K210502 · Stryker Neurovascular · POL · Aug 27, 2021 · Neurology
Device Facts
Record ID
K210502
Device Name
Trevo NXT ProVue Retriever
Applicant
Stryker Neurovascular
Product Code
POL · Neurology
Decision Date
Aug 27, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5600
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K210502 · Aug 27, 2021
Trevo NXT ProVue Retriever
Stryker Neurovascular
Administrative claims database; The Trevo Retriever Registry; The ASSIST Registry; Clinical literature
RWD/RWE was used to assess the safety and effectiveness of using the Trevo Retriever with aspiration catheters having a minimum inner diameter of 0.046 inches. Regression and equivalence analyses of these data sources demonstrated that clinical outcomes and safety endpoints were equivalent when using smaller aspiration catheters compared to larger ones.
Administrative claims; Post-market registry; Literature review; Equivalence analysis; Clinical outcomes
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Analysis of Administrative Claims and Post-market Registries (Trevo Retriever Registry and ASSIST Registry); Retrospective analysis of claims data and registry cohorts; Follow-up/Duration: 90-day mRS
Patients treated with Trevo Retriever for acute ischemic stroke
Trevo Retriever used with larger aspiration catheters
Functional outcome (mRS), revascularization rates (eTICI), and safety endpoints
Indications for Use
The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (0-50cc for age <80 years, 0-20cc for age ≥80 years). Endovascular therapy with the device should start within 6-24 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.
Device Story
Trevo NXT ProVue Retriever is a neurovascular mechanical thrombectomy device. Input: fluoroscopic visualization of device within neurovasculature. Operation: device delivered via microcatheter to thrombus site; microcatheter retracted to deploy shaped section; aspiration catheter advanced over retriever while applying aspiration (syringe/pump); retriever, thrombus, and catheter removed as unit. Used in clinical/interventional settings by physicians. Output: restored blood flow (recanalization). Benefit: reduced disability in acute ischemic stroke patients.
Clinical Evidence
No new clinical trials performed. Evidence includes real-world data (RWD) from administrative claims, Trevo Retriever Registry, and ASSIST Registry. Regression analysis showed equivalence in functional outcomes (OR 0.72; 95% CI: 0.44-1.17) and safety endpoints when using smaller vs. larger aspiration catheters. Registry data confirmed revascularization rates and clinical outcomes (90-day mRS) were equivalent within a ±5% margin.
Indicated for patients with acute ischemic stroke, persistent proximal anterior circulation large vessel occlusion, and smaller core infarcts. Candidates include those ineligible for or failing IV t-PA therapy. Treatment window: 0-6 hours (with IV t-PA), 0-8 hours (ineligible/failed IV t-PA), or 6-24 hours (ICA/MCA-M1 occlusion, specific core infarct volumes based on age).
Regulatory Classification
Identification
A neurovascular mechanical thrombectomy device for acute ischemic stroke treatment is a prescription device used in the treatment of acute ischemic stroke to improve clinical outcomes. The device is delivered into the neurovasculature with an endovascular approach, mechanically removes thrombus from the body, and restores blood flow in the neurovasculature.
Special Controls
In combination with the general controls of the FD&C Act, the Neurovascular Mechanical Thrombectomy Device for Acute Ischemic Stroke Treatment is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient contacting components of the device must be demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including:
(i) Mechanical testing to demonstrate the device can withstand anticipated tensile, torsional, and compressive forces.
(ii) Mechanical testing to evaluate the radial forces exerted by the device.
(iii) Non-clinical testing to verify the dimensions of the device.
(iv) Non-clinical testing must demonstrate the device can be delivered to the target location in the neurovasculature and retrieve simulated thrombus under simulated use conditions.
(v) Non-clinical testing must demonstrate the device is radiopaque and can be visualized.
(vi) Non-clinical testing must evaluate the coating integrity and particulates under simulated use conditions.
(vii) Animal testing must evaluate the safety of the device, including damage to the vessels or tissue under anticipated use conditions.
(3) Performance data must support the sterility and pyrogenicity of the patient contacting components of the device.
(4) Performance data must support the shelf-life of the device by demonstrating continued sterility, package integrity, and device functionality over the specified shelf-life.
(5) Clinical performance testing of the device must demonstrate the device performs as intended for use in the treatment of acute ischemic stroke and must capture any adverse events associated with the device and procedure.
(6) The labeling must include:
(i) Information on the specific patient population for which the device is intended for use in the treatment of acute ischemic stroke, including but not limited to, specifying time from symptom onset, vessels or location of the neurovasculature that can be accessed for treatment, and limitations on core infarct size.
(ii) Detailed instructions on proper device preparation and use for thrombus retrieval from the neurovasculature.
(iii) A summary of the clinical testing results, including a detailed summary of the device- and procedure-related complications and adverse events.
(iv) A shelf life.
{0}------------------------------------------------
August 27, 2021
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Stryker Neurovascular Rebecca Rosman Senior Staff Regulatory Affairs Specialist 47900 Bayside Parkway Fremont, California 94538
### Re: K210502
Trade/Device Name: Trevo NXT ProVue Retriever Regulation Number: 21 CFR 882.5600 Regulation Name: Neurovascular Mechanical Thrombectomy Device for Acute Ischemic Stroke Treatment Regulatory Class: Class II Product Code: POL, NRY Dated: July 22, 2021 Received: July 26, 2021
Dear Rebecca Rosman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaolin Zheng, Ph.D. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210502
Device Name Trevo NXTTM ProVue Retriever
#### Indications for Use (Describe)
1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.
2. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
3. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (0-50cc for age <80 years, 0-20cc for age ≥80 years). Endovascular therapy with the device should start within 6-24 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span> </span> </div> |
|----------------------------------------------|-----------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span> </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary K210502
# Introduction:
According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for a determination of substantial equivalence.
# Submitter Name, Address and Content:
| Submitter: | Stryker Neurovascular<br>47900 Bayside Parkway<br>Fremont, CA 94538-6515<br>(FDA Registration Number: 3008853977) |
|------------|------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Ashley Twitty<br>Manager, Regulatory Affairs<br>Phone: 602-621-3089<br>Fax: 510-413-2588<br>Email: ashley.twitty@stryker.com |
August 23, 2021
Device Name and Classification:
Date Prepared:
| Trade/Proprietary Name: | Trevo NXTTM ProVue Retriever |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Trevo Retriever |
| Classification Name: | Neurovascular Mechanical Thrombectomy Device for Acute Ischemic<br>Stroke Treatment, 21 CFR 882.5600, Class II<br>Percutaneous Catheter, 21CFR 870.1250 – Class II |
| Product Code: | POL. NRY |
{4}------------------------------------------------
# Legally Marketed Predicate Device
| Name of Predicate Device | Name of<br>Manufacturer | 510(k)<br>Number |
|--------------------------------|--------------------------|------------------|
| Trevo NXT™ ProVue<br>Retriever | Stryker<br>Neurovascular | K203219 |
#### Device Description
The Trevo Retriever consists of a flexible, tapered core wire with a shaped section at the distal end. Platinum markers at the distal end allow fluoroscopic visualization. In addition, the shaped section is also radiopaque. Retriever dimensions are indicated on product label. The Retriever delivery wire has a hydrophilic coating on the distal 101cm length to reduce friction during use. The Retriever has a shaft marker to indicate proximity of Retriever tip relative to Microcatheter tip. A Torque Device is provided with the Retriever to facilitate manipulation and retrieval. The Retriever comes preloaded in an insertion tool to introduce the Retriever into a Microcatheter.
### Indications for Use
- 1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.
- 2. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
- 3. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (0-50cc for age <80 years, 0-20cc for age ≥80 years). Endovascular therapy with the device
{5}------------------------------------------------
should start within 6-24 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.
#### Technological Characteristics and Product Feature Comparison
Stryker Neurovascular has demonstrated the Trevo NXT ProVue Retriever with the modified IFU is substantially equivalent to the Predicate device, Trevo NXT ProVue Retriever (K203219) based on the same indications for use, device design, materials, manufacturing, packaging and sterilization methods. A comparison of the Subject device with the Predicate device is summarized in Table 1 below.
| Feature | Predicate Device<br>Trevo NXT ProVue Retriever | Subject Device<br>Trevo NXT ProVue Retriever |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Regulation<br>Number | 21 CFR 882.5600 21 CFR 870.1250 | Same |
| Regulation<br>Name | Neurovascular Mechanical Thrombectomy Device for Acute Ischemic Stroke Treatment Percutaneous Catheter | Same |
| Classification | Class II | Same |
| Product Code | POL NRY | Same |
| Intended Use | Neurovascular mechanical thrombectomy device for acute ischemic stroke treatment used in the treatment of acute ischemic stroke to improve clinical outcomes. | Same |
| Indications for<br>Use | 1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.<br>2. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator | Same |
| Feature | Predicate Device<br>Trevo NXT ProVue Retriever | Subject Device<br>Trevo NXT ProVue Retriever |
| | (IV t-PA) or who fail IV t-PA therapy<br>are candidates for treatment.<br>3. The Trevo Retriever is indicated for use<br>to restore blood flow in the<br>neurovasculature by removing thrombus<br>for the treatment of acute ischemic<br>stroke to reduce disability in patients<br>with a persistent, proximal anterior<br>circulation, large vessel occlusion of the<br>internal carotid artery (ICA) or middle<br>cerebral artery (MCA)-M1 segments<br>with smaller core infarcts (0-50cc for age<br><80 years, 0-20cc for age ≥80 years).<br>Endovascular therapy with the device<br>should start within 6-24 hours of time<br>last seen well in patients who are<br>ineligible for intravenous tissue<br>plasminogen activator (IV t-PA) or who<br>fail IV t-PA therapy. | |
| Target<br>Population | Patients experiencing acute ischemic stroke | Same |
| Anatomical<br>Sites | Neurovasculature | Same |
| | TECHNOLOGICAL CHARACTERISTICS | |
| Device<br>Description | The Retriever consists of a flexible, tapered<br>core wire with a shaped section at the distal<br>end. Platinum markers at the distal end allow<br>fluoroscopic visualization. In addition, the<br>shaped section is also radiopaque. Retriever<br>dimensions are indicated on product label.<br>The Retriever delivery wire has a<br>hydrophilic coating on the distal 101cm to<br>reduce friction during use. The Retriever has<br>a shaft marker to indicate proximity of<br>Retriever tip relative to Microcatheter tip. A<br>torque device is provided with the Retriever<br>to facilitate manipulation and retrieval. The<br>Retriever comes preloaded in an insertion<br>tool to introduce the Retriever into a<br>Microcatheter. | Same |
| Principle of<br>Operation | The Trevo Retriever is delivered to the<br>thrombus using a microcatheter. The<br>Microcatheter is then retracted to deploy the<br>shaped section of the Retriever. If using an<br>Aspiration Catheter, remove Microcatheter.<br>Advance Aspiration Catheter over proximal<br>section of Retriever while applying<br>aspiration using a 60 mL syringe or an<br>aspiration pump. The Retriever with<br>Microcatheter or Aspiration Catheter are | Same |
Table 1. Product Feature Comparison of Subject Device to Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
| Feature | Predicate Device<br>Trevo NXT ProVue Retriever | Subject Device<br>Trevo NXT ProVue Retriever | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| | pulled back to capture the thrombus. The<br>Retriever, thrombus, and Microcatheter or<br>Aspiration Catheter are removed as a unit<br>from the body. | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Procedural<br>Steps<br>Aspiration<br>Source | Syringe, Aspiration pump | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Accessory<br>Devices | Insertion tool and torque device provided<br>within product package | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Sizes | 3x32mm<br>4x28mm<br>4x41mm<br>6x37mm | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Compatibility | Retriever<br>Size Trevo Pro14<br>Microcatheter<br>Inner Diameter<br>0.017 inches<br>(0.43 mm) Trevo Trak 21<br>Microcatheter<br>Inner Diameter<br>0.021 inches<br>(0.53 mm) Trevo Pro18<br>Microcatheter<br>Inner Diameter<br>0.021 inches<br>(0.53 mm) Excelsior®<br>XT-27®<br>Microcatheter<br>(REF XT275081)<br>Inner diameter<br>0.027 inches<br>(0.69 mm) Recommended<br>Minimum<br>Vessel ID (mm) Trevo NXT<br>3x32 ✓ ✓ ✓ ✓ Trevo NXT<br>4x28 ✓ ✓ ✓ 2.5 Trevo NXT<br>4x41 ✓ ✓ ✓ ✓ Trevo NXT<br>6x37 ✓ ✓ ✓ Compatibility of the Retriever with other<br>microcatheters has not been established.<br>Performance of<br>the Retriever device may be impacted if a<br>different microcatheter is used.<br><br>Balloon Guide Catheters and Aspiration<br>Catheters (commercially available<br>aspiration catheters with minimum inner<br>diameter 0.058 inches (1.47mm)) are<br>recommended for use during thrombus<br>removal procedures. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Retriever<br>Size Trevo Pro14<br>Microcatheter<br>Inner Diameter<br>0.017 inches<br>(0.43 mm) Trevo Trak 21<br>Microcatheter<br>Inner Diameter<br>0.021 inches<br>(0.53 mm) Trevo Pro18<br>Microcatheter<br>Inner Diameter<br>0.021 inches<br>(0.53 mm) Excelsior®<br>XT-27®<br>Microcatheter<br>(REF XT275081)<br>Inner diameter<br>0.027 inches<br>(0.69 mm) Recommended<br>Minimum<br>Vessel ID (mm) Trevo NXT<br>3x32 ✓ ✓ ✓ ✓ Trevo NXT<br>4x28 ✓ ✓ ✓ 2.5 Trevo NXT<br>4x41 ✓ ✓ ✓ ✓ Trevo NXT<br>6x37 ✓ ✓ ✓ Compatibility of the Retriever with other<br>microcatheters has not been established.<br>Performance of<br>the Retriever device may be impacted if a<br>different microcatheter is used.<br><br>Balloon Guide Catheters and Aspiration<br>Catheters (commercially available<br>aspiration catheters with minimum inner<br>diameter 0.046 inches (1.17mm)) are<br>recommended for use during thrombus<br>removal procedures. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| MATERIALS | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Core Wire | Nitinol (nickel titanium alloy) | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Shaped Section | Nitinol | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Distal Coil | Platinum/Tungsten | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Shaped Section<br>Radiopaque<br>Wire | Platinum/Tungsten | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Mid Coil | 304 Stainless Steel | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Proximal Coil | Pebax | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Solder | Gold/Tin | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
{8}------------------------------------------------
| Feature | Predicate Device<br>Trevo NXT ProVue Retriever | Subject Device<br>Trevo NXT ProVue Retriever |
|---------------------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------|
| Hydrophilic<br>Coating | Sodium hyaluronate mixture | Same |
| DIMENSIONAL DRAWING | | |
| Dimensional<br>Drawing | All Retriever sizes:<br>Image: Dimensional Drawing with labels A, B, C, D, E, and F | All Retriever sizes:<br>Same |
| Overall Length<br>(A) | 200cm | Same |
| Total Shaped<br>Section Length<br>(B) | 32, 36, 40, 44mm | Same |
| Full Diameter<br>Length (C) | 21, 25, 30, 35mm | Same |
| Shaped Section<br>Diameter (D) | 3, 4, 6mm | Same |
| Delivery Core<br>Wire Outer<br>Diameter (E) | 0.015, 0.019" | Same |
| Cell Coverage<br>Length (F) | 28, 32, 37, 41mm | Same |
| PACKAGING | | |
| Materials and<br>Configuration | Polyethylene Hoop, polycarbonate mounting<br>card, Tyvek/Film Pouch, Chipboard carton | Same |
| Sterilization<br>Method | 100% EtO | Same |
| How Supplied | Sterile/Single Use | Same |
The differences between the devices are not critical as demonstrated above and through the testing referenced below.
#### Risk Assessment
Risk assessment of the Trevo NXT ProVue Retriever has been conducted in accordance with EN ISO 14971. Stryker Neurovascular has determined that the labeling changes to the Trevo NXT ProVue Retriever raise no new questions of safety or effectiveness. Results of testing are appropriate for determining that the Trevo NXT ProVue Retriever with the modified IFU is substantially equivalent to the legally marketed Predicate device.
{9}------------------------------------------------
# Testing Summary
There are no changes to the device intended use or indications for use statement. Other than the proposed labeling changes regarding compatibility, there are no changes in the device design, materials, manufacturing, packaging and sterilization methods; therefore, biocompatibility data, bench performance data, sterilization and stability data from the Predicate device (K203219) are directly applicable. Relevant testing data supporting the Subject device are summarized as follows.
### Performance Data – Bench Testing
Stryker Neurovascular performed the following non-clinical bench test to assess the usability of the Trevo Retriever with 0.046 in. ID aspiration catheters. The additional bench testing is summarized in Table 2 below.
| Test | Test Method Summary | Conclusions |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Simulated Use | Simulated Use testing utilized a<br>neurovascular model to assess the<br>device's ability to retrieve the clot and<br>achieve recanalization. | Simulated Use testing<br>met acceptance criteria. |
| Particulate<br>Characterization | Purpose: The purpose of this test was to<br>document and assess the particulate<br>matter generated from the delivery of<br>neuro-interventional devices (including<br>the Trevo NXT ProVue Retriever)<br>through the inner lumen of the AXS<br>Vecta 46 Intermediate Catheter.<br><br>Method: Particulate testing was<br>conducted with AXS Vecta 46 and Trevo<br>NXT ProVue Retriever in a simulated<br>use condition based on the device IFU<br>and a clinically relevant tortuous model. | All test samples met the<br>applicable user needs<br>and acceptance criteria. |
Table 2. Performance Data - Design Verification Bench Testing
Performance Data – Animal
{10}------------------------------------------------
To support the labeling modification, Stryker Neurovascular leveraged data from an animal study that was conducted in compliance with applicable requirements in the GLP regulation (21 CFR Part 58) to evaluate vascular trauma, recanalization and/or distal emboli of the proposed devices for performing a combined neurothrombectomy procedure.
## Performance Data – Clinical
To support the labeling modifications, Stryker Neurovascular conducted analyses of Real-World Data (RWD) from an administrative claims database, post-market registries (The Trevo Retriever Registry and The ASSIST Registry), as well as a review of relevant literature to assess safety and effectiveness of the recommended use of the Trevo with aspiration catheters with a minimum inner diameter (ID) of 0.046in (1.17mm).
Regression analysis of RWD indicated that using Trevo with smaller aspiration catheters is equivalent to using Trevo with larger aspiration catheters with similar odds of good functional outcome (0.72; 95% CI: 0.44-1.17) and similar odds for multiple safety endpoints. Further analysis of registry data confirmed equivalence of revascularization rates and clinical outcomes independent of catheter size with the 90% CI for the difference in means of the severity adjusted posterior probabilities for eTICI>2c to be from 0.003 to 0.009 (.30% to 0.90%). For 90-day MRS, the CI was from -0.050 to 0.042 (-5.0% to 4.2%). In both cases, the upper and lower bounds of the intervals were within the equivalence margin of ±5%.
#### Shelf Life Testing
Shelf life testing previously conducted for the Trevo NXT ProVue Retriever was leveraged to support the changes to the device and can be found in K192207. Shelf life testing was not performed on the Subject device since there was no impact to device material, design, or safety and efficacy as a result of the labeling changes. As with the Predicate device, the Subject device is labeled with a 2-year shelf life.
# Sterilization
Sterilization evaluation previously conducted for the Trevo NXT ProVue Retriever was leveraged to support the labeling changes to the device and can be found in K192207. The Trevo
{11}------------------------------------------------
NXT ProVue Retrievers are sterilized with 100% Ethylene Oxide and provided sterility assurance level (SAL) of 10 6 has been demonstrated. The Trevo NXT ProVue Retrievers meet EO residuals per EN ISO 10993-7 for limited contacting, externally communicated devices. The Trevo NXT ProVue Retrievers are for single use only.
# Biocompatibilitv
Biocompatibility testing previously conducted for the Trevo NXT ProVue Retriever was leveraged to support the labeling changes to the device and can be found in K192207. The results of biocompatibility testing, and biological safety evaluation of the Trevo NXT ProVue Retrievers demonstrate that the devices meet biological safety requirements per ISO 10993-1 for externally communicating medical devices with circulating blood contact for less than 24 hours. The Trevo NXT ProVue Retrievers are considered to have no residual risk of biological hazards. Also, the devices and its packaging do not contain detectable latex. Therefore, the Trevo NXT ProVue Retriever devices, accessories, and primary packaging are considered biocompatible for their intended use.
# Conclusion
Based on the conclusions drawn from risk assessment, bench testing results, Real World Data from an Administrative Claims, and an analysis of relevant clinical literature summarized above, the Subject device demonstrates substantial equivalence to the legally marketed Predicate device (K203219).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.