Navident

K210947 · Claronav, Inc. · PLV · Jun 27, 2022 · Dental

Device Facts

Record IDK210947
Device NameNavident
ApplicantClaronav, Inc.
Product CodePLV · Dental
Decision DateJun 27, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4120
Device ClassClass 2

Intended Use

Navident is a computerized dental navigational system intended to assist preoperative planning and to guide drilling in a patient jaw during implantation surgery, using pre-acquired CT scan of the jaw. The device is intended for use by a qualified dental surgeon in the treatment of partially or fully edentulous jaws.

Device Story

Navident is an image-guided dental navigation system for implantation surgery. It uses a mobile cart with a stereoscopic video camera and a laptop running proprietary software. Input data includes pre-acquired CT scans and intraoral surface scans. The system tracks the patient's jaw and the dental handpiece in real-time using optical markers (tags) attached to the jaw and handpiece. The software compiles a digital model, allows for treatment planning, registers the model to the patient's anatomy, and provides dynamic visual guidance. During surgery, the surgeon views a 3D display showing the drill tip's position, angle, and depth relative to the planned implant site. This real-time feedback assists the surgeon in aligning the drill with the planned path, potentially improving accuracy and reducing surgical time. The system is used in dental clinics by dental surgeons.

Clinical Evidence

Clinical evidence from three published studies supports the 'Trace' registration method. One study (n=59 patients, 136 implants) showed mean deviations of 0.67mm (entry), 0.9mm (apex), 0.55mm (depth), and 2.50° (angular). A second study (n=39 patients, 63 pterygoid implants) compared dynamic navigation with trace registration to freehand surgery, demonstrating superior accuracy (0.66mm vs 1.54mm coronal; 1.13mm vs 2.73mm apical) and reduced surgical time. Bench testing confirmed system accuracy, sterilization, biocompatibility, and electrical safety.

Technological Characteristics

Image-guided navigation system; stereoscopic triangulation of optical checker-board patterns. Components: mobile cart, stereoscopic camera, laptop, jaw/handpiece trackers, calibrator. Materials: stainless steel, thermoplastic, dental composite resin, PVS. Connectivity: wireless. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 17665-1 (sterilization), ISO 10993-1 (biocompatibility). Software: proprietary, moderate level of concern.

Indications for Use

Indicated for use by qualified dental surgeons to assist in preoperative planning and guide drilling during dental implantation surgery in partially or fully edentulous adult patients, using pre-acquired CT scans.

Regulatory Classification

Identification

A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. ClaroNav Inc. Carly Desmond Director of Regulatory Affairs 1140 Sheppard Avenue West Unit 10 Toronto, Ontario M3K 2A2 CANADA Re: K210947 Trade/Device Name: Navident Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument And Accessories Regulatory Class: Class II Product Code: PLV Dated: June 2, 2022 Received: June 2, 2022 Dear Carly Desmond: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210947 Device Name Navident ### Indications for Use (Describe) Navident is a computerized dental navigational system intended to assist preoperative planning and to guide drilling in a patient jaw during implantation surgery, using pre-acquired CT scan of the jaw. The device is intended for use by a qualified dental surgeon in the treatment of partially or fully edentulous jaws. Type of Use (Select one or both, as applicable) | <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |--------------------------------------------------------------------------------------| | <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) SUMMARY K210947 ClaroNav's Navident | Sponsor: | ClaroNav Inc.<br>1140 Sheppard Avenue West, Unit 10<br>Toronto, Ontario, M3K 2A2<br>Canada | |------------------------------------|--------------------------------------------------------------------------------------------| | Contact Person | Doron Dekel | | Phone<br>e-mail | +1 647 868 6813<br>doron@claronav.com | | Submission Date | June 24, 2022 | | Device Proprietary Name | Navident | | Common Name | Surgical Navigation System | | Classification | PLV | | Class | II | | Regulation<br>Classification Panel | 21 CFR 872.4120; Bone Cutting Instrument and Accessories<br>Dental | ### Predicate Devices Primary predicate: ClaroNav's Navident (K161406) Reference device: X-Nav's X-Guide (K200662) ## Indications for Use Navident is a computerized dental navigational system intended to assist preoperative planning and to guide drilling in a patient jaw during implantation surgery, using pre-acquired CT scan of the jaw. The device is intended for use by a qualified dental surgeon in the treatment of partially or fully edentulous jaws. ### Technological Characteristics Navident is an image-guided dental navigational system intended to assist with preoperative planning and real-time positioning of drilling tools during implantation surgery. In particular, Navident provides visual, real-time feedback on the location of the working tip of a dental handpiece. It shows {4}------------------------------------------------ the location and direction of the tip relative to a volumetric CT image of the patient's jaw registered to that anatomy, and, when available, relative to a path planned on that image. Navident is comprised of the following parts: - The main system is comprised of a cart that carries a stereoscopic video camera and a laptop . with pre-installed proprietary software. - . The Navident system also includes several types of accessories: - Jaw motion Tracking Accessories: Accessories that are rigidly attached to the O patient's jaw and are used to maintain the registration between the jaw and its CT image throughout the procedure via real-time tracking of the jaw. They comprise a tag with optical markings trackable by the stereoscopic video camera and a mechanism for rigid attachment of the tag to the patient's Jaw (for tracking the upper jaw only). - Dental Handpiece Tracking Accessories: Accessories that are rigidly attached to a O dental handpiece to be navigated during the dental procedure. They comprise a tag with optical markings trackable by the stereoscopic video camera and a mechanism for rigid attachment of the tag to the handpiece. - Registration accessories: Accessories that are used in the registration of the patients' O anatomy with the patients' CT scan. - Calibrator: A tool used for calibrating the working tip of the handpiece to determine O the geometric relationship between the optical marking on the handpiece-attached tag and the working tip of the surgical instrument. - Navident's four core functions are: ● - Model: Using data imported from CT and intraoral surface scanners, a digital model of O the jaw anatomy is compiled and presented to the user. - Plan: The user, using functions provided by the Navident application software, O prepares a digital treatment plan for drilling in bone. - Register: At the start of the actual treatment, the digital model is aligned with the real O anatomy it models. The plan, which was created in geometrical reference to the model can now be referenced, or mapped, to the real anatomy. - Guide: The tooltip (the working end) of the treatment tool, usually a handpiece with O an exchangeable tip, is dynamically shown to the user relative to the (registered) model and the treatment plan. {5}------------------------------------------------ The user uses the on-screen quidance to align the tooltip with the planned tooltip path to ensure correct and accurate execution of the plan. During surgery, Navident automatically tracks the handpiece's motions. When the handpiece approaches a pre-planned implant location, Navident provides a cross-hairs dynamic visualization of the drill pose relative to the planned pose of the implant. This visualization quides the hand motions of the surgeon towards precisely locating the planned entry point, adjusting the drill axis to the planned angle, and drilling to the planned depth. When the tip is away from a marked implantation path, Navident provides real-time visual feedback showing the CT image intensities in the region surrounding the drill's tip. The Navident is intended for the drilling of osteotomies for dental implant placement with high-speed and low-speed dental handpieces This submission focuses on modification of the Navident predicate device (K161406) to include additional components including the wireless connection, use of bone screws for edentulous patients, additional patient tracker designs, and a tracing method for registration. ## Performance Data Navident's performance testing included: - Reprocessing validation: . - . Manual and automated cleaning validation: manual and automated cleaning validation was performed on representative devices to confirm that the cleaning processes of Navident's accessories can reduce the organic soil load, per AAMI TIR 30: 2011(R) 2016, to: - o Proteins : < 6.4 µg/cm2 - O Hemoglobin :< 2.2 µg/cm2 The validation demonstrated that the cleaning process included in Navident's labeling can reduce the soil load to an acceptable level. - . Sterilization Validation: Representative master devices of the re-usable Navident accessories were tested to validate that the components can withstand the steam sterilization process and that acceptable sterility is achieved using the recommended sterilization protocols. The sterilization validation testing was conducted according to ISO 17665-1:2006 (R) 2013 and it validated that the re-usable Navident components can be sterilized to reach an acceptable sterility assurance level. #### . Biocompatibility Testing: The patient contacting Navident components are considered tissue contacting for a duration of less than 24 hours. The components were either tested according to the ISO 10993-1 series or a justification for not performing tests was provided per the FDA Guidance Document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." #### . Software: Software documentation for moderate level of concern per the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." {6}------------------------------------------------ #### . Electrical Safety and Electromagnetic Compatibility (EMC): Comprehensive performance testing has been conducted on the Navident device in accordance with the latest recoqnized industry standards, by an accredited NRTL. Product Safety was evaluated for compliance with IEC 60601- 1:2005 Edition 3.1 ANSI/AAMI ES60601-1:2005 / 2012 and C1:2009/ 2012 and A2:2010/ 2012 (Consolidated Text) Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance. Electromagnetic Compatibility was assessed for compliance with IEC 60601-1-2:2014 Edition 4.0. Evaluation assessment of the wireless technology was conducted per FDA Guidance "Radio Frequency Wireless Technology in Medical Devices - Guidance for Industry and Food and Drug Administration Staff." #### . Full system accuracy bench testing: The navigation information provided to the user is based on the mapping of the drill tip pose to the CT image data. The accuracy of this mapping is determined by two key factors: (1) the accumulation of errors through the sequence of real time coordinate system mappings and (2) the stability of the coupling between the patient's reference tag and the patient's anatomy. Accordinqly, to assess the overall system accuracy, the following bench tests were performed: - System accuracy test: to measure the accuracy of the mapping between drill tip poses . and image space. - . Assess the stability and repeatability of the patient reference tag coupling to the patient's jaw: - Stability of the Attachment of the Jaw Tracker S to the patient's Jaw ■ - Stability of the Attachment of the Jaw Tracker C to the patient's Jaw - I Stability of the Attachment of the Jaw Tracker B to the patient's Jaw - Stability of the Attachment of the Jaw Tracker U to the patient's Jaw I In all instances, the Navident device functioned as intended and the results observed was as expected. #### Human factors/usability: . Summative evaluation was conducted on the Navident system within a realistically simulated clinical setting. Fifteen (15) representative users (i.e., oral surgeons, general dental practitioners) were observed performing representative and high-risk tasks with the Navident system while test administrators monitored the proceedings for any use errors, close calls or operational difficulties that may be indicative of use-safety or usability problems. The assessment of the test results demonstrated that the Navident system satisfies the identified user specifications for use by qualified dentists for its intended use. Minor findings of the evaluation were thoroughly analyzed and design improvements to Navident were made to address them prior to the current submission. {7}------------------------------------------------ #### Clinical Literature: . Clinical evidence supporting the performance of the device was collected from published scientific literature. The described studies were conducted outside of the US; however, the patient populations, user profiles, use environment, and clinical practices are considered equivalent and applicable to the US population. Three studies were identified which support the substantial equivalence of the Trace registration method. The first study | reports on the accuracy results of 136 implants placed under dynamic guidance using the trace registration method in 59 partially edentulous patients. Thirty-nine (39) cases involved the maxilla (75 implants) and 36 cases involved the mandible (61 implants). The mean deviation between the planned and actual position for all implants was 0.67 mm at the entry point, 0.9 mm the apex, 0.55 mm in depth and 2.50° as angular deviation. Tracing 5-6 teeth for registration significantly improved all accuracy outcomes compared to tracing only 3-4 teeth. The 95th percentile values were also computed for the present trace registration method and compared with corresponding measurements taken in an earlier study2 using a thermoplastic stent and fiducial marker-based registration (the registration method that was cleared for the predicate Navident device K161406). This data was previously collected and published by the same authors and the same surgeon using the stent approach with the same navigation system. The results demonstrated that using trace registration was favorable compared to fiducial-based registration especially when tracing 5-6 teeth. In another study Stefanelli et al. compared the accuracy of Pterygoid implant placement using dynamic navigation with trace registration versus free hand surgery. In this study 63 pterygoid implants were placed in 39 partially edentulous patients. Thirty-one (31) pterygoid implants were placed using dynamic navigation and trace registration and 32 pterygoid implants were placed using freehand surgery. The mean deviations between the planned and actual position for pterygoid implants placed using dynamic navigation via trace registration was 0.66 mm at coronal level, 1.13 mm at apical level, 0.67 mm in depth, and 2.64° as angular deviation compared to 1.54mm, 2.73mm, 1.17mm and 12.49°, respectively, when freehand surgery alone was used. Pteryqoid implants placed using dynamic navigation technology was more accurate when compared to the prosthetically directed presurgical plan in relation to the greater palatine canal and took less surgical time. In conclusion, one of the main changes introduced by the Navident device in comparison to the predicate Navident device is the addition of another registration method (Trace) on top of the previously available stent and fiducial reqistration. The accuracy of this registration method was evaluated on 136 implants and was found to be as accurate as fiducial base registration. In addition, the trace registration method was used with 31 Pterygoid implants while comparing to free hand surgery. The results were in favor of the use of dynamic navigation with trace registration. References: - 1) Stefanelli LV, Mandelaris GA, DeGroot BS, Gambarini G, De Angelis F, Di Carlo S. Accuracy of a Novel Trace-Registration Method for Dynamic Navigation Surgery. Int J Periodontics Restorative Dent. 2020 May/Jun;40(3):427-435. - 2) Stefanelli LV, DeGroot BS, Lipton Dl, Mandelaris GA. Accuracy of a Dynamic Dental Implant Navigation System in a Private Practice. Int J Oral Maxillofac Implants. 2019 January/February;34(1):205–213. {8}------------------------------------------------ - 3) Stefanelli LV, Graziani U, Pranno N, Di Carlo S, Mandelaris GA. Accuracy of Dynamic Navigation Surgery in the Placement of Pterygoid Implants. Int J Periodontics Restorative Dent. 2020 Nov/Dec;40(6):825-834. ## Substantial Equivalence The current Navident is substantially equivalent to the cleared Navident (K161406). Navident has similar intended use, technological characteristics, and principles of operation as its predicate device. The technological differences between the Navident and its predicate devices raise no new issues of substantial equivalence and its substantial equivalence is further supported by a comparison to a reference device that offers the same intended use and very similar technological characteristics. Performance data demonstrate that Navident is substantially equivalent to the cleared Navident. A comparison of the proposed Navident device to the currently marketed predicate Navident and the reference X-Nav's X-Guide are provided in Table 5-1 below: {9}------------------------------------------------ ## Table 5-1: Comparison of Indications for use and Technological Characteristics | Feature/<br>Characteristic | Proposed Navident | ClaroNav's Navident K161406<br>(Predicate device) | X-Nav's X-guide device<br>K200662<br>(Reference Device) | Justification of difference | |----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | | | | | | Class/Product<br>Code/Classification<br>Name | Class II/ PLV/21 CFR 872.4120<br>(Bone Cutting Instrument and<br>Accessories) | Class II/ PLV/ 21 CFR 872.4120<br>(Bone Cutting Instrument and<br>Accessories) | Class II/ PLV/ 21 CFR 872.4120<br>(Bone Cutting Instrument and<br>Accessories) | Identical | | Indications for Use | Navident is a computerized<br>dental navigational system<br>intended to assist preoperative<br>planning and to guide drilling in<br>a patient jaw during<br>implantation surgery, using pre-<br>acquired CT scan of the jaw. The<br>device is intended for use by a<br>qualified dental surgeon in the<br>treatment of partially or fully<br>edentulous jaws. | Navident is a computerized<br>dental navigational system<br>intended to assist preoperative<br>planning and to guide drilling in<br>a patient jaw during<br>implantation surgery, using pre-<br>acquired CT scan of the jaw. The<br>device is intended for use by a<br>qualified dental surgeon in the<br>treatment of partial edentulism. | The X-Guide® Surgical<br>Navigation System is a<br>computerized navigational<br>system intended to provide<br>assistance in both the<br>preoperative planning phase<br>and intra-operative surgical<br>phase of dental implantation<br>procedures. The system provides<br>software to preoperatively plan<br>dental implantation procedures<br>and provides navigational<br>guidance of the surgical<br>instruments.<br>The device is intended for use<br>for partially edentulous and<br>edentulous adult and geriatric<br>patients who need dental<br>implants as part of their<br>treatment plan. | Navident has the same intended use as the<br>predicate device.<br>Navident expands the cleared indication for<br>use of the predicate device from partially<br>edentulous patients to partially edentulous<br>and edentulous patients. Hardware<br>components (Jaw Tracker B) and alternative<br>registration method (Bone screws) are offered<br>in the Navident to support edentulous<br>patients.<br>The reference device, which is also a dental<br>navigation system that uses similar<br>technology, offers similar technological<br>solutions for edentulous patients and has<br>identical intended use and indications for use<br>as the Navident. | | Main functions | CT-based placement planning of<br>dental procedures. Presentation<br>of position, angle and depth<br>indicators when drilling in or<br>cutting the jaw. | CT-based implant placement<br>planning. Presentation of<br>position, angle and depth<br>indicators when drilling in the<br>jaw. | CT-based implant placement<br>planning. Presentation of<br>position, angle and depth<br>indicators when drilling in the<br>jaw. | Identical | | Use Environment | Dental clinic | Dental clinic | Dental clinic | Identical | | Target Population | Adult patients | Adult patients | Adult patients | Identical | | Users | Dental surgeons | Dental surgeons | Dental surgeons | Identical | | Technological Characteristics | | | | | | Input imaging modality | CT | CT | CT | Identical | | Dynamic object poses<br>measurement<br>technology | Stereoscopic triangulation of<br>checker-board contrast patterns. | Stereoscopic triangulation of<br>checker-board contrast patterns. | Stereoscopic triangulation of<br>checker-board contrast patterns. | Identical | | Handpiece tracking<br>attachment | a visually marked tag. | a visually marked tag. | a visually marked tag. | Identical | | Jaw tracking<br>attachment<br>(Dentate Patients) | <b>"NaviStent"</b> : molded thermoplastic sheet with an integrated arm holding an optically marked plastic tag. OR: <b>"Jaw Tracker"</b> : optically marked flat tag attached to a hand-configurable stainless steel wire attached to the jaw at its other end. The <b>JawTracker C</b> and <b>S</b> version attaches to a tooth, crown or abutment using approved dental composite resin and bonding agent. The <b>JawTracker U</b> version attaches to several adjacent teeth using a U-shaped clamp filled with bite registration material. | <b>"NaviStent"</b> molded thermoplastic sheet with an integrated arm holding an optically marked plastic tag. No support for edentulous jaws. | X-Clip®, is attached to patient teeth using an impression material (low temperature thermoplastic). The Patient Tracker which includes the tracking patterns that is visualized by the tracking camara is firmly attached to the X-Clip. This device remains on the patient teeth for the duration of surgery. | Navident offers an identical solution for Jaw tracking (the NaviStent) as the cleared Navident device. In addition, two alternative jaw attachment solutions are offered (JawTracker C and JawTracker U). While the materials used for the attachment (thermoplastic, PVS, composite) and their hardening methods (cooling, chemical or light curing) vary, they are all moldable dental materials in common usage, with well-known properties. | | Jaw Tracking<br>Attachment<br>(Edentulous Patient) | JawTracker B: optically marked<br>flat tag attached to a hand-<br>configurable stainless steel wire<br>attached to the jaw bone using 3<br>small bone screws. | None | Edentulous Clip® - fixated into<br>position on the patient's jaw<br>using bone screws. The patient<br>tracker, which includes the<br>tracking patterns that is<br>visualized by the tracking<br>camera is firmly attached to the<br>Edentulous Clip. This device<br>remains on the patient teeth for<br>the duration of surgery. | Navident's Jaw Tracker B solution is very<br>similar to the Edentulous clip of the reference<br>X-Guide device, and is based on an optically<br>marked tag that is attached to the patients Jaw<br>using a small plate that is fixated into position<br>on the patient's jaw using bone screws. | | Patient to CT image<br>registration | NaviStent: includes a<br>removable CT Marker part<br>containing a CT-visible<br>fiducial body. Trace Registration:<br>registration based on<br>Anatomical Landmarks on<br>the jaw's surface.. Bone screws: placed in the<br>patient's Jaw prior to CT<br>scan to create fiducials. NaviBite: plastic appliance<br>containing several screws<br>with their heads exposed.<br>Used with impression<br>material to form a bite<br>registration plate that can<br>be repeatably coupled to<br>both jaws. | NaviStent: includes a removable<br>CT Marker part containing a CT-<br>visible fiducial body. | X-Clip: contains embedded<br>radiodense spheres used as<br>fiducial. Bone hole Fiducials: For<br>edentulous patients, the<br>surgeon drills several narrow<br>holes in the bone to serve as<br>fiducials for the CT scan. X-Mark: registration based<br>on Anatomical Landmarks. Bone screws: placed in the<br>patient's Jaw prior to CT<br>scan to create artificial<br>landmarks. | Navident's registration solutions are identical<br>or very similar to the ones offered in the<br>predicate and the reference devices: The Navistent is a fiducial based<br>registration, identical to Navident's cleared<br>registration solution (K161406). The "Trace Registration" is very similar to<br>X-Guide X-mark registration solution. Both<br>are based on existing anatomical<br>landmarks as fiducials. These landmarks<br>are manually marked in the Software and<br>are traced by touching them with the tip<br>of an optically trackable tracer tool. Bone Screw: same as offered by X-Guide<br>reference device, the bone screws are<br>placed in the patient's Jaw prior to CT scan<br>to create artificial landmarks. NaviBite: The screw heads, anchored to<br>the NaviBite plastic shell, function very<br>similarly to the spheres embedded in the<br>X-Clip solution of X-Guide. In the<br>registration process, both are touched by<br>a special pointer to provide their location | | | | | | | | Drill tip calibration | Initial drill axis calibration on a tracked pin, plus tip calibration by placing the tip in a dimple in a special optically marked tool (Calibrator) after each drill change. Support for both low speed and high-speed handpieces | Initial drill axis calibration on a tracked pin, plus tip calibration by placing the tip in a dimple in the Jaw Tag after each drill change. | Initial handpiece calibration using a special optically marked tool, plus tip calibration by touching a point after each drill change. | relative to the optical markings on the tag coupled to the jaw.<br>The calibration steps are similar. In the cleared Navident device, the jaw tag was used to also perform these calibration steps, eliminating the need for a separate tool. The Navident, offers a dedicated calibration tool, same as the reference device (X-Guide). | | Mount for camera and display | Mobile cart weighing 55lb with a folding arm holding camera and screen above patient's chest. | Mobile cart weighing 55lb with a folding arm holding camera and screen above patient's chest. | Mobile cart with a large arm holding camera above patient's head. Screen is attached to cart pole. | Identical to the cleared Navident device | | Presentation of navigation guidance | 3D graphics presentation of drill position, angle and depth relative to planned placement. | 3D graphics presentation of drill position, angle and depth relative to planned placement. | 3D graphics presentation of drill position, angle and depth relative to planned placement. | Identical to the cleared Navident device | | Illumination of tracking targets | Visible light emitted by LED panel (optional) | Visible light emitted by LED panel (optional) | Visible light emitted by LED panel (optional) | Identical to the cleared Navident device | | <b>Performance Characteristics</b> | | | | | | Accuracy at the drill tip | ≤1.0mm | ≤1.0mm | | Identical to the cleared Navident device | | Presentation update rate | Real time | Real time |…
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%