← Product Code [GEI](/productcode/GEI) · K210526

# Medline Cautery Tip Cleaner (K210526)

_Medline Industries, Inc. · GEI · Mar 12, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210526

## Device Facts

- **Applicant:** Medline Industries, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Mar 12, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Medline Cautery Tip Cleaner is a disposable, single use sterile product. It is intended to be used as an electrosurgical accessory. The Cautery Tip Cleaner provides a method of removing eschar build-up from the tip of electrosurgical instruments during the course of a surgical procedure.

## Device Story

Medline Cautery Tip Cleaner is a disposable, sterile, single-use accessory for electrosurgical instruments. Used during surgical procedures to remove eschar build-up from instrument tips. Operated by surgical staff in clinical settings. Device functions as a mechanical cleaning surface to maintain instrument performance and cleanliness during surgery.

## Clinical Evidence

No clinical data provided; device is a simple mechanical accessory.

## Technological Characteristics

Disposable, single-use, sterile electrosurgical accessory. Mechanical cleaning surface for eschar removal.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below.

March 12, 2021

Medline Industries, Inc. Matt Clausen Senior Regulatory Affairs Specialist Three Lakes Drive Northfield, Illinois 60093

Re: K210526

Trade/Device Name: Medline Cautery Tip Cleaner Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: February 16, 2021 Received: February 23, 2021

Dear Matt Clausen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Long H. Chen
Long H. Chen -S -s
Date: 2021.03.12 16:12:49 -05'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K210526

Device Name Medline Cautery Tip Cleaner

Indications for Use (Describe)

The Medline Cautery Tip Cleaner is a disposable, single use sterile product. It is intended to be used as an electrosurgical accessory. The Cautery Tip Cleaner provides a method of removing eschar build-up from the tip of electrosurgical instruments during the course of a surgical procedure.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                 |                                                                                                                                                                                                           |                                                                                                                                                                                                           |                                                                                                                                                               |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <table border="0"><tr><td><table border="0"><tr><td><div align="left"> <input checked="checked" name="form1[type_of_use][]" type="checkbox" value="Residential Use (SINGLE FAMILY Lots Only)"/>Residential Use (SINGLE FAMILY Lots Only)         </div></td></tr></table></td><td><table border="0"><tr><td><div align="left"> <input name="form1[type_of_use][]" type="checkbox" value="Commercial Use (ALL OTHER Lots)"/>Commercial Use (ALL OTHER Lots)         </div></td></tr></table></td></tr></table> | <table border="0"><tr><td><div align="left"> <input checked="checked" name="form1[type_of_use][]" type="checkbox" value="Residential Use (SINGLE FAMILY Lots Only)"/>Residential Use (SINGLE FAMILY Lots Only)         </div></td></tr></table> | <div align="left"> <input checked="checked" name="form1[type_of_use][]" type="checkbox" value="Residential Use (SINGLE FAMILY Lots Only)"/>Residential Use (SINGLE FAMILY Lots Only)         </div>       | <table border="0"><tr><td><div align="left"> <input name="form1[type_of_use][]" type="checkbox" value="Commercial Use (ALL OTHER Lots)"/>Commercial Use (ALL OTHER Lots)         </div></td></tr></table> | <div align="left"> <input name="form1[type_of_use][]" type="checkbox" value="Commercial Use (ALL OTHER Lots)"/>Commercial Use (ALL OTHER Lots)         </div> |
| <table border="0"><tr><td><div align="left"> <input checked="checked" name="form1[type_of_use][]" type="checkbox" value="Residential Use (SINGLE FAMILY Lots Only)"/>Residential Use (SINGLE FAMILY Lots Only)         </div></td></tr></table>                                                                                                                                                                                                                                                               | <div align="left"> <input checked="checked" name="form1[type_of_use][]" type="checkbox" value="Residential Use (SINGLE FAMILY Lots Only)"/>Residential Use (SINGLE FAMILY Lots Only)         </div>                                             | <table border="0"><tr><td><div align="left"> <input name="form1[type_of_use][]" type="checkbox" value="Commercial Use (ALL OTHER Lots)"/>Commercial Use (ALL OTHER Lots)         </div></td></tr></table> | <div align="left"> <input name="form1[type_of_use][]" type="checkbox" value="Commercial Use (ALL OTHER Lots)"/>Commercial Use (ALL OTHER Lots)         </div>                                             |                                                                                                                                                               |
| <div align="left"> <input checked="checked" name="form1[type_of_use][]" type="checkbox" value="Residential Use (SINGLE FAMILY Lots Only)"/>Residential Use (SINGLE FAMILY Lots Only)         </div>                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                 |                                                                                                                                                                                                           |                                                                                                                                                                                                           |                                                                                                                                                               |
| <div align="left"> <input name="form1[type_of_use][]" type="checkbox" value="Commercial Use (ALL OTHER Lots)"/>Commercial Use (ALL OTHER Lots)         </div>                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                 |                                                                                                                                                                                                           |                                                                                                                                                                                                           |                                                                                                                                                               |

> Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/device/K210526](https://fda.innolitics.com/device/K210526)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).