3M Bair Hugger Universal Warming Gown made with Thinsulate Insulation
Applicant
3M Company
Product Code
DWJ · Cardiovascular
Decision Date
Apr 29, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 3M™ Bair Hugger™ Universal Warming Gown made with Thinsulate™ Insulation is part of the 3M patient temperature management system when used with Bair Hugger™ 700 series or 675 warming unit. The system, monitored and controlled by a trained clinician, is used perioperatively to prevent and treat hypothermia.
Device Story
Disposable perioperative warming gown; incorporates Thinsulate™ insulation for added thermal retention. Used as part of 3M Bair Hugger™ temperature management system; requires connection to Bair Hugger™ 700 series or 675 warming unit. Features integrated lower body insert and removable multi-position warming blanket; worn by patient before, during, and after surgery. Clinician operates warming unit to deliver forced-air heat to gown; gown distributes heat to patient to prevent/treat hypothermia. Re-doming of gown allows post-operative warming via lower midline hose port. Benefits include maintenance of normothermia during surgical journey.
Clinical Evidence
Bench testing only. Performance verified via thermal performance testing per IEC 80601-2-35. Biocompatibility testing conducted per ISO 10993-1 for surface-contacting devices with limited duration (<24 hours) contact, including cytotoxicity, sensitization, and irritation tests.
Technological Characteristics
Disposable warming gown with Thinsulate™ insulation. System utilizes forced-air warming via external unit (Bair Hugger™ 700 series or 675). Thermal performance tested per IEC 80601-2-35. Biocompatibility per ISO 10993-1. Available in small, standard, and x-large sizes.
Indications for Use
Indicated for perioperative prevention and treatment of hypothermia in patients requiring temperature management, monitored by a trained clinician.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
Predicate Devices
3M™ Bair Hugger™ Model 675 Total Temperature Management System (K171373)
Reference Devices
Bair Paws® Temperature Management System (K060865)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.
April 29, 2021
3M Company Hilary Hovde Regulatory Affairs Specialist 2510 Conway Ave. St. Paul, Minnesota 55144
#### Re: K210605
Trade/Device Name: 3MTM Bair Hugger™ Universal Warming Gown made with Thinsulate™ Insulation Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II Product Code: DWJ Dated: February 26, 2021 Received: March 1, 2021
### Dear Hilary Hovde:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Nicole Gillette Assistant Director (Acting) DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
#### Device Name
3M™ Bair Hugger™ Universal Warming Gown made with Thinsulate™ Insulation
Indications for Use (Describe)
The 3M™ Bair Hugger™ Universal Warming Gown made with Thinsulate™ Insulation is part of the 3M patient temperature management system') when used with Bair Hugger™ 700 series or 675 warming unit. The system, monitored and controlled by a trained clinician, is used perioperatively to prevent and treat hypothermia.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary for 3MTM Bair Hugger™ Universal Warming Gown made with Thinsulate™ Insulation
3M Company 3M Health Care 2510 Conway Ave. 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000
> Contact: Hilary B. Hovde Regulatory Affairs Specialist Phone Number: (651) 736-0364
Submission Date: February 26, 2021
{4}------------------------------------------------
#### Device Name and Classification
| Trade Name: | 3M™ Bair Hugger™ Universal Warming Gown made with<br>Thinsulate™ Insulation |
|------------------------|-----------------------------------------------------------------------------|
| Common/Usual Name: | Thermal Regulating System |
| Device Classification: | Class II |
| Classification Name: | Thermal Regulating System<br>[21 CFR § 870.5900, DWJ] |
#### Predicate Device
3M™ Bair Hugger™ Model 675 Total Temperature Management System, K171373
#### Reference Devices
Bair Paws® Temperature Management System. K060865
#### Indications for Use
The 3M™ Bair Hugger™ Universal Warming Gown made with Thinsulate™ Insulation is part of the 3M patient temperature management system ('the system') when used with Bair Hugger 700 series or 675 warming unit. The system, monitored and controlled by a trained clinician, is used perioperatively to prevent and treat hypothermia.
### Description of Device
The 3M™ Bair Hugger™ Universal Warming Gown made with Thinsulate™ Insulation is a disposable, single-patient perioperative warming gown with 3M™ Thinsulate™ Insulation. The warming gown is intended to be worn throughout the entire journey: before, during and after surgery.
Bair Hugger™ warming gowns are part of the Bair Hugger™ warming system; there are two parts to the system: the warming gown and warming unit (temperature management unit). The Universal Warming Gown is available in three sizes (small, standard, and x-large) and is intended to be used with the Bair Hugger™ 700 series or 675 warming unit. The system, monitored and controlled by a trained clinician, is used perioperatively to prevent and treat hypothermia.
The warming gown has two patient warming options: an integrated lower body insert and a removable multi-position warming blanket. The lower body insert has a lower midline hose port,
{5}------------------------------------------------
#### TRADITIONAL PREMARKET NOTIFICATION [510(k)] 3M™ Bair Hugger™ Universal Warming Gown made with Thinsulate™ Insulation
and the warming blanket is to be removed from a pocket in the front upper section of the gown. The warming blanket can be used alone or in combination with the warming gown during surgery.
After surgery, the warming gown should be re-domed and used to warm the patient using the lower midline hose port.
#### Comparison of Technological Characteristics with the Predicate and Reference Devices
The 3M™ Bair Hugger™ Universal Warming Gown made with Thinsulate™ Insulation was shown to be substantially equivalent to the 3M™ Bair Hugger™ Model 675 Total Temperature Management System cleared per K171373. 3M™ Bair Hugger™ Warming Gowns are part of the 3M™ Bair Hugger™ Temperature Management System; there are two parts to the system: the warming gown and warming unit (temperature management unit). This 510(k) is for a new warming gown which incorporates a layer of Thinsulate™ Insulation to provide added insulation. There is no change to the patient warming unit part of this system. Safety and efficacy of the new warming gown was demonstrated through biocompatibility testing and performance testing with the gown and unit used as a system as intended.
#### Substantial Equivalence and Summary of Studies
The difference between the subject and predicate device has been evaluated through performance and biocompatibility tests to provide evidence of substantial equivalence for the 3MTM Bair Hugger™ Universal Warming Gown made with Thinsulate™ Insulation.
The device performance was verified through the following tests:
- . Thermal Performance Testing per IEC-80601-2-35
Device samples passed the following biocompatibility tests performed in accordance with ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process as recognized by FDA. The 3M™ Bair Hugger™ Universal Warming Gown made with Thinsulate™ Insulation is categorized as a surface contacting device, with intact skin contact of limited duration (< 24 hours) in accordance with ISO 10993-1 and FDA Guidance, Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process. " The battery of tests included the following:
- Cytotoxicity ●
- . Sensitization
- . Irritation
{6}------------------------------------------------
#### TRADITIONAL PREMARKET NOTIFICATION [510(k)] 3M™ Bair Hugger™ Universal Warming Gown made with Thinsulate™ Insulation
#### Conclusion
Based on the intended use, technological characteristics, performance data, and non-clinical tests performed, the subject device is substantially equivalent to the 3M™ Bair Hugger™ Model 675 Total Temperature Management System (cleared under K171373), Class II, product code DWJ and does not raise new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.