K210541 · Ulrich GmbH & Co. KG · IZQ · Aug 6, 2021 · Cardiovascular
Device Facts
Record ID
K210541
Device Name
ulrichINJECT CT motion
Applicant
Ulrich GmbH & Co. KG
Product Code
IZQ · Cardiovascular
Decision Date
Aug 6, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1650
Device Class
Class 2
Indications for Use
ulrichINJECT CT motion is a contrast media management system that is indicated for the controlled, automatic administration, on the venous side, of contrast media and saline (NaCl), to human subjects undergoing diagnostic examinations in computed tomography (CT) applications. ulrichINJECT CT motion is specifically indicated for use in CT procedures for the delivery of Omnipaque™ (lohexol) Injection, solution - GE Healthcare Inc. contrast media as supplied in Imaging Bulk Packages (IBP), Omnipaque™ (Iohexol) Injection, solution - GE Healthcare Inc., and Visipaque™ (iodixanol) Injection - GE Healthcare Inc. contrast media as supplied in single dose bottles. Pump Tubing-Flex is used for a maximum time of twenty four (24) hours. When used with Omnipaque™ IBP, Omnipaque™ single dose bottles, or Visipaque™ single dose bottles, a maximum of 19 bottles of contrast media can be used or maximum time of twenty four (24) hours of Pump Tubing-Flex, or whichever comes first. Time per contrast media or saline container depends on each contrast media's or saline's use time expiration with a maximum of eight (8) hours per contrast media or saline container. Spike for CT disposable is for single-bottle use only and must be discarded with the media container. The Patient tubing must be discarded after each patient procedure. SYNCopen is indicated for the specific purpose of allowing an injector to interface with a CT scanner. RIS/PACS is indicated for the specific purpose of allowing an injector to interface with a Radiological Information System (RIS) and a Picture Archiving and Communications System (PACS). ulrichINJECT CT motion is to be used only by and under quasi-continuous supervision of trained healthcare professionals in an appropriate licensed healthcare facility, in a room designated for radiological procedures that involve intravasular administration of contrast agent.
Device Story
Syringeless contrast media management system; utilizes peristaltic pump to deliver contrast/saline from bulk/single-dose containers to patient via disposable tubing set (spike, pump tubing, patient tubing). Operated by trained healthcare professionals in radiological suites. Features include ultrasound air detection, pressure monitoring, and check valves. New SYNCopen module enables injector-CT scanner synchronization; RIS/PACS interface allows DICOM worklist retrieval and automated dose report transmission. System prevents disposable reuse via software controls. Benefits include automated, precise contrast delivery and streamlined documentation/workflow integration.
Clinical Evidence
Bench testing only. Includes software verification/validation, electromagnetic compatibility (IEC 60601-1/1-2), sterilization validation (ISO 11135-1), shelf-life (ISO 11607-1), biocompatibility (ISO 10993-1), extractables/leachables, and contamination control (microbial ingress, cross-contamination). Usability studies confirmed safe operation of new software options.
Indicated for human subjects undergoing CT diagnostic examinations requiring controlled, automatic venous administration of contrast media (Omnipaque/Visipaque) and saline. Used by trained healthcare professionals in licensed radiological facilities.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
MEDRAD Stellant FLEX CT Injection System with Certegra Workstation (K182273)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 6, 2021
ulrich GmbH & Co. KG % Rita King CEO MethodSense Inc. 1 Copley Parkway, Suite 410 Morrisville, North Carolina 27560
Re: K210541
Trade/Device Name: ulrichINJECT CT motion Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: IZQ Dated: July 8, 2021 Received: July 9, 2021
Dear Rita King:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210541
Device Name ulrichINJECT CT motion
#### Indications for Use (Describe)
ulrichINJECT CT motion is a contrast media management system that is indicated for the controlled, automatic administration, on the venous side, of contrast media and saline (NaCl), to human subjects undergoing diagnostic examinations in computed tomography (CT) applications.
ulrichINJECT CT motion is specifically indicated for use in CT procedures for the delivery of Omnipaque™ (lohexol) Injection, solution - GE Healthcare Inc. contrast media as supplied in Imaging Bulk Packages (IBP), Omipaque™ (lohexol) Injection, solution - GE Healthcare Inc., and Visipaque™ (lodixanol) Injection, solution - GE Healthcare Inc. contrast media as supplied in single dose bottles.
Pump Tubing-Flex is used for a maximum time of twenty four (24) hours. When used with Omnipaque™ IBP, Omnipaque™ single dose bottles, or Visipaque™ single dose bottles, a maximum of 19 bottles of contrast media can be used or maximum time of twenty four (24) hours of Pump Tubing-Flex, or whichever comes first. Time per contrast media or saline container depends on each contrast media's or saline's use time expiration with a maximum of eight (8) hours per contrast media or saline container.
Spike for CT disposable is for single-bottle use only and must be discarded with the media container. The Patient tubing must be discarded after each patient procedure.
SYNCopen is indicated for the specific purpose of allowing an injector to interface with a CT scanner.
RIS/PACS is indicated for the specific purpose of allowing an injector to interface with a Radiological Information System (RIS) and a Picture Archiving and Communications System (PACS).
ulrichINJECT CT motion is to be used only by and under quasi-continuous supervision of trained healthcare professionals in an appropriate licensed healthcare facility, in a room designated for radiological procedures that involve intravasular administration of contrast agent.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
# ulrich GmbH & Co. KG K210541
This 510(k) Summary is in conformance with 21CFR 807.92
| Submitter: | ulrich GmbH & Co. KG<br>Buchbrunnenweg 12<br>89081 Ulm<br>Germany<br>Phone: +49 731 9654-1714<br>Fax: +49 731 9654-2807 |
|------------------|-------------------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Rita King, CEO<br>MethodSense, Inc.<br>Email: ritaking@methodsense.com<br>Phone: 919-313-3961<br>Fax: 919-313-3979 |
| Company Contact: | Sven Erdmann<br>Vice President Technology & Regulation |
| Date Prepared: | August 6, 2021 |
# Device Name and Classification
| Trade Name: | ulrichINJECT CT motion |
|-----------------------|--------------------------------------------------|
| Common Name: | Contrast Media Management System |
| Classification: | Class II |
| Regulation Number: | CFR 870.1650, Angiographic Injector and Syringe; |
| Classification Panel: | Cardiovascular Panel |
| Product Code: | IZQ |
# Predicate Device:
| Predicate | |
|---------------------------|-------------------------------------------------|
| Trade Name | ulrichINJECT CT motion |
| Common Name | Contrast Media Management System |
| 510(k) Submitter / Holder | ulrich GmbH & Co. KG |
| 510(k) Number | K192872 |
| Classification | Class II |
| Regulation Number | CFR 870.1650, Angiographic Injector and Syringe |
| Classification Panel | Cardiovascular Panel |
| Product Code | IZQ |
{6}------------------------------------------------
| Reference Device | |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | MEDRAD Stellant FLEX CT Injection System with Certegra<br>Workstation, MEDRAD Stellant FLEX Syringe Kits, MEDRAD<br>Stellant CT Injection System with Certegra Workstation,<br>MEDRAD Stellant Syringe Kits, MEDRAD Stellant Connector<br>Tubing, P3T Cardiac, P3T PA, P3T Abdomen, ISI, Connect.CT |
| Common Name | Injector And Syringe, Angiographic |
| 510(k) Submitter / Holder | Bayer Medical Care, Inc. |
| 510(k) Number | K182273 |
| Classification | Class II |
| Regulation Number | CFR 870.1650 |
| Classification Panel | Cardiovascular |
| Product Code | DXT |
{7}------------------------------------------------
### Device Description
ulrichINJECT CT motion is a svringeless contrast media management system that is designed for the controlled, automatic administration, on the venous side, of contrast media and saline, to human subiects undergoing diagnostic examinations in computed tomography (CT) applications.
The ulrichINJECT CT motion system consists of the CT motion terminal, CT motion injector, and the CT motion tubing system. The CT motion tubing system is the only component that comes in contact with the patient and has indirect contact with the blood path of a patient for a limited duration (few minutes). The CT motion tubing system consists of three components:
- Spike for CT
- Pump tubing-flex .
- . Patient Tubing
ulrichINJECT CT motion uses a peristaltic pump as part of the injector which is designed to transport the media fluid through the CT motion tubing system (spikes for CT, CT motion pump tubing-flex, and patient tubing for pump tubing-flex). The ulrichINJECT CT motion system is intended to be used with the following components that are not supplied with the system:
- Multiple patient use saline containers, ●
- . Omnipaque IBP contrast media containers,
- Omnipaque single-dose contrast media bottles. ●
- Visipaque single-dose contrast media bottles, and
- Cannula. .
The ulrichINJECT CT motion system is specifically indicated for use in CT procedures for the delivery of Omnipaque™ (lohexol) Injection, solution - GE Healthcare Inc. contrast media as supplied in Imaging Bulk Packages (IBP), Omnipaque™ (Iohexol) Injection, solution – GE Healthcare Inc., and Visipaque™ (iodixanol) Injection - GE Healthcare Inc. contrast media as supplied in single dose bottles.
ulrichINJECT CT motion is equipped with multiple hardware and software controls that work together for the safe operation of the intended use of the device. Controls include air detectors, which are designed to detect air without direct contact with the medium, pressure controls to manage and regulate pressure inside the tubing system, and check valves to prevent backflow of media and avoid retrograde contamination.
The ulrichINJECT CT motion is provided in three versions:
- Mobile pedestal version ●
- Ceiling version ●
- Wall mounted version .
The mobile pedestal version consists of the injector head and the injector base with rechargeable battery. The ceiling version and the wall-mounted version consist of the injector head, a fixed-height arm, and a movable arm.
The software option SYNCopen is a new software and hardware option which allows a connection between the injector system and a validated CT scanner. Both systems can communicate with each other and thus synchronize time sequences. The software option is only
{8}------------------------------------------------
available if the manufacturer of the CT scanner has enabled the connection to the injector svstem.
The ulrich medical RIS/PACS Interface is a new software option supports transferring documentation-related parameters for a contrast media injection to healthcare IT systems. A worklist can be retrieved from a RIS server by means of the DICOM modality worklist information model. After selecting a patient from the worklist and performing the iniection, a comprehensive contrast media dose report is automatically transmitted to the configured PACS and/or Dose Reporting system.
## Indications for Use
ulrichINJECT CT motion is a contrast media management system that is indicated for the controlled, automatic administration, on the venous side, of contrast media and saline (NaCI), to human subjects undergoing diagnostic examinations in computed tomography (CT) applications.
ulrichINJECT CT motion is specifically indicated for use in CT procedures for the delivery of Omnipaque™ (lohexol) Injection, solution - GE Healthcare Inc. contrast media as supplied in Imaging Bulk Packages (IBP), Omnipaque™ (Iohexol) Injection, solution - GE Healthcare Inc., and Visipaque™ (iodixanol) Injection - GE Healthcare Inc. contrast media as supplied in single dose bottles.
Pump tubing-flex is used for a maximum time of twenty four (24) hours. When used with Omnipaque™ IBP, Omnipaque™ single dose bottles, or Visipaque™ single dose bottles, a maximum of 19 bottles of contrast media can be used or maximum time of twenty four (24) hours of Pump tubing-flex, or whichever comes first. Time per contrast media or saline container depends on each contrast media's or saline's use time expiration with a maximum of eight (8) hours per contrast media or saline container.
Spike for CT disposable is for single-bottle use only and must be discarded with the media container. The Patient tubing must be discarded after each patient procedure.
SYNCopen is indicated for the specific purpose of allowing an injector to interface with a CT scanner.
RIS/PACS is indicated for the specific purpose of allowing an injector to interface with a Radiological Information System (RIS) and a Picture Archiving and Communication System (PACS).
ulrichINJECT CT motion is to be used only by and under quasi-continuous supervision of trained healthcare professionals in an appropriate licensed healthcare facility, in a room designated for radiological procedures that involve intravascular administration of contrast agent.
### Risk Analvsis Method
The ulrichINJECT CT motion was assessed to determine the risks to health associated with the use of the syringeless injector. A risk analysis was conducted in accordance with ISO
{9}------------------------------------------------
14971:2007, Medical devices - Application of risk management to medical devices. Several risks were assessed, including, but not limited to, device malfunction, contamination and infection, and improper use.
## Substantial Equivalence
The ulrichINJECT CT motion is substantially equivalent to the predicate, ulrichINJECT CT motion (K192872) currently on the market. The table below provides a detailed comparison of ulrichINJECT CT motion to the predicate device.
Additionally, the ulrichINJECT CT motion SYNCopen component is substantially equivalent to the reference device, MEDRAD Stellant FLEX CT Injection System with Certegra Workstation, MEDRAD Stellant FLEX Syringe Kits. MEDRAD Stellant CT Injection System with Certegra Workstation, MEDRAD Stellant Syringe Kits, MEDRAD Stellant Connector Tubing, P3T Cardiac, P3T PA, P3T Abdomen, ISI, Connect.CT (K182273), ISI component. The reference device (K182273) ISI functionality allows start of the injector and CT scanner simultaneously and communicates information between the injector and CT scanner, similar to the ulrichINJECT CT motion SYNCopen functionality.
{10}------------------------------------------------
# Detailed Comparison of the Subject and Predicate Device
| Item | Subject Device | Primary Predicate | Comparison |
|----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Indications for<br>Use | ulrichINJECT CT motion<br>ulrichINJECT CT motion is a contrast media<br>management system that is indicated for the<br>controlled, automatic administration, on the<br>venous side, of contrast media and saline<br>(NaCl), to human subjects undergoing<br>diagnostic examinations in computed<br>tomography (CT) applications. | ulrichINJECT CT motion (K192872)<br>ulrichINJECT CT motion is a contrast media<br>management system that is indicated for the<br>controlled, automatic administration, on the<br>venous side, of contrast media and saline<br>(NaCl), to human subjects undergoing<br>diagnostic examinations in computed<br>tomography (CT) applications. | |
| | ulrichINJECT CT motion is specifically<br>indicated for use in CT procedures for the<br>delivery of OmnipaqueTM (Iohexol) Injection,<br>solution - GE Healthcare Inc. contrast media<br>as supplied in Imaging Bulk Packages (IBP),<br>OmnipaqueTM (lohexol) Injection, solution -<br>GE Healthcare Inc., and VisipaqueTM<br>(iodixanol) Injection - GE Healthcare Inc.<br>contrast media as supplied<br>in single dose bottles. | ulrichINJECT CT motion is specifically<br>indicated for use in CT procedures for the<br>delivery of OmnipaqueTM (Iohexol) Injection,<br>solution - GE Healthcare Inc. contrast media<br>as supplied in Imaging Bulk Packages (IBP),<br>OmnipaqueTM (lohexol) Injection, solution -<br>GE Healthcare Inc., and VisipaqueTM<br>(iodixanol) Injection - GE Healthcare Inc.<br>contrast media as supplied<br>in single dose bottles. | Similar. See<br>Note 1. |
| | Pump tubing-flex is used for a maximum time<br>of twenty four (24) hours. When used with<br>OmnipaqueTM IBP, OmnipaqueTM single<br>dose bottles, or VisipaqueTM single dose<br>bottles, a maximum of 19 bottles of contrast<br>media can be used or maximum time of<br>twenty four (24) hours of Pump tubing-flex, or<br>whichever comes first. Time per contrast<br>media or saline container depends on each<br>contrast media's or saline's use time<br>expiration with a maximum of eight (8) hours<br>per contrast media or saline container. | Pump tubing-flex is used for a maximum time<br>of twenty four (24) hours. When used with<br>OmnipaqueTM IBP, OmnipaqueTM single<br>dose bottles, or VisipaqueTM single dose<br>bottles, a maximum of 19 bottles of contrast<br>media can be used or maximum time of<br>twenty four (24) hours of Pump tubing-flex, or<br>whichever comes first. Time per contrast<br>media or saline container depends on each<br>contrast media's or saline's use time<br>expiration with a maximum of eight (8) hours<br>per contrast media or saline container. | |
| Item | Subject Device | Primary Predicate | Comparison |
| | ulrichINJECT CT motion | ulrichINJECT CT motion (K192872) | |
| | Spike for CT disposable is for single-bottle<br>use only and must be discarded with the<br>media container. The Patient tubing must be<br>discarded after each patient procedure.<br>SYNCopen is indicated for the specific<br>purpose of allowing an injector to interface<br>with a CT scanner.<br>RIS/PACS is indicated for the specific<br>purpose of allowing an injector to interface<br>with a Radiological Information System (RIS)<br>and a Picture Archiving and Communication<br>System (PACS).<br>ulrichINJECT CT motion is to be used only<br>by and under quasi-continuous supervision<br>of trained healthcare professionals in an<br>appropriate licensed healthcare facility, in a<br>room designated for radiological procedures<br>that involve intravascular administration of<br>contrast agent. | Spike for CT disposable is for single-bottle<br>use only and must be discarded with the<br>media container. The Patient tubing must be<br>discarded after each patient procedure.<br>ulrichINJECT CT motion is to be used only<br>by and under quasi-continuous supervision<br>of trained healthcare professionals in an<br>appropriate licensed healthcare facility, in a<br>room designated for radiological procedures<br>that involve intravascular administration of<br>contrast agent. | |
| System | Injector Head<br>Touch Terminal | Injector Head<br>Touch Terminal | Same |
| Accessories | Injector Base<br>Wall Mount with moveable arm<br>Ceiling Mount with moveable arm<br>Contrast Media Housing with Heater | Injector Base<br>Wall Mount with moveable arm<br>Ceiling Mount with moveable arm<br>Contrast Media Housing with Heater | Same |
| | Subject Device | Primary Predicate | |
| Item | ulrichINJECT CT motion | ulrichINJECT CT motion (K192872) | Comparison |
| Disposables | ulrichINJECT CT Motion Pump Tubing-flex | ulrichINJECT CT Motion Pump Tubing-flex | |
| | Patient Tubing for Pump Tubing-flex | Patient Tubing for Pump Tubing-flex | Same |
| | ulrichINJECT CT Motion Spike for CT | ulrichINJECT CT Motion Spike for CT | Same |
| Weight | Injector (pedestal version): Approx. 80 kg | Injector (pedestal version): Approx. 80 kg | |
| | Injector (ceiling and wall mount version):<br>Approx. 40 kg | Injector (ceiling and wall mount version):<br>Approx. 40 kg | Same |
| | Terminal: Approx. 3 kg | Terminal: Approx. 3 kg | Same |
| Dimensions | Injector (pedestal version and wall mount<br>version): 64.5 x 64.5 x 144.5 cm | Injector (pedestal version and wall mount<br>version): 64.5 x 64.5 x 144.5 cm | |
| | Injector (ceiling version): Depends on the<br>system selected and the length of the fixed<br>height arm | Injector (ceiling version): Depends on the<br>system selected and the length of the fixed<br>height arm | Same |
| | Terminal: 31 x 27.5 x 17 cm | Terminal: 31 x 27.5 x 17 cm | Same |
| Power<br>Requirement | Rated Voltage: 100 to 240 V AC | Rated Voltage: 100 to 240 V AC | |
| | Rated Current: 1.6 A | Rated Current: 1.6 A | Same |
| | Rated Frequency: 50/60Hz | Rated Frequency: 50/60Hz | Same |
| Battery | Li-Ion battery | Lead gel battery or Li-Ion battery | Same |
| Display Type | Color LCD Terminal with touch screen | Color LCD Terminal with touch screen | Same |
| Syringeless<br>system | Yes | Yes | Same |
| Remote<br>Operation | Yes, via the Touch Terminal | Yes, via the Touch Terminal | Same |
| Single Patient<br>Use Disposable | Patient Tubing for Pump Tubing-flex | Patient Tubing for Pump Tubing-flex | Same |
| Item | Subject Device<br>ulrichINJECT CT motion | Primary Predicate<br>ulrichINJECT CT motion (K192872) | Comparison |
| Designed to<br>Prevent Reuse<br>of Disposables | Yes – via the use of software controls | Yes – via the use of software controls | Same |
| Rotary<br>peristaltic<br>pump<br>Used to<br>administer<br>contrast media<br>and saline | Yes | Yes | Same |
| Disposable<br>uses spikes to<br>spike media<br>container | Yes | Yes | Same |
| Safety Stop<br>Mechanism | Multi-layered software stops; Used Patient<br>Tubing detector and Pump Tubing-flex<br>detector | Multi-layered software stops; Used Patient<br>Tubing detector and Pump Tubing-flex<br>detector | Same |
| Volume<br>Remaining<br>Readout | Yes, displayed on control unit if programmed<br>volume is higher than remaining volume | Yes, displayed on control unit if programmed<br>volume is higher than remaining volume | Same |
| Programmable<br>Pressure Limit | Yes, 195 PSI; user-programmable or<br>automatic | Yes, 195 PSI; user-programmable or<br>automatic | Same |
| Injector-<br>Scanner<br>Interface | SYNCopen functionality allows start of the<br>injector and CT scanner simultaneously and<br>communicates information from the CT<br>motion to the CT scanner. | Not available. | Different. See<br>Note 2. |
| Injection<br>Capabilities | 40 phases per protocol | 40 phases per protocol | Same |
| Injection Rates<br>for Contrast<br>Media | 0.1 mL/s to 10.0 mL/s | 0.1 mL/s to 10.0 mL/s | Same |
| Injection Rates<br>for Saline | 0.1 mL/s to 10.0 mL/s | 0.1 mL/s to 10.0 mL/s | Same |
| Item | Subject Device | Primary Predicate | Comparison |
| Injection<br>Volume per<br>Injection | ulrichINJECT CT motion<br>1 to 200 mL max volume of contrast media<br>per patient with a…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.