← Product Code [DZE](/productcode/DZE) · K210855

# Straumann BLX Implant System (K210855)

_Institut Straumann AG · DZE · Jun 21, 2021 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210855

## Device Facts

- **Applicant:** Institut Straumann AG
- **Product Code:** [DZE](/productcode/DZE.md)
- **Decision Date:** Jun 21, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Straumann BLX Implants are suitable for endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially edentulous patients. BLX implants can placed with immediate function on single-tooth and multi-unit restorations when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are connected to the implants through the corresponding abutment components.

## Device Story

Straumann BLX Implant System consists of fully tapered endosseous dental implants; manufactured from Titanium-13 Zirconium alloy (Roxolid); finished with SLA or SLActive surface. Implants feature TorcFit conical internal connection with Torx-style engagement. Used by dental clinicians in clinical settings for oral rehabilitation; implants surgically placed in upper or lower jaw; prosthetic restorations connected via abutments to restore chewing function. Subject devices include 6mm length implants with 3.75mm and 4.0mm diameters. System provides functional and esthetic benefits for patients requiring tooth replacement. Device is MR Conditional.

## Clinical Evidence

Bench testing only. Dynamic fatigue testing conducted per FDA guidance in saline at 2Hz/37°C for 2 million cycles. Insertion torque testing performed across various bone classes. Biocompatibility evaluated per ISO 10993-1. Sterilization validated to SAL 10^-6 per ISO 11137. No clinical data presented.

## Technological Characteristics

Material: Titanium-13 Zirconium alloy (Roxolid). Surface: SLA or SLActive. Connection: TorcFit (conical). Dimensions: 6mm length; 3.75mm/4.0mm diameter. Design: Tapered body. Sterilization: Irradiation (SAL 10^-6). Connectivity: N/A. Software: N/A.

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- Straumann BLX Line Extension – Implants, SRAs, and Anatomic Abutments ([K181703](/device/K181703.md))

## Reference Devices

- Straumann TLX Implant System ([K200586](/device/K200586.md))
- Straumann Dental Implant System Roxolid SLA Implants ([K150938](/device/K150938.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and consists of the letters "FDA" and the words "U.S. Food & Drug Administration".

Institut Straumann AG % Jennifer Jackson Director, Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810

Re: K210855

Trade/Device Name: Straumann BLX Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: March 22, 2021 Received: March 23, 2021

Dear Jennifer Jackson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

K210855

Device Name

Straumann BLX Implant System

#### Indications for Use (Describe)

Straumann BLX Implants are suitable for endosted implantation in the upper and for the functional and esthetic oral rehabilitation of edentulous and partially edentulous patients. BLX implants can placed with immediate function on single-tooth and multi-unit restorations when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are connected to the implants through the corresponding abutment components.

| Type of Use (Select one or both, as applicable)                   |
|-------------------------------------------------------------------|
| <span>     Prescription Use (Part 21 CFR 801 Subpart D)   </span> |
| <span>     Over-The-Counter Use (21 CFR 801 Subpart C)   </span>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Straumann® BLX Implant System

510(k) Summary

# 510(k) Summary

# Submitter's Contact Information

| Submitter:          | Straumann USA, LLC (on behalf of Institut Straumann AG)<br>60 Minuteman Road<br>Andover, MA 01810<br>Registration No.: 1222315 Owner/Operator No.: 9005052 |  |  |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
|                     | On the behalf of:<br>Institut Straumann AG<br>Peter Merian weg 12<br>4052 Basel, Switzerland                                                               |  |  |
| Contact Person:     | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs<br>Phone Number: +1-978-747-2509<br>Fax Number: +1-978-747-0023                                  |  |  |
| Prepared By:        | Viviana Horhoiu<br>Regulatory Affairs and Compliance Manager<br>Institut Straumann AG<br>Phone number: +41 61 965 1260                                     |  |  |
| Date of Submission: | June 21, 2021                                                                                                                                              |  |  |

# Name of the Device

| Trade Names:           | Straumann® BLX Implant System |
|------------------------|-------------------------------|
| Common Name:           | Endosseous dental implant     |
| Classification Name:   | Endosseous dental implant     |
| Regulation Number:     | §872.3640                     |
| Device Classification: | II                            |
| Product Code(s):       | DZE                           |

{4}------------------------------------------------

Straumann® BLX Implant System

510(k) Summary

## Predicate Device(s)

Primary Predicate:

- . K181703 – Straumann BLX Line Extension – Implants, SRAs, and Anatomic Abutments (Institut Straumann AG)
Reference Devices:

- . K200586 – Straumann TLX Implant System (Institut Straumann AG)
- K150938 Straumann Dental Implant System Roxolid SLA Implants (Institut . Straumann AG)

## Device Description

The BLX Implants are fully tapered implants manufactured utilizing the Roxolid material and finished with the SLA or SLActive surface. The connection is identified as conical fitting with the Torx style engaging feature. The subject BLX implants have endosteal implant diameters of Ø3.75 and Ø4.0 mm with a length of 6 mm and are presented with one prosthetic platform:

- . RB (Regular Base)
The internal connection and the prosthetic platform are identical for all subject devices.

### Indications for Use

Straumann BLX Implants are suitable for endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially edentulous patients. BLX implants can placed with immediate function on single-tooth and multi-unit restorations when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are connected to the implants through the corresponding abutment components.

# Technological Characteristics

The technological characteristics of the subject devices are compared to the primary predicate and reference devices in Table 1.

{5}------------------------------------------------

### Straumann® BLX Implant System

## 510(k) Summary

| Feature                              | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Primary Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Primary Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                      | Subject<br>Straumann BLX Implant<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K181703<br>Straumann BLX Line<br>Extension - Implants, SRAs,<br>and Anatomic Abutments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | K200586<br>Straumann TLX Implant<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | K150938<br>Straumann Dental Implant<br>System - Roxolid SLA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Indications for<br>Use               | Straumann BLX Implants are<br>suitable for endosteal<br>implantation in the upper and<br>lower jaw and for the functional<br>and esthetic oral rehabilitation of<br>edentulous and partially<br>edentulous patients. BLX<br>implants can placed with<br>immediate function on single-<br>tooth and multi-unit restorations<br>when good primary stability is<br>achieved and with appropriate<br>occlusal loading, to restore<br>chewing function. The prosthetic<br>restorations used are connected<br>to the implants through the<br>corresponding abutment<br>components. | Straumann® BLX Implants are<br>suitable for endosteal<br>implantation in the upper and<br>lower jaw and for the functional<br>and esthetic oral rehabilitation of<br>edentulous and partially<br>edentulous patients. BLX<br>Implants can be placed with<br>immediate function on single-<br>tooth, bar and bridge<br>applications when good primary<br>stability is achieved and with<br>appropriate occlusal loading to<br>restore chewing function. The<br>prosthetic restorations are<br>connected to the implants<br>through the corresponding<br>abutment components. | Straumann TLX Implants are<br>suitable for endosteal<br>implantation in the upper and<br>lower jaws and for the functional<br>and esthetic oral rehabilitation of<br>edentulous and partially<br>edentulous patients. TLX<br>Implants can be placed with<br>immediate function on single-<br>tooth and multi-unit restorations<br>when good primary stability is<br>achieved and with appropriate<br>occlusal loading to restore<br>chewing function. The prosthetic<br>restorations are connected to<br>the implants through the<br>corresponding abutment<br>components. | Straumann® dental implants are<br>indicated for oral endosteal<br>implantation in the upper and<br>lower jaw and for the functional<br>and esthetic oral rehabilitation of<br>edentulous and partially dentate<br>patients. Straumann dental<br>implants can also be used for<br>immediate or early implantation<br>following extraction or loss of<br>natural teeth. Implants can be<br>placed with immediate function<br>on single-tooth and/or multiple<br>tooth applications when good<br>primary stability is achieved and<br>with appropriate occlusal<br>loading, to restore chewing<br>function. The prosthetic<br>restorations used are single<br>crowns, bridges and partial or<br>full dentures, which are<br>connected to the implants by the<br>corresponding elements<br>(abutments). |
| Material                             | Titanium-13 Zirconium alloy<br>(Roxolid®)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Titanium-13 Zirconium alloy<br>(Roxolid®)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Titanium-13 Zirconium alloy<br>(Roxolid®)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Titanium-13 Zirconium alloy<br>(Roxolid®)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Surface<br>Treatment                 | Hydrophilic SLActive® and SLA®                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Hydrophilic SLActive®                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Hydrophilic SLActive®                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | SLA®                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Implant to<br>Abutment<br>Connection | TorcFit<br>(with conical fitting)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | TorcFit<br>(with conical fitting)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | TorcFit<br>(with conical fitting)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | CrossFit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Implant<br>Diameter                  | Ø3.75 and 4.0 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Ø3.75 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Ø3.75, 4.0, 4.5, 5.0, 5.5, and 6.5<br>mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Ø 3.3, 4.1, and 4.8 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Feature                              | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Primary Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Primary Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                      | Subject<br>Straumann BLX Implant<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K181703<br>Straumann BLX Line<br>Extension - Implants, SRAs,<br>and Anatomic Abutments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | K200586<br>Straumann TLX Implant<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | K150938<br>Straumann Dental Implant<br>System - Roxolid SLA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Implant Length                       | 6 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 8, 10, 12, 14, 16, and 18 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Ø3.75, 4.0, 4.5, 5.0 mm:<br>6, 8, 10, 12, 14, 16 and 18 mm<br>Ø5.5 and 6.5 mm:<br>6, 8, 10, 12 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 6, 8, 10, 12, 14 and 16 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Implant Design                       | Tapered body                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Tapered body                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Tapered body                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Parallel and tapered body                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Prosthetic<br>platforms              | RB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | RB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | NT, RT, and WT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Sterilization<br>Method              | Irradiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Irradiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Irradiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Irradiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| MR Labeling                          | MR Conditional                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | MR Conditional                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

{6}------------------------------------------------

Straumann® BLX Implant System

510(k) Summary

Table 1 – Comparison of subject device versus predicate devices – BLX Implant System

{7}------------------------------------------------

Straumann® BLX Implant System

510(k) Summary

## Performance Testing

#### Bench Testing

Dynamic fatigue testing was conducted according to the FDA guidance document "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" and demonstrated the subject BLX Implants System is equivalent to the predicate and reference devices. The test was conducted in saline (2 Hz and 37°C) at 2 million cycles covering permanent restoration of the implant without failure.

Insertion tests were performed for the subject devices and it could be proven that there is an adequate insertion torque in different bone classes when the implant is inserted according to the surgical procedure.

Surface area comparison and pull-out testing were leveraged from K200586 as the endosseous surface area of the subject implants is identical to the referenced implants from K200586.

#### Biocompatibility Testing

Biological assessment has been performed according to ISO 10993-1:2009 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" and to the FDA Guidance document "Use of International Standard ISO 10993- 1, 'Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process', Guidance for Industry and Food and Drug Administration Staff, Document issued on: June 16, 2016" for each of the subject devices.

The subject device materials are identical to the predicate and reference device materials, therefore, no new issues regarding biocompatibility were raised.

#### Sterilization Validation and Packaqing

The sterilization process for the subject BLX Implant System as recommended in the labeling was validated to a sterility assurance level (SAL) of 10-6 in accordance with ISO 11137-1:2006, Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices, 2006-04-05. The validation method used was the over kill bioburden method in accordance with ISO 11137-

{8}------------------------------------------------

Straumann® BLX Implant System

510(k) Summary

2:2013, Sterilization of health care products – Radiation – Part 2: Establishing the sterilization dose. The packaging of the subject BLX Implant System is equivalent to the packaging of the predicate and reference device. The shelf life for devices provided sterile is 5 years. The devices will not be marketed as non-pyrogenic. Pyrogenicity information provided is based on FDA Guidance on "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submission for Devices Labeled as Sterile, issued on 21 January 2016." The method used to determine the device meets pyrogen limit specifications is LAL Endotoxin Analysis with testing limit of 20 EU/device, based on a blood contacting and implanted device.

# Conclusion

The documentation submitted in this premarket notification demonstrates the subject Straumann BLX Implant System is substantially equivalent to the primary predicate and reference devices.

---

**Source:** [https://fda.innolitics.com/device/K210855](https://fda.innolitics.com/device/K210855)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
