K210998 · Orthosoft D/B/A Zimmer Cas · LLZ · Aug 17, 2021 · Radiology
Device Facts
Record ID
K210998
Device Name
ROSA Hip System
Applicant
Orthosoft D/B/A Zimmer Cas
Product Code
LLZ · Radiology
Decision Date
Aug 17, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ROSA® Hip System for use with the ROSA® RECON platform, is indicated as a fluoroscopic-guided system for total hip arthroplasty (THA). It is used to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of hip implant components provided that the points of interest can be identified from radiology images. The intraoperative cup placement is performed relative to anatomical landmarks as recorded using the system intraoperatively, and based on preoperative planning values optionally determined using compatible X-Ray based surgical planning tools. The ROSA® Hip System is designed for use on a skeletally mature patient population. The targeted population has the same characteristics as the population that is suitable for the implants compatible with the ROSA® Hip System. The ROSA® Hip System is not for primary image interpretation and is applicable for the direct anterior approach. The ROSA® Hip System is to be used with the following hip replacement systems in accordance with their indications and contraindications: G7® Acetabular System, Avenir® Hip System, Avenir Complete™ Hip System, Taperloc® Complete Hip System, Echo® Hip System.
Device Story
ROSA Hip System assists surgeons during total hip arthroplasty (THA) via direct anterior approach. Inputs: intraoperative fluoroscopic images (C-arm) and preoperative planning values. System captures images via ROSA Tablet; computes instrument orientation relative to patient anatomy; uses ROSA RECON robotic arm to guide cup inserter to planned orientation. Surgeon views measurements and overlays on system display to adjust cup placement, leg length, and femoral offset. Robotic arm remains stationary during acetabular impaction. Used in OR by surgeons; does not perform pre-operative planning or bone cutting. Benefits: improved accuracy of implant component placement and assessment of leg length/offset discrepancies.
Clinical Evidence
Bench testing only. Includes biocompatibility (ISO 10993), electrical safety (IEC 60601-1), and EMC (IEC 60601-1-2). Performance validation included physical/performance tests, engineering analysis, usability engineering, and cadaveric specimen validation lab to confirm safety and effectiveness.
Technological Characteristics
System includes software, reusable/disposable instrumentation, and ROSA RECON robotic arm. Connectivity: C-arm, tablet, and Zimmer Biomet Drive Portal. Standards: ISO 10993 (biocompatibility), IEC 60601-1 (electrical), IEC 60601-1-2 (EMC), IEC 62304 (software lifecycle).
Indications for Use
Indicated for skeletally mature patients undergoing total hip arthroplasty (THA) via direct anterior approach. Contraindicated for patients with significant bone loss (e.g., avascular necrosis, severe dysplasia), active hip joint infections, revision surgery, or presence of radio-opaque elements interfering with imaging.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below.
August 17, 2021
Orthosoft d/b/a Zimmer CAS Paul Hardy Regulatory Affairs Associates Director 75 Queen Street, Suite 3300 Montreal, Quebec H3C 2N6 Canada
# Re: K210998
Trade/Device Name: ROSA Hip System Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ, OLO Dated: July 12, 2021 Received: July 13, 2021
### Dear Paul Hardy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of in vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210998
Device Name ROSA® Hip System
#### Indications for Use (Describe)
The ROSA® Hip System for use with the ROSA® RECON platform, is indicated as a fluoroscopic-guided system for total hip arthroplasty (THA). It is used to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of hip implant components provided that the points of interest can be identified from radiology images.
The intraoperative cup placement is performed relative to anatomical landmarks as recorded using the system intraoperatively, and based on preoperative plannly determined using compatible X-Ray based surgical planning tools.
The ROSA® Hip System is designed for use on a skeletally mature patient population. The targeted population has the same characteristics as the population that is suitable for the implants compatible with the ROSA® Hip System. The ROSA® Hip System is not for primary image interpretation and is applicable for the direct anterior approach.
The ROSA® Hip System is to be used with the following hip replacement systems in accordance with their indications and contraindications: G7® Acetabular System, Avenir Complete™ Hip System, Taperloc® Complete Hip System, Echo® Hip System.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the ROSA® Hip System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on September 13, 2019.
| Sponsor: | Orthosoft, Inc. d/b/a Zimmer CAS<br>75 Queen St., Suite 3300<br>Montreal, QC, H3C 2N6, CANADA<br>Establishment Registration Number: 9617840 |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Paul Hardy<br>Regulatory Affairs Associate Director<br>Telephone: 574-453-6739<br>Paul.Hardy@zimmerbiomet.com |
| Date: | August 17, 2021 |
| Subject Device: | Trade Name: ROSA® Hip System<br>Common Name: ROSA® Hip System, ROSA Hip,<br>ROSA® THA System |
| | Classification Name:<br>• LLZ - Medical Image Management And Processing<br>System (21 CFR 892.2050)<br>• OLO - Stereotaxic Instrument (21 CFR 882.4560) |
### Predicate Devices:
| Manufacturer | Device Name | 510(k) Number |
|------------------|------------------------|---------------|
| JointPoint, Inc. | JointPoint<br>*primary | K160284 |
| Zimmer CAS | ROSA® Knee<br>System | K182964 |
{5}------------------------------------------------
### Purpose and Device Description:
The ROSA® Hip System (RHS) for use with the ROSA® RECON platform is used to assist surgeons in performing Total Hip Arthroplasty (THA) with features to assist in acetabular shell impaction for the direct anterior approach, as well as assessing the leg length discrepancy and the femoral offset.
The ROSA® Hip System uses a Non-Device Medical Device Data System (MDDS) called the Zimmer Biomet Drive Portal, which manages the creation and tracking of surgical cases. The cases reside on the portal until they are uploaded to the ROSA® RECON Platform before surgeries. The ROSA® Hip System utilizes the robotic arm of the ROSA® RECON platform cleared in K182964, but does not add new stereotaxic or robotic components
The system uses fluoroscopic images to determine the instruments' orientation in relation to the patient anatomy and as a guide for acetabular component orientation. The system allows the surgeon to input the case's surgical preoperative planning values and preview the acetabular component orientation intra-operatively. Throughout the surgical workflow, fluoroscopic images are acquired with a C-arm. Fluoroscopic images are then captured with the ROSA® Tablet and transferred onto ROSA®. The current instruments' orientation is computed from the image capture and is adjusted to match the surgeon's planning values using the ROSA® RECON robotic arm. The system provides pre, intra and post-operative measurements relative to patient anatomy and does not provide infrared-based stereotaxic navigation for implant placement. The robotic arm is maintained stationary to keep the instruments in a fixed orientation during acetabular component impaction. The system also provides component selection options based on leg length and offset discrepancies measurements.
The intra-operative workflow and surgical concepts implemented in the system remain close to the conventional THA direct anterior approach workflow. As such, at the time of the surgery, the system mainly assists the surgeon in (1) determining reference alignment axes and cup orientation using image-to-image and robotic
{6}------------------------------------------------
registration, (2) precisely orienting the cup inserter relative to the desired orthopedic implant angle by using a robotic arm, and (3) providing leg length and offset discrepancies measurements based on fluoroscopic image references.
# Indications for Use: The ROSA® Hip System for use with the ROSA® RECON platform, is indicated as a fluoroscopic-guided system for total hip arthroplasty (THA). It is used to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of hip implant components provided that the points of interest can be identified from radiology images. The intraoperative cup placement is performed relative to anatomical landmarks as recorded using the system intraoperatively, and based on preoperative planning values optionally determined using compatible X-Ray based surgical planning tools. The ROSA® Hip System is designed for use on a skeletally mature patient population. The targeted population has the same characteristics as the population
that is suitable for the implants compatible with the ROSA® Hip System. The ROSA® Hip System is not for primary image interpretation and is applicable for the direct anterior approach.
The ROSA® Hip System is to be used with the following hip replacement systems in accordance with their indications and contraindications: G7® Acetabular System, Avenir® Hip System, Avenir Complete™ Hip System, Taperloc® Complete Hip System, Echo® Hip System.
The ROSA® Hip System may not be suitable for use in case of:
- hip pathology with significant bone loss (e.g. avascular o necrosis of the femoral head with collapse, severe dysplasia of the femoral head or the acetabulum)
- active infections of the hip joint area о
- hip replacement revision surgery O
- presence of undesirable radio-opaque element during O intra-operative image acquisition
### Contraindications:
{7}------------------------------------------------
- contraindications for the implant as given by the implant O manufacturer and
- implants that are not compatible with the system O
## Summary of Technological Characteristics:
The rationale for substantial equivalence is based on consideration of the following characteristics:
- · The proposed and predicate devices have the same intended uses and similar indications for use. The devices are intended to assist surgeons in the accurate positioning & placement of hip or knee components.
- · The proposed and JointPoint predicate device utilize a Carm and digital tablet to acquire intraoperative fluoroscopic images for patient registration and positioning of hip implant components.
- · The proposed and ROSA predicate device share the same previously cleared ROSA® RECON Platform, and the proposed and ROSA predicate device consists of major components including a software system, various instrumentation including reusable and disposable. The proposed and ROSA predicate use the robotic arm on the ROSA® RECON Platform to assist the guidance of instruments.
- · The proposed and primary predicate allow users to mark and adjust landmarks within the software for assessment of leg length and offset for the hip implant components
#### Summary of Performance Data (Nonclinical and/or Clinical)
The following performance data was provided in support of the substantial equivalence determination:
# Biocompatibility Testing
The biocompatibility evaluation for ROSA Hip was conducted in accordance with ISO 10993. The evaluation reveals that the ROSA Hip device meets biocompatibility requirements.
{8}------------------------------------------------
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical Safety and EMC testing was conducted on the ROSA RECON Platform. The device complies with recognized electrical safety standards: IEC 60601-1 standard for electrical safety and IEC 60601-1-2 standard for electromagnetic compatibility.
#### Device Performance Testing
Verification and Validation Testing for ROSA Hip was conducted with the following aspects:
- Physical/Performance Tests- to ensure the performance of the implemented features and verify related design inputs
- . Engineering Analysis- to ensure the performance of the implemented features and verify related design inputs
- Usability Engineering- addressed user interactions with ROSA Hip System
- . Validation Lab- performed to validate that using ROSA Hip System is safe and effective and that the performances of the system are acceptable under full simulated use on cadaveric specimens
#### Software Verification and Validation Testing
Software tests were conducted to satisfy requirements of the FDA Guidance for the Content Premarket Submissions for Software Contained in Medical Devices and IEC 62304 (Medical Device Software- Life Cycle Process). The software was considered a "major" level of concern, since a failure of the software could result in serious injury or death to the patient. The testing demonstrates that the ROSA Hip System does not raise any new issues of safety and effectiveness as compared to the predicate and reference device(s).
#### Substantial Equivalence Conclusion
The proposed and predicate devices have the same intended use and similar indications for use as the systems are used to assist surgeons in the accurate positioning & placement of hip or knee components. The proposed and primary predicate device utilize fluoroscopic images acquired from a C-arm for patient registration and
{9}------------------------------------------------
intraoperative hip positioning. The proposed and primary predicate also obtain similar landmarks, allows a user to perform overlays, and provides assessment of leg length and offset.
The proposed device does not perform any pre-operative surgical planning and does not assist with any cutting as compared to the ROSA predicate device. The ROSA predicate and proposed device share the same previously cleared ROSA RECON platform hardware and software core components. In addition, the proposed device and ROSA predicate use the robotic arm to assist the guidance of instruments.
In summary, any differences between the devices does not raise new questions of safety and effectiveness and verification and validation activities demonstrate that the proposed device is at least as safe and effective as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.