central<br>nervous system. The Dynamic |
| Physical Characteristics | Same | |
| Computer | Same | Software package that operates<br>on off-the-shelf workstation<br>PC Compatible |
| | CereFlow™ V1.2 is able to<br>process single-delay and multi-<br>delay ASL data to generate<br>multiple perfusion parameters<br>(Cerebral Blood Flow/CBF,<br>Arterial Transit Time/ATT), and<br>exports the results of ASL image<br>analysis as DICOM images and<br>in a report. | Analysis Module is used for<br>visualization and analysis of<br>dynamic imaging data, showing<br>properties of changes in<br>contrast while repeating<br>acquisitions (e.g. over time with<br>or without variable acquisition<br>parameters) where such<br>techniques are useful or<br>necessary. This functionality is<br>referred to as:<br><br>Perfusion Module – the<br>calculation of parameters<br>related to tissue blood flow<br>(perfusion) and tissue blood<br>volume.<br><br>Permeability Module – the<br>calculation of parameters<br>related to leakage of injected<br>contrast material from<br>intravascular to extracellular<br>space.<br><br>Arterial Spin Labeling (ASL)<br>Module - the calculation of<br>Cerebral Blood Flow/CBF<br>related to tissue flow based on<br>a MR technique using the water<br>in arterial blood as endogenous<br>tracer to evaluate the perfusion.<br><br>Relaxometry Module - the<br>calculation of parameters<br>related to the MR longitudinal<br>and transversal relaxation time<br>and rate.<br><br>Metabolic Module - the<br>calculation of parameters<br>related to the fat signal fraction<br>based on a MR technique using<br>opposed-phase imaging. |
| | | |
| Image Processing Location | Same | Onsite on the desktop<br>computer as local application |
| DICOM Standard Compliance | Same | The software processes DICOM<br>compliant image data |
| Computer OS Compatibility | Windows | Windows, Linux |
| Modalities | Arterial Spin Labeling | Multi-modality |
| User Interface | Same | The software is designed for use<br>on a radiology workstation. |
| Protocols | Same | Predefined or specific<br>acquisition protocol settings |
| Image Processing Algorithm<br>Description | Calculation of parameters<br>related to blood flow based on<br>a MR technique using the water<br>in arterial blood as endogenous<br>tracer to evaluate the perfusion.<br><br>Process single-delay and multi-<br>delay ASL data to quantify<br>multiple perfusion parameters<br>(Cerebral Blood Flow/CBF, and<br>Arterial Transit Time/ATT) | Calculation of parameters<br>related to blood flow based on<br>a MR technique using the water<br>in arterial blood as endogenous<br>tracer to evaluate the perfusion.<br><br>Process single-delay ASL data<br>and quantify Cerebral Blood<br>Flow/CBF |
| Image Acquisition | Same | The acquisition remains the<br>same, i.e. the image processing<br>can be generated from multiple<br>modalities and with predefined<br>or specific acquisition protocol<br>settings. |
| Region of Interest | User cannot select region of<br>interest. | User can customize the<br>segmentation masks based their<br>selected region of interest. |
{8}------------------------------------------------
{9}------------------------------------------------
# Performance Data
The subject device is designed in conformance with:
- ACR/NEMA Digital Imaging Communications in Medicine (DICOM) Version 3.1
- ISO 14971:2019 Medical Devices Application of Risk Management to Medical Devices ●
- IEC 62304:2006/AMD 1:2015 – Medical Device Software – Software Life-Cycle Processes
All specifications of the CereFlow™ V1.2 were verified by a number of tests before release. Non-clinical testing including verification tests on:
- System-level, functional testing .
- O DICOM images importation
- O Case management
- O Automatic semi-quantitative quantification
- O Multiple perfusion parameters calculation
{10}------------------------------------------------
- Visualization of ASL image o
- Report generation o
- Digital phantom validation study
- Instrumental validation studies ●
In addition, retrospective clinical studies were conducted to validate performance of the CereFlow™ V1.2 meeting the acceptance criteria.
Translational MRI, LLC believes that the aforementioned non-clinical testing demonstrates that the subject device is designed in such a way that, when used under the conditions and for the purposes intended, the safety and effectiveness, as well as the performance characteristic of the subject device is substantially equivalent to the predicate device.
# Substantial Equivalence Conclusion
CereFlow™ V1.2 has the same intended use and performance characteristics as the predicate device. Based on the software verification performed, it can be concluded that the differences in technological characteristics between the CereFlow™ V1.2 and the predicate device do not raise different questions of safety and effectiveness under specified use conditions. The indications for use, technological characteristics, and performance characteristics for the CereFlow™ V1.2 are assessed to be substantially equivalent to the predicate device. There is no identified hazard that require additional benefit versus risk analysis in support of substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.