← Product Code [KDI](/productcode/KDI) · K210661

# SC+ Hemodialysis Machine, SC+ Dialysate Cartridge, SC+ Blood Tubeset (K210661)

_Quanta Dialysis Technologies, Ltd. · KDI · Aug 12, 2021 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210661

## Device Facts

- **Applicant:** Quanta Dialysis Technologies, Ltd.
- **Product Code:** [KDI](/productcode/KDI.md)
- **Decision Date:** Aug 12, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5860
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The SC+ Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription, by a trained person who is considered competent in the use of the device. The SC+ Blood Tubeset is a single use, disposable arterial and venous bloodline set intended to provide extracorporeal access during hemodialysis. The Blood Tubeset is compatible only with the SC+ Hemodialysis System.

## Device Story

Hemodialysis delivery system for acute/chronic renal failure; utilizes dialysis water from external RO units to produce dialysate. System comprises SC+ Machine, single-use Dialysate Cartridge, and sterile Blood Tubeset. Machine features water circuit, blood leak/air-in-blood detectors, pneumatic interface, peristaltic blood pump, and sensors. Cartridge performs mixing/proportioning via membrane pumps and monitors conductivity, pressure, and temperature. Tubeset provides extracorporeal access via arterial/venous lines, pressure transducer pods, and venous drip chamber. Used in acute/chronic care facilities by trained operators. Modifications include 3G/4G cellular data transmission to Quanta Service Portal, software update capability, and streamlined priming/air removal workflows. Output informs clinical decision-making regarding fluid removal and dialysis therapy. Benefits include controlled fluid removal and renal replacement therapy.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety, electromagnetic compatibility, software verification, human factors validation, and biological safety testing.

## Technological Characteristics

System uses PVC and PP blood tubing. Machine includes peristaltic blood pump, membrane pumps for dialysate proportioning, conductivity/pressure/temperature sensors, and air/blood leak detectors. Connectivity via 3G/4G cellular and ethernet. Software-controlled with GUI. Sterilization for tubeset is sterile/non-pyrogenic. Complies with consensus standards for hemodialysis systems.

## Regulatory Identification

A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).

## Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”

## Predicate Devices

- SC+ Machine & SC+ Dialysate Cartridge ([K193670](/device/K193670.md))
- SC+ Blood Tubeset ([K193670](/device/K193670.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 12, 2021

Quanta Dialysis Technologies Ltd Sam Drew Risk and Regulatory Manager 1-3 Tything Road Alcester, Warwickshire B49 6EU UNITED KINGDOM

K210661 Re:

> Trade/Device Name: SC+ Hemodialysis Machine, SC+ Dialysate Cartridge, SC+ Blood Tubeset Regulation Number: 21 CFR 876.5860 Regulation Name: High Permeability Hemodialysis System Regulatory Class: Class II Product Code: KDI, FJK Dated: July 12, 2021 Received: July 13, 2021

Dear Sam Drew:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Carolyn Y. Neuland, Ph.D. Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K210661

Device Name SC+ Hemodialysis Machine SC+ Dialysate Cartridge SC+ Blood Tubeset

Indications for Use (Describe) SC+ Hemodialysis Machine/ SC+ Dialysate cartridge

The SC+ Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription, by a trained person who is considered competent in the use of the device.

SC+ Blood Tubeset

The SC+ Blood Tubeset is a single use, disposable arterial and venous bloodline set intended to provide extracorporeal access during hemodialysis. The Blood Tubeset is compatible only with the SC+ Hemodialysis System.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a logo with a dark blue circle that has a break in the lower right quadrant. A small turquoise square fills the break in the circle. The logo is simple and modern, with a clean design.

K210661 510(k) Summary Date of preparation: 12th July 2021

# 510(k) Summary

The content of this 510(k) summary is provided in conformance with 21 CFR Part 807.92

## 1.1 Manufacturer

Quanta Dialysis Technologies Ltd Tything Road, Alcester, Warwickshire, B49 6EU, United Kingdom

## 1.2 Contact

Sam Drew Sam.drew@quantadt.com +44 01789 400043

## 1.3 Device information

The SC+ Hemodialysis system comprises of the SC+ Hemodialysis Machine, the SC+ Dialysate Cartridge and the SC+ Blood Tubeset.

#### 1.3.1 SC+ Machine & Dialysate Cartridge

#### General Device Information 1.3.1.1

| Trade Name          | SC+ Machine & SC+ Dialysate Cartridge                                  |
|---------------------|------------------------------------------------------------------------|
| Common Name         | Hemodialysis Delivery System                                           |
| Product Code        | KDI                                                                    |
| Device              | Dialyzer, High Permeability With Or Without<br>Sealed Dialysate System |
| Classification Name | High permeability hemodialysis system                                  |
| Regulation Number   | 21 CFR §876.5860                                                       |
| Device Class        | Class II                                                               |
| Review Panel        | Gastroenterology/Urology                                               |

Table 1: SC+ Device and Dialysate Cartridge general device information

{4}------------------------------------------------

## 1.3.1.2 Predicate Device Information:

| Trade Name          | SC+ Machine & SC+ Dialysate Cartridge                                  |
|---------------------|------------------------------------------------------------------------|
| Common Name         | Hemodialysis Delivery System                                           |
| Product Code        | KDI                                                                    |
| Device              | Dialyzer, High Permeability With Or Without<br>Sealed Dialysate System |
| Classification Name | High permeability hemodialysis system                                  |
| Regulation Number   | 21 CFR §876.5860                                                       |
| Device Class        | Class II                                                               |
| Review Panel        | Gastroenterology/Urology                                               |
| 510(k) reference    | K193670                                                                |
| Date of Clearance   | 2020-12-23                                                             |

Table 2: SC+ device and Dialysate Cartridge predicate device information

#### SC+ Blood Tubeset 1.3.2

#### 1.3.2.1 General Device Information

| Trade Name          | SC+ Blood Tubeset                                             |
|---------------------|---------------------------------------------------------------|
| Common Name         | Blood Tubing Set                                              |
| Product Code        | FJK                                                           |
| Device              | SC+ Blood Tubeset                                             |
| Classification Name | Set, tubing, blood, with and without anti-regurgitation valve |
| Regulation Number   | 21 CFR §876.5820                                              |
| Device Class        | Class II                                                      |
| Review Panel        | Gastroenterology/Urology                                      |

Table 3: SC+ Blood Tubeset general device information

#### 1.3.2.2 Predicate Device Information:

| Trade Name          | SC+ Blood Tubeset                                                 |
|---------------------|-------------------------------------------------------------------|
| Common Name         | Blood Tubing Set                                                  |
| Product Code        | FJK                                                               |
| Device              | SC+ Blood Tubeset                                                 |
| Classification Name | Set, tubing, blood, with and without anti-<br>regurgitation valve |
| Regulation Number   | 21 CFR §876.5820                                                  |
| Device Class        | Class II                                                          |
| Review Panel        | Gastroenterology/Urology                                          |
| 510(k) reference    | K193670                                                           |
| Date of Clearance   | 2020-12-23                                                        |

Table 4: SC+ Blood Tubeset predicate device information

{5}------------------------------------------------

# 1.4 Device Description

The SC+ is a hemodialysis delivery system intended for acute and chronic dialysis therapy with or without ultrafiltration in an acute of chronic care facility. The system consists of the SC+ Machine, a single use disposable Dialysate Cartridge, and a sterile, single use, disposable Blood Tubeset.

### 1.4.1 SC+ Hemodialysis System

The SC+ Hemodialysis System is intended for acute and chronic dialysis therapy, with or without ultrafiltration, utilizing Dialysis Water (from standalone Reverse Osmosis (RO) units or a central RO ring main) to produce dialysate. The SC+ Hemodialysis system is for use in patients with arteriovenous (AV) fistula or central venous catheter access.

The system consists of the SC+ Machine, a single use disposable Dialysate Cartridge, and a sterile, single use, disposable Blood Tubeset.

The SC+ Machine consists of a water circuit (heater, de-aeration module, etc) blood leak detector, air in blood detector, a pneumatic interface for the dialysate cartridges, a peristaltic blood pump and various other sensors. The dialysate cartridge contains the following; conductivity monitors, interfaces for pressure and temperature measurement, membrane pumps to perform mixing/proportioning in order to produce dialysis fluid and the controlled removal of fluid from a patient with acute and/or chronic renal failure based on a physician's prescription. The dialysate fluid is manufactured using dialysis water purified externally by reverse osmosis that is heated to approximately 37℃ and subsequently deaerated within the machine before entering the cartridge.

The SC+ Blood Tubeset is a single use, sterile device consisting of an arterial line, a venous line, connections to a standard dialyzer, a saline line, three pressure transducer pods integrated into a single unit, a venous drip chamber, and a line for heparin infusion.

# 1.5 Device Modifications

This 510(k) covers the following modifications:

- Addition of two-way wireless transmission capability for machine data ●
- Updated user workflows and interface for ease of use and streamlining ●

## 1.6 Indications for Use

### 1.6.1 SC+ Hemodialysis Machine & Dialysate Cartridge

The SC+ Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription, by a trained person who is considered competent in the use of the device.

### 1.6.2 SC+ Blood Tubeset

The SC+ Blood Tubeset is a single use, disposable arterial and venous bloodline

{6}------------------------------------------------

set intended to provide extracorporeal access during hemodialysis. The Blood Tubeset is compatible only with the SC+ Hemodialysis System.

#### 1.7 Technological Characteristics

#### Changed SC+ device vs SC+ device as cleared in K193670 1.7.1

The technological characteristics of the SC+ Machine and SC+ Dialysate Cartridge are considered to be equivalent to the predicate device, SC+ device and dialysate cartridge (K193670). A summary of the similarities and differences is provided in the table below.

|                                                                                                                                                | Subject Device | Predicate<br>Device |
|------------------------------------------------------------------------------------------------------------------------------------------------|----------------|---------------------|
|                                                                                                                                                | SC+ (K210661)  | SC+ (K193670)       |
| Indicated for use in patients with acute and/or chronic renal failure, with<br>or without ultrafiltration in an acute or chronic care facility | ✓              | ✓                   |
| Use of purified water for dialysate production                                                                                                 | ✓              | ✓                   |
| Use of third-party accessories, including dialyzers, endotoxin retentive<br>filters, acid and bicarbonate                                      | ✓              | ✓                   |
| Application of consensus standards                                                                                                             | ✓              | ✓                   |
| Device is software controlled and utilize Graphic User Interface (GUI).                                                                        | ✓              | ✓                   |
| Design and Construction - Blood pump, alarms, alerts, air detector<br>mechanism, and blood leak detectors.                                     | ✓              | ✓                   |

Table 5: Technological Characteristics of the SC+ Machine and predicate

Differences between the subject device and the predicate device:

- The subject device utilizes 3G/4G cellular communications to transmit machine . data (non-clinical) to the Quanta Service Portal and download software updates (duplicating the predicate device's ethernet connectivity as cleared in K193670)
- The subject device has streamlined process with regards to air in venous line removal and priming

{7}------------------------------------------------

#### Changed SC+ Blood Tubeset vs SC+ Blood Tubeset device as cleared in 1.7.2 K193670

The technological characteristics of the SC+ Blood Tubeset are considered to be equivalent to the predicate device, SC+ Blood Tubeset (K193670). A summary of the similarities and differences is provided in the table below.

|                                                                                                                                                                                                                                 | Subject<br>Device                 | Predicate<br>Device               |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|-----------------------------------|
|                                                                                                                                                                                                                                 | SC+ Blood<br>Tubeset<br>(K210661) | SC+ Blood<br>Tubeset<br>(K193670) |
| Intended Use: To provide extracorporeal access<br>during hemodialysis                                                                                                                                                           | ✓                                 | ✓                                 |
| Materials: Primary fluid path materials are Polyvinyl<br>Chloride (PVC) and Polypropylene (PP).                                                                                                                                 | ✓                                 | ✓                                 |
| Design & Construction: Polyvinyl Chloride (PVC)<br>tubing of various lengths and diameters, with color<br>coded pinch clamps, color coded injection ports,<br>heparin line, saline line, and pressure monitoring<br>components. | ✓                                 | ✓                                 |
| Sterility: Sterile, single use, non-pyrogenic.                                                                                                                                                                                  | ✓                                 | ✓                                 |
| Priming Volume:                                                                                                                                                                                                                 | ≤165ml                            | ≤165ml                            |
| Needle configuration: Double needle                                                                                                                                                                                             | ✓                                 | ✓                                 |

Table 6: Technological Characteristics of the SC+ Blood Tubeset and predicate

Differences between the subject device and the predicate device:

- . The subject device includes a luer activated valve (needless valve/ air removal port) on the venous bloodline for streamlined air removal
- . The saline spike preconnected to the arterial bloodline is replaced with a recirculation luer (this recirculation luer is a component as cleared in K193670)

# 1.8 Summary of Verification and Validation Activities

The following performance testing, developed in accordance with appropriate FDA guidance documents and relevant standards, has been performed on the Modified Device to support the determination of substantial equivalence:

{8}------------------------------------------------

- Performance testing to verify maintenance of the Electrical Safety and Electromagnetic Compatibility profiles
- Software verification
- Human Factors validation testing •
- Biological safety testing •

# 1.9 Conclusion

The modified devices are substantially equivalent to the predicate devices

---

**Source:** [https://fda.innolitics.com/device/K210661](https://fda.innolitics.com/device/K210661)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
