True Tulip, True M.I.S.
K210816 · Innovative Surgical Designs, Inc. · NKB · May 18, 2021 · Orthopedic
Device Facts
| Record ID | K210816 |
| Device Name | True Tulip, True M.I.S. |
| Applicant | Innovative Surgical Designs, Inc. |
| Product Code | NKB · Orthopedic |
| Decision Date | May 18, 2021 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 888.3070 |
| Device Class | Class 2 |
| Attributes | Therapeutic, Pediatric |
Intended Use
True Tulip is intended for posterior, noncervical (T1-S1) pedicle fixation to provide immobilization in skeletally mature patients as an adjunct to fusion for the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, kyphosis, lordosis, spinal tumor, pseudoarthrosis, and failed previous fusion. When used for posterior non-cervical pediation in pediatic patients, the True Tulip is indicated as an adjunct to fusion in the treatment of progressive spinal deformities (i.e., scoliosis, or lordosis) including adolescent idiopathic scoliosis (AIS), neuromuscular scoliosis, and congenital scoliosis. Additionally, the True Tulip is intended to treat pediatric patients diagnosed with spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. Pediatric pedicle screw fixation is limited to a posterior approach. The True Tulip is intended to be used with autograft and/or allograft. True M.I.S. is intended for posterior, noncervical (T1-S1) pedicle fixation to provide immobilization in skeletally mature patients as an adjunct to fusion for the following acute and chronic instabilities or deformities of the thoracic. lumbar. and sacral spine: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, tracture or dislocation), spinal stenosis, scoliosis, kyphosis, spinal tumor, pseudoarthrosis, and failed previous fusion. When used for posterior non-cervical pediatic patients, the True M.I.S. is indicated as an adjunct to fusion in the treatment of progressive spinal deformities (i.e., scoliosis, or lordosis) including adolescent idionathic scoliosis (AIS), neuromuscular scoliosis. Additionally, the True M.I.S. is intended to treat pediatric patients diagnosed with spondylolysis, fracture caused by tumor and/or trauma. pseudarthrosis, and/or failed previous fusion. Pediatric pedicle screw fixation is limited to a posterior approach. The True M.I.S. is intended to be used with autograft and/or allograft.
Device Story
True Tulip and True M.I.S. are thoracolumbosacral pedicle screw systems; provide spinal immobilization; used as adjunct to fusion; indicated for acute/chronic spinal instabilities or deformities; used in thoracic, lumbar, and sacral spine (T1-S1); intended for both skeletally mature and pediatric populations; limited to posterior approach for pediatric patients; used with autograft/allograft; implanted by surgeons in clinical/OR settings; provides mechanical stabilization to facilitate bone fusion; benefits patients by correcting spinal alignment and stabilizing vertebral segments.
Clinical Evidence
No clinical data provided.
Technological Characteristics
Thoracolumbosacral pedicle screw system; posterior, non-cervical (T1-S1) fixation; mechanical stabilization; metallic components for spinal fusion; class II device.
Indications for Use
Indicated for skeletally mature and pediatric patients requiring posterior, non-cervical (T1-S1) pedicle fixation as an adjunct to fusion. Conditions include DDD, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, scoliosis, kyphosis, lordosis, spinal tumor, pseudarthrosis, and failed previous fusion. Pediatric use limited to posterior approach for progressive deformities (AIS, neuromuscular, congenital) and specific pathologies (spondylolysis, trauma/tumor fracture).
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Related Devices
- K220297 — True Tulip, True M.I.S. · Innovative Surgical Designs, Inc. · Sep 16, 2022
- K212627 — True Tulip, True M.I.S. · Innovative Surgical Designs, Inc. · Feb 18, 2022
- K201780 — True Tulip System, True M.I.S. System · Innovative Surgical Designs, Inc. · Jul 22, 2020
- K170710 — True Spinal Fixation System · Innovative Surgical Designs, Inc. · Jul 14, 2017
- K080026 — QUANTUM-MIS CANNULATED SYSTEM · Pioneer Surgical Technology · Jan 28, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the FDA acronym is in blue, followed by the full name of the agency, "U.S. Food & Drug Administration," in a smaller font size.
May 18, 2021
Innovative Surgical Designs, Inc. Jon Rinehart Regulatory 3903 S. Walnut St. Bloomington, Indiana 47401
Re: K210816
Trade/Device Name: True Tulip, True M.I.S. Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWP Dated: May 10, 2021 Received: May 11, 2021
Dear Jon Rinehart:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210816
Device Name True Tulip
#### Indications for Use (Describe)
True Tulip is intended for posterior, noncervical (T1-S1) pedicle fixation to provide immobilization in skeletally mature patients as an adjunct to fusion for the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, kyphosis, lordosis, spinal tumor, pseudoarthrosis, and failed previous fusion.
When used for posterior non-cervical pediation in pediatic patients, the True Tulip is indicated as an adjunct to fusion in the treatment of progressive spinal deformities (i.e., scoliosis, or lordosis) including adolescent idiopathic scoliosis (AIS), neuromuscular scoliosis, and congenital scoliosis. Additionally, the True Tulip is intended to treat pediatric patients diagnosed with spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. Pediatric pedicle screw fixation is limited to a posterior approach. The True Tulip is intended to be used with autograft and/or allograft.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------------------------------------------------------|
| <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
### DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210816
Device Name True M.I.S.
### Indications for Use (Describe)
True M.I.S. is intended for posterior, noncervical (T1-S1) pedicle fixation to provide immobilization in skeletally mature patients as an adjunct to fusion for the following acute and chronic instabilities or deformities of the thoracic. lumbar. and sacral spine: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, tracture or dislocation), spinal stenosis, scoliosis, kyphosis, spinal tumor, pseudoarthrosis, and failed previous fusion.
When used for posterior non-cervical pediatic patients, the True M.I.S. is indicated as an adjunct to fusion in the treatment of progressive spinal deformities (i.e., scoliosis, or lordosis) including adolescent idionathic scoliosis (AIS), neuromuscular scoliosis. Additionally, the True M.I.S. is intended to treat pediatric patients diagnosed with spondylolysis, fracture caused by tumor and/or trauma. pseudarthrosis, and/or failed previous fusion. Pediatric pedicle screw fixation is limited to a posterior approach. The True M.I.S. is intended to be used with autograft and/or allograft.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."