K210871 · Pdo Max, Inc. · NEW · Dec 17, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210871
Device Name
PDO Max Suture with Dual Needle
Applicant
Pdo Max, Inc.
Product Code
NEW · General, Plastic Surgery
Decision Date
Dec 17, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4840
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PDO Max Suture with Dual Needle is indicated for use in soft tissue approximation where use of absorbable sutures is appropriate. The anatomical location(s) of use are on the skin for dermatological applications only. The suture is not intended for interior body cavity applications and the suture is not intended for lifting and supporting tissues.
Device Story
Synthetic polydioxanone (PDO) monofilament suture with bi-directional barbs and pre-attached straight needles at both ends; used for soft tissue approximation in dermatological procedures. Surgeon inserts needles into tissue, pulling suture until transition point reaches tissue; barbs embed in tissue to secure closure without surgical knots. Excess suture trimmed after placement. Device degrades/dissolves in tissue over time. Sterile, EO-gas sterilized. Intended for prescription use by clinicians.
Clinical Evidence
No clinical data. Evidence includes bench testing (biocompatibility, sterilization, shelf-life, USP suture/needle performance) and in vivo animal studies (Sprague-Dawley rat biodegradation/tensile strength) leveraged from reference device K190245. Barb holding strength testing in porcine tissue confirmed non-inferiority to the predicate.
Indicated for soft tissue approximation in dermatological applications. For use where absorbable sutures are appropriate. Not for interior body cavity use or tissue lifting/support.
Regulatory Classification
Identification
An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
PDO MAX, Inc. PDS Barbed Suture, PDO MAXX Threads (K190245)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 17, 2021
PDO MAX, Inc. % Mary Vater Regulatory Consultant Medical Device Academy, Inc. 345 Lincoln Hill Rd. Shrewsbury, Vermont 05738
Re: K210871
Trade/Device Name: PDO Max Suture with Dual Needle Regulation Number: 21 CFR 878.4840 Regulation Name: Absorbable Polydioxanone Surgical Suture Regulatory Class: Class II Product Code: NEW Dated: November 16, 2021 Received: November 16, 2021
Dear Ms. Vater:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Deborah Fellhauer RN, BSN Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K210871
Device Name PDO Max Suture with Dual Needle
#### Indications for Use (Describe)
The PDO Max Suture with Dual Needle is indicated for use in soft tissue approximation where use of absorbable sutures is appropriate. The anatomical location(s) of use are on the skin for dermatological applications only. The suture is not intended for interior body cavity applications and the suture is not intended for lifting and supporting tissues.
Type of Use (Select one or both, as applicable)
| <span style="font-size:12px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
This 510(k) summary of substantial equivalence information is submitted in accordance with the requirements of 21 CFR §807.92:
- l. SUBMITTER PDO Max, Inc. 4971 Bear Road Liverpool, New York 13088 Tel: +1.315.427.8229
Contact Person: Giovanna McCarthy Date Prepared: December 8th, 2021
| II. DEVICE | |
|------------------------------|---------------------------------------------|
| Name of Device: | PDO Max Suture with Dual Needle |
| Classification Name: | Suture, Surgical, Absorbable, Polydioxanone |
| Regulation: | 21 CFR §878.4840 |
| Regulatory Class: | Class II |
| Product Classification Code: | NEW |
= PREDICATE DEVICE Predicate Device Manufacturer: Predicate Device Trade Name:
Predicate Device 510(k):
Reference Device Manufacturer: Reference Device Trade Name: Reference Device 510(k):
Angiotech Quill™ PDO Knotless Tissue-Closure Device (Polydioxanone) K120827
PDO MAX, Inc. PDS Barbed Suture, PDO MAXX Threads K190245
#### DEVICE DESCRIPTION IV.
The synthetic barbed Sutures are made of polydioxanone (PDO) with needles attached to each end of the suture. The needles are straight and pre-loaded onto each end of the suture filament. The pigment for the violet dye is D&C Violet No.2. The sutures are inserted into the tissue using the needles. Each dyed (violet) suture has bi-directional barbs along the long axis of the suture monofilament. The suture is barbed into two sections and each section are in opposite direction. The device is available sterile after ethylene oxide (EO) gas sterilization and degrades or dissolves over time in tissue.
The Synthetic Absorbable suture with needle, approximate tissue without the need to tie surgical knots, by using the opposing barbs on the surface to embed in the tissues after the surgeon precisely places the suture within the tissues. The suture is available in 2-0, 0 and 1, suture sizes, which are the sizes identified in the currently recognized United States Pharmacopeia (USP). The needles are 50mm in length (each) and are reverse cutting type.
#### V. PRINCIPLES OF OPERATION
The PDO Max Suture with Dual Needle has the needles pre-attached to each end of the needle on one end is inserted into the desired tissue location and pushed through to exit the suture following behind. Suture is pulled through until the transition point (non-barbed, middle section) has
{5}------------------------------------------------
reached the tissue. Estimation of the center of the device can be aided by taking a single bite of tissue, then aligning the two needles until both ends of the suture are roughly equal in length. Techniques for soft-tissue approximation may vary depending on type of tissue and closure. Once soft tissue is
approximated, the surgeon ensures that the barbed section of the suture is present where the tissue approximation is desired. The excess suture length is trimmed and the needle and extra suture length are removed. Each suture has bi-directional barbs along the axis of the monofilament that allows the suture to embed in the tissue after the surgeon has placed it; therefore, there is no need to tie a surgical knot.
#### VI. INDICATIONS FOR USE
The PDO Max Suture with Dual Needle is indicated for use in soft tissue approximation where use of absorbable sutures is appropriate. The anatomical location(s) of use are on the skin for dermatological applications only. The suture is not intended for interior body cavity applications and the suture is not intended for lifting and supporting tissues.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVCE
The following characteristics were compared between the subject device and the predicate device in order to demonstrate substantial equivalence:
- . Indications for Use - The predicate and subject device have similar indications for use; both are indicated for use in soft tissue approximation where use of absorbable sutures is appropriate.
- Materials The predicate and subject device are both made of polydioxanone (PDO) sutures and ● medical grade stainless steel needles.
- Design The predicate and subject device are equivalent in design. They are both made of the ● same basic monofilament with similar barbs cut into the monofilament. Both the predicate and subject devices are supplied with attached needles on each end, and have bi-directional barbs.
- . Performance Testing - The predicate and subject device were subjected to performance testing under the NEW product code, including in vivo animal studies to support substantial equivalence in terms of performance.
| Key Feature | PDO Max Suture with Dual Needle | Predicate Device (K120827) |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Suture Material | Same as predicate | Polydioxanone (PDO) |
| Suture Color | Same as predicate | Violet |
| Suture Length | 200mm (USP 2-0, 0 and 1) | 200mm (USP 2-0)<br>205mm (USP 0)<br>215mm (USP 1)<br>Other USP diameter sizes<br>included in the range of 2 to 4-0. |
| Needle Material | Same as predicate | Stainless steel (medical grade) |
| Barb Design | Same as predicate | Bi-directional with transition<br>point in the middle |
| USP Sizes | Three available sizes (USP 2-0 to 1).<br>These sizes are encompassed within<br>the available sizes for the predicate<br>device. | Variety of sizes (USP 2 to 4-0) |
| Needle Length | 50mm (Same as predicate) | According to Quill's online<br>catalog, they offer a variety of<br>needle lengths and types.<br>However, the needles for sutures |
The table below provides comparison of key features of the subject and predicate devices.
{6}------------------------------------------------
| | | USP sizes 2-0, 0 and 1 are<br>available in 50mm lengths (as<br>indicated in the publicly<br>available online catalog). |
|-------------------|-------------------|------------------------------------------------------------------------------------------------------------------------|
| Needle Type | Same as predicate | Straight |
| Needle Attachment | Same as predicate | Needles are pre-attached to each<br>end of the suture. |
Moreover, the previously cleared PDO Max suture (K190245) is cited in this submission as a reference device. The subject device and the reference device are identical and manufacturing process. The design of the sutures is also similar, as they both have cut barbs and are bi-directional. The main difference in design between the reference and subject devices is the barb pattern. Nonetheless, PDO Max has proposed to leverage performance data from K190245 in order to support this submission. Although there are some differences in barb design between both devices, these do not impact any of the leveraged testing data.
# VII. PERFORMANCE DATA
The following performance testing has been conducted (or leveraged from K190245) to support determination of substantial equivalence of the subject device. This includes biocompatibility testing of the suture component, biocompatibility testing of the needle component, sterilization validation, shelf-life study, USP suture and needle performance testing (as required under the NEW product code for absorbable polydioxanone sutures) and in vivo animal testing.
#### Suture Biocompatibility Testing (leveraged from K190245)
- Cytotoxicity
- Acute Systemic Toxicity ●
- Material Mediated Pyrogenicity
- Intracutaneous Reactivity Test .
- Skin Sensitization Test ●
- Genotoxicity
- Implantation Test ●
## Needle Biocompatibility Testing
- Cytotoxicity ●
- Sensitization ●
- Intracutaneous Irritation
- Acute Systemic Toxicity ●
- . Material Mediated Pyrogenicity
## Sterilization Validation (leveraged from K190245)
- EO Sterilization Validation in accordance with ISO 11135:2014 ●
- ucm109897 Sterilization Guidance ●
#### Shelf-Life Study (leveraged from K190245)
- Accelerated Aging ●
#### Suture Performance Testing
- Suture Absorption ●
- Dimension Test ●
{7}------------------------------------------------
K210871
- Tensile Strength
## Needle Performance Testing
- Dimension Test
- . Flexural Rigidity Test
- Pierce Test .
- Corrosion resistance
- Needle Attachment Test
### Animal Testing (leveraged from K190245)
In Vivo Biodegradation in Sprague-Dawley Rat - Absorption and Tensile Strength over time
### Barb Holding Strength Testing
A barb holding strength test was conducted to compare the barb holding strength of the subject device against the predicate device. Suture samples were inserted in porcine tissue to represent skin tissue. Test results demonstrated that the barb holding strength of the subject device sutures is non-inferior to the predicate device.
### Clinical Testing
Clinical testing was not required to support substantial equivalence.
# VIII. CONCLUSIONS
The subject device has similar indications for use to the predicate device. Moreover, both devices also have similar technological characteristics such as barb design and materials of manufacture. Appropriate performance testing was conducted (or leveraged from K190245) to support determination of substantial equivalence of the subject device. The results of this testing demonstrates that the PDO Max Suture with Dual Needle is substantially equivalent in safety and performance to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.