Portex BLUxl Extra Length Tracheostomy Tube, BLUxl Suctionaid Extra Length Tracheostomy Tube, BLUxl Extra Length Tracheostomy Inner Cannula
K210833 · Smiths Medical Asd, Inc. · JOH · Nov 26, 2021 · Anesthesiology
Device Facts
Record ID
K210833
Device Name
Portex BLUxl Extra Length Tracheostomy Tube, BLUxl Suctionaid Extra Length Tracheostomy Tube, BLUxl Extra Length Tracheostomy Inner Cannula
Applicant
Smiths Medical Asd, Inc.
Product Code
JOH · Anesthesiology
Decision Date
Nov 26, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Portex® BLUxl™ Extra Length Tracheostomy Tube is for the adult patient category that require an artificial airway due to trauma or medical condition. Maximum recommended period of use is 29 days. 200/710, 720 and 730/xxx are extra proximal length (y measurement) tracheostomy tubes and are indicated for use in patients with excessive skin surface to anterior tracheal wall distance as frequently seen in obesity. 200/715, 725 and 735/xxx are extra distal length (x measurement) tracheostomy tubes and are indicated for use in patients with normal skin surface to anterior tracheal wall distance and require the distal length of the tracheostomy tube to extend more caudally within the trachea.
Device Story
Extra-length tracheostomy tubes and inner cannulas designed for adult patients with specific anthropometric needs (obesity or need for distal tracheal extension). Tubes provide artificial airway for patients requiring prolonged mechanical ventilation, secretion management, or relief of upper airway obstruction. Suctionaid variants include a vacuum control valve for aspiration of subglottic secretions. Devices are used in hospital or home care settings. Healthcare providers insert the tube to maintain airway patency; the inner cannula is used for maintenance and secretion clearance. Output is a patent airway; clinical decision-making relies on patient respiratory status and anatomical requirements. Benefits include improved airway access for patients with challenging anatomy.
Clinical Evidence
No clinical data provided. Substantial equivalence determined via non-clinical bench testing, including dimensional requirements, kink resistance, tensile strength, cuff performance (inflation/deflation, leak, burst), insertion/removal cycling, and biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, genotoxicity, implantation).
Technological Characteristics
Materials: PVC (tube), PU (cuff), PTFE/LDPE (inner cannula), HDPE (obturator). Sensing/Actuation: Mechanical airway maintenance. Energy: None. Connectivity: None. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 11135:2014, ISO 11138-1:2017, ISO 11137-1:2018, ISO 11737-2:2020, ISO 10993-7:2008, ISO 10993-1:2009, ASTM F2096-11, ASTM F88-15.
Indications for Use
Indicated for airway maintenance in adult tracheostomized patients requiring artificial airways due to trauma or medical conditions, including prolonged mechanical ventilation, inability to manage secretions, or upper airway obstruction. Contraindications: None known.
Regulatory Classification
Identification
A tracheostomy tube and tube cuff is a device intended to be placed into a surgical opening of the trachea to facilitate ventilation to the lungs. The cuff may be a separate or integral part of the tracheostomy tube and is, when inflated, intended to establish a seal between the tracheal wall and the tracheostomy tube. The cuff is used to prevent the patient's aspiration of substances, such as blood or vomit, or to provide a means for positive-pressure ventilation of the patient. This device is made of either stainless steel or plastic.
Predicate Devices
Smiths Medical BLUselect® Tracheostomy Tube, BLUselect® Suctionaid® Tracheostomy Tube and BLUselect® Tracheostomy Tube Inner Cannula (K173384)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 26, 2021
Smiths Medical ASD, Inc. Nancy Deangelo Manager, Regulatory Affairs 6000 Nathan Lane North Minneapolis, Minnesota 55442
Re: K210833
Trade/Device Name: Portex BLUxl Extra Length Tracheostomy Tube, BLUxl Suctionaid Extra Length Tracheostomy Tube, BLUx1 Extra Length Tracheostomy Inner Cannula Regulation Number: 21 CFR 868.5800 Regulation Name: Tracheostomy Tube And Tube Cuff Regulatory Class: Class II Product Code: JOH Dated: July 30, 2021 Received: October 25, 2021
Dear Nancy Deangelo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210833
#### Device Name
Portex® BLUxI™ Extra Length Tracheostomy Inner Cannula
Indications for Use (Describe)
The Portex® BLUxI™ Extra Length Tracheostomy Inner Cannula is intended to be used with the Portex® BLUxI™ Extra Length Tracheostomy Tube indicated for airway maintenance of tracheostomy patients.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210833
#### Device Name
Portex® BLUxITM Extra Length Tracheostomy Tube
#### Indications for Use (Describe)
Portex® BLUxI™ Extra Length Tracheostomy Tube is indicated for airway maintenance of tracheostomised patients.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 5 510(k) Summary
| Sponsor: | Smiths Medical ASD, Inc.<br>6000 Nathan Lane North<br>Minneapolis, MN 55442<br>763.383.3000 |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | March 17, 2021 (revised November 22, 2021) |
| Proprietary Name: | Portex® BLUxl™ Extra Length Tracheostomy Tube, Portex® BLUxl™ Suctionaid® Extra Length Tracheostomy Tube, and Portex® BLUxl™ Extra Length Tracheostomy Inner Cannula |
| Classification Name: | JOH- Tube, Tracheostomy Tube and Tube Cuff |
| Classification: | Class: II<br>Regulation Number: 21 CFR 868.5800<br>Product Code: JOH - Tube, Tracheostomy Tube and Tube Cuff |
| Predicate device: | Smiths Medical BLUselect® Tracheostomy Tube, BLUselect® Suctionaid® Tracheostomy Tube and BLUselect® Tracheostomy Tube Inner Cannula (K173384, cleared on April 10, 2018). |
| Device Description: | Portex® BLUxl™ Extra Length Tracheostomy Tubes and BLUxl™ Suctionaid® Extra Length Tracheostomy Tubes are designed to aid the adult population who require an artificial airway due to trauma or medical condition. Patients who benefit from this procedure are those who: require prolonged intubation for mechanical ventilator support; cannot manage their airway secretions; or have an upper airway obstruction.<br>Portex® BLUxl™ Extra Length Tracheostomy Inner Cannula is intended to be used with the Portex® BLUxl™ Extra Length Tracheostomy Tubes and Portex® BLUxl™ Suctionaid® Extra Length Tracheostomy Tubes. The inner cannulas are single use only.<br>A detailed description of the Portex® BLUxl™ Extra Length Tracheostomy Tube, Portex® BLUxl™ Suctionaid® Extra Length Tracheostomy Tube, and Portex® BLUxl™ Extra Length Tracheostomy Inner Cannula is provided in Section 10 Device Description. |
| Intended Use: | Portex® BLUxl™ Extra Length Tracheostomy Tube is for the adult patient category that require an artificial airway due to trauma or medical condition. Maximum recommended period of use is 29 days.<br>200/710, 720 and 730/xxx are extra proximal length (y measurement) tracheostomy tubes and are indicated for use in patients with excessive skin surface to anterior tracheal wall distance as frequently seen in obesity.<br>200/715, 725 and 735/xxx are extra distal length (x measurement) tracheostomy tubes and are indicated for use in patients with normal skin surface to anterior tracheal wall distance and require the distal length of the tracheostomy tube to extend more caudally within the trachea. |
| Indications for use: | The Portex® BLUxl™ Extra Length Tracheostomy Inner Cannula is intended<br>to be used with the Portex® BLUxl™ Extra Length Tracheostomy Tube for<br>patients that require an artificial airway due to trauma or medical condition.<br>Maximum recommended period of use is 29 days.<br>Portex® BLUXI™ Extra Length Tracheostomy Tube is indicated for airway<br>maintenance of tracheostomised patients.<br><br>The Portex® BLUxl™ Extra Length Tracheostomy Inner Cannula is intended<br>to be used with the Portex® BLUxl™ Extra Length Tracheostomy Tube<br>indicated for airway maintenance of tracheostomy patients. |
| Non-clinical<br>Performance Data: | Non-clinical testing of the components comprised in each configuration of<br>the subject devices, Portex® BLUxl™ Extra Length Tracheostomy Tube,<br>Portex® BLUxl™ Suctionaid® Extra Length Tracheostomy Tube, and Portex®<br>BLUxl™ Extra Length Tracheostomy Inner Cannula were assessed and<br>tested appropriately to design controls, i.e. design verification, design<br>validations. The test results conclude that for Portex® BLUxl™ Extra Length<br>Tracheostomy Tube, Portex® BLUXIT™ Suctionaid® Extra Length<br>Tracheostomy Tube, and Portex® BLUxl™ Extra Length Tracheostomy Inner<br>Cannula are to be substantially equivalent to the predicate device described<br>herein. Testing listed below passed and were verified against their<br>requirements: Mechanical Testing Testing included: Dimensional requirements (such as<br>Tracheostomy tube outer diameter, Tube overall length, Tube<br>inner diameter at the patient end and 15mm connector taper<br>gauging) and mechanical requirements (such as Tube kink<br>resistance, Tensile strength of flange, Cuff inflation/deflation,<br>Cuff resting diameter, Cuff herniation, Cuff attachment, Cuff<br>burst, puncture and leak resistance, Inner Cannula<br>insertion/removal cycling, Obturator insertion and fallout and<br>Robustness of product markings). Artificially Aged Sample Testing Testing included: 6-month, 2-year, 3-year and 5-year artificially<br>aged samples tested to design requirements. Additionally, real<br>time aging will be carried out for the same time periods. Magnetic Resonance Imaging (MRI) Rationale provided utilizing legacy testing data. Cleaning Instruction Validation Cleaning validation of BLUxl™ Extra Length Tracheostomy<br>Inner Cannula. Cleaning validation of external surface of the BLUxI™ Extra<br>Length Tracheostomy Tube. Biocompatibility Testing evaluated: Cytotoxicity, Sensitization, Irritation, Acute<br>System Toxicity, Pyrogenicity, Subchronic Toxicity,<br>Genotoxicity and Implantation. |
| | Sterilization Validations, in accordance with ISO 11135:2014, ISO 11138-1:2017, ISO 11137-1:2018, ISO 11737-2:2020, and ISO 10993-7:2008 have been completed demonstrating results which indicate the subject devices are acceptable and can be distributed sterile to meet with the minimum sterility assurance level (SAL) of 10-6 per ISO 11135. Packaging Testing was performed to ASTM F2096-11 and ASTM F88-15, with passing results. |
| Biocompatibility | Contact classifications: Externally Communicating - Tissue, Surface Device – Mucosal Membrane, Surface Device – Intact Skin<br>Duration: up to 29 days, prolonged |
| Clinical Performance<br>Data: | Clinical testing was not necessary for the determination of substantial equivalence. |
| Substantial<br>Equivalence: | Smiths Medical considers the subject devices, Portex® BLUxl™ Extra Length Tracheostomy Tube, BLUxl™ Suctionaid® Extra Length Tracheostomy Tube, and BLUxl™ Extra Length Tracheostomy Inner Cannula to be substantially equivalent to the predicate device, Smiths Medical BLUselect® Tracheostomy Tube, BLUselect® Suctionaid Tracheostomy Tube and BLUselect® Inner Cannula because there are no significant differences between the devices in intended use, mechanical and functional performance and all utilize the same functional scientific technology. The predicate 510(k) Summary is provided in <b>Attachment 3</b> .<br><br>Smiths Medical has demonstrated there are no new issues of safety and effectiveness raised due to the similarities and/or differences between the subject and predicate/commercialized devices, as each are used to treat the same clinical conditions and represent a similar/basic design concept.<br><br>The table below provides a substantial equivalence summary, between the BLUselect devices and the predicate devices. |
{5}------------------------------------------------
{6}------------------------------------------------
## Substantial Equivalence – Subject vs. Predicate Comparison
| Comparator | Subject Device:<br>Portex® BLUxl™ Extra<br>Length Tracheostomy<br>Tube, Portex® BLUxl™<br>Suctionaid® Extra Length<br>Tracheostomy Tube, and<br>Portex® BLUxl™ Extra<br>Length Tracheostomy Inner<br>Cannula | Predicate Device:<br>BLUselect® Tracheostomy<br>Tube and BLUselect® with<br>Suctionaid Tracheostomy<br>Tube and BLUselect® Inner<br>Cannula | Substantial<br>Equivalence<br>Determination |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Company | Smiths Medical ASD, Inc. | Smiths Medical ASD, Inc. | Identical |
| 510(k) | K210833 | K173384, cleared on April<br>10, 2018 | N/A |
| Comparator | Subject Device:<br>Portex® BLUxI™ Extra<br>Length Tracheostomy<br>Tube, Portex® BLUxI™<br>Suctionaid® Extra Length<br>Tracheostomy Tube, and<br>Portex® BLUxI™ Extra<br>Length Tracheostomy Inner<br>Cannula | Predicate Device:<br>BLUselect® Tracheostomy<br>Tube and BLUselect® with<br>Suctionaid Tracheostomy<br>Tube and BLUselect® Inner<br>Cannula | Substantial<br>Equivalence<br>Determination |
| Image | Image: Portex® BLUxI™ Extra Length Tracheostomy Tube, Portex® BLUxI™ Suctionaid® Extra Length Tracheostomy Tube, and Portex® BLUxI™ Extra Length Tracheostomy Inner Cannula | Image: BLUselect® Tracheostomy Tube and BLUselect® with Suctionaid Tracheostomy Tube and BLUselect® Inner Cannula | N/A |
| Indications | Portex® BLUxI™ Extra<br>Length Tracheostomy Tube<br>is indicated for airway<br>maintenance of<br>tracheostomised patients.<br>The Portex® BLUxI™ Extra<br>Length Tracheostomy Inner<br>Cannula is intended to be<br>used with the Portex®<br>BLUxI™ Extra Length<br>Tracheostomy Tube<br>indicated for airway<br>maintenance of<br>tracheostomy patients. | Smiths Medical Portex®<br>BLUselect® Tracheostomy<br>Tube is indicated for airway<br>maintenance of<br>tracheostomised patients.<br>Smiths Medical Portex®<br>BLUselect® Suctionaid®<br>Tracheostomy Tube is<br>indicated for airway<br>maintenance of<br>tracheostomised patients.<br>Suctionaid® allows<br>aspiration of contaminated<br>mucous and subglottic<br>secretions that collect and<br>build up between the<br>tracheostomy tube cuff and<br>the glottis.<br>The BLUselect® Inner<br>Cannula is intended to be<br>used with the Smiths<br>Medical Portex®<br>BLUselect® Tracheostomy<br>Tube indicated for airway<br>maintenance of<br>tracheostomy patients. | Identical |
| Contraindications<br>in Labeling | There are no known<br>Contraindications<br>associated with this<br>product. | There are no known<br>Contraindications<br>associated with this<br>product. | Identical |
| Product Codes | JOH | JOH | Identical |
| Comparator | Subject Device:<br>Portex® BLUxl™ Extra<br>Length Tracheostomy<br>Tube, Portex® BLUxl™<br>Suctionaid® Extra Length<br>Tracheostomy Tube, and<br>Portex® BLUxl™ Extra<br>Length Tracheostomy Inner<br>Cannula | Predicate Device:<br>BLUselect® Tracheostomy<br>Tube and BLUselect® with<br>Suctionaid Tracheostomy<br>Tube and BLUselect® Inner<br>Cannula | Substantial<br>Equivalence<br>Determination |
| Intended Use | Smiths Medical Portex®<br>BLUxl™ Extra Length<br>Tracheostomy Tube is for<br>the adult patient category<br>that require an artificial<br>airway due to trauma or<br>medical condition.<br>Maximum recommended<br>period of use is 29 days.<br>The Portex® BLUxl™ Extra<br>Length Tracheostomy Inner<br>Cannula is intended to be<br>used with the Portex®<br>BLUxl™ Extra Length<br>Tracheostomy Tube for<br>patients that require an<br>artificial airway due to<br>trauma or medical<br>condition. Maximum<br>recommended period of use<br>is 29 days. | Smiths Medical Portex®<br>BLUselect® Tracheostomy<br>Tube is for patients that<br>require an artificial airway<br>due to trauma or medical<br>condition. Maximum period<br>of use is 29 days.<br>(BLUselect® and<br>BLUselect® Suctionaid®)<br>The BLUselect® Inner<br>Cannula is intended to be<br>used with the Smiths<br>Medical Portex®<br>BLUselect® Tracheostomy<br>Tube for patients that<br>require an artificial airway<br>due to trauma or medical<br>condition. Maximum period<br>of use is 29 days. | Identical |
| Single Patient Use | Yes | Yes | Identical |
| Patient Population | Intended for adults with<br>above average height,<br>weight, and<br>anthropometrics. | Intended for adults with<br>average height, weight, and<br>anthropometrics. | Similar |
| Use Environment | Hospital environments,<br>Home care use | Hospital environments,<br>Home care use | Identical |
| Materials | Tracheostomy Tube:<br>Polyvinyl Chloride<br>(PVC)with Dioctyl<br>Terephthalate (DEHT)<br>plasticizer<br>Tracheostomy Tube Cuff:<br>Polyurethane (PU)<br>Inner Cannula:<br>Polytetrafluoroethylene<br>(PTFE) and Low Density<br>Polyethylene (LDPE)<br>Obturator: High-density<br>Polyethylene (HDPE) | Tracheostomy Tube:<br>Polyvinyl Chloride<br>(PVC)with Dioctyl<br>Terephthalate (DEHT)<br>plasticizer<br>Tracheostomy Tube Cuff:<br>Polyvinyl Chloride (PVC)<br>Inner Cannula: Low<br>Density Polyethylene<br>(LDPE) or LDPE with red<br>colorant (Fenestrated)<br>Obturator: High-density<br>polyethylene (HDPE) | Similar |
| Comparator | Subject Device:<br>Portex® BLUxI™ Extra<br>Length Tracheostomy<br>Tube, Portex® BLUxI™<br>Suctionaid® Extra Length<br>Tracheostomy Tube, and<br>Portex® BLUxI™ Extra<br>Length Tracheostomy Inner<br>Cannula | Predicate Device:<br>BLUselect® Tracheostomy<br>Tube and BLUselect® with<br>Suctionaid Tracheostomy<br>Tube and BLUselect® Inner<br>Cannula | Substantial<br>Equivalence<br>Determination |
| Product<br>Configurations | Uncuffed, Cuffed,<br>Suctionaid® | Uncuffed, Fenestrated<br>Uncuffed, Cuffed,<br>Fenestrated Cuffed,<br>Suctionaid®, Fenestrated<br>Suctionaid® | Similar |
| Sizes (Inner<br>Diameter) | 6.0mm, 7.0mm, 8.0mm,<br>9.0mm | 6.0mm, 7.0mm, 7.5mm,<br>8.0mm, 8.5mm, 9.0mm,<br>9.5mm, 10.0mm | Similar |
| System Overview | | | |
| Standard System<br>Components | Tracheostomy Tube,<br>Obturator, Uncoupling<br>Wedge, Tube Holder, Inner<br>Cannula (x2), Patient<br>Labels, IFU | Tracheostomy Tube,<br>Obturator, Uncoupling<br>Wedge, Tube Holder with<br>Cleaning Brush, Inner<br>Cannula (x2), Patient<br>Labels, IFU | Similar |
| Suctionaid®<br>Standard System<br>Component | Vacuum Control Valve | Vacuum Control Valve | Identical |
| Flange Markings | Laser marked product<br>information with vinyl ink<br>pad printer color coding<br>band | Laser marked product<br>information with vinyl ink<br>pad printer color coding<br>band | Identical |
| Maximum Use | Recommended 29 days | Recommended 29 days | Identical |
| Biocompatibility | Materials compatible to<br>ISO 10993-1:2009 | Materials compatible to<br>ISO 10993-1:2009 | Identical |
| Sterilization | Ethylene Oxide (EO)<br>Sterilized – SAL 10-6 | Ethylene Oxide (EO)<br>Sterilized - SAL 10-6 | Identical |
| Shelf Life | Tracheostomy Tubes: 3-<br>year shelf life<br>Inner Cannula: 5-year shelf<br>life | Tracheostomy Tube and<br>Inner Cannula: 5-year shelf<br>life | Similar |
{7}------------------------------------------------
### K210833
Smiths Medical ASD, Inc.
Traditional 5 10(k) for Portex® BLUx1™ Extra Length Tracheostomy Tube
{8}------------------------------------------------
### K210833
{9}------------------------------------------------
#### K210833
Conclusion: The subject Portex® BLUxI™ Extra Length Tracheostomy Tubes, Portex® BLUxI™ Suctionaid® Extra Length Tracheostomy Tubes, and Portex® BLUxI™ Extra Length Tracheostomy Inner Cannulas has the same indications for use and technological characteristics compared to the predicate devices. The non-clinical performance data included in this submission supports that any differences in technological characteristics from the predicate device do not raise any new questions of safety and effectiveness. It is concluded that the information provided in this submission supports substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.