← Product Code [GCJ](/productcode/GCJ) · K210495

# Port Access System (K210495)

_Surgical Innovations Limited · GCJ · Jun 18, 2021 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210495

## Device Facts

- **Applicant:** Surgical Innovations Limited
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jun 18, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Port Access System is indicated for use in laparoscopic procedures to give access to the abdominal and thoracic cavities while maintaining pneumoperitoneum.

## Device Story

Port Access System provides surgical access to abdominal and thoracic cavities during laparoscopic procedures; maintains pneumoperitoneum. System comprises trocar and cannula assembly; trocar penetrates body wall; cannula serves as conduit for endoscopes and surgical instruments. Device used in hospital settings by surgeons. Single-use components (copolyester) and reusable components (stainless steel, PEEK) available. Single-use parts gamma-sterilized; reusable parts require cleaning/sterilization. System accommodates instrument diameters 5mm-12mm. Facilitates minimally invasive surgery by allowing instrument passage while preventing gas leakage.

## Clinical Evidence

No clinical studies were conducted. Substantial equivalence was demonstrated through non-clinical bench testing, including sealing performance, instrument insertion/friction, insufflation, seal core inversion, tensile/torque strength, fatigue, shelf life, transit studies, cleaning/sterilization validation, and usability testing. Biocompatibility was evaluated per ISO 10993-1:2018.

## Technological Characteristics

Materials: Copolyester (single-use), Stainless Steel (reusable trocar), PEEK plastic (reusable cannula). Dimensions: 5, 10, 12 mm diameters; 55-150 mm working lengths. Components: Trocar, cannula, non-return valve, instrument seal. Sterilization: Gamma irradiation for single-use components. Connectivity: None. Energy: None (mechanical).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- YelloPort Elite Port Access System ([K190592](/device/K190592.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

June 18, 2021

Surgical Innovations Limited Stephen Seed Compliance Director Clayton Wood House, 6 Clayton Wood Bank Leeds, West Yorkshire LS16 6OZ United Kingdom

Re: K210495

Trade/Device Name: Port Access System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: May 27, 2021 Received: June 1, 2021

Dear Stephen Seed:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long H. Chen. Ph.D. Assistant Director Non-Light Based Energy Devices Team DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210495

Device Name Port Access System

Indications for Use (Describe)

The Port Access System is indicated for use in laparoscopic procedures to the abdominal and thoracic cavities while maintaining pneumoperitoneum.

Type of Use (Select one or both, as applicable)

| <span style="font-size: 10px;">☑ Prescription Use (Part 21 CFR 201.5 Subject to Rx)</span> |
|--------------------------------------------------------------------------------------------|
| <span style="font-size: 10px;">☐ Over-The-Counter Use (21 CFR 201.5 Subject to OTC)</span> |

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Surgical Innovations. The word "Surgical" is in a teal color, and the word "Innovations" is in a gray color. The logo is simple and modern.

Port Access System Section 5: 510(k) Summary

#### -Submitter

Surgical Innovations Limited Clayton Wood House 6 Clayton Wood Bank Leeds LS16 6QZ United Kingdom

Phone: +44 (0) 113 230 7597

Contact Person: Stephen Seed Date Prepared: 25 March 2021

#### II. Device

Name of Device: Port Access System Common Name: Laparoscopic Port System Classification Name: Endoscope and Accessories (21 CFR 876.1500) Requlatory Class: II Product Code: GCJ

#### lll. Predicate Device

YelloPort Elite Port Access System - K190592

The predicate has not been subject to any design-related recalls.

#### IV. Device Description

The Port Access System consists of reusable and single use components sold in various configurations. The ports accommodate laparoscopic instrument diameters from 5mm up to the diameter indicated on the chosen accompanying cannula. All single use parts are sterlised using gamma irradiation and supplied to the customer in a sterile condition.

In order to obtain access to the surgical site during laparoscopic surgery, the trocar is introduced into the cannula to accomplish cannula penetration of the abdominal wall. The cannula is connected to the single use seal at its proximal end and once the abdominal/thoracic wall is punctured, the trocar is removed. The cannula acts as a channel for the introduction of the endoscopes and instruments.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for Surgical Innovations. The word "Surgical" is in a teal color, and the word "Innovations" is in a gray color. The logo is simple and modern.

#### V. Indications for Use

The Port Access System is indicated for use in laparoscopic procedures to give access to the abdominal and thoracic cavities while maintaining pneumoperitoneum.

#### VI. Comparison of Technological Characteristics with the Predicate Device

| Property                                 | Proposed Device                                                                                                                                                              | Primary Predicate                                                                                                                                                                                                                                                                                                            |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                          |                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                              |
| Common Name                              | Port Access System<br>Laparoscope, General & Plastic<br>Surgery                                                                                                              | YelloPort Elite Port Access System<br>Laparoscope, General & Plastic<br>Surgery                                                                                                                                                                                                                                              |
| Device Manufacturer                      | Surgical Innovations                                                                                                                                                         | Surgical Innovations                                                                                                                                                                                                                                                                                                         |
| Device Classification                    | II                                                                                                                                                                           | II                                                                                                                                                                                                                                                                                                                           |
| Primary Product Code                     | GCJ                                                                                                                                                                          | GCJ                                                                                                                                                                                                                                                                                                                          |
| 510(k) Number                            | N/A                                                                                                                                                                          | K190592                                                                                                                                                                                                                                                                                                                      |
| Environment                              | Hospital                                                                                                                                                                     | Hospital                                                                                                                                                                                                                                                                                                                     |
| Intended Use/ Indication for Use         | The Port Access System is indicated<br>for use in laparoscopic procedures to<br>give access to the abdominal and<br>thoracic cavities while maintaining<br>pneumoperitoneum. | The YelloPort Elite Port Access<br>System is indicated for use in<br>laparoscopic procedures to give<br>access to the abdominal cavity while<br>maintaining pneumoperitoneum.<br>The YelloPort Elite Port access<br>system is also indicated for use in<br>laparoscopic procedures to give<br>access to the thoracic cavity. |
| Trocar Tip Geometries                    | Bladeless, Pencil point, Locking<br>Shielded, Quill                                                                                                                          | Blunt, Pencil point, Pyramidal,<br>Shielded (locking)                                                                                                                                                                                                                                                                        |
| Trocar / Cannula Diameter and<br>Lengths | Available in 5, 10 and 12 mm<br>diameters; and in 55, 70, 95, and 105<br>and 150 mm working lengths.                                                                         | Available in 10 and 12 mm diameter;<br>and in 75, 105 and 150 mm working<br>lengths.                                                                                                                                                                                                                                         |
| Comparison of Use:                       | Allows instruments from 5mm -12mm<br>in diameter to be used.                                                                                                                 | Allows instruments from 5mm -12mm<br>in diameter to be used.                                                                                                                                                                                                                                                                 |
| Trocar Patient Contacting Material       | Single Use: Copolyester<br>Reusable: Stainless Steel                                                                                                                         | Stainless Steel                                                                                                                                                                                                                                                                                                              |
| Trocar Supplied:                         | Single Use: Sterile<br>Reusable: Non-Sterile                                                                                                                                 | Non-sterile (Reusable)                                                                                                                                                                                                                                                                                                       |
| Cannula Patient Contacting<br>Materials  | Single Use: Copolyester<br>Reusable: PEEK plastic                                                                                                                            | PEEK plastic                                                                                                                                                                                                                                                                                                                 |
| Cannula Supplied:                        | Single Use: Sterile<br>Reusable: Non-Sterile                                                                                                                                 | Non-sterile (Reusable)                                                                                                                                                                                                                                                                                                       |
| Seal                                     | Consists of:<br>Non-Return Valve &<br>Instrument Seal                                                                                                                        | Consists of:<br>Non-Return Valve &<br>Instrument Seal                                                                                                                                                                                                                                                                        |
| Seal Supplied:                           | Sterile                                                                                                                                                                      | Sterile                                                                                                                                                                                                                                                                                                                      |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for Surgical Innovations. The word "Surgical" is in teal, and the word "Innovations" is in gray. The logo is simple and modern.

#### VII. Performance Data

A comparative review of the Port Access System with the predicate devices was conducted and the following performance data are provided in support of determining substantial equivalence of Port Access System to the predicate devices.

## Biocompatibility

Biological evaluation of the single use and reusable components of the Port Access System was conducted in accordance with International Standard ISO 10993-1:2018 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process and the US FDA quidance document Use of International Standard ISO 10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and testing within a risk management process' (2016 & 2020).

## Non-Clinical Testing

The Port Access System was evaluated for its safety and effectiveness based on the following testing:

- Sealing Performance ●
- Instrument Insertion Force (Universal Seal)
- Instrument Shaft Friction (Universal Seal)
- . Instrument Compatibility (Universal Seal)
- Seal Contamination
- Insertion Force
- . Insufflation
- . Seal Core Inversion Testing
- Tensile ●
- . Torque
- Endoscope Insertion/Retraction
- . Fatigue
- Shelf Life
- Transit studies
- . Cleaning & Sterilisation Validation
- . Usability

## Clinical Studies

No Clinical Studies were conducted as part of this submission. Non-Clinical Testing was sufficient to demonstrate substantial equivalence to the predicate device.

#### VIII. Conclusions

The information presented in the 510(k) premarket notifications demonstrates that the Port Access System is considered substantially equivalent to the predicate device. The Non-Clinical Bench Testing performed demonstrates that the Port Access System is safe, effective and performs within its design specifications and is substantially equivalent to the predicate devices. The Port Access System is as safe and effective as the currently marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K210495](https://fda.innolitics.com/device/K210495)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
