K210743 · Siemens Medical Solutions USA, Inc. · IYN · Apr 1, 2021 · Radiology
Device Facts
Record ID
K210743
Device Name
ACUSON Redwood Diagnostic Ultrasound System
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYN · Radiology
Decision Date
Apr 1, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The ACUSON Redwood ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus, and other pelvic structures), Adult, Pediatric and Neonatal Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The systems also provide the ability to measure anatomical structures for fetal, abdominal, pediatric, small organ, neonatal cephalic, cardiac (adult, pediatric and neonatal), trans-esophageal, transrectal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculo-skeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Device Story
Mobile, software-controlled diagnostic ultrasound system; transmits/receives ultrasound echo data; displays B-Mode, M-Mode, PW/CW Doppler, Color Doppler, Color M-Mode, Doppler Tissue, Amplitude Doppler, and Harmonic Imaging. Used in clinical settings by trained healthcare professionals. Provides anatomical measurements and calculation packages for clinical diagnosis. Clinicians view output on integrated 22-inch OLED monitor and 13.3-inch adjustable touch screen. Output assists in clinical decision-making by providing diagnostic information adjunctively with other medical data. Benefits include non-invasive visualization of internal structures and physiological signals.
Clinical Evidence
Bench testing only. No clinical studies were required as the device uses the same technology and principles as existing legally marketed devices.
Technological Characteristics
Mobile diagnostic ultrasound system; 1.0MHz–18MHz frequency range; 22-inch OLED display; 13.3-inch adjustable touch screen. Patient contact materials tested to ISO 10993-1. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, and IEC 62359. Connectivity includes DICOM and wireless. Software-controlled with integrated calculation packages.
Indications for Use
Indicated for fetal, abdominal, pediatric, small parts, OB/GYN, adult/pediatric/neonatal cardiac, pelvic, neonatal/adult cephalic, vascular, musculoskeletal, superficial musculoskeletal, and peripheral vascular imaging and measurements in clinical settings by trained professionals.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol of the Department of Health & Human Services. To the right of the symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of "ADMINISTRATION".
Siemens Medical Solutions USA, Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k SAINT PAUL, MN 55114
## Re: K210743
Trade/Device Name: ACUSON Redwood Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX, OIJ Dated: March 11, 2021 Received: March 12, 2021
## Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
April 1, 2021
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known)
K210743
Device Name
ACUSON Redwood Diagnostic Ultrasound System
#### Indications for Use (Describe)
The ACUSON Redwood ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus, and other pelvic structures), Adult, Pediatric and Neonatal Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The systems also provide the ability to measure anatomical structures for fetal, abdominal, pediatric, small organ, neonatal cephalic, cardiac (adult, pediatric and neonatal), trans-esophageal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculo-skeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Mode of operation: 2D-mode 2D-mode with Harmonics imaging Color flow Doppler Color (velocity) Power (energy) Doppler Tissue Imaging Pulsed Wave Doppler Pulsed Wave Doppler Tissue Imaging High Pulsed Repetition Frequency Pulsed Wave Doppler Continuous Wave Doppler Steerable Continuous Wave Doppler for phased array transducers Auxiliary Continuous Wave Doppler for pencil transducers M-mode M-mode with Harmonics imaging Anatomical M-mode Volume Imaging Combined Modes 2D-mode with color 2D-mode with power 2D/Doppler 2D/Doppler with color 2D/Doppler with power 2D/M-mode 2D/M-mode with color 2D/Anatomical M-mode Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| | 510(k) Summary | K210743 | | | | | | | | | | | |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|----------|------------------|---------------------------------------|-----------------------|-----------|----------------------------------|----------|--------|-------------------------|----------|--------|
| Date: | January 15, 2021 | | | | | | | | | | | | |
| <b>1. Sponsor:</b> | Siemens Medical Solutions USA, Inc.<br>Ultrasound Division<br>22010 South East 51st Street<br>Issaquah, Washington 98029 | | | | | | | | | | | | |
| Contact Person: | Sulgue Choi<br>Tel: (425) 281-9898 | | | | | | | | | | | | |
| <b>2. Device Name:</b> | ACUSON Redwood Diagnostic Ultrasound System | | | | | | | | | | | | |
| Common Name: | Diagnostic Ultrasound System with Accessories | | | | | | | | | | | | |
| Classification: | <table><tr><td>Regulatory Class:</td><td>II</td></tr><tr><td>Review Category:</td><td>Tier II</td></tr><tr><td>Classification Panel:</td><td>Radiology</td></tr></table> | Regulatory Class: | II | Review Category: | Tier II | Classification Panel: | Radiology | | | | | | |
| Regulatory Class: | II | | | | | | | | | | | | |
| Review Category: | Tier II | | | | | | | | | | | | |
| Classification Panel: | Radiology | | | | | | | | | | | | |
| | <table><tr><td>Ultrasonic Pulsed Doppler Imaging System</td><td>892.1550</td><td>90-IYN</td></tr><tr><td>Ultrasonic Pulsed Echo Imaging System</td><td>892.1560</td><td>90-IYO</td></tr><tr><td>Diagnostic Ultrasound Transducer</td><td>892.1570</td><td>90-ITX</td></tr><tr><td>Biopsy Needle Guide Kit</td><td>892.1560</td><td>90-OIJ</td></tr></table> | Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN | Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO | Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX | Biopsy Needle Guide Kit | 892.1560 | 90-OIJ |
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN | | | | | | | | | | | |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO | | | | | | | | | | | |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX | | | | | | | | | | | |
| Biopsy Needle Guide Kit | 892.1560 | 90-OIJ | | | | | | | | | | | |
| <b>Manufacturing Site:</b> | Siemens Healthineers, Ltd.<br>2 <sup>nd</sup> ~ 3 <sup>rd</sup> Floor, 143, Sunhwan-ro, Jungwon-gu, Seongnam-si,<br>Gyeonggi-do, Republic of Korea<br><br>Siemens Medical Solutions USA, Inc.<br>22010 South East 51st Street,<br>Issaquah, Washington 98029, UNITED STATES | | | | | | | | | | | | |
### 3. Legally Marketed Predicate Devices
The ACUSON Redwood Diagnostic Ultrasound System is a multi-purpose diagnostic ultrasound system with accessories and proprietary software, and is substantially equivalent to the company's own products, the ACUSON P200 (K191922) which is a primary predicated device, Sequoia (K201462) and the ACUSON S family (K172162).
{5}------------------------------------------------
### 4. Device Description
The ACUSON Redwood Diagnostic Ultrasound System is a multi-purpose mobile, software controlled, diagnostic ultrasound system with an on-screen display of thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to transmit and receive ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M Mode, Doppler Tissue Mode, Amplitude Doppler Mode, a combination of modes and Harmonic Imaging on a Display.
### 5. Intended Use/Indications for Use
The ACUSON Redwood ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus, and other pelvic structures), Adult, Pediatric and Neonatal Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The systems also provide the ability to measure anatomical structures for fetal, abdominal, pediatric, small organ, neonatal cephalic, cardiac (adult, pediatric and neonatal), trans-esophageal, transrectal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculo-skeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
#### Mode of operation
2D-mode 2D-mode with Harmonics imaging Color flow Doppler Color (velocity) Power (energy) Doppler Tissue Imaging Pulsed Wave Doppler Pulsed Wave Doppler Tissue Imaging High Pulsed Repetition Frequency Pulsed Wave Doppler Continuous Wave Doppler Steerable Continuous Wave Doppler for phased array transducers Auxiliary Continuous Wave Doppler for pencil transducers M-mode M-mode with Harmonics imaging Anatomical M-mode Volume Imaqinq Combined Modes 2D-mode with color 2D-mode with power 2D/Doppler 2D/Doppler with color 2D/Doppler with power 2D/M-mode 2D/M-mode with color 2D/Anatomical M-mode
{6}------------------------------------------------
### Ultrasound Division
### 6. Summary of Technological Characteristics
The modified ACUSON Redwood Ultrasound System is the company's own previously cleared ACUSON P200 (K191922), ACUSON Sequoia (K201462) and the ACUSON S family (K172162) with regard to both intended use and technological characteristics. Both the modified ultrasound system under this review and the predicate ultrasound systems function in the same manner as all diagnostic ultrasound systems and transducers.
The submission device differs from the predicated devices as following:
- . The modified ACUSON Redwood Ultrasound System includes the addition of the 9VE4 transducer previously cleared on the ACUSON Sequoia (K201462).
- . The modified ACUSON Redwood Ultrasound System includes the addition of the 18H6 transducer previously cleared on the ACUSON Sequoia (K201462).
- . The modified ACUSON Redwood Ultrasound System includes the addition of the 7VC2 transducer, which is substantially equivalent to the 7CF2 transducer previously cleared on the ACUSON S family (K172162).
- . The modified ACUSON Redwood Ultrasound System includes the addition of the 3D/4D feature, which is substantially equivalent to the 3D/4D feature previously cleared on the ACUSON Sequoia (K201462).
- The modified ACUSON Redwood Ultrasound System includes the addition of the Freehand 3D . feature, which is substantially equivalent to the 3-Scape real-time 3D Imaging option previously cleared on the ACUSON S family (K172162).
All other hardware and software features of the ACUSON Redwood Diagnostic Ultrasound device remain unchanged. The foundation of the ACUSON Redwood (this submission) is the ACUSON P200(K191922) with features and transducers integrated with the ACUSON P200(K191922) hardware and the ACUSON Redwood (this submission) reuse software developed for P200(K191922) mainly as well as 9VE4 transducer, 18H6 transducer and 3D/4D feature from Sequoia (K201462).
The submission device is substantially equivalent to the predicates with regards to both intended use and technological characteristics.
| Feature /<br>Characteristic | ACUSON Redwood<br>This Submission | ACUSON P200<br>K191922<br>Primary | ACUSON Sequoia<br>K201462<br>Reference | ACUSON S Family<br>K172162<br>Reference |
|-----------------------------|-----------------------------------|-----------------------------------|----------------------------------------|-----------------------------------------|
| | | Predicate device | Predicate device | Predicate device |
### Indications for Use:
| • Fetal | √ | √ | - | - |
|-------------|---|---|---|---|
| • Abdominal | √ | √ | - | - |
| • Pediatric | √ | √ | - | - |
{7}------------------------------------------------
# Siemens Medical Solutions USA, Inc.
## Ultrasound Division
# ACUSON Redwood Diagnostic Ultrasound System
# 510(k) Submission
| Feature /<br>Characteristic | | ACUSON Redwood<br>This Submission | ACUSON P200<br>K191922<br>Primary<br>Predicate device | ACUSON Sequoia<br>K201462<br>Reference<br>Predicate device | ACUSON S Family<br>K172162<br>Reference<br>Predicate device |
|-------------------------------------------|------------------------------------------|-------------------------------------------------------|------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|
| ■ | Small Organ | √ | √ | - | - |
| ■ | Cardiac | √ | √ | - | - |
| ■ | Neonatal Cephalic | | | - | - |
| ■ | Trans-esophageal | | | - | - |
| ■ | Transrectal | √ | √ | - | - |
| ■ | Transvaginal | √ | √ | - | - |
| ■ | Peripheral vessel | √ | √ | - | - |
| ■ | Musculo-skeletal<br>(conventional) | √ | √ | - | - |
| ■ | Musculo-skeletal<br>(superficial) | √ | √ | - | - |
| Frequencies<br>Supported: | | √ | √ | | |
| | | (1.0MHZ~18MHz) | (1.0MHZ~18MHz) | | |
| | Modes: | | | | |
| ■ | B | √ | √ | - | - |
| ■ | M | √ | √ | - | - |
| ■ | PWD (Pulsed Wave<br>Doppler) | √ | √ | - | - |
| ■ | CWD (Continuous<br>Wave Doppler) | √ | √ | - | - |
| ■ | PW DTI (Doppler<br>Tissue Image) | √ | √ | - | - |
| ■ | Color Doppler | √ | √ | - | - |
| ■ | Power Doppler | √ | √ | - | - |
| ■ | Combined (BMDC) | √ | √ | - | - |
| | | | ACUSON P200 | ACUSON Sequoia | ACUSON S Family |
| Feature /<br>Characteristic | | ACUSON Redwood | K191922 | K201462 | K172162 |
| | | This Submission | Primary<br>Predicate device | Reference<br>Predicate device | Reference<br>Predicate device |
| | Features: | | | | |
| ■ | Harmonics | √ | √ | - | - |
| ■ | Compounding | √ | √ | - | - |
| ■ | UltraArt Universal<br>Image processing | √ | √ | - | - |
| ■ | Tissue Equalization<br>(TEQ) | √ | √ | - | - |
| ■ | Biopsy | √ | √ | - | - |
| ■ | Clarify | √ | √ | - | - |
| ■ | Cardiac Imaging | √ | √ | - | - |
| ■ | Speed of Sound | √ | √ | - | - |
| ■ | Protocols | √ | √ | - | - |
| ■ | DICOM | √ | √ | - | - |
| ■ | eSie Calcs | √ | √ | - | - |
| ■ | Wireless | √ | √ | - | - |
| ■ | Virtual Touch Strain<br>Imaging | √ | √ | - | - |
| ■ | Virtual Touch Point<br>Shear Wave (VTIQ) | √ | √ | - | - |
| | Contrast Agent<br>Imaging | √ | √ | - | - |
| ■ | eSie OB | √ | √ | - | - |
| ■ | eSie Left Heart | √ | √ | - | - |
| ■ | eSie Measure | √ | √ | - | - |
| ■ | eSie Follicle | √ | √ | - | - |
| ■ | Panoramic Imaging | √ | √ | - | - |
| Feature /<br>Characteristic | ACUSON Redwood<br>This Submission | ACUSON P200<br>K191922<br>Primary<br>Predicate device | ACUSON Sequoia<br>K201462<br>Reference<br>Predicate device | ACUSON S Family<br>K172162<br>Reference<br>Predicate device | |
| Stress Echo | √ | √ | - | - | |
| Syngo Velocity<br>Vector Imaging<br>(VVI) | √ | √ | - | - | |
| eSieLink | √ | √ | - | - | |
| DICOM SR | √ | √ | - | - | |
| Wireless | √ | √ | - | - | |
| 3D/4D | √ | - | √ | - | |
| Freehand 3D<br>Imaging | √ | - | - | √ | |
| Monitor (OLED) | √ (22") | √ (22") | - | - | |
| Touch Screen:<br>adjustable display | √ (13.3") | √ (13.3") | - | - | |
| Output Display<br>Standard (Track 3) | √ | √ | - | - | |
| Patient Contact<br>Materials | Tested to ISO<br>10993-1 | Tested to ISO<br>10993-1 | - | - | |
| UL 60601-1 Certified | √ | √ | - | - | |
| Indications for Use | √ | √ | - | - | |
{8}------------------------------------------------
# Siemens Medical Solutions USA, Inc.
## Ultrasound Division
# ACUSON Redwood
# Diagnostic Ultrasound System
# 510(k) Submission
{9}------------------------------------------------
## Siemens Medical Solutions USA, Inc.
## Ultrasound Division
# ACUSON Redwood Diagnostic Ultrasound System
## 510(k) Submission
### 7. A brief discussion of nonclinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence
The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and have been found to conform to applicable medical device safety standards. The systems comply with the following voluntary standards:
- l IEC 62359: Edition 2.1 2017-09, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields / Combines IEC 62359 (2010-10) and AMD 1 (2017-09)
- " Safety and EMC Requirements for Medical Equipment
{10}------------------------------------------------
- AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 o (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005)
- IEC 60601-1:2005, Medical electrical equipment Part 1: General requirements for O basic safety and essential performance / This document and its separate amendments continue to be valid together with the consolidated version
- IEC 60601-1-2 Edition 4.0 2014-02, Medical electrical equipment Part 1-2: General O requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-2-18: Edition 3.0 2009-08, Medical electrical equipment Part 2-18: o Particular requirements for the basic safety and essential performance of endoscopic equipment
- IEC 60601-2-37 Edition 2.1 2015, Medical electrical equipment Part 2-37: Particular o requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- ISO 10993-1 Fifth edition 2018-08-01. Biological evaluation of medical devices Part 1: ■ Evaluation and testing within a risk management process / Corrected and reprinted in 2018-10
### 8. A summary discussion of the clinical tests submitted, referenced, or relied on for a determination of substantial equivalence.
Since the ACUSON Redwood Diagnostic Ultrasound System uses the same technology and principles as existing devices, clinical studies were not required to support substantial equivalence.
### 9. Summary
Intended uses and other key features are consistent with traditional clinical practice and FDA guidelines. The design and development process of the manufacturer conforms to 21 CFR 820 Quality System Requlation and ISO 13485:2016 quality system standards. The product is designed to conform to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound system has accumulated a long history of safe and effective performance. Therefore, it is the opinion of Siemens Medical Solutions USA, Inc. that the ACUSON Redwood system is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.