e-Celsius Medical System

K210924 · Bodycap · FLL · Jun 28, 2023 · General Hospital

Device Facts

Record IDK210924
Device Namee-Celsius Medical System
ApplicantBodycap
Product CodeFLL · General Hospital
Decision DateJun 28, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2

Indications for Use

e-Celsius Medical System is indicated for the measurement of core body temperature in patients for whom accurate continuous core body temperatures are clinically necessary for periods exceeding six (6) hours and with whom rectal catheters are not appropriate. e-Celsius Medical System can only be used on the order of a physician who has clinically evaluated the contraindications and warnings associated with the use of and ingestible thermometer and determined that the risk is clinically justified versus an oral thermometer or- rectal catheter. e-Celsius Medical System is indicated for patients 18 years or older as long as their body weight is above 80 pounds. e-Celsius Medical System is not indicated for patients with special needs, esophageal or swallowing disorders, who are psychologically unsound, or have disabilities which may make the patient unable to swallow the device for any reason, these patients should be deemed ineligible for this device. e-Celsius Medical System is a prescription use device primarily (e.g., initial setup, final data assessment, reprocessing, etc.) indicated for use in clinical environments (e.g., hospitals, hospice, clinics, medical offices.)

Device Story

e-Celsius Medical System is an ingestible electronic thermometer pill; measures core body temperature continuously; transmits data wirelessly to a monitor/receiver. Used in clinical settings (hospitals, clinics) under physician order; intended for patients unable to use rectal catheters. Physician evaluates patient suitability; patient swallows pill; pill travels through GI tract; transmits temperature data to external monitor. Healthcare provider reviews continuous temperature data on monitor to guide clinical decisions. Benefits include accurate, non-invasive continuous core temperature monitoring without rectal catheterization.

Clinical Evidence

No clinical data provided; substantial equivalence based on bench testing and comparison to predicate device.

Technological Characteristics

Ingestible electronic thermometer pill; wireless transmission to external monitor. Operates as a continuous core body temperature sensor. Designed for clinical use in patients 18+ years and >80 lbs. Prescription-only device.

Indications for Use

Indicated for patients 18+ years, weight >80 lbs, requiring continuous core body temperature monitoring (>6 hours) where rectal catheters are inappropriate. Contraindicated for patients with swallowing/esophageal disorders, special needs, psychological instability, or inability to swallow the device.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. June 28, 2023 BodyCAP % Ray Kelly Consultant Arazy Group Consultants Inc. 3422 Leonardo Lane New Smyrna Beach, Florida 32168 Re: K210924 Trade/Device Name: e-Celsius Medical System Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: June 17, 2023 Received: June 21, 2023 Dear Mr. Ray Kelly: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting- combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, David Walloschek David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K210924 Device Name e-Celsius Medical System ## Indications for Use (Describe) e-Celsius Medical System is indicated for the measurement of core body temperature in patients for whom accurate continuous core body temperatures are clinically necessary for periods exceeding six (6) hours and with whom rectal catheters are not appropriate. e-Celsius Medical System can only be used on the order of a physician who has clinically evaluated the contraindications and warnings associated with the use of and ingestible thermometer and determined that the risk is clinically justified versus an oral thermometer or- rectal catheter. e-Celsius Medical System is indicated for patients 18 years or older as long as their body weight is above 80 pounds. e-Celsius Medical System is not indicated for patients with special needs, esophageal or swallowing disorders, who are psychologically unsound, or have disabilities which may make the patient unable to swallow the device for any reason, these patients should be deemed ineligible for this device. e-Celsius Medical System is a prescription use device primarily (e.g., initial setup, final data assessment, reprocessing, etc.) indicated for use in clinical environments (e.g., hospitals, hospice, clinics, medical offices.) | Type of Use (Select one or both, as applicable) | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...