← Product Code [LKJ](/productcode/LKJ) · K210902

# EliA Ro52, EliA Ro60 (K210902)

_Phadia AB · LKJ · Jul 27, 2022 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210902

## Device Facts

- **Applicant:** Phadia AB
- **Product Code:** [LKJ](/productcode/LKJ.md)
- **Decision Date:** Jul 27, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5100
- **Device Class:** Class 2
- **Review Panel:** Immunology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K210902 · Jul 27, 2022 | EliA Ro52, EliA Ro60 | Phadia AB | Biobank serum samples; Retrospective clinical patient serum cohorts | Retrospective clinical serum samples were used to determine clinical sensitivity, specificity, and expected values for the EliA Ro52 and EliA Ro60 assays. | Clinical sensitivity; Clinical specificity; Biobank; Retrospective cohort |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective clinical cohort study | Patients with SLE, SS, IIM, SSc, and various control diseases; Sample Size: 755 (EliA Ro52); 713 (EliA Ro60) | Not applicable for this study | Clinical sensitivity and specificity |
| Retrospective biobank analysis | Apparently healthy blood donors (Caucasian, African American, Hispanic, Asian); Sample Size: 640 | Not applicable for this study | Expected values/reference range |

## Indications for Use

EliA Ro52 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Ro52 in human serum as an aid in the diagnosis of Sjögren's syndrome (SS), systemic lupus erythematosus (SLE), systemic sclerosis (SSc) and idiopathic inflammatory myopathies (IIM) in conjunction with other laboratory and clinical findings. EliA Ro52 uses the EliA IgG method. EliA Ro60 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Ro60 in human serum as an aid in the diagnosis of Sjögren's syndrome (SS) and systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA Ro60 uses the EliA IgG method.

## Device Story

EliA Ro52 and EliA Ro60 are automated, solid-phase, fluoroenzyme immunoassays for detecting IgG autoantibodies against SS-A/Ro 52 kDa and 60 kDa proteins. The system uses polystyrene wells coated with human recombinant antigens. Patient serum is added; specific antibodies bind to the antigen. After washing, enzyme-labeled anti-human IgG conjugate is added, followed by a development solution. The instrument measures fluorescence, which is proportional to the amount of bound antibody. Results are compared to a calibration curve. The device is used in clinical laboratories on Phadia 250, 2500, and 5000 series instruments. Healthcare providers use these results alongside clinical findings to diagnose autoimmune conditions. The device benefits patients by providing standardized, automated, and reproducible antibody quantification to support clinical diagnosis.

## Clinical Evidence

Clinical performance was evaluated using 755 (Ro52) and 713 (Ro60) serum samples from target disease groups (SLE, SS, IIM, SSc) and control groups (autoimmune/infectious diseases). For EliA Ro52, sensitivity ranged from 20.9% to 55.0% and specificity from 95.6% to 96.9% depending on the target disease. For EliA Ro60, sensitivity ranged from 48.3% to 71.7% and specificity from 98.4% to 98.6%. Analytical performance included precision (CVs <10%), linearity, and interference testing.

## Technological Characteristics

Solid-phase fluoroenzymeimmunoassay (FEIA). Wells coated with human recombinant SS-A/Ro 52 kDa or 60 kDa protein. Detection via β-Galactosidase labeled anti-human IgG conjugate and 4-Methylumbelliferyl-β-D-galactoside substrate. Fluorescence measurement. Automated system (Phadia 250/2500/5000). Calibrated to WHO IRP 67/86. Analytical measuring range: 0.7–240 EliA U/mL (Ro52) and 0.5–240 EliA U/mL (Ro60).

## Regulatory Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

## Predicate Devices

- Quanta Flash Ro52 ([K141655](/device/K141655.md))
- Quanta Flash Ro60 ([K141328](/device/K141328.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K210902

B Applicant

Phadia AB

C Proprietary and Established Names

EliA Ro52

EliA Ro60

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  LKJ | Class II | 21 CFR 866.5100 - Antinuclear Antibody Immunological Test System | IM - Immunology  |

## II Submission/Device Overview:

A Purpose for Submission:

New device.

B Measurand:

IgG autoantibodies specific to SS-A/Ro proteins (52 kDa and 60 kDa)

C Type of Test:

Automated semi-quantitative solid phase fluoroenzymeimmunoassay

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

K210902 - Page 2 of 23

## III Intended Use/Indications for Use:

### A Intended Use(s):

EliA Ro52 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Ro52 in human serum as an aid in the diagnosis of Sjögren's syndrome (SS), systemic lupus erythematosus (SLE), systemic sclerosis (SSc) and idiopathic inflammatory myopathies (IIM) in conjunction with other laboratory and clinical findings. EliA Ro52 uses the EliA IgG method.

EliA Ro60 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Ro60 in human serum as an aid in the diagnosis of Sjögren's syndrome (SS) and systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA Ro60 uses the EliA IgG method.

### B Indication(s) for Use:

Same as intended use.

### C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

### D Special Instrument Requirements:

For use on the Phadia 250 instrument and the Phadia 2500 and Phadia 5000 instrument series (E-modules).

## IV Device/System Characteristics:

### A Device Description:

The EliA Ro52 and the EliA Ro60 are composed of assay-specific reagents, EliA method-specific reagents, and general reagents.

#### Assay-Specific Reagents include:

- For EliA Ro52
EliA Ro52 Wells: coated with human recombinant SS-A/Ro (52 kDa) protein – 2 carriers (12 wells each), ready to use;

- For EliA Ro60
EliA Ro60 Wells: coated with human recombinant SS-A/Ro (60 kDa) protein – 4 carriers (12 wells each), ready to use;

- EliA ANA Positive Control 250 or 2500/5000: Human serum in PBS containing IgG antibodies to dsDNA, RNP, Sm, Ro, La, Scl-70, CENP and Jo-1 – 6 single use vials, 0.3 mL each, ready to use;

{2}

- EliA ANA 3 Positive Control 250 or 2500/5000: Human monoclonal antibodies in Tris buffer containing IgG antibodies to Ro52, Rib-P and RNA Pol III – 6 single use vials, 0.3 mL each, ready to use;
- EliA IgG/IgM/IgA Negative Control 250 or 2500/5000: Human blood preparation from healthy donors in PBS containing BSA, detergent and 0.095% sodium azide – 6 single-use vials, 0.3 mL each, ready to use.

## EliA Method-Specific Reagents include:

- EliA Sample Diluent: PBS containing BSA, detergent and 0.095% sodium azide – 6 bottles, 48 mL each, ready to use; or 6 bottles, 400 mL each, ready to use.
- EliA IgG Conjugate 50 or 200: β-Galactosidase labeled anti-IgG (mouse monoclonal antibodies) in PBS containing BSA and 0.06% sodium azide – 6 wedge shaped bottles, 5 mL each, ready to use; or 6 wedge shaped bottles, 19 mL each, ready to use.
- EliA IgG Calibrator Strips: Human IgG (0, 4, 10, 20, 100, 600 µg/L) in PBS containing BSA, detergent and 0.095% sodium azide – 5 strips, 6 single-use vials per strip, 0.3 mL each, ready to use.
- EliA IgG Curve Control Strips: Human IgG (20 µg/L) in PBS containing BSA, detergent and 0.095% sodium azide – 5 strips, 6 single-use vials per strip, 0.3 mL each, ready to use.
- EliA IgG Calibrator Well: coated with mouse monoclonal antibodies – 4 carriers (12 wells each), ready to use.

## General Reagents include:

- Development Solution: 0.01% 4-Methylumbelliferyl-β-D-galactoside, &lt;0.0010% preservative
- Stop Solution: 4% Sodium Carbonate
- Washing Solution Additive: detergent, preservative &lt;0.13%
- Washing Solution Concentrate: phosphate buffer

## B Principle of Operation:

The EliA Ro52 and the EliA Ro60 are semi-quantitative solid-phase fluoroimmunoassay for the determination of autoantibodies against SS-A/Ro 52 kDa and 60 kDa proteins. Both assays are fully integrated and automated system which comprises of assay-specific reagents, EliA method-specific reagents, and general reagents.

The specific antigen (SS-A/Ro 52kDa protein or SS-A/Ro 60 kDa protein) is coated to the EliA solid phase component (EliA Well). The EliA wells are molded cups comparable to excised wells from a microtiter plate. If present in the patient's specimen, antibodies to the SS-A/Ro 52 kDa protein or SS-A/Ro 60 kDa protein bind to these specific antigens coated on the EliA well. After washing away non-bound antibodies, enzyme-labeled antibodies against human IgG antibodies (EliA IgG Conjugate) are added to form an antibody-conjugate complex. After incubation, non-bound conjugate is washed away, and the bound complex is incubated with a Development Solution. Following a reaction period, the reaction is stopped and the fluorescence of the reaction mixture is measured. As the extent of the fluorescent signal corresponds to the amount of antibody of interest that is bound to the antigen coated EliA well, the higher the value of fluorescent signal detected by the instrument, the higher the amount of antibody bound and

K210902 - Page 3 of 23

{3}

detected in the sample tested. To evaluate the test results, the fluorescence from patient samples is directly compared to a calibrator control curve.

# V Substantial Equivalence Information:

# A Predicate Device Name(s):

Quanta Flash Ro52, Quanta Flash Ro52 Calibrators, Quanta Flash Ro52 Controls Quanta Flash Ro60, Quanta Flash Ro60 Calibrators, Quanta Flash Ro60 Controls

# B Predicate 510(k) Number(s):

K141655

K141328

# C Comparison with Predicate(s):

EliA Ro52

|  Device & Predicate Device(s): | K210902 | K141655  |
| --- | --- | --- |
|  Device Trade Name | EliA Ro52 | QUANTA Flash Ro52  |
|  General Device Characteristic Similarities  |   |   |
|  Intended Use/Indications For Use | EliA Ro52 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Ro52 in human serum as an aid in the diagnosis of Sjögren's syndrome (SS), systemic lupus erythematosus (SLE), systemic sclerosis (SSc) and idiopathic inflammatory myopathies (IIM) in conjunction with other laboratory and clinical findings. EliA Ro52 uses the EliA IgG method. | QUANTA Flash Ro52 is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-Ro52 autoantibodies in human serum. The presence of anti-Ro52 autoantibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of Systemic Lupus Erythematosus (SLE), Sjögren's Syndrome, Systemic Sclerosis, Idiopathic Inflammatory Myopathies.  |
|  Classification | Class II | Same  |
|  Regulation Number | 21 CFR 866.5100 | Same  |
|  Assay Technique | ELISA | Same  |
|  Type of Test | Semi-quantitative | Same  |
|  Reaction Temperature | 37°C controlled | Same  |
|  General Device Characteristic Differences  |   |   |
|  Antigen | Human recombinant SSA/Ro 52 (kDa) protein | Purified recombinant Ro52 antigen  |

K210902 - Page 4 of 23

{4}

EliA Ro60

|  Device & Predicate Device(s): | K210902 | K141328  |
| --- | --- | --- |
|  Device Trade Name | EliA Ro60 | QUANTA Flash Ro60  |
|  General Device Characteristic Similarities  |   |   |
|  Intended Use/Indications For Use | EliA Ro60 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Ro60 in human serum as an aid in the diagnosis of Sjögren’s syndrome (SS) and systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA Ro60 uses the EliA IgG method. | QUANTA Flash Ro60 is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-Ro60 autoantibodies in human serum. The presence of anti-Ro60 autoantibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of Systemic Lupus Erythematosus and Sjögren’s Syndrome.  |
|  Classification | Class II | Same  |
|  Regulation Number | 21 CFR 866.5100 | Same  |
|  Assay Technique | ELISA | Same  |
|  Type of Test | Semi-quantitative | Same  |

K210902 - Page 5 of 23

{5}

|  Device & Predicate Device(s): | K210902 | K141328  |
| --- | --- | --- |
|  Reaction Temperature | 37°C controlled | Same  |
|  General Device Characteristic Differences  |   |   |
|  Antigen | Human recombinant SS-A/Ro 60 (kDa) protein | Purified recombinant Ro60 antigen  |
|  Product Code | LKJ | LLL  |
|  Sample Dilution | 1:100 | 1:23  |
|  Reporting of Results | EliA U/mL (arbitrary) | Chemiluminescent units (CU)  |
|  Instrumentation | EliA Ro60 uses the EliA IgG method on the instruments Phadia 250 and the E-Modules of the Phadia 2500 and Phadia 5000 series. | Quanta Flash Ro60 is run on the BIO-FLASH instrument.  |
|  Detection Antibody | β-Galactosidase conjugated anti-human IgG (mouse monoclonal antibodies) | Isoluminol conjugated anti-human IgG  |
|  Signal | Fluorescence | Relative Light Units (RLU)  |
|  Calibration | 6-point total IgG Calibration 6 vials of human IgG at concentrations of 0, 4, 10, 20, 100, 600 μg/L | Lot specific Master Curve and two Calibrators  |
|  Interpretation of results | Negative < 7 EliA U/mL Equivocal 7-10 EliA U/mL Positive > 10 EliA U/mL | Negative: < 20 CU Positive: > 20 CU  |
|  Substrate | 0.01% 4-Methylumbelliferyl-β-D-galactoside with preservative | “Trigger” reagents  |

VI Standards/Guidance Documents Referenced:

- CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures, Approved Guideline, Third Edition
- CLSI EP06-Ed2, Evaluation of the Linearity of Quantitative Measurement Procedures
- CLSI EP07, 3rd Edition, Interference Testing in Clinical Chemistry
- CLSI EP09c 3rd Edition, Measurement Procedure Comparison and Bias Estimation Using Patient Samples
- CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures, Approved Guideline - Second Edition
- CLSI EP28-A3c, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory, Approved Guideline - Third Edition
- CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents, Approved Guideline
- CLSI EP37, 1st Edition, Supplemental Tables for Interference Testing in Clinical Chemistry

K210902 - Page 6 of 23

{6}

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility.

Within-lab Imprecision:

Within-lab imprecision of the EliA Ro52 and the EliA Ro60 was assessed by testing four serum samples using one lot of reagent on one Phadia 250 instrument. Samples were tested in duplicate per run, two runs per day for 20 days, for a total of 80 observations per sample. The results are presented in the tables below.

|  Within-Lab Precision of the EliA Ro52  |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | N | Mean (EliA U/mL) | Within-Run |   | Between-Run |   | Between-Day |   | Within-Lab  |   |
|   |   |   |  SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  1 | 80 | 3.0 | 0.1 | 2.9 | 0.1 | 3.6 | 0.0 | 1.0 | 0.1 | 4.6  |
|  2 | 80 | 7.9 | 0.2 | 2.9 | 0.2 | 2.6 | 0.2 | 2.0 | 0.3 | 3.9  |
|  3 | 80 | 10.5 | 0.3 | 2.5 | 0.5 | 4.7 | 0.3 | 3.2 | 0.6 | 5.3  |
|  4 | 80 | 69.1 | 1.9 | 2.7 | 4.6 | 6.6 | 3.6 | 5.2 | 4.9 | 7.1  |
|  Within-Lab Precision of the EliA Ro60  |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | N | Mean (EliA U/mL) | Within-Run |   | Between-Run |   | Between-Day |   | Within-Lab  |   |
|   |   |   |  SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  1 | 80 | 2.6 | 0.1 | 5.4 | 0.2 | 7.1 | 0.1 | 2.2 | 0.2 | 8.9  |
|  2 | 80 | 6.3 | 0.2 | 2.6 | 0.6 | 9.3 | 0.3 | 5.4 | 0.6 | 9.7  |
|  3 | 80 | 10.0 | 0.2 | 2.0 | 0.7 | 6.6 | 0.4 | 3.8 | 0.7 | 6.9  |
|  4 | 80 | 33.3 | 1.3 | 3.8 | 2.2 | 6.8 | 0.0 | 0.0 | 2.6 | 7.7  |

Reproducibility

The reproducibility of the EliA Ro52 and the EliA Ro60 was determined by testing five samples on three Phadia 250 instruments. Each sample was assessed in quadruplicate per run, one run per day for seven days using three reagent lots to generate 84 observations per sample on each instrument, or a total of 252 observations per sample. The results are presented in the tables below.

K210902 - Page 7 of 23

{7}

Reproducibility of EliA Ro52

|  Sample | N | Mean (EliA U/mL) | Within-Run |   | Between-Run/Day |   | Between-Instrument |   | Between-Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  1 | 252 | 4.4 | 0.2 | 3.7 | 0.3 | 5.8 | 0.1 | 1.3 | 0.0 | 0.0 | 0.3 | 6.9  |
|  2 | 252 | 8.3 | 0.3 | 3.0 | 0.3 | 3.2 | 0.0 | 0.0 | 0.2 | 2.2 | 0.4 | 5.0  |
|  3 | 252 | 10.6 | 0.3 | 2.5 | 0.3 | 2.9 | 0.1 | 0.7 | 0.2 | 1.6 | 0.5 | 4.1  |
|  4 | 252 | 32.8 | 0.8 | 2.4 | 1.2 | 3.6 | 0.6 | 1.7 | 1.3 | 3.9 | 2.0 | 5.9  |
|  5 | 252 | 215.4 | 8.3 | 3.9 | 7.9 | 3.7 | 8.2 | 3.8 | 4.2 | 2.0 | 14.7 | 5.7  |

Reproducibility of EliA Ro60

|  Sample | N | Mean (EliA U/mL) | Within-Run |   | Between-Run/Day |   | Between-Instrument |   | Between-Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  1 | 252 | 3.5 | 0.1 | 3.6 | 0.3 | 7.2 | 0.1 | 1.5 | 0.1 | 1.7 | 0.3 | 8.2  |
|  2 | 252 | 6.1 | 0.2 | 3.0 | 0.3 | 4.5 | 0.0 | 0.0 | 0.0 | 0.4 | 0.3 | 5.4  |
|  3 | 252 | 10.2 | 0.2 | 2.3 | 0.3 | 3.2 | 0.1 | 0.7 | 0.2 | 2.1 | 0.5 | 4.4  |
|  4 | 252 | 32.2 | 1.1 | 3.3 | 1.3 | 3.9 | 0.0 | 0.0 | 0.9 | 2.9 | 1.9 | 5.8  |
|  5 | 252 | 201.0 | 9.6 | 4.8 | 11.6 | 5.7 | 2.9 | 1.4 | 10.6 | 5.3 | 18.6 | 9.1  |

Reproducibility on the Phadia 2500E and Phadia 5000E instrument series:

The reproducibility of the EliA Ro52 and the EliA Ro60 was determined by testing five samples on three Phadia 2500E instruments. Each sample was assessed in quadruplicate per run, one run per day for seven days using one reagent lot to generate 84 observations per sample. The results are presented in the tables below.

Reproducibility of EliA Ro52 on Phadia 2500E

|  Sample | N | Mean (EliA U/mL) | Within-Run |   | Between-Run/Day |   | Between-Instrument |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  1 | 84 | 4.7 | 0.2 | 4.0 | 0.1 | 3.0 | 0.1 | 2.2 | 0.3 | 5.0  |
|  2 | 84 | 8.6 | 0.3 | 3.5 | 0.2 | 2.3 | 0.1 | 0.7 | 0.4 | 4.2  |
|  3 | 84 | 11.0 | 0.3 | 2.5 | 0.5 | 4.7 | 0.2 | 1.9 | 0.6 | 5.4  |
|  4 | 84 | 36.3 | 1.5 | 4.0 | 1.0 | 2.8 | 0.0 | 0.0 | 1.8 | 4.9  |
|  5 | 84 | 212.9 | 10.3 | 4.8 | 9.1 | 4.3 | 0.0 | 0.0 | 13.7 | 6.4  |

Reproducibility of EliA Ro60 on Phadia 2500E

|  Sample | N | Mean (EliA U/mL) | Within-Run |   | Between-Run/Day |   | Between-Instrument |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  1 | 84 | 3.6 | 0.2 | 5.3 | 0.3 | 6.9 | 0.1 | 1.4 | 0.3 | 8.7  |
|  2 | 84 | 6.6 | 0.2 | 2.9 | 0.3 | 4.3 | 0.2 | 2.9 | 0.4 | 5.2  |
|  3 | 84 | 10.5 | 0.3 | 2.9 | 0.4 | 3.4 | 0.1 | 0.8 | 0.5 | 4.5  |
|  4 | 84 | 35.8 | 1.2 | 3.3 | 1.3 | 3.7 | 0.0 | 0.0 | 1.8 | 4.9  |
|  5 | 84 | 219.9 | 11.7 | 5.3 | 10.3 | 4.7 | 10.7 | 4.9 | 18.9 | 7.1  |

K210902 - Page 8 of 23

{8}

# 2. Linearity:

The linearity of the EliA Ro52 and the EliA Ro60 was evaluated based on the recommendations of CLSI EP06-Ed2.

Three patient serum samples were serially diluted using a normal blood donor serum sample for a total of 9 dilutions for each dilution series across. The following patient serum sample concentrations were used: 266.1 EliA U/mL, 89.5 EliA U/mL, and 15.9 EliA U/mL were used for EliA Ro52 testing on the Phadia 250 instrument; 275.5 EliA U/mL, 97.7 EliA U/mL, 16.2 EliA U/mL were used for EliA Ro52 testing on the Phadia 2500E instrument; 249.4 EliA U/mL, 88.5 EliA U/mL, and 15.4 EliA U/mL were used for EliA Ro60 testing on the Phadia 250 instrument; and 263.6 EliA U/mL, 92.3 EliA U/mL, and 12.3 EliA U/mL were used for EliA Ro60 testing on the Phadia 2500E instrument. Each sample was then tested in quadruplicate using one lot of EliA Ro52 Well and EliA Ro60 Well, and a single Phadia 250 instrument or Phadia 2500E instrument, respectively. The results were analyzed using by fitting a regression model, assuming proportionality between the observed concentration and the expected concentration, using the weighted least squares method.

EliA Ro52 on Phadia 250:

|  Sample Number | Range [EliA U/mL] | Slope (95% CI) | Intercept (95% CI) | R2 | % Recovery  |
| --- | --- | --- | --- | --- | --- |
|  1 | 8.1 - 266.1 | 0.987 (0.961 - 1.012) | 1.226 (0.865 – 1.587) | 0.999 | 99.7 – 116.7  |
|  2 | 3.2 – 89.5 | 0.994 (0.977 - 1.011) | 0.435 (0.320 – 0.551) | 0.999 | 97.8 – 117.4  |
|  3 | 0.7 - 15.9 | 0.948 (0.884 - 1.013) | 0.091 (-0.053 – 0.235) | 0.990 | 92.4 – 127.5  |
|  Combined | 0.7 – 266.1 | 1.014 (0.983 – 1.045) | 0.025 (-0.08 – 0.130) | 0.993 | 92.3 – 118.5  |

EliA Ro52 on Phadia 2500E:

|  Sample Number | Range [EliA U/mL] | Slope (95% CI) | Intercept (95% CI) | R2 | % Recovery  |
| --- | --- | --- | --- | --- | --- |
|  1 | 7.1 - 275.5 | 0.862 (0.811 – 0.913) | 0.776 (-0.291 – 1.842) | 0.994 | 83.9 – 100  |
|  2 | 2.4 – 97.7 | 0.911 (0.884 – 0.938) | 0.490 (0.265 – 0.715) | 0.998 | 92.8 – 116.7  |
|  3 | 0.7 - 16.2 | 0.904 (0.831 – 0.976) | -0.045 (-0.163 – 0.07) | 0.987 | 82.6 – 104.4  |
|  Combined | 0.7 – 275.5 | 0.928 (0.902 – 0.953) | -0.01 (-0.114 – 0.094) | 0.994 | 80.0 – 114.3  |

For the EliA Ro52 assay, the results support the linearity of the claimed analytical measuring range of  $0.7 - 240$  EliA U/mL.

EliA Ro60 on Phadia 250:

|  Sample Number | Range [EliA U/mL] | Slope (95% CI) | Intercept (95% CI) | R2 | % Recovery  |
| --- | --- | --- | --- | --- | --- |
|  1 | 2.7 – 88.5 | 1.016 (0.967 – 1.064) | 0.674 (0.397 – 0.951) | 0.996 | 100 – 119.3  |
|  2 | 11.0 – 249.4 | 1.047 (1.030 – 1.064) | 1.282 (0.515 – 2.048) | 0.999 | 100 – 116.1  |
|  3 | 0.5 – 15.4 | 0.988 (0.971 – 1.005) | 0.053 (0.20 – 0.086) | 0.999 | 95.1 – 117.1  |
|  Combined | 0.5 – 249.4 | 1.063 (1.044 – 1.081) | -0.004 (-0.076 – 0.07) | 0.998 | 94.0 – 117.4  |

K210902 - Page 9 of 23

{9}

EliA Ro60 on Phadia 2500E:

|  Sample Number | Range [EliA U/mL] | Slope (95% CI) | Intercept (95% CI) | R² | % Recovery  |
| --- | --- | --- | --- | --- | --- |
|  1 | 2.6 – 92.3 | 1.013 (0.968 – 1.058) | 0.136 (-0.125 – 0.396) | 0.997 | 95.9 – 108.1  |
|  2 | 4.8 – 263.6 | 1.033 (0.980 – 1.086) | 0.910 (0.171 – 1.649) | 0.995 | 96.6 – 114.3  |
|  3 | 0.6 - 12.3 | 0.963 (0.942 – 0.984) | 0.040 (0.009 – 0.071) | 0.999 | 95.2 – 104.4  |
|  Combined | 0.6 – 263.6 | 1.069 (1.045 – 1.093) | -0.039 (-0.131– 0.053) | 0.996 | 90.9 – 114.5  |

For the EliA Ro60 assay, the results support the linearity of the claimed analytical measuring range of 0.5 – 240 EliA U/mL.

Hook Effect/Over the Range Results:

Not applicable. Results above the upper limit of the measuring range are reported as “&gt;240” for both assays. No recommendations are made for dilution of samples outside of the analytical measuring range in the Directions for Use of each assay.

3. Analytical Specificity/Interference:

Comparison to Reference Sera:

CDC (Center for Disease Controls and Prevention) ANA human reference sera panel samples #1– #12 were tested using one lot of the Ro52 and one lot of the EliA Ro60 EliA Rib-P assay reagents.

For the EliA Ro52, CDC ANA #2 (speckled pattern in FANA), CDC ANA #7 (SS-A/Ro) and CDC ANA #10 (Jo-1) showed positive results with 13.9 EliA U/mL, 19.5 EliA U/mL and 247.4 EliA U/mL, respectively. CDC ANA #3 (RNP-Sm, SS-A/Ro, SS-B/La) showed an equivocal result with 7.4 EliA U/mL. All other CDC ANA samples remained negative.

For the EliA Ro60, CDC ANA #2 (speckled pattern in FANA), CDC ANA #3 (RNP-Sm, SS-A/Ro, SS-B/La), and CDC ANA #7 (SS-A/Ro) showed positive results with 143.0 EliA U/mL, 146.2 EliA U/mL and 207.2 EliA U/mL, respectively. All other CDC ANA samples remained negative.

Endogenous and Exogenous Interference:

Three serum samples: one negative sample, one sample with a concentration within the equivocal range, and one positive sample, were spiked with the different interfering substances or blank solution. For the EliA Ro52, the concentration of the samples were 1.0 EliA U/mL, 10.4 EliA U/mL, and 116.5 EliA U/mL. For the EliA Ro60, the concentration of the samples were 0.5 EliA U/mL, 8.0 EliA U/mL, and 102.5 EliA U/mL. All samples were tested in triplicate in two runs using one lot of EliA antigen specific wells and one lot of system reagents on the Phadia 250 instrument. The quotient between serum sample spiked with interference substance and serum sample spiked with blank was calculated for each sample spiked with the interfering substance. No interference, the quotient between samples

K210902 - Page 10 of 23

{10}

spiked with interference substance or blank within 0.90 – 1.10, was observed up to the concentrations listed in the table below:

|  Potential Interfering Compound | Concentration with No Interference  |
| --- | --- |
|  *Endogenous substance*  |   |
|  Bilirubin F | 40 mg/dL  |
|  Bilirubin C | 40 mg/dL  |
|  Hemoglobin | 1000 mg/dL  |
|  Lipemic factor | 2000 mg/dL  |
|  Rheumatoid factor | 500 IU/mL  |
|  Human IgG | 3500 mg/dL  |
|  *Exogenous substance*  |   |
|  Ibuprofen | 21.9 mg/dL  |
|  Losartan | 1.14 mg/dL  |
|  Hydroxychloroquine | 0.23 mg/dL  |
|  Azathioprine | 0.26 mg/dL  |
|  Prednisone | 0.01 mg/dL  |
|  Rituximab | 109 mg/dL  |
|  Infliximab | 26.4 mg/dL  |
|  Omeprazole | 0.84 mg/dL  |
|  Diltiazem | 0.09 mg/dL  |

4. Assay Reportable Range:

The reportable range for the EliA Ro52 is 0.7 – 240.0 EliA U/mL.

The reportable range for the EliA Ro60 is 0.5 – 240.0 EliA U/mL.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

Traceability:

The EliA IgG calibrators are traceable (via unbroken chain of calibrations) to the International Reference Preparation (IRP) 67/86 of Human Serum Immunoglobulins A, G and M from WHO. New batches of IgG calibrators are compared to a secondary standard (standardized with the IRP) or the IRP directly and adjusted accordingly to meet the correct concentration.

The instrument measures specific IgG concentrations in µg/L. By using a conversion factor given by the lot-specific code of the EliA test well, the results are automatically converted to EliA U/mL.

Stability:

Stability (open and closed vial, and on-board) of EliA IgG method-specific reagents and general EliA reagents were previously established through the clearance of other EliA tests,

K210902 - Page 11 of 23

{11}

e.g., K141375, K061165. The stability for the EliA Ro52 and the EliA Ro60 assay-specific reagents (i.e., EliA Ro52 Wells and EliA Ro60 Wells) were determined as follows:

## Shelf-life:

The stability of the EliA Ro52 Wells and the EliA Ro60 Wells was evaluated with a real-time study on the Phadia 250 instrument. Three lots EliA Ro52 Wells were stored under standard conditions in sealed foil bags with desiccant bags inside, at 2–8°C, and tested 0, 7, 13, 19, and 25 months. Each lot of wells was initially tested in three identical initial runs and in two identical test runs at the later time points. The wells were tested with four positive samples and with nine negative samples in single determination with a full calibration curve.

The shelf-life was determined to be up to 24 months at 2–8°C for the EliA Ro52 and up to 18 months at 2–8°C for EliA Ro60.

## On-board stability:

The on-board stability the EliA Ro52 Wells and the EliA Ro60 Wells was tested over 8 weeks using three positive and two negative samples on the Phadia 250 instrument. For this study, the carriers were stored at 10°C ±20% and 80% ±20% humidity in the Phadia 250 carrier storage tray in a temperature- and humidity-test cabinet for up to 8 weeks. The following conditions were tested: 8 weeks at 10°C and 80% humidity, 8 weeks at 2–8°C, 4 weeks at 2–8°C and 4 weeks at 10°C and 80% humidity, and 2 weeks at 2–8°C and 6 weeks at 10°C with 80% humidity. At 12 different days during the study, the carriers were removed from the holder, shaken three times to simulate dispensing, and were returned to the carrier storage tray. The wells were measured together after 8 weeks, using three positive and two negative samples, which were assayed in duplicate.

The on-board stability for the Phadia 250 was determined to be 28 days at 2–8°C.

## Open (in-use) Stability:

Stability after first opening of the foil bag containing the EliA Ro52 and EliA Ro60 wells was tested with a real-time study. In this study, EliA Ro52 and EliA Ro60 wells in the carrier (stored at 2–8°C) were removed from the foil bag, stored 5 minutes at room temperature, closed again by the zipper, and stored at +10°C and 80% humidity in a climate chamber until use. This process of opening and sealing the wells in the carrier was repeated six times within 2 months. The wells were stored at +10°C and 80% humidity in a climate chamber for up to 10 months. At 0, 4, 6, and 10 months, the wells were compared to EliA Ro52 and EliA Ro 60 wells stored under standard conditions at +2 to +8°C. Each positive sample, three for EliA Ro52 and two for EliA Ro60, were tested in triplicate. Each negative sample, two for EliA Ro52 and one for EliA Ro60, were tested in single determinations.

The in-use stability was determined to be 6 months at 2–8°C for the EliA Ro52 and EliA Ro60 wells.

## 6. Detection Limit:

Limit of blank (LoB), Limit of detection (LoD) and limit of quantitation (LoQ) of the EliA Ro52 assay were determined on both Phadia 250 and Phadia 2500E.

K210902 - Page 12 of 23

{12}

LoB was determined from the measurement of four blank samples (analyte-free, i.e. IgG depleted sera) tested in five replicates per sample in one run on a Phadia 250 and Phadia 2500E instrument, respectively, over three consecutive days using two different lots of test reagents. The LoB was calculated as the 95th percentile for each lot (60 observations per lot). The LoB was determined as the higher value of these two lots for each type of instrument. The claimed LoB for EliA Ro52 and EliA Ro60 is the highest 95th percentile LoB value across the two lots and instrument types.

LoD was determined by testing four low-level samples in five replicates per sample in one run on a Phadia 250 and Phadia 2500E instrument, respectively, over three consecutive days using two different lots of test reagents. The LoD for each lot was calculated based on 240 determinations with 120 blank and 120 low level replicates per instrument type, following the recommendations of CLSI document EP17-A and with proportions of false positives (α) less than 5% and false negatives (β) less than 5%. The claimed LoD for EliA Ro52 and EliA Ro60 is the greatest LoD across the two lots and both instrument types.

LoQ was determined by testing four low-levels samples in five replicates per sample over three consecutive days, in one run on a Phadia 250 and Phadia 2500E instrument, and two reagent lots. The LoQ was defined as the lowest concentration of measurand that meets the predefined imprecision goal of &lt;20% for each lot. The greatest LoQ across the two lots and both instrument types was set as the claimed LoQ for the EliA Ro52 and EliA Ro60 assays.

|  EliA Ro52  |   |   |   |
| --- | --- | --- | --- |
|  Instrument | LoB EliA U/mL | LoD EliA U/mL | LoQ EliA U/mL  |
|  Phadia 250 | 0.0 | 0.2 | 0.6  |
|  E-module of the Phadia 2500E | 0.1 | 0.3 | 0.7  |
|  The claim for both instrument types | 0.1 | 0.3 | 0.7  |
|  EliA Ro60  |   |   |   |
| --- | --- | --- | --- |
|  Instrument | LoB EliA U/mL | LoD EliA U/mL | LoQ EliA U/mL  |
|  Phadia 250 | 0.1 | 0.2 | 0.4  |
|  E-module of the Phadia 2500E | 0.1 | 0.2 | 0.5  |
|  The claim for both instrument types | 0.1 | 0.4 | 0.5  |

## 7. Assay Cut-Off:

The assay cut-off and the equivocal range for the EliA Ro52 were established by testing a cohort consisting of 69 apparently healthy blood donors, 19 samples from SLE patients and 9 samples from SS patients on a Phadia 250 instrument. The established cut-off was further verified for the additional intended use groups of IIM and SSc patients by testing an additional 10 IIM and 14 SSc patient sera.

The assay cut-off and the equivocal range of the EliA Ro60 was established by testing a cohort consisting of 70 samples from apparently healthy blood donors, 22 samples from SLE patients and 6 samples from SS patients on a Phadia 250 instrument.

K210902 - Page 13 of 23

{13}

The cut-off was set as follows for the EliA Ro52 and the EliA Ro60:

|  Test Result | Interpretation  |
| --- | --- |
|  <7 EliA U/mL | Negative  |
|  7–10 EliA U/mL | Equivocal  |
|  >10 EliA U/mL | Positive  |

In case of equivocal results, it is recommended to retest the patient after 8–12 weeks.

## B Comparison Studies:

### 1. Method Comparison with Predicate Device:

A total of 208 patient serum samples with antibody concentrations covering the measuring range were tested with the EliA Ro52, the EliA Ro60 and their predicates, QUANTA Flash Ro52 and QUANTA Flash Ro60 assays. The tests were performed in single determination according to the Directions for Use for each assay. Only samples that were found within the measuring intervals of the EliA and/or QUANTA Flash tests were used for the method comparison analysis. Positive percent agreement (PPA), negative percent agreement (NPA), and total agreement were calculated with equivocal results considered as negative or as positive, respectively. The results are summarized in the tables below:

**EliA Ro52:**

|  EliA Ro52: equivocal results considered negative | QUANTA Flash Ro52  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive: ≥ 20 Units | Negative: < 20 Units | Total  |
|  EliA Ro52 | Positive: > 10 EliA U/mL | 42 | 2 | 44  |
|   |  Negative: < 10 EliA U/mL | 10 | 127 | 137  |
|   |  Total | 52 | 129 | 181  |
|  Equivocal Results Considered Negative | Calculation | Agreement (%) | 95% CI  |
| --- | --- | --- | --- |
|  PPA | 42 / 52 | 80.8 | 67.5 – 90.4  |
|  NPA | 127 / 129 | 98.4 | 94.5 – 99.8  |
|  Total | (42+127) / 181 | 93.4 | 88.8 – 96.5  |
|  EliA Ro52: equivocal results considered positive | QUANTA Flash Ro52  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive: ≥ 20 Units | Negative: < 20 Units | Total  |
|  EliA Ro52 | Positive: > 7 EliA U/mL | 48 | 12 | 60  |
|   |  Negative: < 7 EliA U/mL | 4 | 117 | 121  |
|   |  Total | 52 | 129 | 181  |

K210902 - Page 14 of 23

{14}

|  Equivocal Results Considered Negative | Calculation | Agreement (%) | 95% CI  |
| --- | --- | --- | --- |
|  PPA | 48 / 52 | 92.3 | 81.5 – 97.9  |
|  NPA | 117 / 129 | 90.7 | 84.3 – 95.1  |
|  Total | (48+117) / 181 | 91.2 | 86.1 – 94.9  |

EliA Ro60

|  EliA Ro60: equivocal results considered negative | QUANTA Flash Ro60  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive: ≥ 20 Units | Negative: < 20 Units | Total  |
|  EliA Ro60 | Positive: > 10 EliA U/mL | 62 | 3 | 65  |
|   |  Negative: < 10 EliA U/mL | 4 | 35 | 39  |
|   |  Total | 66 | 38 | 104  |
|  Equivocal Results Considered Negative | Calculation | Agreement (%) | 95% CI  |
| --- | --- | --- | --- |
|  PPA | 62 / 66 | 93.9 | 85.2 – 98.3  |
|  NPA | 35 / 38 | 92.1 | 78.6 – 98.3  |
|  Total | (62+35) / 104 | 93.3 | 86.6 – 97.3  |
|  EliA Ro60: equivocal results considered positive | QUANTA Flash Ro60  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive: ≥ 20 Units | Negative: < 20 Units | Total  |
|  EliA Ro60 | Positive: > 7 EliA U/mL | 64 | 7 | 71  |
|   |  Negative: < 7 EliA U/mL | 2 | 31 | 33  |
|   |  Total | 66 | 38 | 104  |
|  Equivocal Results Considered Negative | Calculation | Agreement (%) | 95% CI  |
| --- | --- | --- | --- |
|  PPA | 64 / 66 | 97.0 | 89.5 – 99.6  |
|  NPA | 31 / 38 | 81.6 | 65.7 – 92.3  |
|  Total | (64+31) / 104 | 91.3 | 84.2 – 96.0  |

K210902 - Page 15 of 23

{15}

# 2. Matrix Comparison:

Not applicable, serum is the claimed sample type for the EliA Ro52 and the EliA Ro60.

# 3. Instrument Comparison:

# EliA Ro52:

Performance of EliA Ro52 was evaluated on the Phadia 250 and Phadia 2500E instrument using 57 positive, 6 equivocal and 30 negative samples. The samples were analyzed in single determination on one Phadia 250 and one Phadia 2500E instrument each. A Passing-Bablok regression analysis was performed and demonstrated a slope of 0.94 (0.93 - 0.97 95% CI) and an intercept of 0.91 (0.46 - 1.14 95% CI). The PPA, NPA, and total agreement were calculated with equivocal results considered as negative or as positive, respectively. The results are summarized in the table below:

|  Percent Ro52 Agreement | Agreement (%), (95% CI) | Agreement (%), (95% CI)  |
| --- | --- | --- |
|   |  Equivocal = Negative | Equivocal = Positive  |
|  PPA | 80.8% (67.5 – 90.4%) | 80.8% (67.5 – 90.4%)  |
|  NPA | 98.4% (94.5 – 99.8%) | 98.4% (94.5 – 99.8%)  |
|  Total | 93.4% (88.8 – 96.5%) | 93.4% (88.8 – 96.5%)  |

# EliA Ro60:

Performance of EliA Ro60 was evaluated on the Phadia 250 and Phadia 2500E instrument using 42 positive, 9 equivocal and 39 negative samples. The samples were analyzed in single determination on one Phadia 250 and one Phadia 2500E instrument each. A Passing-Bablok regression analysis was performed and demonstrated a slope of 1.01 (0.97 - 1.04 95% CI) and an intercept of 0.24 (0.15 - 0.48 95% CI). The PPA, NPA, and total agreement were calculated with equivocal results considered as negative or as positive, respectively. The results are summarized in the table below:

|  Percent Ro60 Agreement | Agreement (%), (95% CI) | Agreement (%), (95% CI)  |
| --- | --- | --- |
|   |  Equivocal = Negative | Equivocal = Positive  |
|  PPA | 93.9% (85.2 – 98.3%) | 97.0% (89.5 – 99.6%)  |
|  NPA | 92.1% (78.6 – 98.3%) | 81.6% (65.7 – 92.3%)  |
|  Total | 93.3% (86.6 – 97.3%) | 91.3% (84.2 – 96.0%)  |

# C Clinical Studies:

# 1. Clinical Sensitivity and Specificity:

# EliA Ro52:

A total of 755 clinically and ethnically defined serum samples were used to determine the clinical sensitivity and specificity of the EliA Ro52 assay. Samples with a diagnosis of

K210902 - Page 16 of 23

{16}

systemic lupus erythematosus (SLE) (n=120), Sjögren's syndrome (SS) (n=60), idiopathic inflammatory myopathies (IIM) (n=94) and systemic sclerosis (SSc) (n=91) represent the "Target Disease". Samples with a diagnosis of various other autoimmune and infectious diseases are represented as the "Control Disease" group (n=390). The distribution of the diseases and the EliA Ro52 result for each clinical subgroup is summarized in the following table.

|   | N | EliA Ro52  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive n | Equivocal n | Negative n  |
|  Target Disease  |   |   |   |   |
|  Systemic lupus erythematosus (SLE) | 120 | 57 | 4 | 59  |
|  Primary Sjögren's syndrome | 60 | 30 | 3 | 27  |
|  Idiopathic inflammatory myopathies (IIM) | 94 | 34 | 1 | 59  |
|  IIM PM | 40 | 7 | 0 | 33  |
|  IIM DM | 13 | 3 | 1 | 9  |
|  IIM Myositis/CTD-overlap | 18 | 3 | 0 | 15  |
|  IIM Overlap / MCTD | 9 | 9 | 0 | 0  |
|  IIM Jo-1 Positive | 10 | 9 | 0 | 1  |
|  IIM Sporadic inclusion body myositis | 4 | 3 | 0 | 1  |
|  Systemic sclerosis (SSc) | 91 | 19 | 5 | 67  |
|  SSc, diffuse | 56 | 11 | 3 | 42  |
|  SSc, limited | 27 | 8 | 2 | 17  |
|  SSc, various* | 8 | 0 | 0 | 8  |
|  Target Disease (Total) | 365 | 140 | 13 | 212  |
|  |   |   |   |   |
|  Control Disease  |   |   |   |   |
|  Celiac disease | 13 | 0 | 0 | 13  |
|  Crohn's disease | 12 | 0 | 0 | 12  |
|  Ulcerative colitis | 35 | 0 | 1 | 34  |
|  CTD overlap Non-MCTD | 10 | 0 | 1 | 9  |
|  Graves' disease | 12 | 0 | 0 | 12  |
|  Primary antiphospholipid syndrome | 12 | 1 | 0 | 11  |
|  Primary biliary cholangitis | 25 | 1 | 2 | 22  |
|  Primary sclerosing cholangitis | 24 | 0 | 0 | 24  |
|  Type 1 Diabetes | 12 | 0 | 0 | 12  |
|  Lymphoma | 21 | 0 | 0 | 21  |
|  Leukemia | 20 | 0 | 0 | 20  |
|  Varied Cancer | 9 | 0 | 0 | 9  |
|  Mixed connective tissue disease | 10 | 2 | 0 | 8  |
|  Rheumatoid arthritis | 35 | 2 | 0 | 33  |
|  Bacterial infections [total] | 17 | 1 | 0 | 16  |
|  Bacterial infections (Mycoplasma) | 3 | 0 | 0 | 3  |
|  Bacterial infections (Borrelia) | 7 | 0 | 0 | 7  |

K210902 - Page 17 of 23

{17}

|   | N | EliA Ro52  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive n | Equivocal n | Negative n  |
|  Bacterial infections (var. Staphylococcus) | 3 | 1 | 0 | 2  |
|  Bacterial infections (various) | 4 | 0 | 0 | 4  |
|  Viral infections [total] | 80 | 4 | 1 | 75  |
|  Viral infections (Dengue) | 4 | 0 | 0 | 4  |
|  Viral infections (EBV) | 3 | 0 | 0 | 3  |
|  Viral infections (HBV) | 11 | 1 | 0 | 10  |
|  Viral infections (HCV) | 31 | 2 | 1 | 28  |
|  Viral infections (HIV) | 24 | 1 | 0 | 23  |
|  Viral infections (various) | 7 | 0 | 0 | 7  |
|  Hashimoto's disease | 10 | 1 | 0 | 9  |
|  Vasculitis [total] | 33 | 0 | 0 | 33  |
|  Granulomatosis with Polyangiitis | 10 | 0 | 0 | 10  |
|  Eosinophilic granulomatosis with polyangiitis | 4 | 0 | 0 | 4  |
|  Microscopic polyangiitis | 13 | 0 | 0 | 13  |
|  Polyarteritis nodosa | 3 | 0 | 0 | 3  |
|  Giant cell arteritis | 3 | 0 | 0 | 3  |
|  Control Disease (Total) | 390 | 12 | 5 | 373  |
|  |   |   |   |   |
|  TOTAL | 755 | 152 | 18 | 585  |

* The "SSc various" group is composed of the patients having SSc, classified according to the ACR/EULAR van den Hoogen 2013 criterion. The patients were not further subgrouped.

Clinical sensitivity and specificity for the EliA Ro52 as an aid in the diagnosis of SLE were shown in the following tables with excluded SS, SSc and IIM samples from the Target Disease group:

Equivocal = negative

|  EliA Ro52 | Diagnosis  |   |   |
| --- | --- | --- | --- |
|   |  SLE | Controls | Total  |
|  Positive | 57 | 12 | 69  |
|  Negative | 63 | 378 | 441  |
|  Total | 120 | 390 | 510  |

Equivocal = positive

|  EliA Ro52 | Diagnosis  |   |   |
| --- | --- | --- | --- |
|   |  SLE | Controls | Total  |
|  Positive | 61 | 17 | 78  |
|  Negative | 59 | 373 | 432  |
|  Total | 120 | 390 | 510  |
|   | Value | 95% CI  |
| --- | --- | --- |
|  Sensitivity | 47.5% | 38.3 – 56.8  |
|  Specificity | 96.9% | 94.7 – 98.4  |
|   | Value | 95% CI  |
| --- | --- | --- |
|  Sensitivity | 50.8% | 41.6 – 60.1  |
|  Specificity | 95.6% | 93.1 – 97.4  |

Clinical sensitivity and specificity for the EliA Ro52 as an aid in the diagnosis of SS were shown in the following tables with excluded SLE, SSc and IIM samples from the Target Disease group:

K210902 - Page 18 of 23

{18}

Equivocal = negative

|  EliA Ro52 | Diagnosis  |   |   |
| --- | --- | --- | --- |
|   |  SS | Controls | Total  |
|  Positive | 30 | 12 | 42  |
|  Negative | 30 | 378 | 408  |
|  Total | 60 | 390 | 450  |

Equivocal = positive

|  EliA Ro52 | Diagnosis  |   |   |
| --- | --- | --- | --- |
|   |  SS | Controls | Total  |
|  Positive | 33 | 17 | 50  |
|  Negative | 27 | 373 | 400  |
|  Total | 60 | 390 | 450  |
|   | Value | 95% CI  |
| --- | --- | --- |
|  Sensitivity | 50.0% | 36.8 – 63.2  |
|  Specificity | 96.9% | 94.7 – 98.4  |
|   | Value | 95% CI  |
| --- | --- | --- |
|  Sensitivity | 55.0% | 41.6 – 67.9  |
|  Specificity | 95.6% | 93.1 – 97.4  |

Clinical sensitivity and specificity for the EliA Ro52 as an aid in the diagnosis of IIM were shown in the following tables with excluded SLE, SS, and SSc samples from the Target Disease group:

Equivocal = negative

|  EliA Ro52 | Diagnosis  |   |   |
| --- | --- | --- | --- |
|   |  IIM | Controls | Total  |
|  Positive | 34 | 12 | 46  |
|  Negative | 60 | 378 | 438  |
|  Total | 94 | 390 | 484  |

Equivocal = positive

|  EliA Ro52 | Diagnosis  |   |   |
| --- | --- | --- | --- |
|   |  IIM | Controls | Total  |
|  Positive | 35 | 17 | 52  |
|  Negative | 59 | 373 | 432  |
|  Total | 94 | 390 | 484  |
|   | Value | 95% CI  |
| --- | --- | --- |
|  Sensitivity | 36.2% | 26.5 – 46.7  |
|  Specificity | 96.9% | 94.7 – 98.4  |
|   | Value | 95% CI  |
| --- | --- | --- |
|  Sensitivity | 37.2% | 27.5 – 47.8  |
|  Specificity | 95.6% | 93.1 – 97.4  |

Clinical sensitivity and specificity for the EliA Ro52 as an aid in the diagnosis of SSc were shown in the following tables with SLE, SS, and IIM samples excluded from the Target Disease group:

Equivocal = negative

|  EliA Ro52 | Diagnosis  |   |   |
| --- | --- | --- | --- |
|   |  SSc | Controls | Total  |
|  Positive | 19 | 12 | 31  |
|  Negative | 72 | 378 | 450  |
|  Total | 91 | 390 | 481  |

Equivocal = positive

|  EliA Ro52 | Diagnosis  |   |   |
| --- | --- | --- | --- |
|   |  SSc | Controls | Total  |
|  Positive | 24 | 17 | 41  |
|  Negative | 67 | 373 | 440  |
|  Total | 91 | 390 | 481  |
|   | Value | 95% CI  |
| --- | --- | --- |
|  Sensitivity | 20.9% | 13.1 – 30.7  |
|  Specificity | 96.9% | 94.7 – 98.4  |
|   | Value | 95% CI  |
| --- | --- | --- |
|  Sensitivity | 26.4% | 17.7 – 36.7  |
|  Specificity | 95.6% | 93.1 – 97.4  |

K210902 - Page 19 of 23

{19}

EliA Ro60:

A total of 713 clinically and ethnically defined serum samples were used to determine the clinical sensitivity and specificity of the EliA Ro52 assay. Samples with a diagnosis of systemic lupus erythematosus (SLE) (n=120) and Sjögren's syndrome (SS) (n=60) represent the "Target Disease". Samples with a diagnosis of various other autoimmune and infectious diseases are represented as the "Control Disease" group (n=390). The distribution of the diseases and the EliA Ro52 result for each clinical subgroup is summarized in the following table.

|   | N | EliA Ro60  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive n | Equivocal n | Negative n  |
|  Target Disease  |   |   |   |   |
|  Systemic lupus erythematosus (SLE) | 120 | 58 | 3 | 59  |
|  Primary Sjögren's syndrome | 60 | 41 | 2 | 17  |
|  Target Disease (Total) | 180 | 99 | 5 | 76  |
|  |   |   |   |   |
|  Control Disease  |   |   |   |   |
|  Idiopathic inflammatory myopathies (IIM) | 10 | 0 | 0 | 10  |
|  Systemic sclerosis (SSc) [total] | 91 | 21 | 0 | 70  |
|  SSc, diffuse | 56 | 11 | 0 | 45  |
|  SSc, limited | 27 | 10 | 0 | 17  |
|  SSc, various* | 8 | 0 | 0 | 8  |
|  Celiac disease | 13 | 0 | 0 | 13  |
|  Crohn's disease | 12 | 0 | 0 | 12  |
|  Ulcerative colitis | 35 | 0 | 1 | 34  |
|  CTD overlap Non-MCTD | 10 | 0 | 1 | 9  |
|  Graves' disease | 12 | 0 | 0 | 12  |
|  Primary antiphospholipid syndrome | 12 | 1 | 0 | 11  |
|  Primary biliary cholangitis | 25 | 1 | 2 | 22  |
|  Primary sclerosing cholangitis | 24 | 0 | 0 | 24  |
|  Type 1 Diabetes | 12 | 0 | 0 | 12  |
|  Lymphoma | 21 | 0 | 0 | 21  |
|  Leukemia | 20 | 0 | 0 | 20  |
|  Varied Cancer | 9 | 0 | 0 | 9  |
|  Mixed connective tissue disease | 10 | 2 | 0 | 8  |
|  Rheumatoid arthritis | 35 | 2 | 0 | 33  |
|  Bacterial infections [total] | 17 | 1 | 0 | 16  |
|  Bacterial infections (Mycoplasma) | 3 | 0 | 0 | 3  |
|  Bacterial infections (Borrelia) | 7 | 0 | 0 | 7  |
|  Bacterial infections (var. Staphylococcus) | 3 | 1 | 0 | 2  |
|  Bacterial infections (various) | 4 | 0 | 0 | 4  |

K210902 - Page 20 of 23

{20}

|   | N | EliA Ro60  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive n | Equivocal n | Negative n  |
|  Viral infections [total] | 80 | 4 | 1 | 75  |
|  Viral infections (Dengue) | 4 | 0 | 0 | 4  |
|  Viral infections (EBV) | 3 | 0 | 0 | 3  |
|  Viral infections (HBV) | 11 | 1 | 0 | 10  |
|  Viral infections (HCV) | 31 | 2 | 1 | 28  |
|  Viral infections (HIV) | 24 | 1 | 0 | 23  |
|  Viral infections (various) | 7 | 0 | 0 | 7  |
|  Hashimoto's disease | 10 | 1 | 0 | 9  |
|  Vasculitis [total] | 33 | 0 | 0 | 33  |
|  Granulomatosis with Polyangiitis | 10 | 0 | 0 | 10  |
|  Eosinophilic granulomatosis with polyangiitis | 4 | 0 | 0 | 4  |
|  Microscopic polyangiitis | 13 | 0 | 0 | 13  |
|  Polyarteritis nodosa | 3 | 0 | 0 | 3  |
|  Giant cell arteritis | 3 | 0 | 0 | 3  |
|  Control Disease (Total) | 533 | 27 | 1 | 505  |
|  |   |   |   |   |
|  TOTAL | 713 | 126 | 6 | 581  |

* The "SSc various" group is composed of the patients having SSc, classified according to the ACR/EULAR van den Hoogen 2013 criterion. The patients were not further subgrouped.

Clinical sensitivity and specificity for the EliA Ro60 as an aid in the diagnosis of SLE were shown in the following tables with excluded SS, SSc and IIM samples from the Target Disease group:

Equivocal = negative

|  EliA Ro60 | Diagnosis  |   |   |
| --- | --- | --- | --- |
|   |  SLE | Controls | Total  |
|  Positive | 58 | 6 | 64  |
|  Negative | 62 | 436 | 498  |
|  Total | 120 | 442 | 562  |

Equivocal = positive

|  EliA Ro60 | Diagnosis  |   |   |
| --- | --- | --- | --- |
|   |  SLE | Controls | Total  |
|  Positive | 61 | 7 | 68  |
|  Negative | 59 | 435 | 494  |
|  Total | 120 | 442 | 562  |
|   | Value | 95% CI  |
| --- | --- | --- |
|  Sensitivity | 48.3% | 39.1 – 57.6  |
|  Specificity | 98.6% | 97.1 – 99.5  |
|   | Value | 95% CI  |
| --- | --- | --- |
|  Sensitivity | 50.8% | 41.6 – 60.1  |
|  Specificity | 98.4% | 96.8 – 99.4  |

Clinical sensitivity and specificity for the EliA Ro52 as an aid in the diagnosis of SS were shown in the following tables with excluded SLE, SSc and IIM samples from the Target Disease group:

K210902 - Page 21 of 23

{21}

Equivocal = negative

|  EliA Ro60 | Diagnosis  |   |   |
| --- | --- | --- | --- |
|   |  SS | Controls | Total  |
|  Positive | 41 | 6 | 47  |
|  Negative | 19 | 436 | 455  |
|  Total | 60 | 442 | 502  |

Equivocal = positive

|  EliA Ro60 | Diagnosis  |   |   |
| --- | --- | --- | --- |
|   |  SS | Controls | Total  |
|  Positive | 43 | 7 | 50  |
|  Negative | 17 | 435 | 452  |
|  Total | 60 | 442 | 502  |
|   | Value | 95% CI  |
| --- | --- | --- |
|  Sensitivity | 68.3% | 55.0 – 79.7  |
|  Specificity | 98.6% | 97.1 – 99.5  |
|   | Value | 95% CI  |
| --- | --- | --- |
|  Sensitivity | 71.7% | 58.6 – 82.5  |
|  Specificity | 98.4% | 96.8 – 99.4  |

2. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):

Not applicable

D Clinical Cut-Off:

Same as assay cut-off.

E Expected Values/Reference Range:

The expected value in the normal population is negative. Antibody prevalence in autoimmune patients varies widely depending on disease area. The proportion of sera from a normal population found positive for the antinuclear antibodies covered by the EliA Ro52 and EliA Ro60 test is below 1%. Expected values may vary depending on the population tested. A panel of 400 serum samples from apparently healthy Caucasian blood donors were collected from the Biobank at Phadia GmbH (Germany). A supplemental set of 240 serum samples was added to the study to reflect the ethnic composition of the U.S. The 240 supplemental serum samples consisted of 80 African American, 80 Hispanic, and 80 Asian healthy blood donor specimens.

The distribution of samples and the test results are presented in the following tables:

|  Study Demographics | Female (N = 320) |   |   |   |   | Male (N=320)  |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Age  |   |   |   |   |   |   |   |   |   |
|   |  ≤30 | 31–40 | 41–50 | 51-60 | >60 | ≤30 | 31–40 | 41–50 | 51-60 | >60  |
|  Caucasian | 40 | 40 | 40 | 40 | 40 | 40 | 40 | 40 | 40 | 40  |
|  African American | 8 | 8 | 8 | 8 | 8 | 8 | 8 | 8 | 8 | 8  |
|  Hispanic | 8 | 8 | 8 | 8 | 8 | 8 | 8 | 8 | 8 | 8  |
|  Asian | 8 | 8 | 8 | 8 | 8 | 8 | 8 | 8 | 8 | 8  |
|  Total | 64 | 64 | 64 | 64 | 64 | 64 | 64 | 64 | 64 | 64  |

K210902 - Page 22 of 23

{22}

|   | Caucasian |   | African American |   | Hispanic |   | Asian  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Female | Male | Female | Male | Female | Male | Female | Male  |
|  N | 200 | 200 | 40 | 40 | 40 | 40 | 40 | 40  |
|  EliA Ro52 (U/mL)  |   |   |   |   |   |   |   |   |
|  Median | 1.15 | 1.10 | 1.31 | 1.1 | 0.8 | 0.8 | 0.8 | 0.7  |
|  95th Percentile | 1.89 | 1.85 | 1.89 | 1.5 | 1.70 | 1.3 | 2 | 1.52  |
|  EliA Ro60 (U/mL)  |   |   |   |   |   |   |   |   |
|  Median | 0.47 | 0.52 | 0.549 | 0.51 | 0.40 | 0.46 | 0.40 | 0.42  |
|  95th Percentile | 0.78 | 0.87 | 1.77 | 0.92 | 0.73 | 1.10 | 0.86 | 0.74  |

## VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

## IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K210902 - Page 23 of 23

---

**Source:** [https://fda.innolitics.com/device/K210902](https://fda.innolitics.com/device/K210902)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
