OnQ Neuro

K210831 · Cortechs Labs, Inc. · QIH · Nov 19, 2021 · Radiology

Device Facts

Record IDK210831
Device NameOnQ Neuro
ApplicantCortechs Labs, Inc.
Product CodeQIH · Radiology
Decision DateNov 19, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Brain tumor segmentation—Consistent with expert rater manual segmentation performancePassed product acceptance criteria——Independent test dataset>1 (expert rater)

Indications for Use

OnQ Neuro is a fully automated post-processing medical device software intended for analyzing and evaluating neurological MR image data. OnQ Neuro is intended to provide automatic segmentation, quantification, and reporting of derived image metrics. OnQ Neuro is additionally intended to provide automatic fusion of derived parametric maps with anatomical MRI data. OnQ Neuro is intended for use on brain tumors, which are known/confirmed to be pathologically diagnosed cancer. OnQ Neuro is intended for comparison of derived image metrics from multiple time-points. The physician retains the ultimate responsibility for making the final diagnosis and treatment decision.

Device Story

OnQ Neuro is a fully automated post-processing software for neurological MR images; used by radiologists and oncologists in hospitals, clinics, and imaging centers. It accepts DICOM inputs from MRI scanners or PACS; performs automatic segmentation and volumetric quantification of brain tumors; conducts post-acquisition analysis of diffusion-weighted imaging (DWI) data; generates parametric maps (ADC and Restricted Signal Map); and performs longitudinal comparison of metrics. The software operates as a standalone package installed in the cloud or on hospital IT infrastructure; it lacks a user interface, processing images automatically upon receipt. Outputs include numerical volumes, grayscale intensity maps, and graphical color overlays in DICOM format. Clinicians view these outputs on third-party workstations to complement case review, assisting in tumor evaluation and monitoring. The device benefits patients by providing standardized, automated quantitative analysis of tumor progression.

Clinical Evidence

Bench testing only. Performance evaluated by comparing software-derived segmentations to expert-labeled ground truth segmentations of pathologically confirmed brain tumors. Metrics included Dice similarity coefficient and squared correlation coefficient (R2). Results demonstrated accuracy consistent with expert rater performance across various scanner manufacturers, field strengths, and tumor types. Diffusion modeling performance validated using synthetic data with Rician noise and clinical validation testing.

Technological Characteristics

Standalone medical device software; installed in cloud or hospital server/workstation. Performs automated segmentation, volumetric quantification, and diffusion-weighted imaging analysis. Outputs DICOM-compliant reports and image series. No patient-contacting components. Software algorithm utilizes automated segmentation and diffusion modeling (single and multi-compartment).

Indications for Use

Indicated for adult patients with pathologically diagnosed brain tumors. Used for automatic segmentation, quantification, and reporting of MR image metrics, fusion of parametric maps with anatomical MRI, and longitudinal comparison of metrics.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health. November 19, 2021 CorTechs Labs, Inc. % Kora Marinkovic Director of Quality and Regulatory Affairs 5060 Shoreham Place. Suite 240 SAN DIEGO CA 92122 Re: K210831 Trade/Device Name: OnQ Neuro Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: October 11, 2021 Received: October 15, 2021 Dear Kora Marinkovic: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices {2}------------------------------------------------ or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K210831 Device Name OnQ Neuro Indications for Use (Describe) OnQ Neuro is a fully automated post-processing medical device software intended for analyzing and evaluating neurological MR image data. - OnQ Neuro is intended to provide automatic segmentation, quantification, and reporting of derived image metrics. OnQ Neuro is additionally intended to provide automatic fusion of derived parametric maps with anatomical MRI data. OnQ Neuro is intended for use on brain tumors, which are known/confirmed to be pathologically diagnosed cancer. OnO Neuro is intended for comparison of derived image metrics from multiple time-points. The physician retains the ultimate responsibility for making the final diagnosis and treatment decision. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo with a stylized brain graphic above the word "cortechs.ai" in a small, sans-serif font. The brain graphic is rendered in a light teal color and appears to be composed of interconnected lines, giving it a modern, tech-oriented feel. The logo is simple and clean, suggesting a focus on technology related to the brain or artificial intelligence. # 510(k) Summary: OnQ Neuro # 1. Submitter | Name | CorTechs Labs, Inc. | |------------------|----------------------------------------------------| | Address | 5060 Shoreham Place, Suite 240 San Diego, CA 92122 | | Contact Person | Kora Marinkovic | | Telephone Number | (858) 459-9703 | | Fax Number | (858) 459-9705 | | E-mail | koram@cortechslabs.com | | Date Prepared | 11/19/2021 | ### 2. Device | Device Trade Name | OnQ Neuro | |------------------------|------------------------------------------------| | Common Name | Medical Image Processing Software | | Classification Name | System, Image Processing, Radiological | | Regulation Number | 21 CFR 892.2050 | | Regulation Description | Medical image management and processing system | | Product Code | QIH | | Classification Panel | Radiology | # 3. Predicate Devices #### Primary Predicate Device | Device | Multi-Modality Tumor Tracking (MMTT) | |---------------|--------------------------------------| | 510(k) Number | K162955 | | Manufacturer | Philips Medical Systems | | Product Code | LLZ | {5}------------------------------------------------ | Image: cortechsol logo | 510(k) Section/Document Number | 5 | |------------------------|--------------------------------|------------------| | Title | OnQ Neuro: 510(k) Summary | | | Revision: 03 | Pages 2 of 6 | Date: 11/19/2021 | # 4. Device Description OnQ Neuro is a fully automated post-processing medical device software that is used by radiologists, oncologists, and other clinicians to assist with analysis and interpretation of neurological MR images. It accepts DICOM images using supported protocols and performs 1) automatic segmentation and volumetric quantification of brain tumors, which are known/confirmed to be pathologically diagnosed cancer, 2) automatic post-acquisition analysis of diffusion-weighted magnetic resonance imaging (DWI) data and optional automated fusion of derived image data with anatomical MR images, and 3) comparison of derived image metrics from multiple time-points. Output of the software provides values as numerical volumes, and images of derived data as grayscale intensity maps and as graphical color overlays on top of the anatomical image. OnQ Neuro output is provided in standard DICOM format as image series and reports that can be displayed on most third-party commercial DICOM workstations. The OnQ Neuro is a stand-alone medical device software package that is designed to be installed in the cloud or within a hospital's IT infrastructure on a server or PC-based workstation. Once installed and configured, the OnQ Neuro software automatically processes images sent from the originating system (MRI scanner or PACS). The software is configured at installation to receive input DICOM files from a network location, and output DICOM to a network destination. The software is designed without the need for a user interface after installation. Any processing errors are reported either in the output series error report, or system log files. OnQ Neuro software is intended to be used by trained personnel only and is to be installed by trained technical personnel. Quantitative reports and derived image data sets are intended to be used as complementary information in the review of a case. The OnQ Neuro software does not have any accessories or patient contacting components. The OnQ application is intended to be used for the adult population only. # 5. Indications for Use OnQ Neuro is a fully automated post-processing medical device software intended for analyzing and evaluating neurological MR image data. OnQ Neuro is intended to provide automatic segmentation, quantification, and reporting of derived image metrics. OnQ Neuro is additionally intended to provide automatic fusion of derived parametric maps with anatomical MRI data. OnQ Neuro is intended for use on brain tumors, which are known/confirmed to be pathologically diagnosed cancer. OnQ Neuro is intended for comparison of derived image metrics from multiple time-points. The physician retains the ultimate responsibility for making the final diagnosis and treatment decision. {6}------------------------------------------------ Image /page/5/Picture/9 description: The image shows a logo for a company called "cortechs.ai". The logo features a stylized image of a brain above the company name. The brain is drawn with thin, interconnected lines, giving it a modern and technological feel. The company name is written in a lowercase, sans-serif font. | 510(k) Section/Document Number | 5 | | |--------------------------------|---------------------------|------------------| | Title | OnQ Neuro: 510(k) Summary | | | Revision: 03 | Pages 3 of 6 | Date: 11/19/2021 | # 6. Predicate Device Philips Medical Systems' Multi-Modality Tumor Tracking (MMTT, K162955) market-cleared device is identified as the primary predicate device for the OnQ Neuro application. | | 510(k) Number | Product Name | Submitter | |-------------------|---------------|--------------------------------------|-------------------------| | Primary Predicate | K162955 | Multi-Modality Tumor Tracking (MMTT) | Philips Medical Systems | The proposed OnQ Neuro application and its predicate device, MMTT (K162955), are substantially equivalent in regard to their general intended uses, intended users, clinical indications, and principle of operation. They have similar basic design and features. # 7. Comparison to Predicate Device #### Summary Comparison Table for the device (OnQ Neuro) and the predicate device (MMTT) | Feature | OnQ Neuro | MMTT: K162955 (Predicate) | |---------------------------------------------------------------|------------------------------------------------------------------------------------|-----------------------------------------------| | Device Classification Name | System, Image processing,<br>Radiological | System, Image processing,<br>Radiological | | Device Class | Class II | Class II | | Classification Panel | Radiology | Radiology | | Product Code | LLZ | LLZ | | Regulation Description | Picture archiving and<br>communication system | Picture archiving and<br>communication system | | Regulation number | 21 CFR 892.2050 | 21 CFR 892.2050 | | Intended Use | MR imaging data post processing<br>software | MR imaging data post processing<br>software | | Intended users | Radiologists, oncologists | Radiologists, oncologists | | Type of Imaging Scan | MRI | CT, MR PET/CT and SPECT/CT | | Intended Body part | Neurological images | All body images | | Support Longitudinal Analysis | Yes - performs comparison of<br>derived image metrics from<br>multiple time points | Yes | | Support 2D and 3D anatomical<br>sequences | Yes | Yes | | Automatic image registration<br>between series within a study | Yes | Yes | | Segmentation editing tools | No | Yes | {7}------------------------------------------------ | Image: cortechsol logo | 510(k) Section/Document Number<br>5 | |------------------------|-------------------------------------| | Title | OnQ Neuro: 510(k) Summary | | Revision: 03 | Pages 4 of 6 Date: 11/19/2021 | | Tumor Segmentation Type | Fully-automated | Semi-Automated and manual | |-------------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | Findings management of<br>identified tumors | Yes. Prior segmentation results<br>are provided numerically. | Yes. | | Quantitative Analysis of<br>Regions-of-Interest | Yes. Including ROI Volumes, and<br>histogram statistics of optional<br>quantitative maps. | Yes. Including Volume,<br>Min/Max/Mean | | Reporting | Results displayed in tabular and<br>graphical formats. | Results displayed in tabular and<br>graphical formats. | | DICOM Communication | Yes | Yes | | Diffusion Analysis | Yes, including single and multi-<br>compartment diffusion models. | No | | Image Fusion | Automated fusion of segmentation<br>results and parametric maps. | Automated fusion of<br>segmentation results. | | Safety | Display/measurement data can be<br>viewed, accepted, or rejected by a<br>physician. | Display/measurement data can<br>be viewed, accepted, or rejected<br>by a physician. | | Environment for use | Hospital, Clinic, Imaging Center,<br>Medical Offices | Hospital, Clinic, Medical Offices | #### Description of Similarities and Differences: OnQ Neuro and MMTT (Predicate) Both applications, the OnQ Neuro and the MMTT (Predicate), 1) are post-processing software applications for analysis of MR imaging data; 2) have the ability to perform volumetric quantification of MR imaging data; 3) offer the ability to compare medical images and/or multiple time-points; 4) enable visualization of information that would otherwise have to be visually compared disjointedly; 5) have the ability to report derived imaging metrics; and 6) are intended for use on brain tumors, which are known/confirmed to be pathologically diagnosed cancer. They differ in that: 1) The Intended Use of the MMTT application includes language regarding viewing and manipulation of images. Specifically, MMTT provides editing tools for manual and semi-manual tumor annotation and allows loading of multiple concurrent studies for temporal measurements. OnQ Neuro does not include a GUI, so it does not include the ability to manually annotate and manipulate images; 2) While MMTT utilizes a manual, user-dependent feature to outline regions of interest, the OnQ Neuro software provides similar segmentation functionality automatically (i.e., without user intervention); 3) The OnQ Neuro software performs automated analysis of diffusion-weighted images and returns an Apparent Diffusion Coefficient (ADC) map and an enhanced DWI map (the Restricted Signal Map) showing the restricted component of the water diffusion signal, which the MMTT application does not include; 4) The MMTT application provides multi-modality support for CT, MR PET/CT and SPECT/CT scans, which includes support for SUV calculation of PET scans. OnQ Neuro supports only MR scans; and 5) While both applications perform measurements of the ROIs, the MMTT application allows for measurements that support oncology response criteria; OnQ Neuro does not specifically support {8}------------------------------------------------ | Image: cortechs logo | 510(k) Section/Document Number | 5 | |----------------------|--------------------------------|------------------| | Title | OnQ Neuro: 510(k) Summary | | | Revision: 03 | Pages 5 of 6 | Date: 11/19/2021 | oncology response criteria. The CorTechs' OnQ Neuro and the identified Philips Medical Systems Multi-Modality Tumor Tracking (MMTT, K162955) are substantially equivalent in terms of indication for use, intended users, principle of operation and safety and/or effectiveness. In conclusion, CorTechs believes that the proposed OnQ Neuro does not introduce any new potential safety and/or effectiveness issues and is substantially equivalent to the identified predicate device Multi-Modality Tumor Tracking (MMTT, K162955). # 8. Performance and V&V Testing Summary OnQ Neuro software application was tested in accordance with CorTechs verification and validation (V&V) processes. All product and engineering specifications were verified and validated. Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". Verification and Validation tests have been performed to address intended use, the technological characteristics claims, requirement specifications and the risk management results. The V&V and performance data were provided in support of safety and effectiveness for the substantial equivalence determination. #### Performance Testing Summary Performance testing included protocols demonstrating accuracy of automated segmentation compared to manual radiologist segmentations. OnQ Neuro software performs automatic postacquisition analysis of anatomical and diffusion-weighted MRI and coregistration, fusion, and quantification of supported regions. OnQ Neuro automatic segmentation performance is evaluated by comparing the software-derived segmentations to expert-labeled segmentations of brain tumors, which are known/confirmed to be pathologically diagnosed cancer. Comparisons to expert segmentations are quantified using the Dice similarity coefficient (extent of softwarederived vs. ground truth overlap), and squared correlation coefficient (R2) of segmented region of interest volumes. Acceptance criteria are set such that OnQ Neuro v1.1 model performance is consistent (95% percent performance) with expert rater manual segmentation performance and meets minimum clinically acceptable levels. The results of the segmentation performance testing on an independent test dataset demonstrate that OnQ Neuro v1.1.0 segments brain tumor ROIs with an accuracy that passed the product's acceptance criteria. Further, segmentation performance is consistent across scanner manufacturers, field strengths, tumor types, and patient sexes. We conclude that OnQ Neuro v1.1.0 brain tumor segmentation is sufficiently accurate to be used in clinical practice in accordance with the OnQ Neuro v1.1.0 Indications for Use. #### V&V Testing Summary OnQ Neuro tumor segmentation performance is additionally tested using three separate testing methods: 1) objective unit testing comparing the software-derived values to the known ground truth values, 2) system testing to verify that the Tumor Segmentation RGB Overlay and Tumor {9}------------------------------------------------ | Image: cortechsol logo | 510(k) Section/Document Number | 5 | |------------------------|--------------------------------|-------------------------------| | | Title | OnQ Neuro: 510(k) Summary | | | Revision: 03 | Pages 6 of 6 Date: 11/19/2021 | Segmentation Report are correctly generated when compatible anatomical images are input to OnQ Neuro, and 3) clinical validation testing that the Tumor Segmentation RGB Overlay and Tumor Segmentation Report are correct, meet clinical expectations, and are safe and effective. OnQ Neuro diffusion modeling performance is evaluated using three separate testing methods: 1) objective unit testing comparing the software-derived values using synthetic input data, with multiple levels of Rician noise, to the known ground truth values, 2) system testing to verify that the Restricted Signal Map and ADC map are correctly generated when compatible diffusion images are input to OnQ Neuro, and 3) clinical validation testing that the Restricted Signal Map and ADC map are correct, meet clinical expectations, and are safe and effective. The test results in this 510(k) premarket notification demonstrate that the OnQ Neuro: 1) complies with the international and FDA-recognized consensus standards and FDA guidance documents, as listed on the CDRH Premarket Review Submission Cover Sheet Form, and 2) meets the acceptance criteria and is adequate for its intended use and specifications. V&V activities required to establish performance and functionality of OnQ Neuro were performed. Testing performed demonstrated the OnQ Neuro meets all defined functionality requirements and performance claims. ### 9. Conclusions The comparison between the two devices demonstrates that OnQ Neuro is as safe and as effective as its predicate, the Multi-Modality Tumor Tracking application (MMTT). Both OnQ Neuro and its predicate are substantially equivalent in regard to their general intended uses, intended users, and principle of operation. They have similar basic design, functions and technological characteristics. To address the technological differences the appropriate V&V and performance testing has been performed. The V&V testing demonstrates the safety and effectiveness of the device to meet its intended use and specifications. The performance testing shows that the device performs as well as gold standard - computer-aided expert manual segmentation. By virtue of the physical characteristics and intended use, OnQ Neuro is substantially equivalent to its identified predicate device and is as safe and effective as its predicate device without raising any new safety and/or effectiveness concerns.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...