← Product Code [QEW](/productcode/QEW) · K210689

# ClotTriever Thrombectomy System (K210689)

_Inari Medical, Inc. · QEW · May 12, 2021 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210689

## Device Facts

- **Applicant:** Inari Medical, Inc.
- **Product Code:** [QEW](/productcode/QEW.md)
- **Decision Date:** May 12, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5150
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The ClotTriever Thrombectomy System is indicated for: - · The non-surgical removal of thrombi and emboli from blood vessels. - · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT).

## Device Story

ClotTriever Thrombectomy System is a single-use, sterile device for peripheral vascular thrombus removal. System includes a reinforced polymeric sheath with a self-expanding distal mesh funnel, a hemostatic valve, and a catheter with reinforced coaxial shafts, an expandable coring element, and a thrombus collection bag. The device is operated by a physician under fluoroscopic guidance. The catheter is advanced beyond the clot; a self-expanding nitinol wire net is deployed to core, separate, and entrap thrombus. The net is then drawn into the sheath funnel, collapsing the net and capturing the clot for removal. A 60 cc syringe facilitates aspiration and fluid infusion. The system now includes a pre-dilator (16 Fr or 19 Fr) to dilate puncture sites or catheter tracts prior to sheath insertion. The device benefits patients by providing a non-surgical method for clearing thrombi in the peripheral vasculature.

## Clinical Evidence

No clinical data was required for this 510(k) submission. Substantial equivalence was supported by non-clinical bench testing, including visual and dimensional inspection, guidewire compatibility, leak testing, simulated use, insertion/kink radius testing, tensile testing, and in vivo functional testing/radiopacity verification. All acceptance criteria were met.

## Technological Characteristics

System components: reinforced polymeric sheath, self-expanding nitinol wire net, reinforced polymeric coaxial catheter shafts. Pre-dilator material: LDPE/HDPE. Radiopaque distal tips. Single-use, sterile. Guidewire compatibility: 0.035".

## Regulatory Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

## Predicate Devices

- ClotTriever Thrombectomy System ([K210190](/device/K210190.md))

## Reference Devices

- Cook Incorporated Dilator Sets ([K183036](/device/K183036.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 12, 2021

Inari Medical, Inc. Larry Boucher Regulatory Affairs Manager 9 Parker. Suite 100 Irvine, California 92618

Re: K210689

Trade/Device Name: ClotTriever Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW Dated: March 15, 2021 Received: March 16, 2021

Dear Larry Boucher:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K210689

Device Name ClotTriever Thrombectomy System

Indications for Use (Describe) The ClotTriever Thrombectomy System is indicated for:

- · The non-surgical removal of thrombi and emboli from blood vessels.
- · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.

The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT).

|X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) SUMMARY

| Date prepared                            | March 15, 2021                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                                     | Inari Medical, Inc.<br>9 Parker, Suite 100<br>Irvine, CA 92618<br>949.600.8433 x114                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Contact person                           | Larry Boucher<br>Regulatory Affairs Manager                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Trade name                               | ClotTriever Thrombectomy System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Common name                              | Embolectomy catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Regulation name                          | Embolectomy catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification number                    | 21 CFR 870.5150                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Product code                             | QEW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Regulatory class                         | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Predicate device                         | ClotTriever Thrombectomy System (K210190)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Reference Device                         | Cook Incorporated Dilator Sets (K183036)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Description                              | The ClotTriever Thrombectomy System is a single-use, sterile medical device<br>designed for use in the peripheral vasculature. The ClotTriever<br>Thrombectomy System consists of the ClotTriever Sheath and the<br>ClotTriever Catheter. The ClotTriever Sheath comprises a reinforced<br>polymeric sheath equipped with a self-expanding distal mesh funnel, a<br>flush/aspiration port, and a proximal hemostatic valve. The ClotTriever<br>Catheter is comprised of reinforced polymeric coaxial shafts terminating in<br>an expandable coring element and thrombus collection bag. Three ports<br>terminating in stopcocks are provided for de-airing the catheter shafts. To<br>aid in fluoroscopic visualization, the Sheath dilator and ClotTriever Catheter<br>distal tips are radiopaque. Other provided accessories include a pre-dilator,<br>the funnel loading tool and a Large Bore 60 cc syringe |
| Indications for Use                      | The ClotTriever Thrombectomy System is indicated for:<br>The non-surgical removal of thrombi and emboli from blood vessels. Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.  The ClotTriever Thrombectomy System is intended for use in the peripheral<br>vasculature including deep vein thrombosis (DVT).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device modification                      | A pre-dilator that can be inserted into and dilate the target vessel prior to<br>the insertion of the ClotTriever Sheath will be provided with the ClotTriever system:<br>A 16 Fr pre-dilator will be included with the 13 Fr ClotTriever Sheath A 19 Fr pre-dilator will be included with the 16 Fr ClotTriever Sheath          The purpose for providing the pre-dilator is to assure its availability for the ClotTriever thrombectomy procedure.                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Summary of<br>substantial<br>equivalence | The ClotTriever pre-dilator and the predicate device have the same intended use: To be used for dilating puncture sites or catheter tracts for percutaneous placement of devices.<br><br>The addition of the pre-dilator does not change the intended use of the ClotTriever Thrombectomy System: removal of thrombus and emboli from and infusion of fluids into the peripheral vasculature.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                          | <b>Non-Clinical Testing</b><br>In accordance with the Design Failure Modes and Effects Analysis, verification and validation testing were identified to support the substantial equivalence of the pre-dilator. This testing demonstrated compliance with relevant product specifications. These tests included:  Visual and Dimensional Inspection Guidewire Compatibility Leak Testing Simulated Use Insertion/Kink Radius Testing Tensile Testing In Vivo Functional Testing/Radiopacity Verification  The following testing was leveraged from K163549:  Pouch Seal Visual Inspection and Dye Penetration Pouch, Peel, Seal Strength Sterilization Validation  Clinical testing was not required for the determination of substantial equivalence.<br><br>Test results demonstrated that all acceptance criteria were met; therefore, the device conforms to established product specifications.            |
|                                          | <b>Conclusion</b><br>The proposed device modification to the ClotTriever Thrombectomy System does not change its intended use nor does it change the principles of operation. With consideration of the results of the testing, it can be concluded that the proposed pre-dilator is substantially equivalent to the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

{4}------------------------------------------------

{5}------------------------------------------------

| Device              | ClotTriever Thrombectomy System<br>Proposed (TBD)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | ClotTriever Thrombectomy System<br>Predicate (K210190)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Cook Inc. Dilator Sets<br>Reference (K183036)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer        | Inari Medical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Inari Medical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Cook Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product code        | QEW, DRE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | QEW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | DRE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Intended use        | Removal of thrombus and emboli from, and infusion of<br>fluids into, the peripheral vasculature.                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Removal of thrombus and emboli from, and infusion of<br>fluids into, the peripheral vasculature.                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Intended to be used for dilating puncture sites or<br>catheter tracts for percutaneous placement of devices<br>for vascular and non-vascular applications such as in<br>the venous, arterial, biliary and renal systems.                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Indications for use | The ClotTriever Thrombectomy System is indicated for:<br>• The non-surgical removal of thrombi and emboli from<br>blood vessels.<br>• Injection, infusion, and/or aspiration of contrast media<br>and other fluids into or from a blood vessel.<br><br>The ClotTriever Thrombectomy System is intended for use in<br>the peripheral vasculature including deep vein thrombosis<br>(DVT).                                                                                                                                                                                   | The ClotTriever Thrombectomy System is indicated for:<br>• The non-surgical removal of thrombi and emboli from<br>blood vessels.<br>• Injection, infusion, and/or aspiration of contrast media<br>and other fluids into or from a blood vessel.<br><br>The ClotTriever Thrombectomy System is intended for use in<br>the peripheral vasculature including deep vein thrombosis<br>(DVT).                                                                                                                                                                    | Intended to be used for dilating puncture sites or<br>catheter tracts for percutaneous placement of devices<br>for vascular and non-vascular applications such as in<br>the venous, arterial, biliary and renal systems.                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Description  | The ClotTriever Thrombectomy System consists of the<br>ClotTriever Sheath and the ClotTriever Catheter. The<br>ClotTriever Sheath comprises a reinforced polymeric sheath<br>equipped with a self-expanding distal mesh funnel, a<br>flush/aspiration port, and a proximal hemostatic valve. The<br>ClotTriever Catheter is comprised of reinforced polymeric<br>coaxial shafts terminating in an expandable coring element<br>and thrombus collection bag. Other provided accessories<br>include a pre-dilator, the funnel loading tool and a Large<br>Bore 60 cc syringe | The ClotTriever Thrombectomy System consists of the<br>ClotTriever Sheath and the ClotTriever Catheter. The<br>ClotTriever Sheath comprises a reinforced polymeric sheath<br>equipped with a self-expanding distal mesh funnel, a<br>flush/aspiration port, and a proximal hemostatic valve. The<br>ClotTriever Catheter is comprised of reinforced polymeric<br>coaxial shafts terminating in an expandable coring element<br>and thrombus collection bag. Other provided accessories<br>include the funnel loading tool and a Large Bore 60 cc<br>syringe | Cook Dilator Sets are used percutaneously to dilate<br>puncture sites or catheter tracts, thereby facilitating<br>the placement of other therapeutic or diagnostic<br>devices into a natural body space (e.g., the peritoneal<br>cavity, an artery or vein) for various vascular or non-<br>vascular clinical applications. The dilator component of<br>the Dilator Sets is available in diameters ranging from<br>3.0 to 26.0 French and in lengths ranging from 6 to 65<br>cm. The broad range of dilator sizes accommodates the<br>variation in the size of the K183036 devices that could<br>be inserted through the initial access site. |

{6}------------------------------------------------

| Device                  | ClotTriever Thrombectomy System<br>Proposed (TBD)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | ClotTriever Thrombectomy System<br>Predicate (K210190)                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Cook Inc. Dilator Sets<br>Reference (K183036)                                                                                                 |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Principles of operation | The ClotTriever Catheter is advanced into the vessel and<br>beyond the clot. The self-expanding braided nitinol wire net<br>is deployed. The expanded net cores, separates, and entraps<br>thrombus from the vessel as it is being drawn to the funnel<br>opening of the ClotTriever Sheath. The net is collapsed and<br>pulled into and through the ClotTriever Sheath with the<br>entrapped clot. A 60 cc syringe is provided for the aspiration<br>of clot in the sheath and the infusion of contrast media and<br>other fluids. | The ClotTriever Catheter is advanced into the vessel and<br>beyond the clot. The self-expanding braided nitinol wire net<br>is deployed. The expanded net cores, separates, and entraps<br>thrombus from the vessel as it is being drawn to the funnel<br>opening of the ClotTriever Sheath. The net is collapsed and<br>pulled into and through the ClotTriever Sheath with the<br>entrapped clot. A 60 cc syringe is provided for the aspiration<br>of clot in the sheath and the infusion of contrast media and<br>other fluids. | The dilator is advanced over a standard 0.035"<br>guidewire into the access site prior to placement of a<br>therapeutic or diagnostic device. |
| Guidewire compatibility | 0.035"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 0.035"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 0.035"                                                                                                                                        |
| How provided            | Sterile, single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Sterile, single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Sterile, single use                                                                                                                           |
| Pre-dilator             | 13 Fr ClotTriever Sheath:  16 Fr Pre-dilator<br>16 Fr ClotTriever Sheath:  19 Fr Pre-dilator                                                                                                                                                                                                                                                                                                                                                                                                                                        | Recommended accessory                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 3 Fr to 26 Fr                                                                                                                                 |
| Pre-dilator material    | LDPE/HDPE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Not provided                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | LDPE/HDPE                                                                                                                                     |

---

**Source:** [https://fda.innolitics.com/device/K210689](https://fda.innolitics.com/device/K210689)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
