← Product Code KGN · K210580

# InnovaMatrix FS (K210580)

_Triad Life Sciences, Inc. · KGN · Apr 21, 2021 · SU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210580

## Device Facts

- **Applicant:** Triad Life Sciences, Inc.
- **Product Code:** KGN
- **Decision Date:** Apr 21, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Intended Use

InnovaMatrix™ FS is indicated for the management of wounds including: - · Partial and full-thickness wounds - Pressure ulcers - Venous ulcers - · Diabetic ulcers - Chronic vascular ulcers - · Tunneled/undermined wounds - · Surgical wounds (donor sites/grafts, post-Mohs surgery, post- laser surgery, podiatric, wound dehiscence) - · Trauma wounds (abrasions, lacerations, second-degree burns and skin tears) - · Draining wounds The device is intended for one-time use.

## Device Story

InnovaMatrix™ FS is a decellularized extracellular matrix (ECM) topical wound dressing derived from porcine placental tissue. Composed of collagen, elastin, laminin, fibronectin, hyaluronic acid, and sulfated glycosaminoglycans. Provided as sterile, single-use, fenestrated sheets (40-100 microns thick). Applied by clinicians to the wound site to provide a protective, biodegradable cover for wound management. Fenestrations facilitate drainage. Substantially equivalent to the non-fenestrated predicate device.

## Clinical Evidence

Bench testing only. Verification testing confirmed bioburden, endotoxin, and water absorption capacity met specifications. Existing biocompatibility, shelf life, sterilization, human repeat insult patch testing, and human skin prick testing from the predicate device were leveraged.

## Technological Characteristics

Decellularized porcine placental ECM; contains collagen, elastin, laminin, fibronectin, hyaluronic acid, sulfated glycosaminoglycans. Fenestrated sheet form factor, 40-100 microns thick. E-Beam sterilized. Single-use. Biodegradable.

## Predicate Devices

- InnovaMatrix ([K193552](/device/K193552.md))

## Reference Devices

- Oasis® Wound Matrix ([K061711](/device/K061711.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

April 21, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Triad Life Sciences, Inc. % Stephen Rhodes Regulatory Consultant Streamline Regulatory 3502 Dundee Driveway Chevy Chase, Maryland 20815

Re: K210580

Trade/Device Name: InnovaMatrix FS Regulatory Class: Unclassified Product Code: KGN Dated: February 25, 2021 Received: February 26, 2021

Dear Stephen Rhodes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lixin Liu. Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210580

Device Name InnovaMatrix™ FS

Indications for Use (Describe) InnovaMatrix™ FS is indicated for the management of wounds including:

- · Partial and full-thickness wounds
- Pressure ulcers
- Venous ulcers
- · Diabetic ulcers
- Chronic vascular ulcers
- · Tunneled/undermined wounds
- · Surgical wounds (donor sites/grafts, post-Mohs surgery, post- laser surgery, podiatric, wound dehiscence)
- · Trauma wounds (abrasions, lacerations, second-degree burns and skin tears)
- · Draining wounds

The device is intended for one-time use.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary - K210580

In accordance with 21 CFR 807.97(h) and 21 CFR 807.92, the Special 510(k) Summary is provided below.

### 1. SUBMITTER

Triad Life Sciences®, Inc. 1770 Moriah Woods Blvd., Suite 18 Memphis, TN 38117 Registration Number: 3017660750

Contact Person: William J. Willis Phone: (901) 312-7923 Email: bwillis@triadls.com

Prepared By: Stephen P. Rhodes, Streamline Regulatory stephen.rhodes@streamlineregulatory.com Date Prepared: April 21, 2021

### 2. DEVICE

Name of Device: InnovaMatrix™ FS Common Name: Collagen Wound Dressing Classification Regulation/Class: Unclassified Product Code: KGN Panel: General and Plastic Surgery

## 3. PREDICATE AND REFERENCE DEVICE

Predicate Device: InnovaMatrix™ (K193552) Reference Device: Oasis® Wound Matrix (K061711)

## 4. DEVICE DESCRIPTION

### Description

InnovaMatrix™ FS is a decellularized extracellular matrix (ECM) topical wound covering derived from porcine placental tissue. Triad processes the tissue into the ECM topical wound covering. InnovaMatrix™ FS is composed of collagen, elastin, laminin, fibronectin, hyaluronic acid and sulfated glycosaminoglycans. The wound dressing is provided in fenestrated sheets that are approximately 40-100 microns thick in sizes ranging from 2 x 2cm to 5 x 5cm. They are provided as single-use, sterile wound coverings.

{4}------------------------------------------------

# 510(k) Summary - K210580

# 5. INDICATION FOR USE

InnovaMatrix™ FS is indicated for the management of wounds including: partial- and full-thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled / undermined wounds, surgical wounds (donor sites/grafts, post Mohs surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, second-degree burns and skin tears) and draining wounds.

The device is intended for one-time use.

# 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

InnovaMatrix™ FS is a decellularized extracellular matrix (ECM) topical wound covering derived from porcine placental tissue. Triad processes the tissue into the ECM topical wound coverings. InnovaMatrix™ FS is composed of collagen, elastin, laminin, fibronectin, hyaluronic acid and sulfated glycosaminoglycans. The biodegradable wound matrix provides a protective cover to the wound.

The only modification made to the device since its previous clearance in K193552 is the addition of fenestrations.

Table 1 provides a summary comparison between the subject and predicate devices.

|                      | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Device                                               |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|
| 510(k) Number        | TBD                                                                                                                                                                                                                                                                                                                                                                                                                                                               | K193552                                                        |
| Applicant            | Triad Life Sciences, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                         | Triad Life Sciences, Inc.                                      |
| Device Name          | InnovaMatrix™ FS                                                                                                                                                                                                                                                                                                                                                                                                                                                  | InnovaMatrix™                                                  |
| Regulation/Class     | Unclassified Collagen Wound<br>Dressing                                                                                                                                                                                                                                                                                                                                                                                                                           | Unclassified Collagen Wound<br>Dressing                        |
| Product Code         | KGN                                                                                                                                                                                                                                                                                                                                                                                                                                                               | KGN                                                            |
| Prescription or OTC? | Prescription                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Prescription                                                   |
| Indications for Use  | InnovaMatrix™ FS is indicated<br>for the management of wounds<br>including: partial- and full-<br>thickness wounds, pressure<br>ulcers, venous ulcers, diabetic<br>ulcers, chronic vascular ulcers,<br>tunneled/undermined wounds,<br>surgical wounds (donor<br>sites/grafts, post-Mohs surgery,<br>post-laser surgery, podiatric,<br>wound dehiscence), trauma<br>wounds (abrasions, lacerations,<br>second-degree burns and skin<br>tears) and draining wounds. | same                                                           |
| Design               | Single layer, porcine<br>extracellular matrix from<br>placenta (fenestrated)                                                                                                                                                                                                                                                                                                                                                                                      | Single layer, porcine<br>extracellular matrix from<br>placenta |
| Sizes                | 2 x 2 cm, 4 x 4 cm, 5 x 5 cm                                                                                                                                                                                                                                                                                                                                                                                                                                      | 1 x 1 cm to 5 x 5 cm                                           |
| Thickness            | 40-100 microns                                                                                                                                                                                                                                                                                                                                                                                                                                                    | same                                                           |
| Single Use?          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                               | same                                                           |
| Sterile              | E-Beam-sterilized                                                                                                                                                                                                                                                                                                                                                                                                                                                 | same                                                           |

### Table 1: Device Comparison Table

{5}------------------------------------------------

# 510(k) Summary - K210580

Overall, the difference in technological characteristics of the subject and predicate device, i.e., the addition of fenestrations, does not raise any different questions of safety and effectiveness.

### 7. PERFORMANCE DATA

Verification testing was conducted to confirm that bioburden, endotoxin and water absorption capacity were within the specified values in accordance with the risk assessment.

As InnovaMatrix™ FS is comprised of the same materials and undergoes the same manufacturing processing steps with the inclusion of the creations, existing biocompatibility, shelf life, sterilization, human repeat insult patch testing and human skin prick testing for the predicate InnovaMatrix™ remains applicable to InnovaMatrix™ FS.

### 8. CONCLUSIONS

InnovaMatrix™ FS has the identical indications for use as the predicate device InnovaMatrix™. The technological characteristics are similar to the technological characteristics of the predicate wound dressing. The only modification for the subject device is the addition of fenestrations to the predicate device. The fenestrations of the subject device are features of the reference device, Cook Biotech's OASIS® Wound Matrix. The reference device has the same indications for use as the subject device. Both the predicate and subject devices are placental porcine-derived, decellularized dressings that are comprised primarily of collagen. The dressings are both intended for the management of wounds. Based on the indications for use, technological characteristics, and performance test results, InnovaMatrix™ FS is substantially equivalent to the predicate device InnovaMatrix™ (K193552). In addition, the Design Controls demonstrates that the modified device met the pre-determined acceptance criterion for the verification activity to substantiate the addition of fenestrations to the wound dressing.

---

**Source:** [https://fda.innolitics.com/device/K210580](https://fda.innolitics.com/device/K210580)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
