← Product Code [NXC](/productcode/NXC) · K210763

# Inman Digital Clear Aligners (K210763)

_Inman Orthodontic Laboratories, Inc. · NXC · Sep 29, 2021 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210763

## Device Facts

- **Applicant:** Inman Orthodontic Laboratories, Inc.
- **Product Code:** [NXC](/productcode/NXC.md)
- **Decision Date:** Sep 29, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5470
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Inman Digital Clear Aligners are indicated for the treatment of tooth malocclusions in patients with permanent dentition (i.e. all second molars). The Inman Digital Clear Aligners position teeth by way of continuous gentle force.

## Device Story

Inman Digital Clear Aligners are customized, removable, clear thermoplastic polyurethane orthodontic trays. Dental clinicians provide intraoral scans or physical impressions to Inman Orthodontic Lab. Lab technicians use 3Shape Ortho System software to plan tooth movements and design aligner series based on clinician prescriptions. Clinicians review and approve digital treatment plans. Lab produces aligners via press thermoforming on molds. Patients wear trays ~22 hours/day, changing sequentially to apply continuous gentle force for tooth repositioning. Used in clinical settings under dental supervision. Benefits include gradual, incremental correction of malocclusions.

## Clinical Evidence

No clinical data. Bench testing only. Mechanical properties (tensile strength, elongation, modulus) verified per ASTM D638. Biocompatibility testing performed per ISO 10993 (Ames assay, cytotoxicity, irritation, sensitization, subacute toxicity). Manufacturing process validated for design consistency.

## Technological Characteristics

Material: Thermoplastic polyurethane. Form factor: Customized clear removable aligner trays. Manufacturing: Press thermoforming. Software: 3Shape Ortho System for 3D model management, diagnosis, and treatment planning. Mechanical testing: ASTM D638. Biocompatibility: ISO 10993 compliant.

## Regulatory Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

## Predicate Devices

- ClearCorrect System ([K113618](/device/K113618.md))

## Reference Devices

- 3Shape OrthoSystem™ ([K152086](/device/K152086.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name in a sans-serif font. The logo is simple and professional, conveying the agency's authority and mission.

September 29, 2021

Inman Orthodontic Laboratories, Inc. % Patsy Trisler Regulatory Consultant Trisler Consulting 7949 Beaumont Green East Drive Indianapolis, Indiana 46250

Re: K210763

Trade/Device Name: Inman Digital Clear Aligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: August 30, 2021 Received: September 1, 2021

Dear Patsy Trisler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210763

Device Name

Inman Digital Clear Aligners

#### Indications for Use (Describe)

Inman Digital Clear Aligners are indicated for the treatment of tooth malocclusions in patients with permanent dentition (i.e. all second molars). The Inman Digital Clear Aligners position teeth by way of continuous gentle force.

| Type of Use (Select one or both, as applicable)                                                                 |
|-----------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">✓</span> Prescription Use (Part 21 CFR 801 Subpart D)  </span> </div> |
| <div> <span> <span style="font-size:16px">❍</span> Over-The-Counter Use (21 CFR 801 Subpart C)  </span> </div>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary – K210763

| Submitter Name:                                                           | Inman Orthodontic Laboratories, Inc.                                                                                                                                                                                                                                                                                                                                                                                           |
|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address:                                                        | 3953 NW 126th Avenue<br>Coral Springs, FL 33065                                                                                                                                                                                                                                                                                                                                                                                |
| Phone Number:                                                             | 800-289-0118                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Contact Person:                                                           | Donal P Inman, President                                                                                                                                                                                                                                                                                                                                                                                                       |
| Date Prepared:                                                            | September 21, 2021                                                                                                                                                                                                                                                                                                                                                                                                             |
| Device Trade Name:                                                        | Inman Digital Clear Aligners                                                                                                                                                                                                                                                                                                                                                                                                   |
| Common Name                                                               | Aligner, Sequential                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification Name<br>Number                                             | Orthodontic Plastic Bracket<br>21 CFR 872.5470                                                                                                                                                                                                                                                                                                                                                                                 |
| Product Code                                                              | NXC                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Regulatory Class                                                          | 2                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Primary Predicate Device:                                                 | K113618, ClearCorrect System, ClearCorrect, LLC                                                                                                                                                                                                                                                                                                                                                                                |
| Reference Device                                                          | K152086, 3Shape OrthoSystem™, 3Shape A/S                                                                                                                                                                                                                                                                                                                                                                                       |
| Indications for Use                                                       | Inman Digital Clear Aligners are indicated for the treatment of<br>tooth malocclusions in patients with permanent dentition (i.e. all<br>second molars). The Inman Digital Clear Aligners position teeth<br>by way of continuous gentle force.                                                                                                                                                                                 |
| Device Description,<br>and Summary of<br>Technological<br>Characteristics | The Inman Digital Clear Aligners consist of a series of dental-<br>clinician prescribed customized clear plastic removable<br>orthodontic aligner trays that are made from a clear, thin<br>thermoformed polyurethane.                                                                                                                                                                                                         |
|                                                                           | A dental clinician prescribes the aligners based on an evaluation<br>of the patient's teeth. Either intraoral scans or physical<br>impressions of the patient's teeth are provided to Inman<br>Orthodontic Lab by the dental clinician, along with the<br>prescription. The series of plastic aligner trays are designed<br>according to the prescription using standard dental software for<br>planning the tooth alignments. |
|                                                                           | The software used is 3Shape Ortho System (Reference device<br>K152086). It is used for managing 3D scanned orthodontic<br>models, orthodontic diagnosis by measuring, analyzing, inspecting<br>and visualizing 3D scanned orthodontic models, virtual planning of<br>orthodontic treatments by simulating tooth movements and design<br>of a variety of orthodontic appliances based on 3D scanned<br>orthodontic models.      |
|                                                                           | After the plan is developed by Inman Orthodontic Lab, the<br>prescribing dental clinician reviews and approves the model                                                                                                                                                                                                                                                                                                       |

{4}------------------------------------------------

scheme before the molds are produced. Following approval, the customized aligner travs are fabricated by Inman Orthodontic Lab by press thermoforming of the material on the molds using standard thermoforming equipment. The trays are provided to the dental clinician who provides them to the patient assuring fit and function during the entire treatment period. Mechanism of Action The aligners are designed to gradually move the patient's teeth incrementally, repositioning them from their original misaligned state to a more aligned state. This occurs through forces applied by the appliance to the teeth over time until final correction, according to the dental clinician's prescription, has been achieved. Each aliqner tray is worn by patients for approximately 22 hours per day for the prescribed period, until it is time to move to the next tray in the series. Performance Testing Laboratory Testing Test data were submitted to: ■ assure the mechanical properties of the aligner material meet specifications for up to 5 years shelf life [according to ASTM D638 standard's testing for: tensile strength (PSI), elongation (%), tensile@ yield (PSI), elongation @ yield (PSI) and tensile modulus (PSI)]; . assure the aligner material packaging retains the required moisture barrier properties; . validate the processes used for the design and manufacture of the customized aligners, to ensure consistency between the aligner's digital design and the manufactured aligners. All testing met the pre-determined acceptance criteria. Biocompatibility The thermoplastic polyurethane used for making the aligner series has been tested according to Good Laboratory Practices for its biocompatibility according to ISO 10993, as follows: Part 3 (Bacterial Mutagenicity - Ames Assay), Part 5 (Cytotoxicity Elution - MEM), Part 10 (Intracutaneous/Intradermal Reactivity), Part 10 (Oral Mucosa; Irritation), Part 10 (Maximization for Delaved-Type Hvpersensitivity). Part 11 (Subacute Intraperitoneal Toxicity) All testing showed that the material met the requirements of the test methods and is safe and biocompatible for the stated intended use.

{5}------------------------------------------------

Animal | Human Testing

Neither animal nor human testing are needed for this device because it is composed of the same materials, is designed similarly, and is manufactured by a similar method as the predicate device.

Comparison to There are no notable differences comparing the Inman Digital Predicate Device: Clear Aligners to the predicate ClearCorrect aligner.

- The intended use is the same. 트
- . The mechanism of action is similar.
- . The polyurethane material used to make the aligners is the same.
- . The method of manufacture for producing the customized clear aligners is similar.
- . The use of software for planning and manufacturing are similar.

Substantial Based on the documentation presented in the 510(k), as summarized above and illustrated in the following comparison Equivalence Conclusion table, it can be concluded that Inman Digital Clear Aligners medical device is substantially equivalent to the predicate device.

{6}------------------------------------------------

| Trade Name:                                                   | Proposed: Inman Digital Clear<br>Aligner                                                                                                                                                                                                                   | Predicate: ClearCorrect<br>System                                                                                                                                                                                                         |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                                                 | ТВС                                                                                                                                                                                                                                                        | K113618                                                                                                                                                                                                                                   |
| Manufacturer                                                  | Inman Orthodontic Laboratory,<br>Inc.                                                                                                                                                                                                                      | ClearCorrect, LLC                                                                                                                                                                                                                         |
| Classification #<br>and Name<br>Product Code<br>Class         | 21 CFR 872.5470<br>Orthodontic Plastic Bracket<br>NXC<br>2                                                                                                                                                                                                 | 21 CFR 872.5470<br>Orthodontic Plastic Bracket<br>NXC<br>2                                                                                                                                                                                |
| Indications for<br>Use                                        | Inman Digital Clear Aligners are<br>indicated for the treatment of<br>tooth malocclusions in patients<br>with permanent dentition (i.e. all<br>second molars). The Inman<br>Digital Clear Aligners position<br>teeth by way of continuous gentle<br>force. | The ClearCorrect System<br>indicated for the treatment of<br>tooth malocclusions in patients<br>with permanent dentition (i.e. all<br>second molars). The<br>ClearCorrect System positions<br>teeth by way of continuous gentle<br>force. |
| Mechanism of<br>Action                                        | Alignment of teeth by application<br>of continuous gentle force, by<br>sequential use of preformed<br>plastic trays.                                                                                                                                       | Alignment of teeth by application<br>of continuous gentle force, by<br>sequential use of preformed<br>plastic trays.                                                                                                                      |
| Description of<br>Use                                         | Each preformed plastic tray is<br>worn by the patient as prescribed<br>by the dental practitioner, usually<br>a few weeks prior to using the<br>next sequential aligner tray.                                                                              | Each preformed plastic tray is<br>worn by the patient as prescribed<br>by the dental practitioner, usually<br>a few weeks prior to using the<br>next sequential aligner tray.                                                             |
| Material                                                      | Thin thermoplastic polyurethane                                                                                                                                                                                                                            | Thin thermoplastic polyurethane                                                                                                                                                                                                           |
| Manufacturing<br>Method                                       | Thermoforming                                                                                                                                                                                                                                              | Thermoforming                                                                                                                                                                                                                             |
| Biocompatible                                                 | Yes, meets ISO 10993<br>requirements                                                                                                                                                                                                                       | Yes, meets ISO 10993<br>requirements                                                                                                                                                                                                      |
| Prescription Use                                              | Yes, Rx Only                                                                                                                                                                                                                                               | Yes, Rx Only                                                                                                                                                                                                                              |
| Software Used<br>for Treatment<br>Planning /<br>Manufacturing | Yes                                                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                       |
| Single Patient<br>Use                                         | Yes                                                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                       |
| Non-Sterile<br>Packaging                                      | Yes                                                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                       |

### Substantial Equivalence Comparison

---

**Source:** [https://fda.innolitics.com/device/K210763](https://fda.innolitics.com/device/K210763)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).