← Product Code [MQB](/productcode/MQB) · K210985

# 1717FCC (K210985)

_Rayence Co., Ltd. · MQB · Apr 28, 2021 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210985

## Device Facts

- **Applicant:** Rayence Co., Ltd.
- **Product Code:** [MQB](/productcode/MQB.md)
- **Decision Date:** Apr 28, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

1717FCC is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

## Device Story

1717FCC is a digital flat-panel X-ray detector; captures X-ray radiation using a scintillator (CsI:Tl) coupled to an amorphous silicon (a-Si) or Indium Gallium Zinc Oxide (IGZO) TFT sensor. Device connects to a user PC via wired LAN (Ethernet); does not control X-ray generators. Captures and digitizes X-ray images into RAW files; RAW files processed into DICOM images by separate console software (Xmaru RF). Used in clinical settings for general radiography; operated by trained healthcare professionals (technicians/radiologists). Output provides diagnostic images for clinical decision-making; enables transition from analog to digital radiography; improves workflow and image quality.

## Clinical Evidence

No clinical trials performed. Substantial equivalence supported by non-clinical bench testing. Performance metrics (MTF, DQE, NPS) compared against predicate 1717SCC using IEC 62220-1 standards. Qualitative image review confirmed equivalent or superior spatial and soft tissue contrast resolution compared to the predicate. Testing confirms device meets safety and performance requirements for general radiographic imaging.

## Technological Characteristics

Digital flat-panel detector; Scintillator: CsI:Tl; Sensor: a-Si or IGZO TFT + PIN photodiode; Imaging area: 17x17 inches; A/D conversion: 14/16-bit; Connectivity: Wired LAN (Ethernet); Power: 24VDC adapter; Standards: IEC 62220-1 (DQE), IEC 60601-1 (Safety), IEC 60601-1-2 (EMC), IEC 60601-1-6 (Usability).

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- 1717SCC_127μm and 1717SCC_140μm ([K171420](/device/K171420.md))

## Reference Devices

- DRF 4343 (Pixium RF 4343) ([K080859](/device/K080859.md))
- SkyPlate ([K171461](/device/K171461.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

April 28, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Rayence Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 7505 Fannin St., Suite 610 HOUSTON TX 77054

Re: K210985

Trade/Device Name: 1717FCC Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB, JAA Dated: March 26, 2021 Received: April 1, 2021

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K210985

Device Name 1717FCC

#### Indications for Use (Describe)

1717FCC is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

| Type of Use (Select one or both, as applicable)                                      |                                               |
|--------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K210985

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date 510k summary prepared: April 16, 2021

Submitter's Name, address, telephone number, a contact person:

| Submitter's Name :      | Rayence Co., Ltd.                                       |
|-------------------------|---------------------------------------------------------|
| Submitter's Address:    | 14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea |
| Submitter's Telephone:  | +82-31-8015-6459                                        |
| Contact person:         | Mr. Kee Dock Kim / RA Team Manager / +82-31-8015-6459   |
| Official Correspondent: | Dave Kim (davekim@mtech-inc.net)                        |
| Address:                | 7505 Fannin St. Ste 610-V111, Houston, TX 77054         |
| Telephone:              | +713-467-2607                                           |

Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:

| Trade/proprietary name | : 1717FCC                           |
|------------------------|-------------------------------------|
| Common Name            | : Digital Flat Panel X-ray Detector |
| Regulation Number      | : 21 CFR 892.1680                   |
| Regulation Name        | : Stationary X-ray System           |
| Regulatory Class       | : Class II                          |
| Product Code           | : MQB, JAA                          |

### Primary Predicate Device :

| Trade/Device Name | : 1717SCC_127μm and 1717SCC_140μm   |
|-------------------|-------------------------------------|
| Common Name       | : Digital Flat Panel X-ray Detector |
| 510(k) Number     | : K171420                           |
| Regulation Number | : 21 CFR 892.1680                   |
| Regulation Name   | : Stationary X-ray System           |
| Regulatory Class  | : Class II                          |

{4}------------------------------------------------

| Product Code      | : MQB                                         |
|-------------------|-----------------------------------------------|
| Reference Device: |                                               |
| Trade/Device Name | : DRF 4343 (Pixium RF 4343)                   |
| 510(k) Number     | : K080859                                     |
| Regulation Number | : 21 CFR 892.1650                             |
| Regulation Name   | : Image intensified Fluoroscopic X-ray system |
| Regulatory Class  | : Class II                                    |
| Product Code      | : JAA and MQB                                 |

## 2. Device Description

1717FCC is a digital solid state X-ray detector that is based on flat-panel technology. This radiographic image detector and processing unit consists of a scintillator coupled to amorphous silicon (a-Si) / Indium Gallium Zinc Oxide (IGZO) on TFT sensor. This device is connected to the user PC via wired LAN (ethernet cable) and it needs to be integrated with a radiographic imaging system. It does not operate as an X-ray generator controller but can be utilized to capture and digitalize X-ray images for radiographic diagnosis. The RAW files can be further processed as DICOM compatible image files by separate console SW(Xmaru RF) for a radiographic diagnosis and analysis.

## 3. Indication for use

1717FCC is indicated for digital imaging solution designed for general radiographic system for human It is intended to replace film or screen based radiographic systems in all general purpose anatomy. diagnostic procedures. Not to be used for mammography.

## 4. Summary of Design Control Risk management

1717FCC digital X-ray detector is a modification of 1717SCC (K171420). 1717FCC was developed for the purpose of retrofitting the stationary X-ray system with a film detector.

The risks and the hazardous impact of the device modification were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design change were successfully mitigated and accepted.

{5}------------------------------------------------

# 5. Summary of the technological characteristics of the device compared to the predicate device:

1717FCC detector described in this 510(k) has the same indications for use and similar technical characteristics as its predicate devices, 1717SCC (K171420).

## 5.1 Scintillator layer

*scintillator layer. ( * scintillator : a phosphor that produces scintillations)

| Scintillator Type   | Proposed | Predicate                          |
|---------------------|----------|------------------------------------|
| CsI (Cesium Iodide) | 1717FCC  | 1717SCC 127um and<br>1717SCC 140um |

### 5.2 Power source

|       |            | Proposed<br>1717FCC                                            | Predicate<br>1717SCC                                    |
|-------|------------|----------------------------------------------------------------|---------------------------------------------------------|
| Power | Type       | Power adapter                                                  | Power supply                                            |
|       | Model name | AHM85PS24                                                      | RP003A                                                  |
|       | Dimension  | 150 X 64 X 37<br>(cable length: 900 mm)                        | 188 X 92 X 41.5                                         |
|       | Weight     | 0.4                                                            | 0.5                                                     |
|       | Rating     | Input: 100-240 V, 1.0 A, 50/60 Hz<br>Output: 24VDC (Max 3.54A) | Input: 100-240VAC (50/60Hz)<br>Output: 24VDC (Max 1.7A) |

### 5.3 Generator specifications

| Model          | Manufacture                          | Specification |                    |        |        |
|----------------|--------------------------------------|---------------|--------------------|--------|--------|
|                |                                      |               | 32kW               | 40kW   | 50kW   |
| CMP 200        | Communications & Power<br>Industries | kVp           | 40-125             | 40-150 | 40-150 |
|                |                                      | mA            | 10-400             | 10-500 | 10-630 |
| EDITOR HFe 501 | Rontgenwerk Bochum                   | kVp           | 40-150             |        |        |
|                |                                      | mA            | 10-630             |        |        |
| UD150L-40E/40F | Shimadzu                             | kVp           | 40-150             |        |        |
|                |                                      | mA            | @100 kVp- 500(320) |        |        |
|                |                                      |               | @80 kVp- 630(400)  |        |        |
| PXR-321B       | Poskom Co.,Ltd.                      | kVp           | 125/150            |        |        |
|                |                                      | mA            | 500                |        |        |

### 5.4 Comparison with the predicate device

| Characteristic | Proposed Device (K210985) | Predicate Device (K171420) |
|----------------|---------------------------|----------------------------|
|----------------|---------------------------|----------------------------|

{6}------------------------------------------------

| Manufacturer           |          |                                                                                                                                                                                                                                                                                      | Rayence Co.,Ltd.                                                                                                                                           |                | Rayence Co.,Ltd.                                                                                                                                                                                                                                                                                                     |                                                                                                  |         |         |
|------------------------|----------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|---------|---------|
| Product Name           |          |                                                                                                                                                                                                                                                                                      | 1717FCC                                                                                                                                                    |                | 1717SCC                                                                                                                                                                                                                                                                                                              |                                                                                                  |         |         |
| Feature                |          | Image: 1717FCC                                                                                                                                                                                                                                                                       |                                                                                                                                                            | Image: 1717SCC |                                                                                                                                                                                                                                                                                                                      | Similar                                                                                          |         |         |
| Indications<br>for use |          | 1717FCC is indicated for digital<br>imaging solution designed for<br>general radiographic system for<br>human anatomy. It is intended to<br>replace film or screen based<br>radiographic systems in all general<br>purpose diagnostic procedures. Not<br>to be used for mammography. |                                                                                                                                                            |                | 1717SCC_127um and<br>1717SCC_140um, are indicated<br>for digital imaging solution<br>designed for general radiographic<br>system for human anatomy. It is<br>intended to replace film or screen<br>based radiographic systems in all<br>general purpose diagnostic<br>procedures. Not to be used for<br>mammography. |                                                                                                  | Same    |         |
| Detector Type          |          |                                                                                                                                                                                                                                                                                      | Amorphous Silicon (a-Si) TFT + PIN<br>type photodiode<br>IGZO TFT + PIN type photodiode<br>(option)                                                        |                |                                                                                                                                                                                                                                                                                                                      | Amorphous Silicon (a-Si) TFT                                                                     | Similar |         |
| Scintillator           |          |                                                                                                                                                                                                                                                                                      | CsI:Tl                                                                                                                                                     | 1717FCC        | CsI:Tl                                                                                                                                                                                                                                                                                                               | 1717SCC                                                                                          | Similar |         |
| Imaging Area           |          |                                                                                                                                                                                                                                                                                      | 17 x 17 inches                                                                                                                                             |                |                                                                                                                                                                                                                                                                                                                      | 17 x 17 inches                                                                                   |         | Similar |
| Pixel matrix           |          |                                                                                                                                                                                                                                                                                      | 140 type: 3000 x 3000 (Full resolution)<br>280 type: 1500 x 1500 (2x2 binning)<br>420 type: 1000 x 1000 (3x3 binning)<br>560 type: 750 x 750 (4x4 binning) |                |                                                                                                                                                                                                                                                                                                                      | 140 type : 3072 x 3072                                                                           |         | Similar |
| Pixel pitch            |          |                                                                                                                                                                                                                                                                                      | 140 μm / 280 μm/ 420 μm/ 560 μm                                                                                                                            |                |                                                                                                                                                                                                                                                                                                                      | 140 μm                                                                                           |         | Similar |
| A/D conversion         |          |                                                                                                                                                                                                                                                                                      | 14 / 16 bit                                                                                                                                                |                |                                                                                                                                                                                                                                                                                                                      | 14 bit / 16 bit                                                                                  | Same    |         |
| MTF                    | a-Si TFT |                                                                                                                                                                                                                                                                                      | 1.0 lp/mm, Typ. 0.535<br>2.0 lp/mm, Typ. 0.220<br>3.0 lp/mm, Typ. 0.099<br>3.5 lp/mm, Typ. 0.073                                                           |                |                                                                                                                                                                                                                                                                                                                      | 1.0 lp/mm, Typ. 0.580<br>2.0 lp/mm, Typ. 0.283<br>3.0 lp/mm, Typ. 0.158<br>3.5 lp/mm, Typ. 0.120 |         | Similar |
|                        | IGZO TFT |                                                                                                                                                                                                                                                                                      | 1.0 lp/mm, Typ. 0.525<br>2.0 lp/mm, Typ. 0.209<br>3.0 lp/mm, Typ. 0.092<br>3.5 lp/mm, Typ. 0.064                                                           |                |                                                                                                                                                                                                                                                                                                                      | -                                                                                                |         |         |
|                        | DQE (0)  | a-Si TFT                                                                                                                                                                                                                                                                             |                                                                                                                                                            | Typ. 0.751     |                                                                                                                                                                                                                                                                                                                      | 0 lp/mm, Typ. 0.740                                                                              |         | Similar |
|                        |          | IGZO TFT                                                                                                                                                                                                                                                                             |                                                                                                                                                            | Typ. 0.766     |                                                                                                                                                                                                                                                                                                                      | -                                                                                                |         |         |
| Preview time           |          |                                                                                                                                                                                                                                                                                      | <2 seconds                                                                                                                                                 |                |                                                                                                                                                                                                                                                                                                                      | <2 seconds                                                                                       |         | Same    |
| Data output            |          |                                                                                                                                                                                                                                                                                      | RAW                                                                                                                                                        |                | RAW                                                                                                                                                                                                                                                                                                                  |                                                                                                  | Same    |         |

{7}------------------------------------------------

|            | *The RAW files are convertible into<br>DICOM 3.0 by console S/W | *The RAW files are convertible into<br>DICOM 3.0 by console S/W |         |
|------------|-----------------------------------------------------------------|-----------------------------------------------------------------|---------|
| Dimensions | 470 x 470 x 50 mm                                               | 460 × 460 × 15.5 mm                                             | Similar |
| Weight     | 6.0 kg                                                          | 4 kg (incl. battery)                                            | Similar |

## 5.5 Comparison with Reference Device

| Manufacturer               | Rayence Co.,Ltd.                                                                                                                                                                                                                                                                     |                                                     | Villa Sistemi Medicali                                                                                                                                                                                                                                                                                                                                                                                    |         |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Product Name               | 1717FCC                                                                                                                                                                                                                                                                              |                                                     | DRF 4343 (Pixium RF 4343)                                                                                                                                                                                                                                                                                                                                                                                 |         |
| Feature                    | Image: 1717FCC                                                                                                                                                                                                                                                                       |                                                     | Image: DRF 4343                                                                                                                                                                                                                                                                                                                                                                                           | Similar |
| 510(k) number              | K210985                                                                                                                                                                                                                                                                              |                                                     | K080859                                                                                                                                                                                                                                                                                                                                                                                                   |         |
| Indications<br>for<br>use  | 1717FCC is indicated for digital<br>imaging solution designed for<br>general radiographic system for<br>human anatomy. It is intended to<br>replace film or screen based<br>radiographic systems in all general<br>purpose diagnostic procedures. Not<br>to be used for mammography. |                                                     | The DRF 4343 is intended to<br>capture digital images from a<br>radiographic / fluoroscopic system<br>through a dynamic digital flat<br>panel, to digitalize, archive and<br>review images and to provide a<br>network connection via DICOM<br>protocol to various output (e.t.<br>hardcopy, softcopy and archive)<br>devices which uses a device.<br>This device is not intended for<br>mammography use. | Similar |
| Detector Type              | Amorphous Silicon (a-Si) TFT + PIN<br>type photodiode<br>IGZO TFT + PIN type photodiode                                                                                                                                                                                              |                                                     | Amorphous Silicon (a-Si) TFT                                                                                                                                                                                                                                                                                                                                                                              | Similar |
| Scintillator               | CsI:Tl                                                                                                                                                                                                                                                                               |                                                     | CsI:Tl                                                                                                                                                                                                                                                                                                                                                                                                    | Same    |
| Imaging Area               | 17 x 17 inches                                                                                                                                                                                                                                                                       |                                                     | 17 x 17 inches                                                                                                                                                                                                                                                                                                                                                                                            | Similar |
| Frame rate                 | GigE                                                                                                                                                                                                                                                                                 | 6 @ (1x1)<br>25 @ (2x2)<br>45 @ (3x3)<br>60 @ (4x4) | 8 f/s (RAD), 15 f/s for fluoroscopy                                                                                                                                                                                                                                                                                                                                                                       | Better  |
|                            | Camera link                                                                                                                                                                                                                                                                          | 9 @ (1x1)<br>30 @ (2x2)<br>45 @ (3x3)<br>60 @ (4x4) | 12 f/s (RAD), 18 f/s for fluoroscopy<br>(large field)                                                                                                                                                                                                                                                                                                                                                     |         |
|                            | 5GigE                                                                                                                                                                                                                                                                                | 15 @ (1x1)                                          |                                                                                                                                                                                                                                                                                                                                                                                                           |         |
|                            |                                                                                                                                                                                                                                                                                      | 30 @ (2x2)<br>45 @ (3x3)<br>60 @ (4x4)              |                                                                                                                                                                                                                                                                                                                                                                                                           |         |
| <i>Image</i><br>resolution | Upto 3.5 lp/mm                                                                                                                                                                                                                                                                       |                                                     | Upto 3.4 lp/mm                                                                                                                                                                                                                                                                                                                                                                                            | Better  |

{8}------------------------------------------------

## 6. Summary of Performance Testing

1717FCC Digital Flat Panel X-Ray Detector has the same indications for use, the same scintillator material (CsI:Tl ), the same generator specifications and the same risk analysis characteristics compared to 1717SCC, the predicate devices (K171420). The pixel matrix and pixel pitch sizes are different imaging areas but the differences do not raise new concerns for the safety and effectiveness of the subject device.

The clinical and non-clinical test report for the subject device were prepared and submitted to FDA to demonstrate the substantial equivalency of the subject device performance compared to the predicate device.

After comparing a broad review of plain radiographic images taken with 1717FCC and 1717SCC_140μm, images obtained with 1717FCC and 1717SCC were equivalent quality for the same view obtained from a similar patient with the 1717SCC 140μm. In general, both the spatial and soft tissue contrast resolution are slightly superior using the 1717FCC. Specifically, the soft tissues on extremity films were seen with better clarity. There is little difficulty in evaluating a wide range of anatomic structures necessary to provide a correct conclusion.

The non-clinical test report contains the MTF, DQE and NPS performance test comparison between the subject device (1717FCC), and the predicate device (1717SCC), by using the identical test equipment and same analysis method described by IEC 62220-1.

The MTF and DQE testing represent the ability to visualize object details of a certain size and contrast. 1717FCC demonstrated equivalent or better performance in terms of MTF and DQE as well as NPS compared to 1717SCC, the predicate device, at all spatial frequencies.

The frame rate and image resolution for 1717 FCC, the subject device, perform better than the specification of the reference device, DRF 4343 (K080859) which are referenced in K131766.

Based on the non-clinical consideration evaluation, the sponsor can claim the substantial equivaleny between the subject device and the predicate device in terms of diagnostic image quality.

{9}------------------------------------------------

The manufacturing facility is in conformance with the design control procedure requirements specified in 21 CFR 820.30 and the relevant 21CFR820 standards as the records are available for review. Furthermore, the device conforms with all labeling requirements as per 21 CFR Subchapter J, specifically, 21 CFR 10120.30 and 1020.32.

## 7. Summary for any testing and reference guidance:

- A IEC 6220-1-1 Edition 1.0 2015-03 Medical electrical equipment-Characteristics of digital X-ray imaging devices Part 1-1: Determination of the detective quantum efficiencyDetectors used in radiographic imaging
- A EMC testing were conducted in accordance with standard IEC 60601-1-2: 2014
- A IEC 60601-1-6:2010 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- ♪ Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1: 2005, COR1:2006, COR2:2007, AMD1:2012 (Medical electrical equipment Part 1:General requirements for basic safety and essential performance) and IEC 60601-1-6:2010 (Third edition) +A1:2013
- A Non-clinical consideration according to FDA Guidance "Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices"
- > "Guidance for the Contents of Premarket Submission for Software Contained in Medical Device".
- > Pediatric Information for X-ray Imaging Device Premarket Notifications
- > Content of Premarket Submissions for Management of Cybersecurity in Medical Devices

### 8. Conclusions:

In accordance with the performance outcomes. 1717FCC demonstrated equivalent or better performance compared to 1717SCC(K171420) and SkyPlate (K171461). Therefore, Rayence claims the substantial equivalency between the subject device and the predicate device in terms of diagnostic image quality about safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K210985](https://fda.innolitics.com/device/K210985)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
