← Product Code [OWB](/productcode/OWB) · K210900

# Alphenix, INFX-8000C/B, INFX-8000C/S, V9.2 (K210900)

_Canon Medical Systems Corporation · OWB · Apr 12, 2021 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210900

## Device Facts

- **Applicant:** Canon Medical Systems Corporation
- **Product Code:** [OWB](/productcode/OWB.md)
- **Decision Date:** Apr 12, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1650
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for the heart, brain, abdomen and lower extremities.

## Device Story

Alphenix is a digital radiography/fluoroscopy system for interventional angiography. It utilizes a C-arm with an X-ray tube, beam limiter, receptor, and patient table. The system captures X-ray images for diagnostic and interventional procedures in the heart, brain, abdomen, and lower extremities. Operated by clinicians in interventional settings, it transforms X-ray inputs into real-time fluoroscopic or radiographic outputs. These outputs assist physicians in visualizing vasculature and guiding interventional tools, facilitating minimally invasive procedures. The system includes software for image processing, dose optimization, and system control. Modifications in V9.2 include a new Angio X-ray tube (DSRX-T7735GFS), updated system software, and usability enhancements. The device benefits patients by providing high-quality imaging for precise diagnostic and therapeutic interventions.

## Clinical Evidence

Bench testing only. No clinical data was required. Verification and validation testing included phantom testing to verify image metrics and conformity testing to IEC standards (e.g., IEC 60601-1, 60601-2-43, 60601-2-28). Testing demonstrated that system modifications result in performance equal to or better than the predicate device.

## Technological Characteristics

Digital radiography/fluoroscopy system with C-arm, X-ray tube, beam limiter, and FPD (TFP-1216C/A1). Energy source: Ionizing radiation. Connectivity: Networked. Software: V9.2, Windows 10. Standards: IEC 60601-1, IEC 60601-2-43, IEC 60601-2-28, IEC 62304, IEC 62366. Features: Pulsed fluoroscopy, DSA, ABC control, and one-shot radiography. Materials: Standard medical-grade components per ISO 13485.

## Regulatory Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Alphenix, INFX-8000C/B, V8.0 ([K181804](/device/K181804.md))

## Reference Devices

- Cardiac X-ray Tube, DSRX-T7634GFS ([K203551](/device/K203551.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

April 12, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square. The full name is "U.S. FOOD & DRUG ADMINISTRATION".

Canon Medical Systems Corporation % Ms. Janine Reyes Manager, Regulatory Affairs Canon Medical Systems USA, INC. 2441 Michelle Drive TUSTIN CA 92780

Re: K210900

Trade/Device Name: Alphenix, INFX-8000C/B, INFX-8000C/S, V9.2 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB Dated: March 23, 2021 Received: March 26, 2021

Dear Ms. Reyes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products mav instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael D. O'Hara For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

K210900

Device Name

Alphenix, INFX-8000C/B, INFX-8000C/S, V9.2

#### Indications for Use (Describe)

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for the heart, brain, abdomen and lower extremities.

| Type of Use (Select one or both, as applicable)                                                          |                                                                                                         |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">❌</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">▢</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Canon Medical Systems USA, Inc. The word "Canon" is written in red, with the rest of the text in black. The text is aligned to the left.

# Made For life

K210900

# 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92

# 1. CLASSIFICATION and DEVICE NAME

| Classification Name    | Image-Intensified Fluoroscopic X-ray System |
|------------------------|---------------------------------------------|
| Product Code           | OWB                                         |
| Regulation Number      | 21 CFR 892.1650                             |
| Regulatory Class       | Class II                                    |
| Trade Proprietary Name | Alphenix, INFX-8000C/B, INFX-8000C/S, V9.2  |

# 2. SUBMITTER'S NAME

Canon Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550

# 3. OFFICIAL CORRESPONDENT

Naofumi Watanabe Senior Manager, Regulatory Affairs and Vigilance

# 4. CONTACT PERSON, U.S. AGENT and ADDRESS

#### Contact Person

Janine F. Reyes Manager, Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 669-7853 Fax: (714) 730-1310 jfreyes@us.medical.canon

# 5. MANUFACTURING SITE

Canon Medical Systems Corporation (TMSC) 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan

- 6. ESTABLISHMENT REGISTRATION 9614698
- 7. DATE PREPARED March 23, 2021
- 8. TRADE NAME(S) Alphenix, INFX-8000C/B, INFX-8000C/S, V9.2

#### Official Correspondent/U.S. Agent

Paul Biggins Sr. Director, Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 669-7808 Fax: (714) 730-1310 pbiggins@us.medical.canon

{4}------------------------------------------------

# 9. DEVICE NAME

Interventional Fluoroscopic X-ray System

# 10. CLASSIFICATION PANEL

Radiology

# 11. DEVICE CLASSIFICATION

Class II (per 21 CFR 892.1650)

# 12. PRODUCT CODE / DESCRIPTION

Product Code: OWB - Image-Intensified Fluoroscopic X-ray System

# 13. PERFORMANCE STANDARD

This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard].

#### 14. PREDICATE DEVICE

Alphenix, INFX-8000C/B, V8.0 (K181804)

TABLE 1: Predicate Device

| Trade Proprietary Name | Alphenix, INFX-8000C/B, V8.0                |
|------------------------|---------------------------------------------|
| Marketed by            | Canon Medical Systems USA, Inc.             |
| 510(k) Number          | K181804                                     |
| Clearance Date         | August 2, 2018                              |
| Classification Name    | Image-Intensified Fluoroscopic X-ray System |
| Product Code           | OWB                                         |
| Regulation Number      | 21 CFR 892.1650                             |
| Regulatory Class       | Class II                                    |

#### 15. REASON FOR SUBMISSION

Modification of a cleared device

#### 16. SUBMISSION TYPE

Special 510(k)

# 17. DEVICE DESCRIPTION

The Alphenix, INFX-8000C/B, INFX-8000C/S, V9.2, is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm which is equipped with an X-ray tube, beam limiter and X-ray receptor, X-ray controller, computers with system and processing software, and a patient radiographic table.

#### 18. INDICATIONS FOR USE

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

{5}------------------------------------------------

# 19. SUMMARY OF CHANGE(S)

This submission is to report the following items have been changed:

#### Summary of Software Changes:

- V9.2 Software: System software changed from V8.0 to V9.2 for improved usability and . to support the new Angio X-ray tube, DSRX-T7735GFS.
- . Summary of other software changes since K181804:

#### V8.0 to V8.3

- o Pulsed fluoroscopy improvement
- o Usability/operability improvement
- o Service improvements (non-clinical)

#### V8.3 to V9.0

- o Operating system update from Windows 7 to Windows 10
- o AlphaCT IQ improvement
- o 2D IQ Improvement
- o Dose optimization for DSA and spot fluoroscopy
- o Usability/operability improvement
- o Service improvements (non-clinical)
- · Cybersecurity improvement

#### V9.0 to V9.1

- o ABC control is optimized for the Cardiac X-ray Tube, DSRX-T7634GFS
- o Changes related to the combination of the Cardiac X-ray Tube, DSRX-T7634GFS
- Changes related to the combination of the Hi-Def (aSi/CMOS) FPD, TFP-1216C/A1

#### V9.1 to V9.2

- o ABC control is optimized for the Angio X-ray Tube, DSRX-T7735GFS
- o Changes related to the combination of the Angio X-ray Tube, DSRX-T7735GFS
- o Multi gain for the Hi-Def (aSi/CMOS) FPD, TFP-1216C/A1
- o Usability/operability improvement
- o Functionality improvement
- o Service improvements (non-clinical)
- o Support of new dosimeter

#### Summary of Hardware Changes:

- . Angio X-ray Tube, DSRX-T7735GFS (subject modification)
This X-ray tube replaces the previous Angio X-ray tube, DSRX-T7345GFS. The changes to this tube are as follows:

- o A change in the Radiography small focal spot maximum mA from 180 mA to 250 mA.
- A change in the Pulsed Fluoroscopy small focal spot maximum mA from 50 mA to 120 mA.
- o A change in the Radiography 0.1s small focal spot maximum input from 17 kW to 28 kW.
- o A change in the Pulsed Fluoroscopy medium focal spot maximum input from 16 kW to 20 kW.
- A change in the Pulsed Fluoroscopy small focal spot maximum input from 5.5 kW to 14.4 kW.

{6}------------------------------------------------

#### . Summary of other hardware changes since K181804:

#### Cardiac X-ray Tube, DSRX-T7634GFS (K203551)

This X-ray tube replaces the previous X-ray tube, DSRX-T7444GDS. The changes to this tube are as follows:

- o A change in the instantaneous output at 10% power to 28.8 kW from the previous 22kW.
- o A change in the pulsed fluoroscopy maximum mA from 200 mA to 320 mA.

# X-ray Generator Modification

- o Increase in pulsed fluoroscopy peak power (kW) for cardiac X-ray tube DRSX-T7444GDS from 17 kW to 22 kW at duty factor of 10% or less.
- o X-ray generator modification to reduce the heat generation in the power circuit.
- o Change of the supplemental symbol for the DSRX-T7444GDS from RUB to RXB when used in conjunction with this X-ray generator modification.

# Usability/Operability Improvements

- Enabled one shot radiography at every acquisition program (DA, DSA, etc.) using footswitch.
- o Modified table side consoles are available.

# Combination with the Hi-Def (aSi/CMOS) FPD, TFP-1216C/A1

# 20. SAFETY

The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1 standards, its collateral standards and particular standards; IEC 60601-2-43 and IEC60601-2-28. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR §1020, that apply to this device, will be met and reported via product report.

# LIST OF APPLICABLE STANDARDS

- . ANSI AAMI ES 60601-1:2005/(R)2012+A1:2012
- IEC 60601-1-2:2014 .
- IEC 60601-1-3:2008 +A1:2013 ●
- IEC 60601-1-6:2010 +A1:2013
- . IEC 60601-2-28:2017
- . IEC 60601-2-43:2010 +A1:2017
- IEC 62304:2006 +A1:2015 .
- IEC 62366:2007 +A1:2014

# 21. TESTING

Risk analysis and verification/validation testing conducted through bench testing demonstrate that the established specifications for the device have been met. Testing included conformity testing to IEC standards and phantom testing was conducted to verify image metrics related to improvements and changes to the predicate device. Clinical images were deemed not necessary for the aforementioned improvements via design control and risk management activities.

This submission contains test data that demonstrates that the system modifications result in performance that is equal to or better than the predicate system. Testing of the modified system was conducted in accordance with the applicable standards published by the International Electromechanical Commission (IEC) for Medical Devices and XR Systems.

{7}------------------------------------------------

Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.

Additionally, the design controls used for this device included risk management and all known risks were mitigated to an acceptable level.

# 22. SUBSTANTIAL EQUIVALENCE

The Alphenix, INFX-8000C/B, INFX-8000C/S, V9.2, is substantially equivalent to the Alphenix, INFX-8000C/B, V8.0, (181804), marketed by Canon Medical Systems USA, Inc. The Alphenix, INFX-8000C/B, /S, V9.2, includes system software change from V8.0 to V9.2, new Angio X-ray tube, usability improvements and operability improvements. The basic system configuration, method of operation, base software and manufacturing process remain unchanged from the cleared device. There are no new indications for use or intended use of the device.

# 23. CONCLUSION

The Alphenix, INFX-8000C/B, INFX-8000C/S, V9.2, performs in a manner similar to and is intended for the same use as the predicate device, as indicated in product the labeling. The modifications incorporated do not change the indications for use or the intended use of the device. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate device. It is the contention of Canon Medical Systems Corporation that all new safety issues have been addressed in the design of this change and that adequate evidence of this is presented with this submission.

---

**Source:** [https://fda.innolitics.com/device/K210900](https://fda.innolitics.com/device/K210900)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
