Single-Use Medical Face Mask

K210622 · Jiangxi Hongda Medical Equipment Group Co., Ltd. · FXX · Oct 8, 2021 · General, Plastic Surgery

Device Facts

Record IDK210622
Device NameSingle-Use Medical Face Mask
ApplicantJiangxi Hongda Medical Equipment Group Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateOct 8, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Intended Use

The single-use medical face masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided sterile and non-sterile.

Device Story

Single-use, flat-pleated medical face mask; three-layer construction (polypropylene spunbonded outer/inner layers; polypropylene melt-blown middle layer). Available in Level 1, 2, and 3 (ASTM F2100-19). Features ear loops (polyester/spandex) or ties (polypropylene nonwoven) and adjustable nose clip (polypropylene/iron). Provided sterile (EO) or non-sterile. Used in clinical settings for infection control; worn by healthcare personnel and patients. Acts as physical barrier to reduce exposure to blood, body fluids, and particulates. Benefits patient/staff by mitigating cross-contamination risk.

Clinical Evidence

No clinical data. Bench testing only. Compliance verified via ASTM F2100-19, F2101-19 (BFE), F2299-03 (PFE), F1862-17 (synthetic blood penetration), and EN 14683:2019 (differential pressure). Biocompatibility confirmed per ISO 10993-5, -7, and -10. Accelerated aging (3 years) confirmed performance retention.

Technological Characteristics

Three-layer polypropylene nonwoven construction; melt-blown middle layer (30g/m² for Level 1; 40g/m² for Levels 2/3). Nose clip: polypropylene/iron. Ear loops: polyester/spandex. Ties: polypropylene nonwoven. Sterilization: Ethylene Oxide (EO). ASTM F2100-19 compliant. Dimensions: 175x95mm.

Indications for Use

Indicated for use by patients and healthcare personnel to protect against transfer of microorganisms, body fluids, and particulate material in infection control practices to reduce exposure to blood and body fluids.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. October 8, 2021 Jiangxi Hongda Medical Equipment Group Co., Ltd. Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box. 120-119 Shanghai. 200120 China Re: K210622 Trade/Device Name: Single-Use Medical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: September 04, 2021 Received: September 14, 2021 Dear Diana Hong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K210622 Device Name Single-use Medical Face Mask ## Indications for Use (Describe) The single-use medical face masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided sterile and non-sterile. Level 1 mask Models: Sterile Tie-on, Sterile Ear loop, Non-sterile Tie-on, Non-sterile Ear loop Level 2 mask Models: Sterile Tie-on. Sterile Ear loop. Non-sterile Tie-on. Non-sterile Ear loop Level 3 mask Models: Sterile Tie-on, Sterile Ear loop, Non-sterile Tie-on, Non-sterile Ear loop Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) |X Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K210622 This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. The assigned 510(k) Number: K210622 - 1. Date of Preparation: 10/07/2021 - 2. Sponsor Identification # Jiangxi Hongda Medical Equipment Group Co., Ltd. 39 South Shengli Road, Jinxian County, Nanchang City, Jiangxi Province, China. Establishment Registration Number: 3007392007. Contact Person: Xiaohua Yang Position: Management representative Tel: +86-791-85660368 Fax: +86-791-85660368 Email: jsb@jxhd.cn - 3. Designated Submission Correspondent Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person) # Mid-Link Consulting Co., Ltd. P.O. Box 120-119, Shanghai, 200120, China Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net {4}------------------------------------------------ - 4. Identification of Proposed Device Trade Name: Single-use Medical Face Mask Common Name: Surgical Mask Regulatory Information Classification Name: Mask, Surgical Classification: II; Product Code: FXX: Regulation Number: 21CFR 878.4040 Review Panel: General Hospital Indication for use: The single-use medical face masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided sterile and non-sterile. Device Description: The proposed device is a three-layer, single-use, flat- pleated mask. The inner and outer layers of the mask are made of polypropylene spunbonded nonwoven, and the middle layer is made of polypropylene melt-blown nonwoven. The proposed devices are available in two types, ear loop and tie-on. The ear loops are made of polyester and spandex, and the ties are made of polypropylene nonwoven. The ear loops/ties are held in place over the users' mouth and nose by ear loops/ ties welded to the mask. The nose clip is made of polypropylene and iron. Users can adjust the nose clip according to the shape of the bridge of the nose, and fix the mask on the bridge of the nose to prevent the mask from falling off. Both the ear loop and tie-on masks are available in level 1, level 2 and level 3 masks based on ASTM F2100-19, which are provided in following table. The difference of the level 1, level 2 and level 3 masks is the gram weight of the melt-blown nonwoven. The gram weight of the melt-blown nonwoven for level 1 is 30g/m², and that for level 2 and level 3 masks is 40g/m². Level 2 and level 3 masks are exactly the same and are divided into level 2 and level 3 just for marking. The proposed devices are provided in sterile and non-sterile two types. | ASTM | Specification | Ear loop | Tie-on | Color | Sterility | |----------|----------------------|----------|--------|-------|-------------| | F2100-19 | | √ | | | Sterile | | Level 1 | $17.5 \times 9.5$ cm | √ | | White | Sterile | | | | √ | | White | Non-sterile | | | | | √ | White | Sterile | | | | | √ | White | Non-sterile | {5}------------------------------------------------ | Level 2 | | √ | | White | Sterile | |---------|--|---|---|-------|-------------| | | | √ | | White | Non-sterile | | | | | √ | White | Sterile | | | | | √ | White | Non-sterile | | Level 3 | | √ | | White | Sterile | | | | √ | | White | Non-sterile | | | | | √ | White | Sterile | | | | | √ | White | Non-sterile | #### 5. Identification of Predicate Device 510(k) Number: K160269 Product Name: Surgical Face Masks (Ear loops and Tie-on) #### 6. Non-Clinical Test Conclusion Non clinical tests were conducted to verify that the proposed device met all design specifications as was same/similar to the predicate device. The test results demonstrated that the proposed device complies with the following standards: - A ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritation and skin sensitization: - > ISO 10993-5:2009 Biological evaluation of medical device-Part 5: Tests for in vitro cytotoxicity; - > ASTM F2100-19 Standard Specification for Performance of Materials Used in Medical Face Masks - A ASTM F2101-19 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials; - > ASTM F2299/F2299M-03 (2017) Standard Test Method for Determining the Initial Efficiency of Material Used in medical Face Masks to Penetration by Particulates using Latex Spheres; - > ASTM F1862/F1862M-17 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity); - > 16 CFR Part 1610 Standard for the Flammability of Clothing Textiles; - > EN 14683: 2019 Medical face masks- Requirements and test methods; - > ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection; - > ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials; - > ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration: - > ISO 10993-7:2008 Biological Evaluation of Medical Device- Part 7: Ethylene Oxide Sterilization Residuals; {6}------------------------------------------------ | Models/types | Test Method | Acceptance Criteria | Test Result | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sterile Tie-on | Resistance to Penetration<br>by Synthetic blood from 3<br>non-consecutive lots<br>ASTM F1862/F1862M-17<br>Standard Test Method for<br>Resistance of Medical Face<br>Masks to Penetration by<br>Synthetic Blood<br>(Horizontal Projection of<br>Fixed Volume at a Known<br>Velocity) | Level 1: No penetration<br>at 80 mmHg;<br>Level 2: No penetration<br>at 120 mmHg;<br>Level 3: No penetration<br>at 160 mmHg | Sterile Tie-on<br>Level 1: Pass at 80 mmHg;<br>Level 2: Pass at 120 mmHg;<br>Level 3: Pass at 160 mmHg<br>Sterile Ear loop<br>Level 1: Pass at 80 mmHg;<br>Level 2: Pass at 120 mmHg;<br>Level 3: Pass at 160 mmHg<br>Non-sterile Tie-on<br>Level 1: Pass at 80 mmHg;<br>Level 2: Pass at 120 mmHg;<br>Level 3: Pass at 160 mmHg<br>Non-sterile Ear loop<br>Level 1: Pass at 80 mmHg;<br>Level 2: Pass at 120 mmHg;<br>Level 3: Pass at 160 mmHg | | Sterile Ear loop | | | | | Non-sterile Tie-on | | | | | Non-sterile Ear loop | | | | | | Bacterial Filtration<br>Efficiency from 3<br>non-consecutive lots<br>ASTM F2101-2019<br>Standard Test Method for<br>Evaluating the Bacterial<br>Filtration Efficiency (BFE)<br>of Medical Face Mask<br>Materials, Using a<br>Biological Aerosol of<br>Staphylococcus aureus | Level 1: ≥95%<br>Level 2: ≥98%<br>Level 3: ≥98% | Sterile Tie-on<br>Level 1: 98.48%<br>Level 2: 99.84%<br>Level 3: 99.85%<br>Sterile Ear loop<br>Level 1: 98.38%<br>Level 2: 99.86%<br>Level 3: 99.89%<br>Non-sterile Tie-on<br>Level 1: 98.39%<br>Level 2: 99.82%<br>Level 3: 99.87%<br>Non-sterile Ear loop<br>Level 1: 98.67%<br>Level 2: 99.86%<br>Level 3: 99.88% | | | Particulate Filtration<br>Efficiency from 3<br>non-consecutive lots<br>ASTM F2299/F2299M-03 | Level 1: ≥95%<br>Level 2: ≥98%<br>Level 3: ≥98% | Sterile Tie-on<br>Level 1: 99.22%<br>Level 2: 99.85%<br>Level 3: 99.88% | | | | | | | (2017) Standard Test<br>Method for Determining the<br>Initial Efficiency of<br>Material Used in medical<br>Face Masks to Penetration<br>by Particulates using Latex<br>Spheres | | Sterile Ear loop<br>Level 1: 99.23%<br>Level 2: 99.86%<br>Level 3: 99.88%<br>Non-sterile Tie-on<br>Level 1: 99.23%<br>Level 2: 99.87%<br>Level 3: 99.88%<br>Non-sterile Ear loop<br>Level 1: 99.23%<br>Level 2: 99.87%<br>Level 3: 99.87% | | | Differential Pressure from 3<br>non-consecutive lots<br>EN 14683:2019 Medical<br>face masks- Requirements<br>and test methods | Level 1 <5.0 mm H2O/cm²<br>Level 2: <6.0 mm H2O/cm²<br>Level 3: <6.0 mm H2O/cm² | Sterile Tie-on<br>Level 1: 4.00 mm H2O/cm²<br>Level 2: 4.83 mm H2O/cm²<br>Level 3: 4.90 mm H2O/cm²<br>Sterile Ear loop<br>Level 1: 3.83 mm H2O/cm²<br>Level 2: 4.77 mm H2O/cm²<br>Level 3: 4.90 mm H2O/cm²<br>Non-sterile Tie-on<br>Level 1: 4.10 mm H2O/cm²<br>Level 2: 4.90 mm H2O/cm²<br>Level 3: 4.93 mm H2O/cm²<br>Non-sterile Ear loop<br>Level 1: 4.00 mm H2O/cm²<br>Level 2: 4.90 mm H2O/cm²<br>Level 3: 4.97 mm H2O/cm² | | | Flammability from 3<br>non-consecutive lots<br>16 CFR 1610 Standard for<br>the Flammability of<br>Clothing Textiles<br>Corrections | Class I | Class I | | | EO ECH Residual Test<br>ISO 10993-7:2008 | EO residual shall not<br>exceed 4mg/device<br>ECH residual shall not<br>exceed 9mg/device | EO residual was less than<br>4mg/device<br>ECH residual was less than<br>9mg/device | | | Cytotoxicity<br>ISO 10993-5: 2009 | The viability should be<br>>70% of the blank | The viability was ≥70% of<br>the blank. And the 50% | | | Biological Evaluation of<br>Medical Devices-Part 5:<br>Tests for in Vitro<br>Cytotoxicity | And the 50% extract of<br>the test sample should<br>have at least the same<br>or a higher viability<br>than the 100% extract. | extract of the test sample<br>had a higher viability than<br>the 100% extract.<br><br>Under the conditions of the<br>study, the proposed device<br>was non-cytotoxic. | | | Irritation<br>ISO 10993-10: 2010<br>Biological Evaluation of<br>Medical Devices-Part 10:<br>Tests for Irritation and Skin<br>Sensitization | Non-irritating | Under the conditions of the<br>study, the proposed device<br>was non-irritating. | | | Sensitization<br>ISO 10993-10: 2010<br>Biological Evaluation of<br>Medical Devices-Part 10:<br>Tests for Irritation and Skin<br>Sensitization | Non-sensitizing | Under the conditions of the<br>study, the proposed device<br>was non-sensitizing. | | Table 1. Summary of the Non-clinical Testing {7}------------------------------------------------ {8}------------------------------------------------ - 7. Clinical Test Conclusion No clinical study is included in this submission. {9}------------------------------------------------ - 8. Summary of Technological characteristics | ITEM | Proposed Device K210622 | Predicate Device K160269 | Remark | |-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------| | Product Code | FXX | FXX | Same | | Regulation No. | 21 CFR 878.4040 | 21 CFR 878.4040 | Same | | Class | II | II | Same | | Indication for Use | The single-use medical face masks are intended to<br>be worn to protect both the patient and healthcare<br>personnel from transfer of microorganisms, body<br>fluids, and particulate material. These face masks<br>are intended for use in infection control practices<br>to reduce the potential exposure to blood and<br>body fluids. This is a single-use, disposable<br>device, provided sterile and non-sterile. | The surgical face masks are<br>intended to be worn to protect both<br>the patient and healthcare personnel<br>from transfer of microorganisms,<br>body fluids, and particulate<br>material. These face masks are<br>intended for use in infection control<br>practices to reduce the potential<br>exposure to blood and body fluids.<br>This is a single-use, disposable<br>device, provided non-sterile. | Similar | | Mask style | Flat-pleated | Flat-pleated | Same | | Design feature | Ear loop or tie-on | Ear loop or tie-on | Same | | Color | White | Blue, White | Different | | Dimension | Mask body: 175×95mm<br>Nose Clip: 80~105mm<br>Ear loop: 140~150mm<br>Ties: 330~400mm | 175×90mm<br>180×90mm | Different | | ASTM<br>F2100<br>Level | Level 1, Level 2, Level 3 | Level 1, Level 2, Level 3 | Same | | Sterility | Non sterile/Sterile<br>Sterilization method: Ethylene Oxide (EO)<br>Packaging material: Paper and Polyethylene film | Non-sterile | Different | | EO ECH Residual | EO residue: <4mg/device<br>ECH residue: <9mg/device | The predicate device is non-sterile | | | Use | Single Use, Disposable | Single Use, Disposable | Same | | Fluid resistance | Level 1: Pass at 80 mmHg<br>Level 2: Pass at 120 mmHg<br>Level 3: Pass at 160 mmHg | Level 1: Pass at 80 mmHg<br>Level 2: Pass at 120 mmHg<br>Level 3: Pass at 160 mmHg | Same | | Particulate<br>filtration<br>efficiency | Level 1:<br>Ear loop (Non-sterile): ≥95%<br>Ear loop (Sterile): ≥95%<br>Tie-on (Non-sterile): ≥95% | Level 1: Pass at 99.6%<br>Level 2: Pass at 99.6%<br>Level 3: Pass at 99.7% | Different | | | Tie-on (Sterile): ≥95% | | | | | Level 2: | | | | | Ear loop (Non-sterile): ≥98% | | | | | Ear loop (Sterile): ≥98% | | | | | Tie-on (Non-sterile): ≥98% | | | | | Tie-on (Sterile): ≥98% | | | | | Level 3: | | | | | Ear loop (Non-sterile): ≥98% | | | | | Ear loop (Sterile): ≥98% | | | | | Tie-on (Non-sterile): ≥98% | | | | | Tie-on (Sterile): ≥98% | | | | | Level 1: | | | | | Ear loop (Non-sterile): ≥95% | | | | | Ear loop (Sterile): ≥95% | | | | | Tie-on (Non-sterile): ≥95% | | | | | Tie-on (Sterile): ≥95% | | | | | Level 2: | | | | | Ear loop (Non-sterile): ≥98% | Level 1: Pass at 98% | | | Bacterial filtration<br>efficiency | Ear loop (Sterile): ≥98% | Level 2: Pass at 98% | Different | | | Tie-on (Non-sterile): ≥98% | Level 3: Pass at 99% | | | | Tie-on (Sterile): ≥98% | | | | | Level 3: | | | | | Ear loop (Non-sterile): ≥98% | | | | | Ear loop (Sterile): ≥98% | | | | | Tie-on (Non-sterile): ≥98% | | | | | Tie-on (Sterile): ≥98% | | | | | Level 1: | | | | | Ear loop (Non-sterile): <5mmH2O/cm2 | | | | | Ear loop (Sterile): <5mmH2O/cm2 | Level 1: Pass at 2.0 mmH2O/cm2<br>Level 2: Pass at 1.6 mmH2O/cm2<br>Level 3: Pass at 2.5 mmH2O/cm2 | | | | Tie-on (Non-sterile): <5mmH2O/cm2 | | | | Differential<br>pressure | Tie-on (Sterile): <5mmH2O/cm2 | | | | | Level 2: | | | | | Ear loop (Non-sterile): <6mmH2O/cm2 | | Different | | | Ear loop (Sterile): <6mmH2O/cm2 | | | | | Tie-on (Non-sterile): <6mmH2O/cm2 | | | | | Tie-on (Sterile): <6mmH2O/cm2 |…
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%